K Number
K093570

Validate with FDA (Live)

Date Cleared
2009-12-04

(16 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Vinyl Patient Examination Gloves, Light Yellow Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves, Light Yellow Color that meets all of the requirements of ASTM standard D 5250-06e1.

AI/ML Overview

The provided text describes the acceptance criteria and performance of Powder Free Vinyl Patient Examination Gloves, Light Yellow Color, as submitted in a 510(k) summary (K093570) to the FDA. The study presented a non-clinical evaluation to demonstrate substantial equivalence to a predicate device.

Here's the breakdown of the acceptance criteria and study information:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-06 el.Meets
Physical PropertiesASTM standard D 5250-06 el.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 el and D6124-06Meets (< 2mg/glove)
BiocompatibilityPrimary Skin Irritation in rabbits (AAMI / ANSI / ISO ISO 10993-10)Passes (Not a Primary Skin Irritation)
Dermal sensitization in the guinea pig (AAMI / ANSI / ISO ISO 10993-10)Passes (Not a Dermal sensitization)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set for each characteristic. However, it indicates that the tests were performed according to specified ASTM and CFR standards, which would inherently include appropriate sample size requirements for those tests. The data provenance is not specified beyond the tests being conducted for the Jiangsu Jaysun Glove Co.,Ltd., located in China. The study is a non-clinical evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This section is not applicable as the device is a medical glove, and the evaluation involves objective performance testing against established physical and chemical standards, not expert interpretation of outputs like in imaging or diagnostic AI.

4. Adjudication Method for the Test Set

This section is not applicable for the same reasons as point 3. The evaluation relies on objective measurements against predefined standards, not subjective assessment requiring adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

This section is not applicable. The device is a medical glove, not an AI system. Therefore, an MRMC study and the concept of human readers improving with or without AI assistance are irrelevant.

6. Standalone (Algorithm Only) Performance Study

This section is not applicable. The device is a medical glove. The evaluation focuses on the physical and biological properties of the glove itself, not on an algorithm's performance.

7. Type of Ground Truth Used

The ground truth used for the device's performance evaluation is based on established industry and regulatory standards. These include:

  • ASTM standard D 5250-06e1 (for dimensions and physical properties)
  • 21 CFR 800.20 (for freedom from pinholes, commonly referred to as the "waterleak test" for AQL)
  • ASTM standard D 6124-06 (for powder residual)
  • AAMI / ANSI / ISO 10993-10 (for biocompatibility, specifically primary skin irritation and dermal sensitization)

These standards define the acceptable range or threshold for each characteristic.

8. Sample Size for the Training Set

This information is not applicable. The device is a medical glove, and its evaluation does not involve a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Section C 510(k) Summary (21 CFR 807.92)

K093570

DEC - 4 2009

510(k) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: " (applicant leave blank)

Premarket Notification [510(k)] Summary

[(a){1)]. The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :Jiangsu Jaysun Glove Co.,Ltd.
Submitter's address :No.199 Jiangling Avenue, Suqian EconomicDevelopment Zone,Suqian 223800, Jiangsu,China
Phone number :86-527-84568809
Fax number :86-527-84568806
Name of contact person:Ms.Sea Xu
Date the summary was prepared:Oct.30, 2009

[(a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves,Light Yellow Color
Proprietary/Trade name:Powder Free Vinyl Patient Examination Gloves,Light Yellow ColorOther clients private labeling
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital(80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

{1}------------------------------------------------

K093570

Class I* Powder Free Vinyl Patient Examination Gloves, Light Yellow Color that mee the requirements of ASTM standard D 5250-06e1.

Predicate device: Powder-Free Vinyl Patient Examination Gloves, JIANGSU JAYSUN GLOVE CO., LTD., K062522 .

[(a)(4)] A description of the device

Device Description : Powder Free Vinyl Patient Examination Gloves, Light Yellow Color that meets all of the requirements of ASTM standard D 5250-06e1.

[(a)(5)] The summary describes the intended use of the device

Device Intended Use: Powder Free Vinyl Patient Examination Gloves, Light Yellow Color is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The Powder Free Vinyl Patient Examination Gloves, Light Yellow Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06 el.Meets
Physical PropertiesASTM standard D 5250-06 el.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 eland D6124-06Meets<2mg/glove
BiocompatabilityPrimary Skin Irritation in rabbitsAAMI / ANSI / ISO ISO 10993-10Dermal sensitization in the guinea pigAAMI / ANSI / ISO ISO 10993-10PassesNot a Primary Skin IrritationPassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder Free Vinyl Patient Examination Gloves, Light Yellow Color, meet requirements per ASTM D5250-06 el, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-87: AAMI / ANSI / ISO 10993-10:2002.

[(b)(2)] A brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical tests that demonstrate that the device is as safe, as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Light Yellow Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/11 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing the department's mission to protect the health of all Americans. The department's name is inscribed around the seal's perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

DEC - 4 2009 Jiangsu Jaysun Glove Company, Limited C/O Mr. Chu Xiaoan Official Correspondent Beijing Easy-Link Company Room 1606 Building, 1 Jianxiang Yuan, No. 209 Bei Si Huan Zhong Road Hai Di, Beijing China 100083

Re: K093570

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Light Yellow Color Regulation Number: 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: November 8, 2009 Received: November 18, 2009

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Antan O. ant h
Susan Dunner, D.D.S., M.S.

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K093576

INDICATIONS FOR USE

Jiangsu Jaysun Glove Co.,Ltd. Applicant:

510(k) Number (if known): * Kの9357 O

Powder Free Vinyl Patient Examination Gloves, Light Yellow Color Device Name:

Indications For Use:

Powder Free Vinyl Patient Examination Gloves, Light Yellow Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clannie-Williams

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K093570

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.