K Number
K092353

Validate with FDA (Live)

Date Cleared
2009-10-29

(86 days)

Product Code
Regulation Number
866.3310
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MONOLISA™ Anti-HAV IgM EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD). This assay is indicated for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis. Assay results, in conjunction with other serological or clinical information, may be used for the laboratory diagnosis of individuals with acute or recent hepatitis A. The MONOLISA™ Anti-HAV IgM EIA is intended for manual use and with the Evolis™ Automated Microplate System in the detection of IgM antibodies to hepatitis A virus.

Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, and cord blood or neonatal specimens.

Warning: This assay is not intended for screening blood or solid or soft tissue donors.

Device Description

The MONOLISA™ Anti-HAV IgM EIA is an enzyme immunoassay (IgM antibody capture format) for the detection of IgM antibodies to hepatitis A virus. In the assay procedure, patient specimens, a calibrator, and controls are incubated with antibodies coated on the microwells. If IgM antibodies to HAV are present in a specimen or control, they bind to the antibody. Excess sample is removed by a wash step. The HAV Viral Antigen and the Conjugate (containing horseradish peroxidase - labeled mouse monoclonal antibody to HAV) are successively added to the microwells and allowed to incubate. The presence of anti-HAV IgM in the sample enables the HAV Viral Antigen and the Conjugate to bind to the solid phase. Excess Conjugate and HAV Viral Antigen are removed by a wash step, and a TMB Chromogen /Substrate solution is added to the microwells and allowed to incubate. If a sample contains anti-HAV IgM, the bound enzyme (HRP) causes the colorless tetramethylbenzidine (TMB) in the Chromogen solution to change to blue. The blue color turns yellow after the addition of a Stopping Solution. If a sample does not contain anti-HAV IgM, the Chromogen/Substrate solution in the well remains colorless during the substrate incubation, and after the Stopping Solution. The color intensity is measured spectrophotometrically. Absorbance value readings for patient specimens are compared to the cutoff value.

The performance of the MONOLISA™ Anti-HAV IgM EIA was evaluated in conjunction with the EVOLIS™ Automated Microplate System. The EVOLIS™ Automated Microplate System is a fully automated microplate analyzer that performs all functions necessary for the complete processing of microplate assays. Functions include: barcode scanning, sample pre-dilutions, sample and reagent dispensing, plate incubations, plate wash cycles, photometric measurement of completed assay plates and results evaluation. The analyzer instrument is controlled via the EVOLIS™ software, a Windows® 2000 application running on a separate dedicated PC. An operator loads the appropriate microplates, assay reagents, and patient and control samples, then selects assay parameters, loads sample information, initiates instrument processing, and generates result reports.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (Implicitly Derived)Reported Device Performance (MONOLISA™ Anti-HAV IgM EIA on EVOLIS™ vs. Manual)
Correlation/Method ComparisonOverall Agreement: High percentage agreement with the manual method.Overall Percent Agreement: 99.7% (with 95% CI of 98.9 - 99.9%)
Positive Agreement: High percentage agreement for reactive samples.Positive Percent Agreement: 100% (with 95% CI of 96.1 - 100%)
Negative Agreement: High percentage agreement for nonreactive samples.Negative Percent Agreement: 99.7% (with 95% CI of 98.8 - 99.9%)
Reproducibility (Combination Plate)Overall Agreement: High percentage agreement with the manual method (individual plate format).Overall Percent Agreement: 99.4% (with 95% CI of 97.7 - 99.8%)
Positive Agreement: High percentage agreement for reactive samples (combination plate).Positive Percent Agreement: 100% (with 95% CI of 92.7 - 100%)
Negative Agreement: High percentage agreement for nonreactive samples (combination plate).Negative Percent Agreement: 99.2% (with 95% CI of 97.3 - 99.8%)
PrecisionLow coefficients of variation (%CV) for within-run, between-run, between-day variability across different panel members (positive, high negative, negative, and controls), and different sample types (serum, various plasma types).Within-run %CV: Ranged from 1.3% to 6.1% (Panel Set 1, 2, 3) Between-Run %CV: Ranged from 3.5% to 14.9% Between-Day %CV: Ranged from 4.4% to 14.7% Total %CV: Ranged from 6.8% to 21.8%
Reproducibility (Multi-site)Low coefficients of variation (%CV) for within-run, between-day, between-site, and total variance across different panel members.Within-run %CV: Ranged from 2.3% to 20.5% (across all sites and panel members) Between-Day %CV: Ranged from 0.4% to 19.4% Between-Site %CV: Ranged from 0.0% to 2.9% (for P2 and P5 only, others 0.0%) Total %CV: Ranged from 2.5% to 23.9%
Pipettor and Washer Carry-overNo significant carry-over between samples/wells.Verified that disposable tip pipettes and washer do not carry residuals.
Pipetting AccuracyAchieve a coefficient of variation (CV) of ≤7.7% across the microwell plate for sample and reagent dispensing.Demonstrated pipetting accuracy with a CV of ≤7.7%.

