K Number
K092253

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-10-08

(72 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The fMRI Hardware System is a stimulus presentation and response collection system intended to be used by trained professionals to facilitate auditory and visual stimulation to be used in functional MR Imaging (fMRI) based on BOLD contrast.

Device Description

The system presents auditory and visual stimulus to the patient gives feedback through a pair of handheld grips. A synchronization module synchronizes the nordicAktiva stimulus presentation software with the MR scanner. The System consists of five subsystems: AudioSystem, VisualSystem, ResponseGrip, SyncBox and nordicAktiva.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the fMRI Hardware System:


The provided document, a 510(k) Summary for a device modification, offers limited details regarding specific quantitative acceptance criteria or a dedicated study report to "prove" the device meets them in the traditional sense of a clinical trial. Instead, the submission emphasizes substantial equivalence to an existing cleared device (K080515) and highlights function and safety testing to confirm the modified device (adding an in-room LCD monitor option) maintains performance.

Therefore, the information below is constructed based on the available text and highlights where specific details are not provided.

1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of this 510(k) for a device modification (adding an LCD monitor option) and the focus on substantial equivalence, the "acceptance criteria" are implied to be adherence to the original device's functionality and safety, as well as general expectations for such a system. The document does not provide numerical performance metrics or specific criterion thresholds in a table format.

Acceptance Criterion (Implied)Reported Device Performance
Functional Equivalence: Visual stimulus presentation (video)"The same video signal can be presented on the coil-mounted display and the in-room LCD-monitor." (Section 3.4)
Functional Equivalence: Auditory stimulus presentationUnchanged. The AudioSystem is not part of the modification.
Functional Equivalence: Response collectionUnchanged. The ResponseGrip is not part of the modification.
Functional Equivalence: Synchronization with MR scannerUnchanged. The SyncBox and nordicAktiva are not part of the modification.
Safety: No new safety risks introduced by the modification"The LCD monitor is placed outside magnets bore behind the MR scanner and is in no physical contact with the patient." (Section 3.4) "The fMRI Hardware System has been tested for function and safety and fulfills all requirement specifications." (Section 3.6)
Intended Use: Consistent with the original device"The intended use for the modified fMRI Hardware System have remained unchanged." (Section 3.5) "This is the same intended use as previously cleared for the fMRI Hardware System, K080515." (Section 3.1)
Overall Performance: Meets all requirement specifications"The fMRI Hardware System has been tested for function and safety and fulfills all requirement specifications." (Section 3.6)

2. Sample Size for the Test Set and Data Provenance

The document does not specify a sample size for a test set in the context of a clinical performance study. The testing mentioned in Section 3.6 ("The fMRI Hardware System has been tested for function and safety...") appears to be internal verification and validation testing of the hardware and software components, rather than a study involving patient data or expert interpretation.

Therefore, there is no explicit data provenance (e.g., country of origin, retrospective/prospective) related to a clinical test set.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

The document does not mention the use of experts to establish ground truth for a test set. This type of submission, concerning a hardware accessory modification, typically focuses on engineering verification and validation rather than clinical performance studies that require expert-derived ground truth.

4. Adjudication Method for the Test Set

Since there is no mention of a test set requiring expert ground truth or interpretation, there is no adjudication method described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC comparative effectiveness study was NOT done or reported. The submission is for a hardware system modification, not an AI-powered diagnostic tool where human reader performance would be a primary metric. Therefore, there is no effect size reported for human readers with vs. without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance Study

No standalone performance study was conducted or reported in the context of a diagnostic algorithm. The device is a stimulus presentation and response collection system for fMRI studies, not an algorithm that provides diagnostic outputs.

7. Type of Ground Truth Used

Given that no clinical performance study or diagnostic algorithm evaluation is described, there is no mention of a specific type of ground truth (e.g., expert consensus, pathology, outcomes data) being used. The "ground truth" for the device's function would be its ability to accurately present stimuli and record responses, as determined by engineering tests and adherence to specifications.

8. Sample Size for the Training Set

The document does not mention a training set sample size. This device is a hardware system, not a machine learning or AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Since there is no training set, there is no ground truth establishment method for it described.

{0}------------------------------------------------

NordicNeuroLab AS Bergen, Norway

K092253

510(k) Summary

Special: Device Modification 510(k) Type: July 20th, 2009 Submission Date:

  • Submitter:
    NordicNeuroLab AS · Møllendalsveien 65C N-5009 Bergen Norway

Phone: +47 55 70 70 95 +47 55 70 70 96 Fax: E-mail: chandana@nordicneurolab.com Establishment Registration Number: 3006738448

Contact:

Chandana Gurung Bhandari Møllendalsveien 65C N-5009 Bergen Norway

Direct: +47 95 20 37 87 Phone: +47 55 70 70 95 Fax: +47 55 70 70 96

E-mail: chandana@nordicneurolab.com

Legally marketed Device name and 510(k) number: fMRI Hardware System. K080515

Modified Device Name Device Common Name: Basis for Submission:

fMRI Hardware System Accessory to MRI System, Nuclear Magnetic Resonance Imaging Device Modification

Classification Regulation:

21 CFR 892.1000

Class: Panel: Product Code:

Radiology LNH

ll

Trade/Proprietary Name:

fMRI Hardware System

page 1 of 4

510(k) Summary Attachment 1

Special 510(k) fMRI Hardware System

OCT - 8 2009

:

{1}------------------------------------------------

Device Description

3.1. Intended Use

The fMRI Hardware System is a stimulus presentation and response collection system intended to be used by trained professionals to facilitate auditory and visual stimulation to be used in functional MR Imaging (fMRI) based on BOLD contrast.

