(221 days)
Not Found
No
The summary describes standard PACS functionalities and image processing without mentioning AI, ML, or related concepts. The performance studies section only mentions software verification and validation, not AI/ML model training or testing.
No
This device is a software program that processes and displays spinal X-ray images and aids physicians in calculations and reporting, but it does not directly treat or alleviate a medical condition.
Yes
The device is described as an aid in the calculation of alteration of motion segment integrity (AOMSI) and generates reports on measurements, which are functions associated with aiding in the diagnosis of a medical condition.
Yes
The device description explicitly states it is a "software computer program installed on regular computers" and describes its functionality solely in terms of software operations (acceptance, display, processing, review, editing, measurements, report generation, storage, teleradiology exchange). There is no mention of any hardware components being part of the device itself.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The Spinal Guides PACS device processes and displays medical images (spinal X-rays) and patient information. It aids physicians in viewing and analyzing these images for diagnostic purposes related to the spine.
- Lack of Biological Sample Analysis: The device does not analyze any biological samples from the patient. Its input is image data, not biological material.
Therefore, the function and intended use of the Spinal Guides PACS device clearly fall outside the scope of In Vitro Diagnostics. It is a medical image processing and viewing system.
N/A
Intended Use / Indications for Use
Spinal Guides PACS is intended to perform: operations relating to the acceptance of spinal X-ray medical images and patient demographic information, their display, digital processing, review and editing, measurements report generation with respective storage and teleradiology exchange capabilities. They are intended to be used by a physician to view the images and use as an aid in calculation of alteration of motion segment integrity (AOMSI) of human spine. This device is not intended to be used for mammography images and does not require usage of 5 Mega pixel monitors.
Product codes
LLZ
Device Description
The device is represented by a software computer program installed on regular computers. The device has two types of users: a web and a local user. The web user submits an image set for processing to the local user via internet to the device's database. The local user is viewing the images of the image set, process the image set and sends a PDF report with calculations for the web user.
Workflow details: Working image is transferred to the device database by a web user. An operator (a local user) then uses a workstation to analyze AOMSI and generate a PDF, which is then available for viewing on the web for the web user. The local user/operator/server is installed at only one core location and X-ray images from different users are sent to the location for analyses. The device is primarily intended for viewing digital spinal X-ray images and to prepare analyses of alteration of motion segment integrity (AOMSI).
Automated part of the reporting process:
- Automatic limitation of points placed respective to the number of vertebras
- Automation of 4th point positioning.
- Automatic mathematical calculations
- Automatic notice to calibrate an image upon placement of first point on an image in image editor window
- Automation of notification of an improper order of points
- Automatic user password retrieval
- Automatic submission of PDF report via e-mail to the web user
What is needed to perform the analysis:
- The analyses are to be performed only by a licensed medical practitioner who is knowledgeable of specific spinal anatomy and specific protocols for image analysis
- Medical Device Spinal Guides PACS
- Spinal pure lateral X-ray views in flexion and extension taken with the calibrator in place
- Computer hardware
- Internet with enough available speed
Processing: For all square vertebras the first point (I) is placed on the superior anterior external line angle of the vertebral body. For all square vertebras the second point (II) is placed on the superior posterior external line angle of the vertebral body. For all square vertebras the third point (III) is placed on the inferior posterior external line angle of the vertebral body. For all square vertebras the forth point is computer generated. The 4th point (IV) is placed to the location to make the inferior side of the figure equal and parallel to the same of the upper.
For the vertebra without vertebral body (C1) the point locations are as follows:
- The first point is placed on the Superior Anterior External Line Angle formed by dens of C2 and the most proximal point to it on C1
- The second point is placed on the Superior Posterior Internal Line Angle formed by radiographic laminar-pedicle junction of C1 and the superior border of C1 at the end point of that junction
- The third point is placed on the Inferior Posterior Internal Line Angle formed radiographically by laminar-pedicle junction of C1 and the inferior border of C1 at the very end point of that junction.
- The forth point is computer generated, as described above
For triangular vertebra (C2):
4. The first point is placed on the Inferior Anterior External Line Angle of the vertebral body.
5. The second point is placed on the Inferior Posterior External Line Angle of the vertebral body.
6. The third point is placed on the next Superior Posterior External Line Angle formed between transverse processus of C2 and the dens of C2 vertebrae
7. The forth point is computer generated
Measuring: Before placement of points the local user has to calibrate the image. Once we have placed the points we can relate them to each other by calculating degrees of their planes. After the points are placed and proper geometrical figures are created the device will measure relative metric discrepancies. The device measures relative angulation as well as specific metric relationship of shifts in millimeters and in percentages relative specific points. Then, the data is compared with known norms and abnormalities respective to the specific points and spinal parts.
