(33 days)
Powdered vinyl patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
powdered vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06.
The provided text describes the acceptance criteria and performance of "Powdered Vinyl Patient Examination Gloves, Clear (Non-colored)." It does not pertain to an AI/ML device or software. Therefore, many of the requested categories in the prompt are not applicable.
Here's the information that can be extracted relevant to the prompt, acknowledging the nature of the device:
1. A table of acceptance criteria and the reported device performance
Characteristics | Acceptance Criteria (Standard) | Reported Device Performance |
---|---|---|
Dimension | ASTM standard D 5250-06 | Meets |
Physical Properties | ASTM standard D 5250-06 | Meets |
Freedom from pinholes | 21 CFR 800.20 (Waterleak test on pinhole AQL) | Meets |
Powder Amount | ASTM standard D 5250-06 | Meets (` |
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.