2. Sample Sizes and Data Provenance

  • Test set for Correlation/Method Comparison:
    • Study 1 (EVOLIS™ vs. Manual): 691 retrospective samples.
    • Study 2 (EVOLIS™ Combination Plate vs. Manual): 313 samples.
  • Data Provenance: The document does not explicitly state the country of origin. The samples were retrospective.

3. Number of Experts and Qualifications for Ground Truth

  • The document does not mention the use of experts to establish ground truth for the test set.
  • The "ground truth" for the comparison studies was the result of the MONOLISA™ Anti-HAV IgM EIA tested manually (the predicate device). The assumption is that the manual method itself, having received prior FDA clearance (K063319), represents the established truth for HAV IgM detection.

4. Adjudication Method

  • The document does not describe a specific adjudication method using experts.
  • For borderline results in the correlation studies, the following rule was applied: "specimens that were borderline with the reference assay and negative with EVOLIS™ were considered as false negative for the EVOLIS™." This represents a specific rule for handling discordant borderline results rather than an expert adjudication process.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done.
  • This study pertains to the performance of an automated assay system compared to a manual assay method, not a comparison of human readers with and without AI assistance. Therefore, there is no effect size reported for human readers improving with AI.

6. Standalone Performance Study

  • Yes, a standalone (algorithm only) performance study was effectively done. The performance studies (correlation/method comparison, precision, reproducibility, carry-over, pipetting accuracy) evaluate the MONOLISA™ Anti-HAV IgM EIA when used with the EVOLIS™ Automated Microplate System as a fully automated system, without direct human intervention in the result determination process itself (beyond loading, selecting parameters, and generating reports). The "algorithm" here refers to the automated system and its integrated processes compared to the manual method.

7. Type of Ground Truth Used

  • The ground truth used for the performance comparison studies was the result from the predicate device (Manual MONOLISA™ Anti-HAV IgM EIA assay). This is a comparative ground truth established by another existing, cleared diagnostic method, rather than direct pathology, expert consensus on clinical findings, or outcomes data.

8. Sample Size for the Training Set

  • The document does not specify a separate "training set" in the context of an algorithm or AI model development. This submittal is for an automated system running an existing assay, not a novel AI algorithm that requires training on a data set. Therefore, this information is not applicable or provided.

9. How Ground Truth for Training Set Was Established

  • As noted above, there is no explicit "training set" for an AI algorithm mentioned. The performance characteristics of the assay itself (MONOLISA™ Anti-HAV IgM EIA) were established through its prior 510(k) clearance (K063319) when used manually. The current submission focuses on demonstrating that combining this assay with the EVOLIS™ Automated Microplate System yields equivalent performance.

