K Number
K092301
Date Cleared
2009-10-22

(85 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Gloves, (Yellow, white, blue, pink), 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Better Care Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves, based on the provided K092301 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Test PerformedPerformance Standard / AQLReported Device Performance (Meets/Fails)
Physical and Dimensions TestingASTM-D-5250-06e1Meets requirements
Pinhole Detection (Water Fill Test)AQL 2.5, Inspection Level IMeets requirements
Primary Skin IrritationNo primary skin irritant reactionsNo primary skin irritant reactions
Skin Sensitization (Allergic Contact Dermatitis)No sensitization reactionsNo sensitization reactions
Residual PowderNo more than 2 mg powder per glove (ASTM D-6124-06)Meets "powder-free" claims

2. Sample Size Used for the Test Set and Data Provenance

  • Physical and Dimensions Testing: Inspection level S-2, AQL 4.0. The specific number of gloves tested is not explicitly stated, but it would be determined by the sampling plan outlined in ASTM-D-5250-06e1 for an AQL of 4.0 at an S-2 inspection level.
  • Pinhole Detection (Water Fill Test): AQL 2.5, Inspection Level I. The specific number of gloves tested is not explicitly stated, but it would be determined by the sampling plan for an AQL of 2.5 at an Inspection Level I.
  • Primary Skin Irritation and Skin Sensitization: The document states "testing was conducted" but does not specify the sample size (e.g., number of subjects or samples) used for these biocompatibility tests.
  • Residual Powder Test: Based on ASTM D-6124-06. The number of gloves tested for residual powder is not explicitly stated.
  • Data Provenance: The document does not specify the country of origin for the data or if the studies were retrospective or prospective. Given that this is a 510(k) for a manufacturing company, it's likely that the tests were conducted by or on behalf of the manufacturer to demonstrate compliance with standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This type of device (patient examination gloves) does not typically involve human experts establishing "ground truth" in the way an AI diagnostic device would. The "ground truth" for these tests is established by objective, standardized laboratory measurements and compliance with industry standards (e.g., ASTM standards, AQL limits for defects). There are no "experts" in the sense of clinicians interpreting results for a test set.

4. Adjudication Method for the Test Set

Not applicable. The tests performed are objective laboratory tests with defined pass/fail criteria based on standards. There is no subjective interpretation requiring adjudication among multiple readers.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

No. An MRMC study is relevant for AI devices that assist human readers in tasks like image interpretation. This device is a physical product (gloves), and its performance is evaluated through material and quality control tests, not through human reader interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in a sense. The tests performed (physical, dimensions, pinhole, biocompatibility, residual powder) evaluate the intrinsic properties and performance of the glove itself, without any human-in-the-loop or interpretive component in the way an algorithm would be evaluated. The "device" (glove) is tested in a standalone manner against defined physical and chemical parameters.

7. The Type of Ground Truth Used

The ground truth used for these tests is objective, standardized measurements and compliance with established industry standards.

  • For physical properties and dimensions, the ground truth is defined by the tolerances and specifications within ASTM-D-5250-06e1.
  • For pinholes, the ground truth is defined by the FDA 1000 ml. Water Fill Test and its AQL 2.5 acceptance criterion.
  • For biocompatibility (skin irritation and sensitization), the ground truth is established by the absence of adverse reactions in standardized biological tests.
  • For residual powder, the ground truth is the quantitative measurement of powder content against the 2 mg per glove limit specified by ASTM D-6124-06.

8. The Sample Size for the Training Set

Not applicable. This device is a physical product, not an algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

K092301

OCT 2 2 2009

. EXHIBIT #1 Page I of 2

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(K) number is:

l. Submitter's Identification:

Mr. Guixi Liu Better Care Plastic Products Co., Ltd. Fugian Xi Road, West district of Shenze Industrial Base, Shenze County, Hebei Province 050000 Date Summary Prepared: June 30, 2009

2. Name of the Device:

Better Care Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves, (Yellow, white, blue, pink)

3. Predicate Device Information:

Shijiazhuang Great Eagle Plastic Products Co., Ltd Powder-Free (Yellow) Vinyl Patient Examination Gloves (K992861)

4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Gloves, (Yellow, white, blue, pink), 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

5. Intended Use:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

{1}------------------------------------------------

6. Comparison to Predicate Devices:

Better Care Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, (Yellow, white, blue, pink) is substantially equivalent in safety and effectiveness to the Shijiazhuang Great Eagle Plastic Products Co., Ltd.'s Powder-Free (yellow) Vinyl Patient Examination Gloves.

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:

The standards used for Better Care Plastic Products Co., Ltd.'s glove production are based on ASTM-D-5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.

Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

A Residual Powder Test that based on ASTM D-6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

8. Labeling:

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

9. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

10. Conclusions:

Better Care Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves, (Yellow, white, blue, pink) conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a seal for the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

OCT 22 2009

BetterCare Plastic Products Company, Limited C/O Ms. Jie Liu Project Manager Surprotect, Incorporated 3973 Schaefer Avenue Chino, California 91710

Re: K092301

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink) Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: October 14, 2009 Received: October 15, 2009

Dear Ms. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Ms. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Antwerp D, asmen ben
S. P. R. S. M.

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment A

INDICATION FOR USE

510 (k) NUMBER (IF KNOW): APPLICANT: DEVICE NAME:

Better Care Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves, (Yellow, White, Blue, Pink)

INDICATIONS FOR USE:

A patient examination glove is disposable device intended for medical purpose that is A patient exammation Elove is alop seat of the entamination between patient and examiner.

Shiela H. Mingo Lugo

(Division Sign-Off) Division of Anesthesiology. General Hospital infection Control, Dental Devices

510(k) Number: K092301

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter-Use 1 (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrent of CDRH, Office of Device Evaluation (ODE)

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.