K Number
K093923

Validate with FDA (Live)

Device Name
CEROX MODEL 3210
Date Cleared
2010-01-19

(28 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The noninvasive Or-Nim CerOx Model 3210 monitor is intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain of an adult. It is also intended for use as an adjunct monitor of hemoglobin oxygen saturation of blood in a region of skeletal muscle tissue beneath the sensor in adults. The prospective clinical value of data from the CerOx monitor has not been demonstrated in disease states. The CerOx monitor should not be used as the sole basis for diagnosis or therapy.

Device Description

The CerOx Model 3210 uses the well-established principles of near infrared spectroscopy (NIRS) to monitor the concentration of oxygenated hemoglobin relative to the total concentration of hemoglobin in the blood. CerOx Model 3210 comprises a display and processing unit and probes that are coupled to the patient using a single-use biocompatible adhesive. When the probes are attached to the patient, the system is operated to monitor the tissue blood oxygen saturation level. The CerOx Model 3210 can be operated in two modes: Cerebral Mode and Muscle Mode.

AI/ML Overview

The provided document, K093923, is a Special 510(k) Premarket Notification for the CerOx 3210 device. This type of submission is for modifications to a previously cleared device. Therefore, the focus of the submission is primarily on demonstrating that the modified device remains as safe and effective as, and substantially equivalent to, the predicate device, rather than proving the original device's performance from scratch or conducting extensive new clinical studies.

Based on the provided text, a detailed study proving the device meets specific acceptance criteria, with the requested information (sample sizes, expert qualifications, etc.), is not explicitly described in the provided sections. The summary focuses on comparing the modified device to its predicate and the design control processes.

Here's a breakdown of the available information and what is missing based on your questions:

1. Table of Acceptance Criteria and Reported Device Performance

This information is not provided in the document. The submission states: "The results of the design control process and the tests performed on the modified device (CerOx Model 3210) support the conclusion that it remains as safe and effective as, and remains substantially equivalent to, the cleared predicate device (Pacifica 01)." This indicates that substantial equivalence to the predicate, which presumably met its own acceptance criteria, is the primary "acceptance criterion" for this special 510(k). No specific quantitative performance metrics or acceptance criteria for the device itself are listed.

2. Sample Size Used for the Test Set and Data Provenance

This information is not provided. The document refers to "Hardware and software tests" but does not detail the sample sizes for these tests, nor the nature of the data (e.g., patient data, simulated data).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided. Given that the submission focuses on modifications and substantial equivalence to a predicate, it's unlikely a new, extensive clinical validation with expert-established ground truth was conducted for this specific 510(k) modification.

4. Adjudication Method for the Test Set

This information is not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

This information is not provided. An MRMC study is typically for evaluating the impact of AI assistance on human reader performance, which doesn't seem relevant for this device (a cerebral oximeter) as described. The device acts as an "adjunct monitor" for regional hemoglobin oxygen saturation and is not described as an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device is an oximeter, which inherently operates as a "standalone" sensor, providing measurements directly. The document states it "uses the well-established principles of near infrared spectroscopy (NIRS) to monitor the concentration of oxygenated hemoglobin relative to the total concentration of hemoglobin in the blood." Therefore, its performance is inherently standalone in terms of data acquisition and processing to produce a saturation value. However, the exact performance metrics and how they were tested beyond basic hardware/software validation are not detailed.

7. The Type of Ground Truth Used

This information is not explicitly stated for any performance evaluation. For oximeters, ground truth is typically established through co-oximetry measurements from arterial blood samples. However, this submission primarily relies on substantial equivalence to the predicate and compliance with voluntary consensus standards rather than a new clinical performance study.

8. The Sample Size for the Training Set

This information is not applicable/provided. The CerOx 3210 is a NIRS-based medical device, not an AI/ML algorithm that requires a training set in the conventional sense. It uses established physical principles, not machine learning, to derive its measurements.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided for the same reasons as #8.


Summary of what can be inferred:

  • Acceptance Criteria for THIS submission: The primary acceptance criterion for this Special 510(k) is demonstrating substantial equivalence to the predicate device (Pacifica 01, K073407) and adherence to recognized standards.
  • Study: The "study" mentioned consists of design control processes, risk analysis (FMECA), and hardware/software tests to validate mitigations identified during the FMECA. This is not a clinical performance study with patient data and expert ground truth as typically understood for new device submissions.
  • Ground Truth: The device relies on "well-established principles of near infrared spectroscopy (NIRS)." The performance is validated through testing against these principles and ensuring the modifications did not negatively impact the device's ability to operate safely and effectively, as compared to the predicate.
  • AI/ML: The device description does not indicate the use of AI or machine learning; therefore, concepts like training sets, MRMC studies, or "algorithm only" performance (beyond its core function as a sensor) are not relevant to this submission.

In conclusion, the document focuses on demonstrating that modifications to an existing device (CerOx 3210 as a modification of Pacifica 01) did not alter its safety or effectiveness in a way that would preclude substantial equivalence to the original. It does not provide the detailed clinical study data often found in original 510(k) submissions, particularly concerning specific acceptance criteria with quantitative performance metrics, sample sizes, or expert adjudication.

{0}------------------------------------------------

K093923
Or-Nim Medical Ltd.

ತಿ

Special 510(k) Premarket Notification CerOx 3210 (Pacifica 01 Modification)

9. 510(K) SUMMARY

Submission Date:

ﺗﺮ

December 14, 2009

Submitter Information:

Company Name: Or-Nim Medical Ltd.

