K Number
K093976
Date Cleared
2010-03-25

(91 days)

Product Code
Regulation Number
870.2920
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Raisin™ Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relatively to gravity, and time-stamped, patient-logged events. The Raisin™ Personal Monitor enables unattended data collection for clinical and research applications. The Raisin™ Personal Monitor may be used in any instance where quantifiable analysis of eventassociated heart rate, activity, and body position is desirable.

Device Description

The Raisin100 Personal Monitor (RPM) is a miniaturized, ambulatory, battery-operated datalogging device that is worn on the torso to record heart rate, activity, and patient-logged events. Patient-logged events can be extrinsic (e.g., dosing of a medication) or intrinsic (e.g., a symptom) and are time-stamped using a manual button on the device, in order to contextualize the physiologic measures. Subjective meaning of these events is assigned by the user. In addition to quantification of physical motion, signals from the device's accelerometer are used to determine body position relative to gravity. Electrode-to-electrode impedance is also measured to assess whether the device is attached properly to the user. RPM recorded data are transferred via Bluetooth telemetry to a general computing device for display and conversion for export to other programs. The RPM is available in two form factors to accommodate individual comfort preferences: one-piece and two-piece. The functionality, intended use, duration and location of wear, and fundamental scientific technologies are exactly the same between the two RPM form factors.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Raisin™ Personal Monitor, based on the provided text:

Acceptance Criteria and Device Performance

ParameterAcceptance Criteria (Expected Results)Reported Device Performance (Algorithm Results)
Heart Rate R-wave Detection (Bench Testing - ANSV/AAMI EC 13 standard)
Default ECG waveform80 bpm80.0 bpm
T-wave rejection (R-wave 1 mV, T-wave 0.4 mV)80 bpm80.0 bpm
Ventricular bigeminy80 bpm79.9 bpm
Slow alternating ventricular bigeminy60 bpm60.5 bpm
Rapid alternating ventricular bigeminy120 bpm119.8 bpm
Bidirectional systoles90 bpm90.1 bpm
Default ECG waveform (Pacing pulse 2 mV, 2 ms width)80 bpm80.0 bpm
Heart Rate R-wave Detection (Arrhythmia Database)
Positive Detection AccuracyNot explicitly stated (implied high)Median: 99.7%, Standard Deviation: 5.9%
False Positive RateNot explicitly stated (implied low)Median: 0%, Standard Deviation: 1.7%
Heart Rate R-wave Detection (Clinical Setting - Various Body Locations)Not explicitly stated (implied near 100%)Anterior Chest: 99.40%Xyphoid: 99.17%Stomach: 99.07%Lateral Chest: 98.82%(Average R-wave detection accuracy)
Accelerometer (Bench Validation)Known acceleration applied against each of its three axes to demonstrate linearityShown in scatter plots with strong linear correlation between measured and applied acceleration for X, Y, and Z axes.
Accelerometer (Clinical Validation)Capture expected features of subject movement (e.g., walking)Demonstrated data from a representative walking test showing expected acceleration fluctuations.

Study Information

  1. Sample size used for the test set and the data provenance:

    • Heart Rate R-wave Detection (Arrhythmia Database): All 48 test files from the MIT-BIH arrhythmia database.
    • Heart Rate R-wave Detection (Clinical Setting - Various Body Locations): Data from 4 subjects (Subject 1, Subject 2, Subject 4, Subject 5) across different body locations. The provenance is not explicitly stated but implies prospective clinical data collection for these subjects.
    • Accelerometer (Clinical Validation): Data from a "representative walking test" and "a representative subject." The exact number of subjects or tests is not explicitly stated.
    • Accelerometer (Bench Validation): Not applicable for test set size as it's a bench validation with known inputs.
    • Data Provenance: The MIT-BIH arrhythmia database is a publicly available, retrospective database. Clinical data for the heart rate and accelerometer validation appears to be prospectively collected (e.g., "representative walking test," "subject 1, chest location, sitting").
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Heart Rate R-wave Detection (Arrhythmia Database): Not specified in the provided text. The MIT-BIH arrhythmia database has widely accepted expert annotations, but the number and qualifications of the original annotators are not detailed here.
    • Heart Rate R-wave Detection (Clinical Setting - Various Body Locations): Not specified. The reference is to "automatically identified R-waves highlighted" against the captured ECG waveform, implying a comparison against the raw ECG, but the method for establishing the true R-wave locations for accuracy calculation (e.g., manual expert review) is not detailed.
    • Accelerometer (Bench Validation): Not applicable, as ground truth is the "known acceleration applied."
    • Accelerometer (Clinical Validation): Not specified. The "expected features" of walking would likely derive from general biomechanical understanding, not specific expert annotations for each test.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not specified in the provided text for any of the studies.

  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not performed or reported. This device is primarily a data logger and not an AI-assisted diagnostic tool for human readers.

