K Number
K091953
Device Name
PRO-PICC
Manufacturer
Date Cleared
2009-09-16

(77 days)

Product Code
Regulation Number
880.5970
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PRO-PICC® C catheter is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media and allows for central venous pressure monitoring when a 20gauge or larger lumen is used. For blood sampling, infusion or therapies use a 4F or larger catheter. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.

Device Description

The PRO-PICC catheter is an open-ended lumen catheter designed for power injection. The catheters are an extension of the Medcomp® PRO-LINE® CT Power Injectable CVC (K053345) and PRO-PICC® CT (K081904) catheter line. The PRO-PICCo CT catheter is comprised of a soft radiopaque polyurethane material. The lumen has a reverse taper design and is connected to the extensions via a soft pliable hub with suture wing for secure placement. Clamps are provided on the extension tubes to prevent and fluid communication. Female luer connectors provide the connection for intravenous administration. The power iniectable extension lines are purple in color to differentiate it from non-power injectable catheters. The extensions are also is printed with the words power injectable. The I.D. Ring within the clamp contains information regarding checking for blood return and flushing along with rate of infusion for power injection. The dual lumen catheter has one purple colored clamp and one natural colored clamp allowing units to designate power injection on one side and infusion; blood withdraw and pressure monitoring on the other although both lumens are capable of power injection.

The PRO-PICC catheter is available in additional French sizes and either a single or double lumen. The catheter has a usable length of 50cm to 60 cm depending on French size with depth markings in 5 cm increments. Stylet and adaptor sideport are provided to assist in catheter insertion.

The catheter is packaged sterile in a variety of tray configurations with the necessary accessories to facilitate catheter insertion.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Medcomp PRO-PICC CT device:

1. Table of Acceptance Criteria and Reported Device Performance:

The provided document describes a traditional 510(k) submission, which primarily focuses on demonstrating substantial equivalence to predicate devices rather than establishing novel performance acceptance criteria for a new device type. Therefore, explicit, quantitative, and standalone acceptance criteria in the typical sense (e.g., "sensitivity must be > X%") are not present.

Instead, the acceptance criteria are implicitly met by demonstrating that the PRO-PICC CT is substantially equivalent to established predicate devices in terms of:

Acceptance Criteria Category (Implicit)Reported Device Performance (Summary)
Intended UseSubstantially equivalent to predicate devices.
Anatomical LocationSubstantially equivalent to predicate devices.
Basic DesignSubstantially equivalent to predicate devices.
MaterialsOne material change, but overall biocompatibility tested and meets ISO 10993. Most materials are equivalent to predicate devices.
Performance (Functional)Functional performance is "substantially equivalent." Bench testing conducted based on internal engineering methods and applicable international standards.
LabelingSubstantially equivalent to predicate devices.
Manufacturing ProcessSubstantially equivalent to predicate devices.
Method of SterilizationSubstantially equivalent to predicate devices.
BiocompatibilityAll materials demonstrate compliance with ISO 10993.
Safety and EffectivenessNo new issues of safety and effectiveness are raised by the differences from predicate devices.

Important Note: The document explicitly states: "Performance standards for pressure injection have not been established by FDA under section 514 of the Federal Food, Drug and. Cosmetic Act. Testing is based upon internal engineering testing methods." This reinforces that quantitative, pre-defined FDA-mandated performance metrics for the specific function of "power injection" were not applicable as acceptance criteria in this submission.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not applicable. This submission relies on substantial equivalence and bench testing, not a clinical study with a "test set" of patients or data in the way an AI/ML device would.
  • Data Provenance: Not applicable. The "data" refers to engineering bench test results and biocompatibility testing of materials, not clinical data from specific geographic regions.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

  • Number of Experts: Not applicable. There is no "ground truth" establishment by external medical experts for a clinical test set in this 510(k) submission.
  • Qualifications of Experts: Not applicable. The evaluation is based on engineering testing and comparison to existing predicate devices.

4. Adjudication Method for the Test Set:

  • Adjudication Method: Not applicable. There is no "test set" requiring adjudication by multiple readers or experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • MRMC Study: No. This device is a physical catheter, not an AI/ML-driven diagnostic or assistive tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Standalone Performance: Not applicable. This device is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used:

  • Ground Truth: In the context of this 510(k), the "ground truth" for the device's acceptable performance is primarily derived from:
    • Predicate Device Performance: The established safety and effectiveness of the existing predicate devices (K053345, K081904, K070996).
    • Bench Test Results: Internal engineering testing demonstrating the device meets its design specifications and performs as intended, consistent with the predicates.
    • Biocompatibility Standards: Compliance with ISO 10993 for materials used.

8. The Sample Size for the Training Set:

  • Sample Size: Not applicable. This is not an AI/ML device that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Ground Truth Establishment: Not applicable. This is not an AI/ML device requiring a training set with established ground truth.

In summary, the provided document details a traditional 510(k) submission for a physical medical device (catheter) that demonstrates substantial equivalence to predicate devices through bench testing and material biocompatibility, rather than through clinical studies with defined test sets, expert-driven ground truth, or AI/ML performance metrics.