Company Address:

1 Yodfat Street Lod, 71291 Israel

Contact Person:

Michal Balberg, PhD Chief Technical Officer Or-Nim Medical Ltd. Tel: +972-8-9282801 Fax: +972-8-9282805 michalb@ornim.com

Device Information:

Trade Name:CerOx Model 3210
CommonName:Oximeter/Cerebral Oximeter/Tissue Oximeter
ClassificationName:Oximeter, Tissue Saturation (21 CFR 870.2700)
ProductCode:MUD
RegulatoryClass:II

Predicate Device: The Or-Nim CerOx Model 3210 is a modification of the cleared Or-Nim Pacifica 01 (K073407), which serves as the predicate device.

Device Description: The CerOx Model 3210 uses the well-established principles of near infrared spectroscopy (NIRS) to monitor the concentration of oxygenated hemoglobin relative to the total concentration of hemoglobin in the blood.

CerOx Model 3210 comprises a display and processing unit and probes that are coupled to the patient using a single-use biocompatible adhesive.

CONFIDENTIAL

'JAN 1 9 2010

21

{1}------------------------------------------------

093923

Special 510(k) Premarket Notification CerOx 3210 (Pacifica 01 Modification)

・・・・

Page 2 of (2)
Or-Nine Medical Ltd.

When the probes are attached to the patient, the system is operated to monitor the tissue blood oxygen saturation level.

The CerOx Model 3210 can be operated in two modes: Cerebral Mode and Muscle Mode.

Intended Use: The CerOx Model 3210 is intended to monitor oxygen saturation of blood in the body.

Note: the Intended Use of the modified device has not changed as a result of the modifications outlined in this submission.

Indications for Use: The noninvasive Or-Nim CerOx Model 3210 monitor is intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain of an adult. It is also intended for use as an adjunct monitor of hemoglobin oxygen saturation of blood in a region of skeletal muscle tissue beneath the sensor in adults. The prospective clinical value of data from the CerOx monitor has not been demonstrated in disease states. The CerOx monitor should not be used as the sole basis for diagnosis or therapy.

Note: the Indications for Use statement for the modified device has not changed as a result of the modifications outlined in this submission.

Comparison to Predicate Device: ﻟﺴﻌﺔ

The Or-Nim CerOx Model 3210 represents a modification to the cleared Or-Nim Pacifica 01 (K073407). The original and modified devices have the same intended use, the same indications for use, the same technological characteristics, and are used on the same sites of the body.

Summary of Design Control and Test Activities:

The risk analysis method used to assess the impact of the modifications was Failure Modes, Effects and Criticality Analysis (FMECA). For risks which were identified as a result of the modifications, mitigations were designed and implemented in accordance with the Quality System Regulations. Hardware and software tests were then performed to validate the mitigations.

Conclusions: The results of the design control process and the tests performed on the modified device (CerOx Model 3210) support the conclusion that it remains as safe and effective as, and remains substantially equivalent to, the cleared predicate device (Pacifica 01).

CONFIDENTIAL

{2}------------------------------------------------

3 3 edical Ltd.

Special 510(k) Premarket Notification CerOx 3210 (Pacifica 01 Modification)

10. STANDARDS DATA REPORT FOR 510(K)S

FDA Recognized Voluntary Consensus Standards 10.1.

Both the original Pacifica and the modified CerOx are in compliance with FDA-recognized performance and safety standards. Form 3654 (Appendix H) lists the standards cited and the table attached to it specifies sections to which the CerOx conforms. For each standard, sections other than those listed in the table attached to Form 3654 were deemed Not Applicable.

Testing of acoustic parameters and display was performed in accordance with standards from the American Institute of Ultrasound in Medicine (AIUM) (Table 11.2). Design and testing of the Cer Ox was performed in accordance with 21 CFR§1040.10 and 21 CFR§1040.11.

CONFIDENTIAL

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the HHS logo, which is a stylized human figure with outstretched arms, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the logo. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Or-Nim Medical, Ltd. % Michal Balberg, Ph.D. Chief Technical Officer 1 Yodfat Street Lod, 71291 Israel

JAN 1 9 2010

Re: K093923

Trade/Device Name: CerOx Model 3210 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: December 17, 2009 Received: December 22, 2009

Dear Dr. Balberg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Mr. Michal Balberg, Ph.D.

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

8. STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known): K093923

CerOx Model 3210 Device Name:

Indications for Use:

The noninvasive Or-Nim CerOx Model 3210 monitor is intended for use as an adjunct monitor of regional hemoglobin oxygen saturation of blood in the brain of an adult. It is also intended for use as an adjunct monitor of hemoglobin oxygen saturation of blood in a region of skeletal muscle tissue beneath the sensor in adults. The prospective clinical value of data from the CerOx monitor has not been demonstrated in disease states. The CerOx monitor should not be used as the sole basis for diagnosis or therapy.

ﺗﺮ ﺇﻟﻰ ﺍﻟﻤﺆﻟﻔﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﺍﺻﻞ ﺍﻟﻤ

A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ro. Rr. mo Des. R.R.

(Division Sign-Off) (Division of Surgical, Orthopedic, Division of Surgious.
and Restorative Devices

098923

510(k) Number

CONFIDENTIAL

Page 1 of 1

20

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).