{0}------------------------------------------------

510(k) Summary

OCT 29 2009

1. Company:Bio-Rad Laboratories6565 185th Avenue NERedmond, WA 98052Phone: 425 881-8300Fax: 425 498-1651
Contact:Linda Staswick
Date Summary Prepared:October 22, 2009

2. Device Name:

Device Trade Name:MONOLISA™ Anti-HAV IgM EIAEVOLIS™ Automated MicroplateSystem
Common Name:IgM Antibody to Hepatitis A Virus Automated Laboratory Analyzer
Classification Name:Hepatitis A Virus (HAV)serological assaysDiscrete photometric chemistryanalyzer for clinical use
Product Code:LOLJJE
Regulation Number:21 CFR 866.331021 CFR 862.2160
Regulatory Class:Class IIClass I
Panel:MicrobiologyChemistry

3. Substantial Equivalence:

The MONOLISA™ Anti-HAV IgM EIA used with the EVOLIS™ Automated Microplate System is substantially equivalent to the MONOLISA™ Anti-HAV IgM EIA using the manual method (KO63319)

Description of the Device: 4.

The MONQLISA™ Anti-HAV IgM EIA is an enzyme immunoassay (IgM antibody capture format) for the detection of IgM antibodies to hepatitis A virus. In the assay procedure, patient specimens, a calibrator, and controls are incubated with antibodies coated on the microwells. If IgM antibodies to HAV are present in a specimen or control, they bind to the antibody. Excess sample is removed by a wash step. The HAV Viral Antigen and the Conjugate (containing horseradish peroxidase - labeled mouse monoclonal antibody to HAV) are successively added to the microwells and allowed to incubate. The presence of anti-HAV IgM in the sample enables the HAV Viral Antigen and the Conjugate to bind to the solid phase. Excess Conjugate and HAV Viral Antigen are removed by a wash step, and a TMB Chromogen /Substrate solution is added to the microwells and allowed to incubate. If a sample contains anti-HAV IgM, the bound enzyme (HRP) causes the colorless tetramethylbenzidine (TMB) in the Chromogen solution to change to blue. The blue color turns vellow after the addition of a Stopping Solution. . If a sample does not contain anti-HAV IgM, the Chromogen/Substrate solution in the well remains colorless during the substrate incubation, and after the Stopping Solution. The color intensity is measured spectrophotometrically. Absorbance value readings for patient specimens are compared to the cutoff value.

The performance of the MONOLISA™ Anti-HAV IgM EIA was evaluated in conjunction with the EVOLIS™ Automated Microplate System. The EVOLIS™ Automated Microplate System is a fully automated microplate analyzer that performs all functions necessary for the complete processing of microplate assays. Functions include: barcode scanning, sample

{1}------------------------------------------------

pre-dilutions, sample and reagent dispensing, plate incubations, plate wash cycles, photometric measurement of completed assay plates and results evaluation. The analyzer instrument is controlled via the EVOLIS™ software, a Windows® 2000 application running on a separate dedicated PC. An operator loads the appropriate microplates, assay reagents, and patient and control samples, then selects assay parameters, loads sample information, initiates instrument processing, and generates result reports.

5. Intended Use:

The MONOLISA™ Anti-HAV IgM EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD). This assay is indicated for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis. Assay results, in conjunction with other serological or clinical information, may be used for the laboratory diagnosis of individuals with acute or recent hepatitis A. The MONOLISA™ Anti-HAV IqM EIA is intended for manual use and with the Evolis™ Automated Microplate System in the detection of IgM antibodies to hepatitis A virus.

Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, and cord blood or neonatal specimens.

Warning: This assay is not intended for screening blood or solid or soft tissue donors.

6. Technological Characteristics

The following tables summarize similarities and differences between the MONOLISA™ Anti-HAV IqM EIA tested manually and the MONOLISA™ Anti-HAV IqM EIA tested with the EVOLIS™ Automated Microplate System.