This is the same intended use as previously cleared for the fMRI Hardware System, K080515.

3.2. System Description

The system presents auditory and visual stimulus to the patient gives feedback through a pair of handheld grips. A synchronization module synchronizes the nordicAktiva stimulus presentation software with the MR scanner. The System consists of five subsystems: AudioSystem, VisualSystem, ResponseGrip, SyncBox and nordicAktiva.

The intended use of the system is to support fMRI studies. fMRI stands for functional Magnetic Resonance Imaging. This technique is useful when determining certain diseases, gaining more information about a patient's condition or investigating cognitive functions. The technique is also used to patients suffering from a brain tumor in both the pre-operative stage by examining the area of the brain affected.

The System is used to present the stimulus necessary to provoke physiological processes in the brain. Visual [VisualSystem] and auditory [AudioSystem] stimulus and manual responses from the patient [ResponseGrips] are of primary interest. The timing of the visual and audio stimulation is critical to make sure that the correct MR image of the brains activity is linked to the stimulus presented. A synchronization unit [SyncBox], connected between the MR-scanner and the stimulus presentation software (nordicAktiva), is included in the system to make sure that the synchronization is correct.

page 2 of 4

510(k) Summary Attachment 1

Special 510(k) fMRI Hardware System

{2}------------------------------------------------

NordicNeuroLab AS Bergen, Norway

Image /page/2/Figure/1 description: The image shows a diagram of a system with several components. The components include a SyncBox, Nordic Aciva, Communication Console, ResponseGrip Interface, and an LCD screen. The diagram also shows the flow of audio and video signals between the components, as well as a trigger pulse from an MR scanner.

Figure 1 presents the complete configuration of the fMRI Hardware System. All signals entering or leaving the scanner room are received and transmitted by use of fiber optics. The system allows video and audio signals from the stimulus PC to enter the shielded scanner room and to be presented to the MR. The subject responds to the stimulus by using the handheld grips.

3.3. Sub-system components and description

3.3.1. VisualSystem

The VisualSystem allows video signals from the stimulus presentation PC to enter the shielded scanner room and to be presented to the patient trough a set of coil-mounted displays or an in-room LCD monitor.

3.3.2. AudioSystem

The AudioSystem allows auditory signals from the stimulus presentation PC to enter the scanner room and to be presented to the patient wearing a set of headphones. A communication console allows the operator to adjust the sound from the PC and to speak directly to the patient through a built-in microphone.

3.3.3. ResponseGrip

The purpose of this component is to collect patient responses during an fMRI study. The ResponseGrip consists of two hand-held grips with two buttons each. By pressing the patient can respond to the presented stimulus. The ResponseGrip is connected to an optical-electrical adapter which converts

Special 510(k) fMRI Hardware System

510(k) Summary

{3}------------------------------------------------

light to electrical signals. The electrical signal is fed to the Stimulus PC by using standard PC communication interfaces.

3.3.4. SyncBox

The SyncBox is connected directly to the MRI scanner where it receives timing pulses sent out with each image series and demodulates this signal before it's forwarded to the stimulus PC. In this way one can ensure that the stimulus presentation software is synchronized with the MRI image recordings.

3.3.5. nordicAktiva

nordicAktiva is a software that generates visual and auditory stimulus to the patient. The stimulus presentation is synchronized with the scanner through the SyncBox, and presented to the patient through the VisualSystem and AudioSystem. nordicAktiva records the responses fed to the Stimulus PC from the ResponseGrip, as well as the synchronization pulses from the SyncBox.

3.4. Identification of Change to Unmodified Device

The VisualSystem will be expanded with the option of using an in-room LCD monitor instead of the coilmounted display. The LCD monitor is placed outside magnets bore behind the MR scanner and is in no physical contact with the patient. The same video signal can be presented on the coil-mounted display and the in-room LCD-monitor. The operator has the choice to use either the coil mounted display or the in-room LCD monitor.

3.5. Statement of Substantial Equivalence

To summarize, the modified fMR! Hardware System is found substantial equivalent to the previously cleared device. The modified system only has a new device added to display video. The indications for use for the modified fMRI Hardware System have remained unchanged.

3.6. Summary of Testing

The fMRI Hardware System has been tested for function and safety and fulfills all requirement specifications.

page 4 of 4

510(k) Summary Attachment 1

Special 510(k) fMRI Hardware System

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with three curved lines stacked on top of each other. The lines are thicker at the bottom and become thinner as they rise. To the left of the lines, there is a partial word that appears to be vertically oriented. The logo is simple and abstract.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Mrs. Chandana Gurung Bhandari Quality Manager NordicNeuroLab Møllendalsveien, Bergen, Hordaland, N-5009 NORWAY

OCT - 8 2009

Re: K092253

Trade/Device Name: fMRI Hardware System Regulation Number: 21 CFR 892,1000 Regulation Name: Magnetic resonance diagnostic device . Regulatory Class: II Product Code: LNH Dated: September 9, 2009 Received: September 14, 2009

Dear Mrs. Bhandari:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{5}------------------------------------------------

Page 2 -

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

NordicNeuroLab AS Bergen, Norway

Indications for Use

510(k) Number (if known): K092253

Device Name: fMRI Hardware System

Indications for Use:

The fMRI Hardware System is a stimulus presentation and response collection system intended to be used by trained professionals to facilitate auditory and visual stimulation to be used in functional MR Imaging (fMRI) based on BOLD contrast.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)

510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Norm M. Khan

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device

Special 510(k) fMRI Hardware System

Indications for Use Attachment 2

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.