K Number
K100908
Device Name
UNICEM HM
Date Cleared
2010-06-22

(82 days)

Product Code
Regulation Number
872.3275
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Final cementation of ceramic, composite or metal inlays, onlays, crowns, bridges, 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis); Final cementation of post and screws; Final cementation of ceramic, composite or metal restorations on implant abutments; Cementation of abutments made of Lava™ zirconium oxide ceramic.

Device Description

Unicem HM in the Clicker™ dispenser is a dual-curing, self-adhesive resin cement for hand mixing. It contains bi-functional (meth)acrylate. The proportion of inorganic fillers is about 70% by weight; the grain size (D 90%) is about 12.5 um. The mixing ratio; based on volume, is 1 part base paste : 1 part catalyst. The cement is available in various shades.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the Unicem HM device, based on the provided 510(k) summary:

Acceptance Criteria and Device Performance

The device, Unicem HM, was compared to its predicate device, Unicem Aplicap, across several performance metrics. The acceptance criteria were established by relevant ISO and DIN standards.

Table 1: Acceptance Criteria and Reported Device Performance for Unicem HM

performance MetricMethodLimit (Acceptance Criteria)Unicem HM PerformanceUnicem Aplicap Performance (Predicate)
Film thickness [µm]ISO 4049< 50$9 \pm 3$$12 \pm 2$
Working time [sec]ISO 4049> 60passpass
Setting time [min:s]ISO 4049< 1003:4505:00
Radiopacity mmISO 4049> 1.02.32.4
Flexural strength [MPa] (dark cured)ISO 4049> 50$57 \pm 11$$53 \pm 7$
Flexural strength [MPa] (light cured)ISO 4049> 50$87 \pm 9$$64 \pm 6$
Compressive strength [MPa] (dark cured)ISO 9917> 70$247 \pm 16$$209 \pm 15$
Compressive strength [MPa] (light cured)ISO 9917> 70$281 \pm 15$$218 \pm 13$
Surface hardness [MPa] (dark cured)DIN 2039-1n.a. (Not Applicable)$202 \pm 8$$209 \pm 13$
Surface hardness [MPa] (light cured)DIN 2039-1n.a. (Not Applicable)$280 \pm 35$$151 \pm 10$

Summary of Device Performance against Acceptance Criteria:

Based on the table above, Unicem HM met all the specified limits for film thickness, working time, setting time, radiopacity, flexural strength (both dark and light cured), and compressive strength (both dark and light cured). For surface hardness, which had no specified numerical limit ("n.a."), the values are provided for comparison. The study concluded that Unicem HM is "safe and effective as as the predicate devices."


Study Details

The provided document describes a bench testing study comparing Unicem HM to a predicate device, Unicem Aplicap, and other predicate devices (Panavia F 2.0, Maxcem 2) were also referenced for substantial equivalence. This is not a clinical study involving human patients.

  1. Sample size used for the test set and the data provenance:

    • The document does not explicitly state the sample sizes (number of specimens tested) for each performance metric. It provides mean values and standard deviations ($ \pm $ values), which implies multiple samples were tested for each metric to calculate these statistics.
    • Data Provenance: The company submitting the 510(k) is 3M ESPE AG, based in Germany. Therefore, the data likely originated from Germany. The study is retrospective in the sense that the data appears to be from laboratory tests conducted to demonstrate compliance with standards for the premarket notification.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This question is not applicable (N/A) to this type of study. Ground truth, in the context of expert consensus, refers to clinical assessments or interpretations. As this is a bench testing study evaluating material properties against established standards, it does not involve expert readers establishing ground truth. The "ground truth" here is objective physical and chemical properties measured according to standardized methods.
  3. Adjudication method for the test set:

    • N/A. Adjudication methods (like 2+1, 3+1) are used to resolve disagreements among human experts in clinical studies or image interpretation. This study involves objective measurements of material properties, not human interpretation, so no adjudication method was used.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an MRMC study. It is a bench testing study for a dental cement, not a diagnostic device involving human readers or AI assistance.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • No. This is not an algorithm or AI device. It is a material science study.
  6. The type of ground truth used:

    • The "ground truth" in this context is based on established international standards (ISO and DIN) for dental materials. The device's performance was measured against these objective standards. Additionally, the performance of the predicate device serves as a benchmark for comparison to demonstrate substantial equivalence.
  7. The sample size for the training set:

    • N/A. This is a material testing study, not a machine learning study, so there is no concept of a "training set."
  8. How the ground truth for the training set was established:

    • N/A. As there is no training set for this type of study, this question is not applicable.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

JUN 2 2 2010

Submitter

)

g

Company:3M ESPE AG
Street:ESPE Platz
ZIP-Code, City:D-82229 Seefeld
Federal State:Bavaria
Country:Germany
Establishment Registration Number9611385
Official Correspondent:Dr. Desi W. Soegiarto,Regulatory Affairs Specialist
Phone:011-49-8152-700 1169
Fax:011-49-8152-700 1869
E-mail:desi.soegiarto@mmm.com
Date:June 09, 2010

Name of Device

Proprietary Name:Unicem HM
Classification Name:Dental cement other than zinc oxide- eugenol
Common Name:Self adhesive cement

Predicate Devices

Unicem Aplicap/Maxicap by 3M ESPE AG, Germany.........K094007, K020256 Panavia F 2.0 by Kuraray Medical Inc., Japan ................................................................................................................................. Maxcem 2 by Kerr Corporation, U.S.A., presumably marketed as Maxcem Elite ..........................................................................................................................................