{2}------------------------------------------------

Table 1: Similarities between devices
ParameterMONOLISA™ Anti-HAV IgM EIA tested with the EVOLIS™ Automated Microplate SystemMONOLISA™ Anti-HAV IgM EIA tested manually
Intended Use/Indications for UseThe MONOLISA™ Anti-HAV IgM EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD).The MONOLISA™ Anti-HAV IgM EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD).
Assay procedurePer the instructions in the package insertPer the instructions in the package insert
Plate incubation$60 \pm 5$ minutes at 37°C + 2°C$60 \pm 5$ minutes at 37°C + 2°C
Plate washingWash with ≥ 370 µL of Working Wash Solution per well, and 30 - 60 second soak between each wash cycle for a total of 5 cycles.Wash with ≥ 370 µL of Working Wash Solution per well, and 30 - 60 second soak between each wash cycle for a total of 5 cycles.
Result interpretationResult interpretations, based on sample O.D.s, are determined according to package insert criteria.Result interpretations, based on sample O.D.s, are determined according to package insert criteria.
Photometric measurement of completed assay platesRead absorbance using 450 nm filter with 620 nm as the referenceRead absorbance using 450 nm filter with 615 to 630 nm as the reference

Table 1· Similarities between devices

Table 2: Differences between devices
ParameterMONOLISA™ Anti-HAV IgM EIAtested with the EVOLIS™Automated Microplate SystemMONOLISA™ Anti-HAV IgM EIAtested manually
Sample and reagentdispensingSamples and reagents are dispensedby the automated systemSamples and reagents are dispensedmanually
Barcode readingSample and reagent ID are verifiedautomaticallyNA or can be performed manuallywith barcode wand
Plate incubationPlates are automatically moved tothe incubation chamberPlates are moved manually to theincubation chamber
Plate wash cyclesPlates are automatically washedPlates are moved manually to anautomated plate washer
Data managementArchives and retrieves data andsample informationNA
Spectrophotometricverification of sample andreagent pipetingPerformed automaticallyOptional verification visually or withmicroplate reader

7. Performance Characteristics:

The performance of the MONOLISA™ Anti-HAV IgM EIA with the EVOLIS™ Automated Micropiate System was compared to the MONOLISA™ Anti-HAV IgM EIA tested manually, which wad previously received marketing clearance from the Agency. Substantial equivalence of the MONOLISA™ Anti-HAV IgM EIA, using manual equipment, was determined May 3, 2007 (K063319).

{3}------------------------------------------------

Correlation/method comparison

Studies have been performed with the MONOLISA™ Anti-HAV IgM EIA on the EVOLIS™ Automated System and compared to the results of testing the same kits and samples with the manual method. In this study 691 retrospective samples were tested on the MONOLISA™ Anti-HAV IgM assay using four (4) EVOLIS™ instruments at three sites. The same samples were tested manually (reference method) on the MONOLISA™ Anti-HAV IdM assay. The positive, neqative and overall percent along with the 95% confidence interval are presented below. In determining the percent agreement on borderline results, specimens that were borderline with the reference assay and negative with EVOLIS™ were considered as false negative for the EVOLIS ™.

EVOLIS™ Anti-HAV IgM Results
Manual Anti-HAV IgM ResultsReactiveBorderlineNonreactiveTotal
Reactive940094
Borderline1001
Nonreactive10595596
Total960595691
Table 2: MONOLISA™ Anti-HAV IgM EIA on EVOLIS™ vs. Manual Results

The positive percent agreement with the reference method, manual testing, is 100% (94/94) with a 95% confidence interval of 96.1 - 100%. The negative percent agreement with the reference method is 99.7% (595/597) with a 95% confidence interval of 98.8 - 99.9%. The overall percent agreement is 99.7% (689/691) with a 95% confidence interval of 98.9 - 99.9%

The EVOLIS™ was also evaluated by performing a combination of 2 assays on the same plate. In this study 313 samples were tested with the MONOLISA™ Anti-HAV IgM EIA on a combination plate format on EVOLIS™ (two separate MONOLISA™ hepatitis assays run in a single microplate frame). Results were compared to the same samples tested manually (the reference method, individual plate format) on the MONOLISA™ Anti-HAV IgM assay. In determining the percent agreement on borderline results, specimens that were borderline with the reference assay (manual individual plate) and negative with EVOLIS™ (combination plate) were considered as false negative for the EVOLIS™ (combination plate).

