K Number
K093836
Manufacturer
Date Cleared
2010-04-15

(121 days)

Product Code
Regulation Number
862.1660
Panel
CH
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Elecsys TSH CalCheck 5 is an assayed control for use in calibration verification and for use in the verification of the assay range established by the Elecsys TSH reagent on the indicated Elecsys cobas e immunoassay analyzers.

Device Description

The Elecsys TSH CalCheck 5 is a lyophilized product consisting of TSH in equine serum matrix. During manufacture, the analyte is spiked into the matrix at the desired concentration levels.

AI/ML Overview

The provided 510(k) summary for the device, "Elecsys TSH CalCheck 5," is for an in-vitro diagnostic control material. As such, it does not involve clinical studies with human patients, image analysis, or expert-based ground truth establishment in the same way a diagnostic imaging device would. Instead, its "performance" is assessed based on its chemical and physical characteristics and how accurately it can verify instrument calibration and assay ranges.

Therefore, many of the requested categories in the prompt are not applicable to this type of device submission.

Here is an analysis based on the information provided in the K093836 document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state formal "acceptance criteria" in a quantitative, pass/fail table as would be expected for a diagnostic device. Instead, it states that the device was evaluated for "value assignment and stability." The implication is that the performance characteristics (e.g., how well the assigned values match expectations, and the stability over time) were deemed acceptable for substantial equivalence to the predicate device.

CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Value AssignmentAssumed to be accurate and consistent with intended use as an assayed control for calibration verification.Evaluated for value assignment. (Specific data not provided in summary).
Stability (Unopened)Stable at 2-8°C until expiration date.Same as predicate.
Stability (Reconstituted)Stable at 20-25°C for 4 hours.Same as predicate.
FormatLyophilizedSame as predicate.
MatrixEquine serum matrixSame as predicate.
Levels of ControlFive distinct levels for calibration verification.Five levels, an increase from the predicate's three.

2. Sample Size Used for the Test Set and Data Provenance

  • Test set: Not applicable in the traditional sense of patient data. The "test set" would be samples of the manufactured control material itself, used for evaluation. The document does not specify the number of batches or individual vials tested.
  • Data Provenance: The exact location or nature of the testing is not specified, but it would have been conducted by Roche Diagnostics as part of their product development and validation process. This would be considered prospective manufacturing and testing data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not Applicable. For an in-vitro diagnostic control, the "ground truth" is established through highly controlled analytical methods and established reference materials/methods, not through expert human interpretation. The concentrations of TSH in the control material are assigned through a rigorous internal process, not "established by experts."

4. Adjudication Method for the Test Set

  • Not Applicable. No human adjudication is involved for this type of chemical control material. The "adjudication" is metrological, relating to the accuracy and precision of the analytical measurements used to assign values and assess stability.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

  • No. An MRMC study is relevant for diagnostic devices that involve human interpretation of medical images or data. This device is a quality control material for an automated immunoassay analyzer.

6. If a Standalone (Algorithm only without human-in-the-loop performance) was done

  • Not Applicable / Inherent. This device itself is not an algorithm or a diagnostic tool; it's a calibrator/control. Its performance is entirely "standalone" in the sense that it is a physical product with defined chemical characteristics. Its interaction is with the immunoassay analyzer, not a human interpreter directly.

7. The Type of Ground Truth Used

  • Assigned Values based on Analytical Methods and Reference Materials: The "ground truth" for the Elecsys TSH CalCheck 5 is the assigned TSH concentration values for each of its five levels. These values are determined by the manufacturer (Roche Diagnostics) using highly precise and accurate analytical methods, typically traceable to international reference standards and/or internal established reference methods. This ensures that the control material has known concentrations to verify the performance of the Elecsys TSH reagent and analyzer.

8. The Sample Size for the Training Set

  • Not Applicable. This device is a chemical control product, not a machine learning algorithm that requires a "training set" of data.

9. How the Ground Truth for the Training Set was Established

  • Not Applicable. See point 8.

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.