  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the reported performance metrics (e.g., R-wave detection accuracy, false positive rate, accelerometer linearity) are all indicative of standalone algorithm performance. The device is described as an "unattended data collection" system.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Heart Rate R-wave Detection (Arrhythmia Database): R-wave locations were "annotated R-wave locations" from the MIT-BIH arrhythmia database, which typically involves expert-reviewed annotations.
    • Heart Rate R-wave Detection (Bench Testing): "Expected Results (bpm)" based on the ANSV/AAMI EC 13 standard, implying predefined reference values for specific ECG waveforms.
    • Heart Rate R-wave Detection (Clinical Setting - Various Body Locations): "Automatically identified R-waves highlighted" implies comparison against captured ECG, but the true ground truth for accuracy calculation is not explicitly stated (e.g., manual expert annotation of raw ECG).
    • Accelerometer (Bench Validation): "Known acceleration applied" (physical inputs).
    • Accelerometer (Clinical Validation): "Expected features" of movement (e.g., walking), based on general physiological understanding of movement patterns.
  7. The sample size for the training set: Not specified. The studies describe validation testing but do not provide details on the training set used for developing the device's algorithms.

  8. How the ground truth for the training set was established: Not specified, as training set details are not provided.

{0}------------------------------------------------

510(k) Summary

1.4.1 -- 510(k) Owner

Proteus Biomedical 2600 Bridge Parkway, Suite 101 Redwood City, CA 94065 (650) 632-4031 (tel) (650) 632-4071 (fax)

MAR 2 5 2010

193976 777

1.4.2 -- Contact Person

Gregory Moon, MD

1.4.3 -- Date Summary Prepared

February 10, 2010

1.4.4 -- Name of Device

Trade name:Raisin™ Personal Monitor
Common name:Physiological Data and Event Logging Device
Classification name:Cardiovascular Transmitter and Receiver (Product Code DXH)

1.4.5 -- Predicate Devices

HealthePod™ (K083174) Actiheart® (K052489) Actiwatch-Score® (K991033)

1.4.6 -- Device Description and Technologic Characteristics

.. .

The Raisin100 Personal Monitor (RPM) is a miniaturized, ambulatory, battery-operated datalogging device that is worn on the torso to record heart rate, activity, and patient-logged events.

{1}------------------------------------------------

Patient-logged events can be extrinsic (e.g., dosing of a medication) or intrinsic (e.g., a symptom) and are time-stamped using a manual button on the device, in order to contextualize the physiologic measures. Subjective meaning of these events is assigned by the user. In addition to quantification of physical motion, signals from the device's accelerometer are used to determine body position relative to gravity. Electrode-to-electrode impedance is also measured to assess whether the device is attached properly to the user. RPM recorded data are transferred via Bluetooth telemetry to a general computing device for display and conversion for export to other programs. The RPM is available in two form factors to accommodate individual comfort preferences: one-piece and two-piece. The functionality, intended use, duration and location of wear, and fundamental scientific technologies are exactly the same between the two RPM form factors.

1.4.6.1 -- Basic technologies

ParameterSensor TechnologyMethod
Heart rateBiopotential low-frequency amplifierDigitized R wave
ActivityAccelerometerDigitized accelerometer output
Body angleAccelerometerDouble integration of accelerometer output
Patient event loggingPatient activated buttonDigital pulse
Inter-electrode impedanceBiopotential high-frequency amplifierDigitized impedance from small auxiliary current

Proteus Biomedical, Inc.

{2}------------------------------------------------

1.4.6.2 -- Physical Characteristics

ParameterValue
ShapeOne-piece: ovoid
Two-piece: triangular
SizeOne-piece: 115 x 54 x 12 mm
Two-piece: 95 x 84 x 10 mm
Weight50 g
20 g
Battery typeRechargeable lithium ion
Moisture susceptibilityWaterproof
Memory4 MB
Storage temperature-25 °C to +75 °C
Relative humidity10% to 90%, not condensing

1.4.6.3 – Theory of Operation

The Raisin™ Personal Monitor acquires, time-stamps and logs digital data corresponding to physiologic signals and patient-marked events. Heart rate, quantified using R-wave frequency, is sensed via three adhesive skin electrodes on the base of the data recorder. Activity data are provided by a 3-axis accelerometer integrated into the RPM. Subjects can mark subjectively defined, personally relevant events by depressing a button on the data recorder. Raisin™ Personal Monitor data are periodically uploaded to a general computing device via Bluetooth telemetry for display and export.

1.4.7 -- Intended Use

The Raisin" Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relatively to gravity, and time-stamped, patient-logged events. The Raisin™ Personal Monitor enables unattended data collection for clinical and research

Proteus Biomedical, Inc.

{3}------------------------------------------------

applications. The Raisin™ Personal Monitor may be used in any instance where quantifiable analysis of event-associated heart rate, activity, and body position is desirable.

1.4.8 -- Summary of Non-Clinical Performance Data

The three-axis accelerometer provides motion and position data and is validated against a known acceleration applied against each of its three axes.

The figure below shows bench validation of the accelerometer in all three of its axes.

