(104 days)
The Diamond Anterior Cervical Plate System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
- Degenerative Disc Disease (as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies)
- Trauma (including fractures) .
- Tumors .
- Deformities or curvatures (including kyphosis, lordosis, or scoliosis) ●
- Pseudarthrosis .
- . Failed previous fusion
- . Spondylolisthesis
- . Spinal Stenosis.
Warning: This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine
The Diamond Anterior Cervical Plate System is made of a medical grade titanium alloy and Nitinol, and consists of plates and screws of various lengths to accommodate single or multilevel fusions and variations in patients' anatomy. One level plates range from 22mm to 34mm. Two level plates range from 36mm. Three level plates range from 53mm to 77mm. Four level plates range from 68mm to 92mm. All plates are 18mm wide and have an average thickness of 2.5mm. The screws range in sizes from 12 mm to 18 mm and come in diameters of 4.0mm and 4.3mm.
This document describes the 510(k) summary for the Diamond Anterior Cervical Plate System.
Here's an analysis of the provided text, focusing on the requested information regarding acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical integrity tested according to ASTM F1717 standard for static and dynamic tests. | "The Diamond Anterior Cervical Plate System was mechanically tested using ASTM 1717. Tests included were: Static Compression Bending, Static Torsion, Static Tension Bending, and Dynamic Compression Bending." |
| Device performance is substantially equivalent to predicate devices. | "The pre-clinical testing performed indicated that the Diamond Anterior Cervical Plate is substantially equivalent to the predicate devices and is adequate for the intended use." |
| Material composition equivalent to predicate devices. | "The Diamond Anterior Cervical Plate System and Medtronic Orion Anterior Cervical Plating System are both composed of Titanium Alloy and Nitinol. Synthes Spine CSLP and X-Spine ACT system are both composed of Titanium Alloy." |
| Dimensions (width, thickness, lengths) equivalent to predicate devices. | "The dimensions of Synthes Spine CSLP variable plate is 18mm wide and 2.5mm thick which is roughly equivalent to the Diamond Anterior Cervical Plate which is 18mm wide and has an average thickness of 2.5mm. The Diamond Anterior Cervical Plate System is also provided in the same lengths as the X-Spine ACT system." |
2. Sample size used for the test set and the data provenance
The document only mentions "pre-clinical testing" and does not specify the sample size for the mechanical tests (number of plates/screws tested). Data provenance (e.g., country of origin, retrospective/prospective) is not provided as the study is a mechanical test rather than a clinical study with patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This is a mechanical device performance study, not a study requiring expert-established ground truth on clinical images or patient outcomes.
4. Adjudication method for the test set
Not applicable. This is a mechanical device performance study.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a mechanical device, not an AI-powered diagnostic or assistive tool.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Not applicable. This is a mechanical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for this device's performance is established by the ASTM F1717 standard physical performance requirements for intervertebral body fixation orthoses. Compliance with these established engineering standards serves as the ground truth.
8. The sample size for the training set
Not applicable. There is no training set mentioned, as this is a mechanical device and not an AI/machine learning model.
9. How the ground truth for the training set was established
Not applicable, as there is no training set. The "ground truth" for evaluating the mechanical device is based on established engineering standards (ASTM F1717).
{0}------------------------------------------------
510(k) Summary according to 807.92(c)
MAY 1 3 2010
| Date: | April 9, 2010 |
|---|---|
| Contact: | Tim Lusby |
| Amendia™, LLC | |
| 1155 Allgood Road, Suite 6 | |
| Marietta, GA 30062 | |
| 770-874-0935 | |
| Trade Name: | Diamond Anterior Cervical Plate System |
| Product Class: | Class II |
| Classification: | 888.3060 |
| Product Codes: | KWQ |
| Panel Code: | 87 Orthopedics |
Indications for Use: The Diamond Anterior Cervical Plate System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
- Degenerative Disc Disease (as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies)
- Trauma (including fractures) .
- Tumors .
- Deformities or curvatures (including kyphosis, lordosis, or scoliosis) ●
- Pseudarthrosis .
- . Failed previous fusion
- . Spondylolisthesis
- . Spinal Stenosis.
Warning: This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine
Device Description: The Diamond Anterior Cervical Plate System is made of a medical grade titanium alloy and Nitinol, and consists of plates and screws of various lengths to accommodate single or multilevel fusions and variations in patients' anatomy. One level plates range from 22mm to 34mm. Two level plates range from 36mm. Three level plates range from 53mm to 77mm. Four level plates range from 68mm to 92mm. All plates are 18mm wide and have an average thickness of 2.5mm. The screws range in sizes from 12 mm to 18 mm and come in diameters of 4.0mm and 4.3mm.
Predicate Device(s): Predicate devices previously cleared by FDA include the X-Spine ACP (K041469), the Synthes CSLP (K000536) and the Medtronic Orion Anterior Cervical Plate System K042499).
{1}------------------------------------------------
Performance Testing: The Diamond Anterior Cervical Plate System was mechanically tested using ASTM 1717. Tests included were: Static Compression Bending, Static Torsion, Static Tension Bending, and Dynamic Compression Bending. The pre-clinical testing performed indicated that the Diamond Anterior Cervical Plate is substantially equivalent to the predicate devices and is adequate for the intended use.
Comparison with Predicate: The Diamond Anterior Cervical Plate System and Medtronic Orion Anterior Cervical Plating System are both composed of Titanium Alloy and Nitinol. Synthes Spine CSLP and X-Spine ACT system are both composed of Titanium Alloy. The dimensions of Synthes Spine CSLP variable plate is 18mm wide and 2.5mm thick which is roughly equivalent to the Diamond Anterior Cervical Plate which is 18mm wide and has an average thickness of 2.5mm. The Diamond Anterior Cervical Plate System is also provided in the same lengths as the X-Spine ACT system.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like strokes. The bird is positioned to the right of a circular arrangement of text. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 1 3 2010
Amendia™, LLC % Mr. Tim Lusby 1155 Allgood Road, Suite 6 Marietta, Georgia 30062
Re: K100265
Trade/Device Name: Diamond Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWO Dated: April 09, 2010 Received: April 12, 2010
Dear Mr. Lusby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. Tim Lusby
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Harbare Brielm
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
3. Statement of Indications for Use
510(k) Number (if known): K100265
Indications for Use:
The Diamond Anterior Cervical Plate System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The system is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusions in patients with the following indications:
- Degenerative Disc Disease (as defined by neck pain of discogenic origin with . degeneration of the disc confirmed by patient history and radiographic studies)
- Trauma (including fractures) .
- Tumors .
- Deformities or curvatures (including kyphosis, lordosis, or scoliosis) ●
- . Pseudarthrosis
- Failed previous fusion
- . Spondylolisthesis
- Spinal Stenosis. .
Warning: This device is not approved for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine
Prescription Use デ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
§ 888.3060 Spinal intervertebral body fixation orthosis.
(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.