{1}------------------------------------------------

Description for the Premarket Notification

i

Unicem HM is classified as a Dental cement other than zinc oxide-eugenol (21 C.F.R. § 872.3275 [b]) because it is a device composed of various materials other than zinc oxideeugenol.

Unicem HM in the Clicker™ dispenser is a dual-curing, self-adhesive resin cement for hand mixing. It contains bi-functional (meth)acrylate. The proportion of inorganic fillers is about 70% by weight; the grain size (D 90%) is about 12.5 um. The mixing ratio; based on volume, is 1 part base paste : 1 part catalyst. The cement is available in various shades.

Unicem HM has been cleared under K062292. This 510(k) Premarket Notification has been submitted in order to seek clearance for additional new Indications for Use listed hereafter:

Final cementation of 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis);

Final cementation of ceramic, composite or metal restorations on implant abutments Cementation of abutments made of Lava™ zirconium oxide ceramic.

The predicate devices to which Unicem HM has been compared are Unicem Aplicap/Maxicap by 3M ESPE AG, Germany (K094007, K020256), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455) and Maxcem 2 by Kerr Corporation, U.S.A., presumably marketed as Maxcem Elite (K073209).

As its predicate devices, Unicem HM is a dual-curing (chemical and light) resin based cement system containing methacrylate to be used with metal, composite, and porcelain restorations. Maxcem Elite and Unicem HM are both paste/paste systems.

The intended use of Unicem HM is comparable to the area of the intended use of the predicate devices of Unicem HM.

In this 510(k) premarket notification Unicem HM has been compared to its predicate devices with regard to chemical composition, performance data and indications for use. The comparison for chemistry, performance data and indications for use shows that Unicem HM is substantially equivalent to the predicate devices: Unicem Aplicap/Maxicap by 3M ESPE AG, Germany (K094007, K020256), Panavia F 2.0 by

{2}------------------------------------------------

Kuraray Medical Inc., Japan (K032455) and Maxcem 2 by Kerr Corporation, U.S.A., presumably marketed as Maxcem Elite (K073209).

The following table shows the performance data of Unicem HM and its predicate device Unicem Aplicap:

MethodLimitUnicemHMUnicemAplicap
Film thickness [µm]ISO 4049< 50$9 \pm 3$$12 \pm 2$
Working time [sec]ISO 4049> 60passpass
Setting time [min:s]ISO 4049< 1003:4505:00
Radiopacity mmISO 4049> 1.02.32.4
Flexural strength [MPa]dark cured ISO 4049> 50$57 \pm 11$$53 \pm 7$
Flexural strength [MPa]light cured ISO 4049> 50$87 \pm 9$$64 \pm 6$
Compressive strength [MPa]dark cured ISO 9917> 70$247 \pm 16$$209 \pm 15$
Compressive strength [MPa]light cured ISO 9917> 70$281 \pm 15$$218 \pm 13$
Surface hardness [MPa]dark cured DIN 2039-1n.a.$202 \pm 8$$209 \pm 13$
Surface hardness [MPa]light cured DIN 2039-1n.a.$280 \pm 35$$151 \pm 10$

Biocompatibility testing was carried out.

ﺮﺓ ﻣﺴﺘ

In summary, it can be concluded that Unicem HM is a safe and effective as as the predicate devices: Unicem Aplicap/Maxicap by 3M ESPE AG, Germany (K094007, K020256), Panavia F 2.0 by Kuraray Medical Inc., Japan (K032455) and Maxcem 2 by Kerr Corporation, U.S.A., presumably marketed as Maxcem Elite (K073209).

{3}------------------------------------------------

Indications for Use

Final cementation of ceramic, composite or metal inlays, onlays, crowns, bridges, 2-3unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis);

Final cementation of post and screws;

Final cementation of ceramic, composite or metal restorations on implant abutments; Cementation of abutments made of Lava™ zirconium oxide ceramic.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized emblem consisting of three curved shapes, possibly representing human figures or abstract forms. The emblem is black, and the text is also in black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Dr. Desi W. Soegiarto Regulatory Affairs Specialist 3M ESPE AG Dental Products ESPE Platz Seefeld, Bavaria Germany D-82229

JUN 2 2 2010

Re: K100908

Trade/Device Name: Unicem HM Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: March 29, 2010 Received: April 1, 2010

Dear Dr. Soegiarto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2- Dr. Soegiarto

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Susan Gunnes

Anthony D. Watson, B.S., M.S.. M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Indications For Use:

K100908

Unicem HM

Final cementation of ceramic, composite or metal inlays, onlays, crowns, bridges, 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis) Final cementation of post and screws Final cementation of ceramic, composite or metal restorations on implant abutments Cementation of abutments made of Lava™ zirconium oxide ceramic

X Prescription Use (Part 21 CFR 801 Subpart Di

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Kein Muluy for MSR
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of _

510(k) Number: K100908

§ 872.3275 Dental cement.

(a)
Zinc oxide-eugenol —(1)Identification. Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 872.9.(b)
Dental cement other than zinc oxide-eugenol —(1)Identification. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.(2)
Classification. Class II.