Image /page/3/Figure/5 description: The image contains three scatter plots, each titled "Measured Acceleration vs. Applied Acceleration." The plots show the relationship between measured acceleration and applied acceleration for the X, Y, and Z axes. In each plot, the data points form a linear pattern, indicating a strong correlation between measured and applied acceleration. The axes are labeled with values ranging from -1 to 1.

Proteus Biomedical, Inc.

Raisin System 510k

1-21

{4}------------------------------------------------

The biopotential low-frequency amplifier is used to quantify heart rate by measuring R-wave frequency. The table below shows R-wave detection validation results, based upon a modified Hamilton-Tompkins algorithm, tested using guidelines set forth in the ANSV/AAMI EC 13 standard.

Test DescriptionExpected Results (bpm)Algorithm Results (bpm)
Default ECG waveform8080.0
T-wave rejectionR-wave amplitude of 1 mVT-wave amplitude of 0.4 mV8080.0
Ventricular bigeminy8079.9
Slow alternating ventricularbigeminy6060.5
Rapid alternating ventricularbigeminy120119.8
Bidirectional systoles9090.1
Default ECG waveformPacing pulse with 2 mVamplitude, 2 ms width8080.0

The table below shows validation testing results of R-wave detection during arrhythmia. Raisin™ Personal Monitor-reported R-wave locations were compared with annotated R-wave locations in all 48 test files from the MIT-BIH arrhythmia database.

MetricMedianStandard Deviation
Positive Detection Accuracy99.7%5.9%
False Positive Rate0%1.7%

Proteus Biomedical, Inc.

{5}------------------------------------------------

K093976
P6/7

1.4.9 -- Summary of Clinical Performance Data

The three-axis accelerometer was also validated clinically by assessing subject movement, in this case walking, to assess capture of expected features. The figure below demonstrates data from a representative walking test.

Image /page/5/Figure/3 description: The image shows three line graphs representing a walking test across the X, Y, and Z axes. Each graph plots 'time(sec)' on the x-axis, ranging from 0 to 3.5, against 'Acceleration (G)' on the y-axis. The X-axis graph shows acceleration fluctuating between approximately 0.5 and 1.5 G, the Y-axis graph shows acceleration fluctuating between approximately -1 and 0 G, and the Z-axis graph shows acceleration fluctuating between approximately -0.6 and 0.2 G.

The figure below shows a representative subject ECG captured by the RPM, with the automatically identified R-waves highlighted.

Image /page/5/Figure/5 description: The image shows an example captured and processed ECG waveform for subject 1, chest location, sitting. The x-axis represents time in seconds, ranging from 40 to 55. The y-axis represents ECG raw amplitude, ranging from -100 to 100. The plot shows the ECG waveform and R-waves, with the R-waves marked by diamond symbols.

Proteus Biomedical, Inc.

{6}------------------------------------------------

The table below shows robust R-wave detection accuracy when heart rate data were collected from different body locations.

AnteriorChestXyphoidStomachLateralChest
Subject 110099.72--
Subject 299.3099.0099.2499.61
Subject 499.1498.5899.3198.05
Subject 599.1499.3798.6698.81
Average R-wave detectionaccuracy99.4099.1799.0798.82

1.4.10 -- Conclusions

The Raisin" Personal Monitor (RPM) is a small, ambulatory, battery-operated data-logging device that is worn on the chest surface to record heart rate, activity, body angle relative to gravity, and patient-logged events. Patient-logged events are used to contextualize the physiologic measures. The RPM's functionality has been validated in non-clinical and clinical testing as summarized above.

Proteus Biomedical, Inc.

.

1-24

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo of the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure with outstretched arms, suggesting care and protection. The overall design is simple and conveys a sense of governmental authority and public service.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

MAR 2 5 2010

Proteus Biomedical, Inc. c/o Gregory Moon, M.D. Director of Clinical Affairs 2600 Bridge Parkway, Suite 101 Redwood City, CA 94065

Re: K093976

Trade/Device Name: Raisin™ Personal Monitor Regulatory Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitters and Receivers Regulatory Class: II (two) Product Code: 74 DXH Dated: February 26, 2010 Received: March 2, 2010

Dear Dr. Moon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{8}------------------------------------------------

Page 2 - Gregory Moon, M.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K093976

Device Name: Raisin™ Personal Monitor

Indications for use:

The Raisin™ Personal Monitor is a miniaturized, wearable data-logger for ambulatory recording of heart rate, activity, body angle relatively to gravity, and time-stamped, patient-logged events. The Raisin™ Personal Monitor enables unattended data collection for clinical and research applications. The Raisin™ Personal Monitor may be used in any instance where quantifiable analysis of eventassociated heart rate, activity, and body position is desirable.

Prescription Use _ J (21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)Division of Cardiovascular Devices
510(k) NumberK093976

§ 870.2920 Telephone electrocardiograph transmitter and receiver.

(a)
Identification. A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.(b)
Classification. Class II (performance standards).