(127 days)
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion.
The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The Accu Vision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), turnor. stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
This submission is a line extension to the Polaris Spinal System to add new titanium dominoes and washers to the system.
The provided document is a 510(k) summary for the Polaris Spinal System, which is a medical device, and not an AI/ML device. Therefore, the common acceptance criteria for AI/ML devices such as accuracy, sensitivity, specificity, AUC, etc., and the associated study designs (MRMC, standalone, human-in-the-loop) are not applicable.
Instead, this document describes a line extension to an existing spinal fixation system, focusing on its substantial equivalence to predicate devices based on mechanical performance testing.
Here's an analysis based on the information provided regarding the device's performance:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Mechanical) | Reported Device Performance (Mechanical) |
|---|---|
| Static Transverse Rotation (per ASTM 1798-97) | Met all mechanical test requirements based on worst-case construct testing. |
| Static Axial Slip (per ASTM 1798-97) | Met all mechanical test requirements based on worst-case construct testing. |
| Dynamic Axial Compression Bending-Fatigue (per ASTM 1717-04) | Met all mechanical test requirements based on worst-case construct testing. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample size for the mechanical tests (e.g., number of constructs or replicates tested).
The data provenance is not explicitly stated as country of origin, but it refers to "Mechanical testing was conducted in accordance with FDA's Guidance for Industry and FDA Staff - Spinal System 510(k)s dated May 3, 2004." This suggests in-house or contracted laboratory testing following US regulatory guidelines. The testing is prospective in the sense that the tests were performed specifically for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is not applicable as the "ground truth" for a mechanical test is defined by the physical limits and standards set within the ASTM specifications, not by human expert assessment. The "experts" would be the engineers and technicians conducting the tests and interpreting the results against the ASTM standards.
4. Adjudication method for the test set
Not applicable for mechanical testing. The results are quantitative and compared against pre-defined ASTM standards.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical spinal fixation system, not an AI diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable for a physical medical device.
7. The type of ground truth used
The "ground truth" for the performance evaluation is based on established industry standards and guidance documents for mechanical testing of spinal systems:
- FDA's Guidance for Industry and FDA Staff - Spinal System 510(k)s dated May 3, 2004
- ASTM 1798-97 (reapproved 2003) for Static Transverse Rotation and Static Axial Slip
- ASTM 1717-04 for Dynamic Axial Compression Bending-Fatigue
8. The sample size for the training set
Not applicable. The concept of a "training set" is relevant for AI/ML models, not for mechanical testing of a medical device.
9. How the ground truth for the training set was established
Not applicable for mechanical testing.
In summary, the study described in the K100409 document is a mechanical performance study designed to demonstrate substantial equivalence of the new components (titanium dominoes and washers) of the Polaris Spinal System to existing predicate devices. The acceptance criteria are defined by industry-standard ASTM tests and FDA guidance for spinal systems. The device "meets the acceptance criteria" by satisfying the requirements of these mechanical tests, which are intended to ensure the device's safety and effectiveness for its intended use as a spinal fixation system.
{0}------------------------------------------------
Kloo4og
Image /page/0/Picture/1 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in large, bold, sans-serif font, with each letter outlined in black. Below the word "BIOMET" is the word "SPINE" in a smaller, sans-serif font, also outlined in black. The logo is simple and clean, with a focus on the company name.
JUN 2 8 2010
510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.92.
| Preparation Date: | June 22, 2010 |
|---|---|
| Applicant/Sponsor: | Biomet Spine100 Interpace ParkwayParsippany, NJ 07054 |
| Contact Person: | Vivian KellyPhone: 973-299-9300 x2214Fax: 973-257-0232 |
| Trade name: | Polaris Spinal System |
| Common Name: | Non-cervical spinal fixation system |
| Classification Name(Product Code): | Posterior, noncervical, nonpedicle use (KWP)Anterior/anterolateral noncervical use (KWQ)Noncervical pedicle applications (MNI, MNH and NKB) |
| Device Panel - Regulation No.: | Orthopedic - 21 CFR 888.3050, 888.3060 and 888.3070 |
Device Description:
This submission is a line extension to the Polaris Spinal System to add new titanium dominoes and washers to the system.
Indications for Use:
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion.
The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
Page 1 of 2
{1}------------------------------------------------
100409
The Accu Vision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), turnor. stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
Summary of Technologies:
The technological characteristics of the new components are the same as, or similar to, the predicate devices.
Performance Data:
Mechanical testing was conducted in accordance with FDA's Guidance for Industry and FDA Staff - Spinal System 510(k)s dated May 3, 2004. Per the guidance document, the following testing was conducted: Static Transverse Rotation and Static Axial Slip per ASTM 1798-97 (reapproved 2003) and Dynamic Axial Compression Bending-Fatigue per ASTM 1717-04. The mechanical testing verifies that the subject device is substantially equivalent to other spinal systems currently on the market and has met all mechanical test requirements based on the worst-case construct testing.
Substantial Equivalence:
The Polaris Spinal System is substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. Examples of predicates include the Polaris Spinal System (K090523), the Synergy Spinal System (K950099) and the Altius OCT System (K033961).
Conclusion:
The subject device is substantially equivalent to its predicate devices when used as a spinal fixation device. The indications for use and fundamental technology of the device remain unchanged. Furthermore, mechanical testing and other supporting information sufficiently demonstrate the substantial equivalence of the subject device to the other components in the Polaris Spinal System. Based on this information, the subject device does not raise any new issues regarding the safety or efficacy.
Page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Biomet Spine % Ms. Vivian Kelly, MS, RAC Regulatory Affairs Specialist 100 Interpace Parkway Parsippany, New Jersey 07054
TIM 8 8 5010
Re: K100409
Trade/Device Name: Polaris Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB,MNH, MNI, KWP, KWQ Dated: May 21, 2010 Received: May 24, 2010
Dear Ms. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Vivian Kelly, MS, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm11.5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Barbara Brennan
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known): K100409
Device Name: Polaris Spinal System
Indications for Use:
The Polaris Spinal System is a non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion as a pedicle screw fixation system, a posterior hook and sacral/iliac screw fixation system, or as an anterolateral fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft. The device is indicated for all the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation); deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion.
The Ballista instruments are intended to be used with the 5.5 Polaris implants. The Ballista instruments when used with the Ballista cannulated screws and percutaneous rods, are indicated to provide the surgeon with a percutaneous approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, and lordosis), tumor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a non-cervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The Accuvision Instruments, when used with the Polaris Spinal System implants are indicated to provide the surgeon with a minimally invasive approach for posterior spinal surgery for the following indications: degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma, (i.e., fracture or dislocation), deformity or curvature (i.e., scoliosis, kyphosis, and lordosis), turnor, stenosis, pseudarthrosis, and failed previous fusion that warrant the use of a noncervical spinal fixation device intended for use as a pedicle screw fixation system or sacral/iliac screw fixation system. Pedicle screw fixation is limited to skeletally mature patients and for use with autograft.
The dominos in the Polaris Spinal System can be used to connect the Polaris Spinal System to the Altius Spinal System, The Array Spinal System, the Biomet Omega21 Spinal System, or the Synergy Spinal System to achieve additional levels of fixation. Please refer to the individual system's Package Insert for a list of the indications for use for each system.
Prescription Use X
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
KI00409 510(k) Number_
Page 9 of 74
N/A