EVOLIS™ Anti-HAV IgM ResultsIndividual Plate
Manual Anti-HAV IgM ResultsCombination PlateReactiveBorderlineNonreactiveTotal
Reactive490049
Borderline1001
Nonreactive01262263
Total501262313

Table 3: MONOLISA™ Anti-HAV IgM ElA on EVOLIS™ Combination Plate Testing vs. Manual Results

The positive percent agreement with the reference method, manual testing, is 100% (49/49) with a 95% confidence interval of 92.7 – 100%. The negative percent agreement with the reference method is 99.2% (262/264) with a 95% confidence interval of 97.3 - 99.8%. The overall percent agreement is 99.4% (311/313) with a 95% confidence interval of 97.7 - 99.8%

{4}------------------------------------------------

Precision Study (Within-Laboratory)

A 21-member panel was tested: three (3) serum samples with six (6) corresponding plasma samples (EDTA K2, EDTA K3, Sodium Citrate, Sodium Heparin, Lithium Heparin, ACD) at three (3) different levels [1 low positive near the cutoff (Panel Set 1), 1 negative near the cutoff (Panel Set 2) and 1 negative (Panel Set 3)]. Two replicates each of the twenty-four (24) member panel were assayed twice a day for 20 days. The data were analyzed following the CLSI guidance EP5-A2, Evaluation of Precision Performance of Quantitative Measurement Methods. The mean ratio, the Standard Deviation (SD) and percent coefficient of variation (%CV) were calculated for each panel member.

The data summary is shown in the following tables, which summarize testing with the EVOLIS™ Automated System:

NMeanWithin run¹Between Run ²Between Day ³Total ⁴
Panel MemberS/COSDCV (%)SDCV (%)SDCV (%)SDCV (%)
Positive Control801.970.0351.80.0914.60.1638.30.1909.7
High Negative800.100.0066.10.01514.90.01514.70.02221.8
Cutoff Control803.780.1463.90.1323.50.1664.40.2586.8
Serum (1)801.550.0362.30.0764.90.1388.90.16110.4
EDTA K2 (1)801.440.0201.40.0755.20.1319.10.15210.6
EDTA K3 (1)801.490.0302.00.0835.60.1268.50.15410.3
Sodium Citrate (1)801.480.0332.20.0865.80.1409.50.16811.3
Sodium Heparin (1)801.410.0241.70.0805.70.1329.40.15611.1
Lithium Heparin (1)801.390.0261.90.0775.50.1208.70.14510.5
ACD (1)801.640.0211.30.1076.60.1448.80.18111.0
Serum (2)800.620.0162.70.0315.00.0599.50.06811.1
EDTA K2 (2)800.690.0162.30.0345.00.07711.30.08612.5
EDTA K3 (2)800.690.0142.00.0466.60.07310.50.08712.5
Sodium Citrate (2)800.740.0141.90.0445.90.07510.10.08811.9
Sodium Heparin (2)800.660.0111.60.0416.20.0619.20.07411.2
Lithium Heparin (2)800.660.0203.00.0406.10.0588.90.07311.1
ACD (2)800.780.0121.50.0526.70.0729.20.09011.5
Serum (3)800.100.0043.60.0109.70.01010.10.01514.5
EDTA K2 (3)800.110.0054.70.01110.30.0098.20.01514.0
EDTA K3 (3)800.100.0044.20.0109.50.01110.60.01514.8
Sodium Citrate (3)800.100.0032.90.0099.20.0109.80.01413.8
Sodium Heparin (3)800.100.0043.80.0098.70.0109.90.01413.7
Lithium Heparin (3)800.100.0154.50.0109.50.0109.20.01414.0
ACD (3)780.100.0054.30.01010.00.0098.70.01513.9

