(61 days)
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.
This document is a 510(k) summary for a "Powder Free Vinyl Patient Examination Glove," which is a Class I medical device. The document primarily focuses on establishing substantial equivalence to a predicate device and demonstrating compliance with relevant ASTM standards and FDA requirements. It does not present a study proving the device meets acceptance criteria in the typical sense of a clinical or AI-based diagnostic device study.
Instead, the "acceptance criteria" here refer to the performance standards and tests required for this type of medical glove to be marketed. The "study" refers to the non-clinical tests performed to demonstrate compliance with these standards.
Here's the breakdown of the information requested, tailored to the context of this device:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Standard/Test) | Reported Device Performance |
|---|---|
| ASTM D5250-06e1 Standard for Vinyl Patient Examination Gloves (General Requirements) | All testing meets requirements for physical and dimension testing conducted on gloves. |
| FDA 1000 ml. Water Fill Test (Pinhole integrity) | Samplings (AQL 2.5, inspection level I) meet requirements. |
| Primary Skin Irritation Test | No primary skin irritant reactions. |
| Skin Sensitization (Allergic Contact Dermatitis) Test | No sensitization reactions. |
| Residual Powder Test (ASTM D6124-06) | Gloves meet "powder-free" claims (contain no more than 2 mg powder per glove). |
| Biocompatibility Requirements | Met (implied by skin irritation/sensitization tests). |
| Labeling Claims | No special labeling claims; not claimed as hypoallergenic. |
2. Sample Size Used for the Test Set and Data Provenance
- ASTM D5250-06e1 (Physical and Dimensions Testing): Inspection level S-2, AQL 4.0. (Specific sample size not explicitly stated, but these AQL values dictate the sampling plan).
- FDA 1000 ml. Water Fill Test: Inspection level I, AQL 2.5. (Specific sample size not explicitly stated, but these AQL values dictate the sampling plan).
- Biocompatibility Tests (Skin Irritation and Sensitization): Sample size for testing on subjects is not specified.
- Residual Powder Test (ASTM D6124-06): Sample size not specified.
- Data Provenance: The document does not specify the country of origin for the testing or if it was retrospective or prospective. Given it's a 510(k) submission, the testing would be conducted prospectively to demonstrate compliance before market entry.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- Not Applicable. This is a physical product (medical glove) and not a diagnostic device that requires expert interpretation for a "ground truth." The "ground truth" for these tests comes from objective measurements against established ASTM standards and FDA requirements.
4. Adjudication Method for the Test Set
- Not Applicable. As explained above, this is about objective measurements against standards, not subjective interpretations requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, What Was the Effect Size of How Much Human Readers Improve With AI vs. Without AI Assistance
- Not Applicable. This is a medical glove, not an AI-assisted diagnostic device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done
- Not Applicable. This is a medical glove, not an algorithm/AI.
7. The Type of Ground Truth Used
- The "ground truth" for this device is based on objective measurements and adherence to established industry standards (ASTM D5250-06e1, ASTM D6124-06) and regulatory requirements (FDA 1000 ml. Water Fill Test, biocompatibility criteria).
8. The Sample Size for the Training Set
- Not Applicable. This is a physical product, not a machine learning model, so there is no "training set."
9. How the Ground Truth for the Training Set Was Established
- Not Applicable. As above, there is no "training set" for this type of device.
{0}------------------------------------------------
EXHIBIT #I Page 1 of 2
510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(K) number is: _ K | D1137
l . Owner's Identification:
Mr. Jack Wang Hongze Plastic Technology Co., Ltd. West District of Shenze Industrial Zone, Fugian West Road, Shenze County, Shijiazhuang, Hebei, China, 050000
Date Summary Prepared: March 22, 2010
-
- Name of the Device: Hongze Plastic Technology Co., Ltd.
Powder Free Vinyl Patient Examination Glove
- Name of the Device: Hongze Plastic Technology Co., Ltd.
3. Predicate Device Information:
Better Care Plastic Products Co., Ltd Powder Free Vinyl Patient Examination Gloves (K092296)
4. Device Description:
Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.
5. Intended Use:
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
6. Comparison to Predicate Devices:
Hongze Plastic Technology Co., Ltd.'s Powder Free Vinyl Patient Examination Glove is substantially equivalent in safety and effectiveness to Better Care Plastic Products Co., Ltd's Powder Free Vinyl Patient Examination Gloves.
1:
{1}------------------------------------------------
KI01137
EXHIBIT #1 Page 2 of 2
7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:
The standards used for Hongze Plastic Technology Co., Ltd.'s glove production are based on ASTM D5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.
Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
A Residual Powder Test that based on ASTM D6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).
8. Labeling:
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
9. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic Claim.
10. Conclusions:
Hongze Plastic Technology Co., Ltd.'s Powder Free Vinyl Patient Examination Glove conform fully to ASTM D5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center is a stylized symbol that resembles three overlapping human figures or abstract shapes.
Public Health Service
JUN 2 2 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Hongze Plastic Technology Company, Limited C/O Ms. Kathy Liu Treeline Trading Incorporated 1930 Brea Canyon Road Diamond Bar, California 91765
Re: K101137
Trade/Device Name: Powder Free Vinyl Patient Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: March 22, 2010 Received: April 22, 2010
Dear Ms. Kathy Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Liu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Susan Runne
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Attachment A
INDICATION FOR USE
510 (k) NUMBER (IF KNOW): APPLICANT: DEVICE NAME:
Hongze Plastic Technology Co., Ltd. Powder Free Vinyl Patient Examination Glove
INDICATIONS FOR USE:
A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Prescription Use (Part 21 CFR 801 Subpart D) AND/ OR Over-The-Counter-Use 4 (21CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Division Sign-Off Concurrent of CDRH, Office of De ്വുടെ പ്രദേശം പ്രദേശീർമ്മിക്കും, General Hospital Infection Control, Dental Devices
K101134 510(k) Number: _
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.