The device does not read pathologies. It is an AOMSI (alteration of motion segment integrity) specific.
Quality control functionality:
- If the order of the points placed are not followed the above protocol (not as 1, 2, 3, but 1, 3, 2) the 4th point will be placed away from the field of the column by the computer making the operator aware of the error made.
- The error figure will be seen by the local user immediately during processing.
- No points can be placed on an image without prior calibration
- X-ray images have to be taken with the calibrator, a small quadrangular metal plate with known dimensions. The calibrator will be affixed on the patient's skin in the designated area with the help of special, made out of paper, holder. The holder will position the metal plate vertically. The plate will be shown on an image as a quadrangle. A local user (operator) will be adjusting the image's size according to the calibrator's image and its known dimensions. The image set will be reviewed for quality before the processing can be done. If images sent are not in the appropriate quality, the image set will be rejected by the local user. The web user will see the specific image set rejected and why and will consider for a resubmission a better quality image.
- Buttons: 1) "Undo"- returns to the previous state, 2) "Zoom In"- increases the image in size, 3) "Zoom Out"- decreases the image in size, 4) "Reset to Original"- returns the image to the original size, 5) "Trace Image" guides the operator through the processing of the image, 6) "Test Points" retrieves the points from the database on the processed image from the database.
Mentions image processing
Yes
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Spinal X-ray
Anatomical Site
Human spine
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Physician / Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Nonclinical tests for Spinal Guides PACS (software verification and validation) have been performed in conformance to specifications and the application is in conformance to DICOM standard PS3 NEMA 3.0 1/1/2008.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
MAR ] 3 2010
Page 1 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
Revised 510(k) Summary
Spinal Guides PACS (PACS Medical Imaging Display Workstation)
Picture Archiving and communications Common/Classification Name: system (PACS)
LLZ
Classification Product Code
Proprietary Name:
510 (K) Letter Owner
Address
Phone number Fax e-mail
109-33 71st Road, #11B Forest Hills, NY 11375 718-261-4812 718-261-4812 albertydavydov@yahoo.com
Spinal Guides PACS
Albert Davydov, DDS, PC
Albert Davydov, DDS Submitter's Contact person:
August 1, 2009 Date the Summary was prepared:
A. LEGALLY MARKETED PREDICATE DEVICE
The predicate device name | Opal-RadTM |
---|---|
Applicant Name | Viztek, Inc. |
6491 Powers Ave. | |
Jacksonville, FL 32217 | |
510(K) Number | K063337 |
Regulation # 21 CFR | 892.2050 |
1
Page 2 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
LLZ Classification Product Code
B. DEVICE DESCRIPTION
The device is represented by a software computer program installed on regular computers.
The device has two types of users: a web and a local user.
The web user submits an image set for processing to the local user via internet to the device's database.
The local user is viewing the images of the image set, process the image set and sends a PDF report with calculations for the web user.
Workflow details
Working a mage is transferred to the device database by a web user. An operator (a local user) then uses a workstation to analyze AOMSI and generate a PDF (a look abol ations, which is then available for viewing on the web for the web user.
The local user/operator/server is installed at only one core location and X-ray images from different users are sent to the location for analyses.
The device is primarily intended for viewing digital spinal X-ray images and to prepare analyses of alteration of motion segment integrity (AOMSI).
Automated part of the reporting process
-
- Automatic limitation of points placed respective to the number of vertebras
-
- Automation of 4th point positioning.
-
- Automatic mathematical calculations
-
- Automatic notice to calibrate an image upon placement of first point on an image in image editor window
-
- Automation of notification of an improper order of points
-
- Automatic user password retrieval
-
- Automatic submission of PDF report via e-mail to the web user
What is needed to perform the analysis?
2
Page 3 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
-
- The analyses are to be performed only by a licensed medical practitioner who is knowledgeable of specific spinal anatomy and specific protocols for image analysis
-
- Medical Device Spinal Guides PACS
- Spinal pure lateral X-ray views in flexion and extension taken with the calibrator 3. in place
-
- Computer hardware
-
- Internet with enough available speed
Processing
For all square vertebras the first point (I) is placed on the superior anterior
external line angle of the vertebral body.
