K Number
K092530
Manufacturer
Date Cleared
2010-04-02

(227 days)

Product Code
Regulation Number
872.4880
Panel
Dental
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sprytie™ wires are indicated for use in the jaws in adults and children to directly wire bony segments together, for the fixation of arch bars or splints to the teeth, for stabilization of bony segments, or for wiring the teeth together (maxillomandibular fixation).

Device Description

The Sprytie™ is a 25, 26, or 27 gauge annealed 316L stainless steel wire 7 inches in length, with a hollow 22TW, 23TW, or 24TW sieeve crimped over one end of the wire, respectively. The Sprytie™ is used in place of traditional light gauge stainless steel surgical wire during the placement of arch bars and splints, direct intraosseous wiring, bridal wire placement, and direct maxillomandibular fixation (ivy loops, risden wires, etc.). The sleeve end is inserted around the teeth first. The stiffness of the sleeve prevents the flexure of the wire.

AI/ML Overview

The provided document is a 510(k) summary for the Sprytie™ device (K092530), which describes a stainless steel wire with a crimped sleeve for maxillary and mandibular fixation. It states that the device is substantially equivalent to predicate devices based on intended use and technical specifications.

However, the document does not contain any information regarding acceptance criteria, device performance studies, test set sample sizes, data provenance, expert ground truth, adjudication methods, multi-reader multi-case studies, standalone performance, training set details, or how ground truth was established.

Based on the content of the provided text, I can only provide the following information from the request:

1. A table of acceptance criteria and the reported device performance

  • Acceptance Criteria: Not mentioned in the document.
  • Reported Device Performance: The document claims the Sprytie™ "performs as well as or better than predicate devices" and "improves the ease of use of the wire." However, no specific performance metrics or data from a study are provided to substantiate these claims.

Summary of what cannot be provided given the input:

  • Sample sized used for the test set and the data provenance.
  • Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
  • Adjudication method for the test set.
  • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or its effect size.
  • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done. (This is irrelevant for a physical wire device).
  • The type of ground truth used.
  • The sample size for the training set.
  • How the ground truth for the training set was established.

Reasoning:

The document focuses on demonstrating substantial equivalence to existing predicate devices (Aragon Wiring System, K022821; Erich Arch Bar, K061271) for regulatory purposes rather than presenting a performance study with detailed acceptance criteria and results.

The key statements are:

  • "The Sprytie™ has the same intended use as predicate devices, as well as substantially equivalent technical specifications."
  • "The modification to a traditional 316L stainless steel wire with the addition of a 304 stainless steel sleeve improves the ease of use of the wire, and does not change the intended use of the wire."
  • "The Sprytie™ is as safe, as effective, and performs as well as or better than predicate devices."

These statements are conclusions drawn by the submitter for the 510(k) process, not data points from a specific performance study against defined acceptance criteria. The FDA letter confirms the substantial equivalence determination but does not refer to a performance study with specific metrics.

{0}------------------------------------------------

K092530

5-1
1082

510(k) Summary

807.92(c)

Sponsor

APR - 2 2010

807.92(a)(1)

Submitter Russell Walther 6026 Martel Avenue Dallas, Texas 75206

Contact Person: Russell Walther Telephone: 817-992-0232 Fax: 817-249-0771

Date Prepared: August 10, 2009

Device Name

807.92(a){2)

Proprietary Name:Sprytie™ (models ST001, ST002, ST003)
Common/Usual Name:Stainless Steel Wire for Maxillary andMandibular Fixation
Classification:Class IIDental Classification PanelWire, Fixation, Intraosseous
Regulation Number:872.4880
Product Code:DZK

Predicate Devices

807.92(a)(3)

Predicate Device #1:Aragon Wiring System, K022821
Predicate Device #2:Erich Arch Bar, K061271

Device Description

807.92(a){4)

The Sprytie™ is a 25, 26, or 27 gauge annealed 316L stainless steel wire 7 inches in length, with a hollow 22TW, 23TW, or 24TW sieeve crimped over one end of the wire, respectively.

{1}------------------------------------------------

A.K092530

The Sprytie™ is used in place of traditional light gauge stainless steel surgical wire during the placement of arch bars and splints, direct intraosseous wiring, bridal wire placement, and direct maxillomandibular fixation (ivy loops, risden wires, etc.). The sleeve end is inserted around the teeth first. The stiffness of the sleeve prevents the flexure of the wire.

Device Intended Use

Sprytie™ wires are indicated for use in the jaws in adults and children to directly wire bony segments together, for the fixation of arch bars or splints to the teeth, for stabilization of bony segments, or for wiring the teeth together (maxillomandibular fixation).

Device Technological Characteristics

The wire component of the Sprytie™ is identical to predicate devices. The Sprytie™ includes a ¾ inch 304 stainless steel sleeve crimped over one end. In predicate devices the wire is packaged in 75-675 ft length reusable spools. The predicate devices do not contain the 304 stainless steel portion.

Conclusion

The Sprytie™ has the same intended use as predicate devices, as well as substantially equivalent technical specifications. The modification to a traditional 316L stainless steel wire with the addition of a 304 stainless steel sleeve improves the ease of use of the wire, and does not change the intended use of the wire. The Sprytie™ is as safe, as effective, and performs as well as or better than predicate devices.

807.92(b)(3)

807.92(a)(6)

807.92(a)(5)

5-2

2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR - 2 2010

Mr. Russell B. Walther Russell B. Walther 6026 Martel Avenue Dallas, Texas 75206

Re: K092530

Trade/Device Name: Sprytie™M Regulation Number: 21CFR 872.4880 Regulation Name: Intraosseous Fixation Screw or Wire Regulatory Class: II Product Code: DZK Dated: March 18, 2010 Received: March 24, 2010

Dear Mr. Walther:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Walther

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

hn for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K092530

4-1

Indications for Use

510(k) Number (if known):

Device Name: Sprytie™

Indications for Use:

Sprytie™ wires are indicated for use in the jaws in adults and children to directly wire bony segments together, for the fixation of arch bars or splints to the teeth, for stabilization of bony segments, or for wiring the teeth together (maxillomandibular fixation).

Prescription Use _X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

rrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) (Orvision of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

§ 872.4880 Intraosseous fixation screw or wire.

(a)
Identification. An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.(b)
Classification. Class II.