Table 4: MONOLISA™ Anti-HAV IgM EIA Precision Results by Panel Member Signal to Cutoff (S/CO)

Within Run: variability of the assay performance from replicate to replicate

Between Run: variability of the assay performance from run to run

3 Between Day: variability of the assay performance from day to day

4 Total: total variability of the assay performance includes within run, between run and between day.

{5}------------------------------------------------

Reproducibility Study


A 6-member panel consisting of diluted plasma specimens (negative and different levels of positive) was tested in triplicate, once a day for 5 days with the MONOLISA™ Anti-HAV IgM EIA at 3 separate clinical trial sites. Each panel was coded with a different number on each day tested in order to blind the operator to the expected value of the sample. One (1) lot was used at each of 3 sites.

The data from all sites were combined to obtain standard deviation (SD) and percent coefficient of variation (CV) for within run, between day, between site and total variance. The data were analyzed according to the principles described in the Clinical Laboratory Standards Institute guidance EP15-A2, User Protocol for Evaluation of Qualitative Test Performance. The summaries are shown in the following tables:

TestSiteID #Panel MemberNMean(S/CO)Within Run1Between Day2Total3
SD%CVSD%CVSD%CV
Site #1P1Negative300.160.03320.50.0149.00.03622.4
P2High Negative300.740.0233.00.0243.20.0334.4
P3Low Positive301.210.0363.00.0342.80.0504.1
P4Low Positive301.210.0514.20.0312.50.0594.9
P5Positive303.130.0822.60.1173.70.1434.6
P6Positive293.680.1103.00.1363.70.1754.8
P7Positive Control301.950.0794.10.0934.80.1226.3
P8Negative Control300.110.0086.90.01110.40.01412.5
P9Cutoff Calibrator303.280.1394.20.1594.90.2116.4
Site #2P1Negative300.170.0137.70.0085.00.0169.2
P2High Negative300.720.0192.70.0101.40.0223.0
P3Low Positive301.180.0292.50.0040.40.0292.5
P4Low Positive301.170.0363.10.0221.80.0423.6
P5Positive303.060.0702.30.0521.70.0872.9
P6Positive303.650.0932.50.0711.90.1163.2
P7Positive Control301.910.0733.80.0231.20.0764.0
P8Negative Control300.120.0097.40.0097.70.01310.7
P9Cutoff Calibrator303.540.2456.90.1183.30.2727.7
Site #3P1Negative300.160.0159.00.02113.00.02515.8
P2High Negative290.690.0294.20.0405.70.0497.1
P3Low Positive301.140.0514.40.0474.20.0696.1
P4Low Positive301.140.0484.20.0504.40.0696.1
P5Positive292.940.0822.80.1113.80.1384.7
P6Positive293.570.2697.60.2085.80.3409.5
P7Positive Control301.840.0784.20.0754.10.1085.9
P8Negative Control280.110.01514.10.02119.40.02623.9
P9Cutoff Calibrator283.560.2787.80.3439.60.44112.4
Table 5: MONOLISA ™ Anti-HAV IgM EIA Reproducibility Results by Panel Member Signal to Cutoff (S/CO)
----------------------------------------------------------------------------------------------------------------

1 Within Run: variability of the assay performance from replicate to replicate

2 Between Day: variability of the assay performance from day to day

3 Total: total variability of the assay performance includes within run and between day

{6}------------------------------------------------

Panel MemberNMeanS/COWithin Run1Between Day2Between Site3Total4
SD%CVSD%CVSD%CVSD%CV
P1900.160.02213.50.0148.90.00050.00.02616.1
P2890.720.0243.30.0273.80.0212.90.0425.8
P3901.180.0403.40.0342.90.0312.60.0615.2
P4901.170.0453.90.0363.10.0322.80.0665.7
P5893.050.0782.60.0983.20.0842.80.1515.0
P6883.630.1764.80.1444.00.00050.00.2276.3
P7901.900.0774.00.0703.70.0442.30.1135.9
P8880.110.0119.70.01513.00.00050.00.01816.2
P9883.460.2276.60.2296.60.1063.10.3399.8