{0}------------------------------------------------

SEP 1 6 2009

Section 5510(k) SUMMARYTraditional 510K
A.Submitter Information:
Submitter:MEDCOMP®1499 Delp DriveHarleysville, PA 19438Tel: (215) 256-4201Fax: (215) 256-9191
Contact:Jean CallowRegulatory Specialist
Date Prepared:June 30, 2009
B.Trade Name:Medcomp® PRO-PICCCT
Common Name:Catheter, Intravascular, Therapeutic,Long-Term
Classification Name:Long Term Intravascular Catheter (80 LJS)
Regulation Name:Percutaneous, implanted, long-term intravascularcatheter
C.F.R. Section:880.5970
Class:II
C.Predicate Devices:K053345 Medcomp, PRO-LINE® CT PressureInjectable CVC, class II 880.5970K081904 Medcomp, PRO-PICC® CT, class II 880.5970K070996 Bard Access Systems, Inc., 4Fr SL PowerPICC®, class II 880.5970

Device Description:

The PRO-PICC catheter is an open-ended lumen catheter designed for power injection. The catheters are an extension of the Medcomp® PRO-LINE® CT Power Injectable CVC (K053345) and PRO-PICC® CT (K081904) catheter line. The PRO-PICCo CT catheter is comprised of a soft radiopaque polyurethane material. The lumen has a reverse taper design and is connected to the extensions via a soft pliable hub with suture wing for secure placement. Clamps are provided on the extension tubes to prevent and fluid communication. Female luer connectors provide the connection for intravenous administration. The power iniectable extension lines are purple in color to differentiate it from non-power injectable catheters. The extensions are also is printed with the words power injectable. The I.D. Ring within the clamp contains information regarding checking for blood return and flushing along with rate of infusion for power injection. The dual lumen catheter has one purple colored clamp and one natural colored clamp allowing units to designate power injection on one side and infusion; blood withdraw and pressure monitoring on the other although both lumens are capable of power injection.

The PRO-PICC catheter is available in additional French sizes and either a single or double lumen. The catheter has a usable length of 50cm to 60 cm depending on French size with depth

Medcomp PRO-PICC o catheter Summary

Page 1 of 2

{1}------------------------------------------------

markings in 5 cm increments. Stylet and adaptor sideport are provided to assist in catheter insertion.

Kos 1983

The catheter is packaged sterile in a variety of tray configurations with the necessary accessories to facilitate catheter insertion.

Indications for Use:

The PRO-PICC catheter is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media and allows for central venous pressure monitoring when a 20gauge or larger lumen is used. For blood sampling, infusion or therapies use a 4F or larger catheter. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.

Comparison to Predicate Devices:

The PRO-PICC catheters are substantially equivalent to the predicate devices in terms of intended use, anatomical location, basic design, most materials, performance, labeling, manufacturing process and method of sterilization.

The differences between the PRO-PICC and the predicate devices are one material change and expansion of product line with additional French sizes and the addition of single and double lumens.

Bench / Performance Data:

Performance testing of the proposed devices was conducted in accordance with applicable international standards and FDA guidance documents. Performance standards for pressure injection have not been established by FDA under section 514 of the Federal Food, Drug and. Cosmetic Act. Testing is based upon internal engineering testing methods.

The results of these tests in conjunction with the substantial equivalence claims effectively demonstrate the proposed devices are equivalent to the predicate devices.

Biocompatibility:

Testing for all materials used for the PRO-PICC o has been submitted in previously cleared Medcomp devices. All biocompatibility testing demonstrates the materials used meet the requirements of ISO 10993.

Technological Characteristics:

Technological similarities between the proposed devices and predicate devices remain the same.

Summary of Substantial Equivalence:

The proposed devices meet the performance criteria of design verification as specified by ISO standards, guidance documents and internal test protocols. The proposed device has the same intended use, operation and function as the predicates. There are no differences that raise new issues of safety and effectiveness. The proposed devices are substantially equivalent to the legally marketed predicate devices.

Medcomp PRO-PICC o catheter Summary

Page 2 of 2

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center is an emblem featuring a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus has a staff with a snake winding around it, and a pair of wings at the top.

Food and Drug Administration 10903 New Hampshire Avenne Document Mail Conter - WO66-G609 Silver Spring, MD 20993-0002

Ms. Jean Callow Regulatory Specialist Medcomp 1499 Delp Drive Harleysville, Pennsylvania 19438

Re: K091953

Trade/Device Name: PRO-PICCo CT Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term intravascular Catheter Regulatory Class: II Product Code: LJS Dated: June 29, 2009 Received: July 13, 2009

SEP 1 6 2009

Dear Ms. Callow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Callow

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH

/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony V. Watson br

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

PRO-PICC ® CT Device Name: ------------

Indications for Use:

The PRO-PICC® C catheter is indicated for short term or long term peripheral access to the central venous system for intravenous therapy and power injection of contrast media and allows for central venous pressure monitoring when a 20gauge or larger lumen is used. For blood sampling, infusion or therapies use a 4F or larger catheter. The maximum recommended infusion rate varies by catheter French size and is printed on the catheter.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE),

Charles Zeller
(Division Sign-Off)

Division of Anesthesiology, General Hospital Infection Control, Dental Devices

Page 1 of 1

1841

510(k) Number: K091953

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”