K Number
K092956
Date Cleared
2009-12-04

(70 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder free vinyl patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

AI/ML Overview

The provided text describes the acceptance criteria and performance of the Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored). However, it does not include information about a "study" in the traditional sense of a clinical trial or a machine learning model's performance study with separate training and test sets, and expert ground truth establishment. This is a 510(k) premarket notification for a medical device (examination gloves), which primarily relies on conformance to established standards and comparison to a predicate device, rather than a novel algorithm's performance.

Here's a breakdown of the requested information based on the provided text, with explanations for what is not applicable or not provided:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-06e1Meets
Physical PropertiesASTM standard D 5250-06e1Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06e1 and D6124-06Meets (<2mg/glove)
Biocompatibility - Primary Skin IrritationAAMI / ANSI / ISO 10993-10 (in rabbits)Passes (Not a Primary Skin Irritation)
Biocompatibility - Dermal SensitizationAAMI / ANSI / ISO 10993-10 (in guinea pig)Passes (Not a Dermal sensitization)

2. Sample Size for the Test Set and Data Provenance

The document does not specify a "test set" in the context of a dataset for evaluating an algorithm. For physical device testing (dimension, physical properties, freedom from pinholes, powder residual), sample sizes would be defined by the respective ASTM standards and 21 CFR regulations, but these specific numbers are not provided in the summary. For biocompatibility, the tests were conducted in animals (rabbits for irritation, guinea pigs for sensitization), but the number of animals is not specified.

Data Provenance: The tests were conducted according to international standards (ASTM, ISO) and US regulations (21 CFR). The location of the testing laboratories is not specified, but the manufacturer is based in China. The "retrospective or prospective" classification is not applicable here as these are laboratory and animal tests, not human studies.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This concept is not applicable to the evaluation of examination gloves. Ground truth for these tests is established by adhering to standardized testing protocols and objective measurements (e.g., measuring dimensions, weight of powder, visual inspection for pinholes, observing animal reactions). There are no "experts" establishing a "ground truth" through interpretation similar to medical imaging.

4. Adjudication Method for the Test Set

Not applicable. The tests involve objective measurements against predefined criteria/standards rather than subjective interpretations requiring adjudication.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is typically relevant for interpretative devices or technologies (e.g., AI in radiology) where human-in-the-loop performance is being assessed. This document concerns a physical device (examination gloves).

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, this is not applicable. The device is a physical examination glove, not an algorithm.

7. The Type of Ground Truth Used

The ground truth used for the device's conformance is defined by:

  • Standard Specifications: Requirements outlined in ASTM standard D 5250-06e1 and ASTM D6124-06.
  • Regulatory Requirements: Specifications from 21 CFR 800.20.
  • Biocompatibility Standards: Criteria from AAMI / ANSI / ISO 10993-10 for irritation and sensitization.
    These are objective, measurable standards, not expert consensus or pathology in the typical sense for diagnostic devices.

8. The Sample Size for the Training Set

Not applicable. There is no "training set" as this is not an AI/machine learning device. The product development and manufacturing processes are controlled to meet the specified standards.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Section C 510(k) Summary (21 CFR 807.92)

510(k) Summary

DEC - 4 2009

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K092956 " (applicant leave blank)

Premarket Notification [510(k)] Summary

((a)(1)). The summary contains on the first page, preferably on your letterhead paper, the submitter's name, address, phone and fax numbers, name of contact person, and date the summary was prepared :

Submitter's name :Shandong Aobo Medical Co.,Ltd.
Submitter's address :Xingshan Laboder Road,Economic DevelopmentZone of Zichuan,Zibo City, ShandongProvince, 255100, China.
Phone number :86-533-2557777
Fax number :86-533-6023666
Name of contact person:Mr.He lin
Date the summary was prepared:Sep 10.2009

((a)(2)]. The name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known

Device Name:Powder Free Vinyl Patient Examination Gloves,Clear(non-colored)
Proprietary/Trade name:Powder Free Vinyl Patient Examination GlovesOther clients private labeling
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital(80)
Product Code:LYZ

[(a)(3)]. An identification of the legally marketed device to which your firm is claiming substantial equivalence .

Class I* powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

Section C

{1}------------------------------------------------

5/0(k): K092956

the requirements of ASTM standard D 5250-06e1.

Predicate device: FUGUAN (Brand) Powder-Free Vinyl Patient Examination Gloves, Shijiazhuang Fuguan Plastic Products Co., Ltd.. K032908 .

[(a)(4)] A description of the device

Device Description : powder free vinyl patient examination gloves, Clear(non-colored) that meets all of the requirements of ASTM standard D 5250-06e1.

[{a)(5)] The summary describes the intended use of the device

Device Intended Use: powder free vinyl patient examination glove, Clear(non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

[(a)(6)] A summary of the technological characteristics of new device compared to the predicate device.

The powder free vinyl patient examination gloves, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D 5250-06 el.Meets
Physical PropertiesASTM standard D 5250-06 el.Meets
Freedom from pinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06 eland D6124-06Meets<2mg/glove
BiocompatabilityPrimary Skin Irritation in rabbitsAAMI / ANSI / ISO ISO 10993-10PassesNot a Primary Skin Irritation
Dermal sensitization in the guinea pigAAMI / ANSI / ISO ISO 10993-10PassesNot a Dermal sensitization

[(b)(1)] A brief discussion of the nonclinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Powder free vinyl patient examination gloves , Clear(non-colored) meet requirements per ASTM D5250-06 e1, per ASTM D6124-06, per 21 CFR 800.20 and FDA recognition number 2-87: AAMI / ANSI / ISO 10993-10:2002.

[(b)(2)] A_brief discussion of the clinical submitted, reference, or relied on in the premarket notification submission for a determination of substantial equivalence .

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

[(b)(3)] The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe,as effective, and performed as well or better than the legally marketed device identified in (a)(3).

It can be concluded that the Powder Free Vinyl Patient Examination Gloves meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of a human figure embracing a bird, which is meant to symbolize the department's mission of protecting and promoting the health and well-being of Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Shandong Aobo Medical Company, Limited C/O Mr. Chu Xiaoan Official Correspondent Beijing Easy-Link Company, Limited Room 1606 Building, 1 Jian Xiang, Yuan No. 209 Bei Si Huan Zong Road Hai Di, Beijing China 100083 DEL - 4 cuu9

· Re: K092956

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (Non-colored) Regulation Number: 21CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: 1 Product Code: LYZ Dated: November 2, 2009 Received: November 12, 2009

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

{3}------------------------------------------------

Page 2- Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Noss Jr.
S. D. D. S., M. A.

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Section B Indications for Use

INDICATIONS FOR USE

Applicant:Shandong Aobo Medical Co.,Ltd.
510(k) Number (if known): *K092956
Device Name:Powder Free Vinyl Patient Examination Gloves, Clear(Non-colored)

Indications For Use:

Powder free vinyl patient examination glove is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

currence of CDRH, Office of Device Evaluation (ODE)

510( ( 1 1 1 1 ( rev. 01 )

Page 1/1

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.