For all square vertebras the second point (II) is placed on the superior
posterior external line angle of the vertebral body.
For all square vertebras the third point (III) is placed on the inferior
posterior external line angle of the vertebral body.
For all square vertebras the forth point is computer generated. The 4th
point (IV) is placed to the location to make the inferior side of the figure equal and
parallel to the same of the upper.
For the vertebra without vertebral body (C1) the point locations are as follows:
- The first point is placed on the Superior Anterior External Line 1.
Angle formed by dens of C2 and the most proximal point to it on C1
3
Page 4 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
The second point is placed on the Superior Posterior Internal Line 2. Angle formed by radiographic laminar-pedicle junction of C1 and the superior border of C1 at the end point of that junction
The third point is placed on the Inferior Posterior Internal Line 3. Angle formed radiographically by laminar-pedicle junction of C1 and the inferior border of C1 at the very end point of that junction. The forth point is computer generated, as described above
For triangular vertebra (C2):
The first point is placed on the Inferior Anterior External Line 4. Angle of the vertebral body.
The second point is placed on the Inferior Posterior External Line 5. Angle of the vertebral body.
The third point is placed on the next Superior Posterior External 6. Line Angle formed between transverse processus of C2 and the dens of C2 vertebrae
- The forth point is computer generated
4
Page 5 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
Once the third point is placed, the fourth one is generated by a computer.
The fourth point is placed to complete a proper quadrangle to form square, parallelogram or rhombus.
The fourth point is placed by a computer to make the lower side equal and
parallel to the upper side and the opposing internal line angles equal to each other.
Measuring
Before placement of points the local user has to calibrate the image.
Once we have placed the points we can relate them to each other by
calculating degrees of their planes.
After the points are placed and proper geometrical figures are created the
device will measure relative metric discrepancies.
The device measures relative angulation as well as specific metric relationship of shifts in millimeters and in percentages relative specific points.
Then, the data is compared with known norms and abnormalities respective to the specific points and spinal parts.
Significance of the study
In order to treat the patient better, estimate prognosis of the patient's condition and quantify the patient's impairment due to damages to the spine the AMA published certain criteria by which a doctor has to formulate a diagnosis
5 of 11
5
Page 6 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
and treatment plan, as well as quantify the patient's permanent impairment rating. The last edition, the 6th Edition of AMA published in 2008 specific limitations in norm and abnormalities for spinal evaluations of alteration of motion segment integrity calculations.
Using computerized technology installed within the device a doctor can spend almost no time in completion of otherwise lengthy mathematical calculations and get precise diagnostic report ready to apply for the best patient's care.
Once specific measurements of relationship of one vertebra to another are obtained, they are compared to norms and abnormalities.
The patient will be treated better and treatments will get better outcomes.
The individual and third party payers will save money when treatments are not indicated.
The device does not read pathologies. It is an AOMSI (alteration of motion segment integrity) specific.
Quality control functionality
If the order of the points placed are not followed the above protocol (not as 1, 2, 1. 3, but 1, 3, 2) the 4th point will be placed away from the field of the column by the computer making the operator aware of the error made.
6 of 11
6
Page 7 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
The error figure will be seen by the local user immediately during
processing.
No points can be placed on an image without prior calibration
X-ray images have to be taken with the calibrator, a small quadrangular metal plate with known dimensions. The calibrator will be affixed on the patient's skin in the designated area with the help of special, made out of paper, holder. The holder will position the metal plate vertically. The plate will be shown on an image as a quadrangle. A local user (operator) will be adjusting the image's size according to the calibrator's image and its known dimensions. The image set will be reviewed for quality before the processing can be done. If images sent are not in the appropriate quality, the image set will be rejected by the local user. The web user will see the specific image set rejected and why and will consider for a resubmission a better quality image
7 of 11
7
Page 8 of 11
Traditional 510 (k) Submission
న.
Submission Date: K092427
3/2/2010
Buttons: 1) "Undo"- returns to the previous state,
-
"Zoom In"- increases the image in size,
-
"Zoom Out"- decreases the image in size,
-
"Reset to Original"- returns the image to the
original size,
- "Trace Image" guides the operator through the
processing of the image
- "Test Points" retrieves the points from the
database on the processed image from the database
The architecture of the device is represented by the following diagram:
8
Page 9 of 11
Image /page/8/Figure/2 description: This image shows a diagram of a traditional 510(k) submission process. The submission date is 3/2/2010, and the submission number is K092427. The diagram includes a web browser, a web server, a database server, and a local application. The web browser sends image set data to the web server via HTTP, and the web server generates an image set and MSI report. The web server also gets completed image set calculations from the SQL server via TCP/IP and inserts the image set into the image storage.
