K Number
K092454
Manufacturer
Date Cleared
2010-05-21

(284 days)

Product Code
Regulation Number
866.1640
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

VITEK® 2 Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species. VITEK 2 Yeast Voriconazole is a quantitative test intended for use with the VITEK 2 and VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Voriconazole has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.

Active in vitro and in clinical infections Candida albicans Candida krusei Candida parapsilosis Candida tropicalis

Active in vitro but their clinical significance is uknown: Candida lusitaniae Candida guilliermondii

The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.

Device Description

The antifungal presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antifungal medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours for yeast). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antifungal contained on the card.

AI/ML Overview

This report describes the acceptance criteria and study proving the VITEK® 2 Yeast Voriconazole device meets these criteria.

1. Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Target)Reported Device Performance
Essential Agreement (EA) with CLSI reference method incubated for 24 hrs. (Not explicitly stated as a numerical target, but implied to be high for "acceptable performance")99.2% Overall Essential Agreement
Category Agreement (CA) with CLSI reference method incubated for 24 hrs. (Not explicitly stated as a numerical target, but implied to be high for "acceptable performance")99.2% Overall Category Agreement
Essential Agreement (EA) with CLSI reference method incubated for 48 hrs. (Not explicitly stated as a numerical target, but implied to be high for "acceptable performance")96.9% Overall Essential Agreement
Category Agreement (CA) with CLSI reference method incubated for 48 hrs. (Not explicitly stated as a numerical target, but implied to be high for "acceptable performance")98.7% Overall Category Agreement
ReproducibilityAcceptable
Quality ControlAcceptable

2. Sample Size Used for the Test Set and Data Provenance

The document states "An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains." However, the exact sample size (number of isolates) for the test set is not explicitly provided in the furnished text.

The data provenance is not explicitly stated in terms of country of origin, but it is implied to be from an "external evaluation," generally suggesting data collected in a clinical or laboratory setting. The use of "fresh and stock clinical isolates" indicates a mix of prospective (fresh clinical isolates) and retrospective (stock clinical isolates and challenge strains) data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of experts to establish ground truth. Instead, the ground truth was established by comparing the device's performance against a "CLSI reference method." This method would involve standardized laboratory procedures rather than expert consensus on individual cases.

4. Adjudication Method for the Test Set

Since the ground truth was established by the CLSI reference method, there was no adjudication method involving multiple human readers as would be typical in medical imaging or interpretation studies. The comparison was directly against the results generated by the reference method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This type of study typically evaluates human reader performance with and without AI assistance, which is not applicable to an automated antimicrobial susceptibility test (AST) system.

6. Standalone Performance Study

Yes, a standalone performance study was done. The entire study described evaluates the VITEK® 2 Yeast Voriconazole device as a standalone algorithm (automated system) without human intervention in the interpretation process once the card is processed. The performance metrics (Essential Agreement and Category Agreement) directly reflect the algorithm's accuracy compared to the reference method.

7. Type of Ground Truth Used

The type of ground truth used was the CLSI reference method (Clinical and Laboratory Standards Institute). This is a standardized laboratory methodology for determining antimicrobial susceptibility, considered the gold standard for AST systems.

8. Sample Size for the Training Set

The document does not provide information regarding a specific training set or its sample size. This implies that the device's development might have used a different type of underlying data or that the training data details are proprietary and not included in this summary. For biological assays, often historical data and laboratory standards guide algorithm development.

9. How the Ground Truth for the Training Set Was Established

Since a specific training set is not mentioned, the method for establishing its ground truth is also not provided.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for bioMerieux. The logo consists of a stylized circle that is split into two halves. The left half of the circle has horizontal lines, while the right half is solid black. The text "BIOMERIEUX" is written in a stylized font below the circle.

X092454

510(k) SUMMARY

VITEK® 2 Yeast Voriconazole

510(k) Submission Information:

MAY 2 1 2010

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:Nancy WeaverAssociate Director, Regulatory Affairs
Phone Number:314 -731-8695
Fax Number:314-731-8689
Date of Preparation:August 06, 2009
B. Device Name:
Formal/Trade Name:VITEK® 2 Yeast Voriconazole
Classification Name:21 CFR 866.1640Antimicrobial Susceptibility TestProduct Code NGZ
Common Name:VITEK® 2 AST-YS Voriconazole
C. Predicate Device:VITEK® 2 Yeast Fluconazole (K061945)

D. 510(k) Summary:

VITEK® 2 Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species. VITEK® 2 Yeast Voriconazole is a quantitative test intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a Íaboratory aid in the determination of in vitro susceptibility to antifungal agents. Voriconazole has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.

Candida albicansCandida tropicalis
Candida kruseiCandida lusitaniae
Candida parapsilosisCandida guilliermondii

The antifungal presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mcg/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antifungal medium within the card. The VITEK® 2 automatically fills, seals and places the card into the incubator/reader. The VITEK® 2 Compact has a manual filling and sealing operation. The VITEK® 2 monitors the growth of each well in the card over a defined period of time (up to 36 hours for

bioMérieux, Inc. 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 800/638-4835 Fax: 314/731-8700 http://www.biomerieux-usa.com

{1}------------------------------------------------

2

yeast). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antifungal contained on the card.

VITEK 2 Yeast Voriconazole demonstrated substantially equivalent performance when compared with the CLSI reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA. Issued Feb. 5, 2003.

The Premarket Notification (510(k)) presents data in support of VITEK® 2 Yeast Voriconazole. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK® 2 Yeast Voriconazole by comparing its performance with the CLSI reference method incubated at 24 and at 48 hrs. The data is representative of performance on both the VITEK® 2 and VITEK® 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK® 2 Yeast Voriconazole demonstrated acceptable performance of 99.2% overall Essential Agreement and 99.2% overall category agreement with the reference method incubated for 24 hrs. With the reference method incubated 48 hrs the overall Essential Agreement was 96.9% and overall category agreement was 98.7%. Reproducibility and Quality Control demonstrated acceptable results.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

Biomerieux, Inc. c/o Nancy Weaver Associate Director, Regulatory Affairs 595 Anglum Rd. Hazelwood, Missouri 63042

MAY 2 1 2010

Re:K092454
Trade/Device Name:Vitek ®2 Yeast Voriconazole
Regulation Number:21CFR §866.1640
Regulation Name:Antifungal susceptibility of Candida species to Voriconazole
Regulatory Class:Class II
Product Code:NGZ
Dated:May 10, 2010
Received:May 11, 2010

Dear Ms. Weaver:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section

{3}------------------------------------------------

Page 2 - Nancy Weaver

510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Saly aath

Sally A. Hojvat, M.Sc., Ph.I Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):_ KO9 2 459

Device Name: VITEK 2 Yeast Voriconazole (0.12 - 8 µg/ml)

Indications For Use:

VITEK® 2 Yeast Voriconazole is designed for antifungal susceptibility testing of Candida species. VITEK 2 Yeast Voriconazole is a quantitative test intended for use with the VITEK 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antifungal agents. Voriconazole has been shown to be active against most strains of the following microorganisms listed below according to the FDA label for the antifungal.

Active in vitro and in clinical infections Candida albicans Candida krusei Candida parapsilosis Candida tropicalis

Active in vitro but their clinical significance is uknown: Candida lusitaniae Candida guilliermondii

The VITEK 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 System for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gram-negative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, S. pneumoniae and clinically significant yeast.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD)

Freddie M. Cooke

Luvision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) KD92454

Page 1 of 1

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).