Table 6: MONOLISA™ Anti-HAV IgM EIA Reproducibility Summary by Panel Member Signal to Cutoff (S/CO)

1 Within Run: variability of the assay performance from replicate to replicate

2 Between Day: variability of the assay performance from day to day

3 Between Site: variability of the assay performance from site to site

4 Total: total variability of the assay performance includes within run and between site

5 Negative variances were rounded to zero, per statistical convention

Pipettor and washer carry-over

The pipette carryover study verified that the disposable tip pipettes on the EVOLIS™ do not carry residuals from one sample or well to another. In a washer carryover study, it was verified that the washer on the EVOLIS™ does not carry residuals from one well to another during the washing steps.

Pipetting accuracy

Dye studies were performed to determine pipetting accuracy for samples and reagents. These studies were conducted using 2 different volumes for samples and controls, and demonstrated pipetting accuracy with a CV of ≤7.7% across the microwell plate.

8. Conclusion

The MONOLISA™ Anti-HAV IgM EIA tested with the EVOLIS™ Automated Microplate System demonstrated equivalent performance to the MONOLISA™ Anti-HAV IgM EIA tested with the manual assay method, which had previously received FDA 510(k) clearance.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

Bio-Rad Laboratories Attn: Linda Staswick 6565 185" Ave. NE Redmond, WA 98052

CT 9 9 2009

Re: K092353

Trade/Device Name: MONOLISA™ Anti-HAV IgM ELA with the EVOLIS™ Automated Microplate System Regulation Number: 21 CFR 866.3310 Regulation Name: Hepatitis A virus (HAV) serological assays Regulatory Class: Class II Product Codes: LOL, JJE Dated: July 30, 2009 Received: August 4, 2009

Dear Ms. Staswick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{8}------------------------------------------------

Page 2 - Linda Staswick

CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

foqatra

Sally Hojvat, Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Indications for Use

510(k) Number (if known): K092353

Device Name: MONOLISA™ Anti-HAV IgM EIA

Indication For Use:

The MONOLISA™ Anti-HAV IgM EIA is an in vitro enzyme immunoassay kit intended for use in the qualitative detection of IgM antibodies to hepatitis A virus (anti-HAV IgM) in human (adult and pediatric) serum or plasma (EDTA, Heparin, Citrate, ACD). This assay is indicated for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis. Assay results, in conjunction with other serological or clinical information, may be used for the laboratory diagnosis of individuals with acute or recent hepatitis A. The MONOLISA™ Anti-HAV IgM EIA is intended for manual use and with the Evolis™ Automated Microplate System in the detection of IgM antibodies to hepatitis A virus.

Assay performance characteristics have not been established for immunocompromised or immunosuppressed patients, and cord blood or neonatal specimens.

Warning: This assay is not intended for screening blood or soft tissue donors.

Prescription Use X (21 CFR Part 801 Subpart D) Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

And/Or

Uve Schuf

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

4092353 510(k)

§ 866.3310 Hepatitis A virus (HAV) serological assays.

(a)
Identification. HAV serological assays are devices that consist of antigens and antisera for the detection of hepatitis A virus-specific IgM, IgG, or total antibodies (IgM and IgG), in human serum or plasma. These devices are used for testing specimens from individuals who have signs and symptoms consistent with acute hepatitis to determine if an individual has been previously infected with HAV, or as an aid to identify HAV-susceptible individuals. The detection of these antibodies aids in the clinical laboratory diagnosis of an acute or past infection by HAV in conjunction with other clinical laboratory findings. These devices are not intended for screening blood or solid or soft tissue donors.(b)
Classification. Class II (special controls). The special control is “Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Hepatitis A Virus Serological Assays.” See § 866.1(e) for the availability of this guidance document.