9
Page 10 of 11
Traditional 510 (k) Submission
Submission Date: K092427
3/2/2010
C. INTENDED USE
ー、
Spinal Guides PACS is intended to perform: operations relating to the acceptance of spinal X-ray medical images and patient demographic information, their display, digital processing, review and editing, measurements report generation with respective storage and teleradiology exchange capabilities.
They are intended to be used by a physician to view the images and use as an aid in calculation of alteration of motion segment integrity (AOMSI) of human spine. This device is not intended to be used for mammography images and does not require usage of 5 Mega pixel monitors.
D. SUBSTANTIAL EQUIVALENCE SUMMARY
"Spinal Guides" PACS has substantially same indications for use as the legally marketed predicate device Opal-Rad manufactured by Viztek, Inc., K063337 with major differences being:
-
- It does not process images for mammography and does not require 5M pixel monitors, nor does it display 3D images acquired from CT and MR modalities.
-
- It was created in C# computer language
-
- Spinal Guides PACS were specifically made viewing of the images and for calculation of Alteration of Motion Segment Integrity (AOMS) of human spine.
This indication difference does not modify the intended use- to assist users performing quantifications and qualifications common to the general medicine, orthopedic, radiologic, and chiropractic specialty.
The new device has substantially the same control methods and operating principles as the predicate (both are computer systems and software installed in them, both intended for viewing of images) with the following major differences:
-
- Images for Spinal Guides are to be taken with the calibrator affixed on the patient's skin
- Web users are opening the images and enter all required demographic information 2. and complete an image set for submission to database in "Pending" status
-
- Local users are able to open image sets and view the images as well as process the images using automatic calculation process and formulate a PDF report turning "In process" and "Waiting for Completion" and "Complete", or turn the image sets into "Reject" to reject the image set.
10
Page 11 of 11
Traditional 510 (k) Submission
Submission Date: K092427
.3/2/2010
-
- Once the image set turned to "Complete" by the local user the PDF report is automatically available to the web user
Descriptive characteristics and data provided in this submission are sufficiently precise, to the best of our knowledge, to assure substantial equivalence.
- Once the image set turned to "Complete" by the local user the PDF report is automatically available to the web user
E. TECHNOLOGICAL CHARACTERISTICS
Both the new device and the predicate are software applications installed on standard computers and servers, and utilize teleradiology via internet.
The technological characteristics are substantially the same in the proposed and predicate devices, with major differences are as follows:
-
- Visual Studio 3.5.net C# programming language
-
- The new device does not require usage of 5 megapixel monitors since the application is not intended to use mammography images
F. TESTING
Nonclinical tests for Spinal Guides PACS (software verification and validation) have been performed in conformance to specifications and the application is in conformance to DICOM standard PS3 NEMA 3.0 1/1/2008
G. CONCLUSIONS
This 510(k) has demonstrated Substantial Equivalence of Spinal Guides PACS and Opal-Rad ™ as defined in the Federal Food, Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
11
Image /page/11/Picture/0 description: The image shows the logo for the Department of Health & Human Services (USA). The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles an abstract bird or a flowing design, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring. MD 20993-0002
MAR 1 3 2010
Albert Davydov, DDS 109-33 7156 Road, #11B FOREST HILLS NY 11375
Re: K092427
Trade/Device Name: Spinal Guides PACS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 2, 2010 Received: March 9, 2010
Dear Dr. Davydov:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
12
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Donald J. Trump
Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
13
Page 1 of 1
Submission Date:
3/2/2010
Indications for use Statement
510(k) Number:
Spinal Guides PACS
Device Name:
Indications For Use:
Spinal Guides PACS is intended to perform: operations relating to the acceptance of spinal Xray medical images and patient demographic information, their display, digital processing, review and editing, measurements report generation with respective storage and teleradiology exchange capabilities. They are intended to be used by a physician to view the images and use as an aid in calculation of
alteration of motion segment integrity (AOMSI) of human spine.
This device is not intended to be used for mammography images and does not require usage of 5 Mega pixel monitors.
Prescription Use AND/OR Over-The-Counter Use No Yes (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Beview Evaluation Page 1 of
(Division Sign-On)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K092427