K Number
K091723
Date Cleared
2009-07-06

(25 days)

Product Code
Regulation Number
890.3860
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the Freedom Sciences Version 1.5 Motion Control Module is to allow for remote motion control of a powered wheelchair that uses the Invacare MK6 Electronics while the wheelchair is unoccupied. It is not intended for use when a person is seated in the wheelchair.

The MCM Version 1.5 is compatible with powered wheelchairs that use Invacare MK6 Electronics.

Device Description

The Freedom Sciences Version 1.5 Motion Control Module for powered wheelchairs is a wireless, remote control product designed for use with powered wheelchairs which use the Invacare MK6 Electronics . Its intended function and use is to allow for remote motion control of the powered wheelchair only while the wheelchair is unoccupied. It is not intended for use when a person is seated in the wheelchair.

The MCM v1.5 allows for remote motion control using high level motion control commands. It interfaces to the power wheelchair using the host wheelchair attendant control interface on the motor controller. This method of interfacing with the powered wheelchair retains all configured safety interlocks inherent to the host motor controller itself and is the standard means for integrating auxiliary input devices.

AI/ML Overview

The provided text describes a Special 510(k) Premarket Notification for the "Freedom Sciences LLC Version 1.5 Motion Control Module (MCM)". This device is intended for remote motion control of unoccupied powered wheelchairs that use Invacare MK6 Electronics.

However, the document is a premarket notification for a medical device and not a study report. It does not contain the detailed information necessary to answer all the questions regarding acceptance criteria and performance studies in the requested format for an AI/device performance study.

Here's an attempt to answer the questions based on the limited information provided in the document:

1. Table of acceptance criteria and the reported device performance

Acceptance Criteria (Inferred)Reported Device Performance
Functional Performance: - Ability to provide remote motion control of a powered wheelchair. - Compatibility with Invacare MK6 Electronics. - Retention of host motor controller's safety interlocks. - Operation only when the wheelchair is unoccupied.The MCM v1.5 allows for remote motion control using high-level motion control commands. It interfaces to the power wheelchair using the host wheelchair attendant control interface on the motor controller. This method of interfacing with the powered wheelchair retains all configured safety interlocks inherent to the host motor controller itself and is the standard means for integrating auxiliary input devices. "The Freedom Sciences MCM v1.5 met the applicable required performance criteria and functioned as intended." "It is not intended for use when a person is seated in the wheelchair."
Safety: - Adherence to safety standards. - Prevention of use when a person is seated.The device's interface method retains all configured safety interlocks inherent to the host motor controller. Its intended use explicitly states, "It is not intended for use when a person is seated in the wheelchair." The study mentioned was "as required by FDA's July 26, 1995, draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles"". This guidance would implicitly cover safety aspects.
Substantial Equivalence: - Equivalence to a predicate device (Freedom Sciences Motion Control Module for Powered Wheelchairs, K073330).The Freedom Sciences MCM v1.5 Motion Control Module is "substantially equivalent to the Freedom Sciences Motion Control Module (MCM) for powered wheelchairs (K073330)."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "test set" sample size or data provenance in the context of an AI/algorithm performance study. The testing refers to the "MCM v1.5 was tested as required by FDA's July 26, 1995, draft publication entitled 'Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles'". This guidance would relate to engineering and functional testing of the physical device, not an AI model requiring a test dataset.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and not provided. The device is a motion control module for a wheelchair, not an AI diagnostic or analytical tool that requires expert-established ground truth for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable and not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable and not provided. This device is not an AI assistance tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device itself is a "standalone" system in the sense that it operates independently to control the wheelchair remotely. However, it's not an "algorithm only" in the context of typical AI performance studies. Its performance is about the module's ability to transmit commands and the wheelchair's response, which is a functional test. The document states it allows for "remote motion control using high level motion control commands," implying the module's algorithms are responsible for interpreting and executing these commands. The testing "met the applicable required performance criteria and functioned as intended." This confirms its standalone functional performance was evaluated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance would be objective measurements of its functional capabilities (e.g., successful remote control of movement, adherence to safety interlocks, reliability of communication) against engineering specifications and regulatory guidance for powered wheelchairs. It is not based on expert consensus on diagnoses, pathology, or outcomes data.

8. The sample size for the training set

Not applicable and not provided. This is not an AI model that undergoes "training" with a dataset in the conventional sense. The "training" would be more akin to software development and hardware testing.

9. How the ground truth for the training set was established

Not applicable and not provided.

{0}------------------------------------------------

FreedomSciences
Special 510(k) Premarket Notification

JUL - 6 2009

Version 1.5 Motion Control Module

1691723

510(k) SUMMARY

Freedom Sciences LLC Version 1.5 Motion Control Module

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared.

Freedom Sciences, LLC The Navy Yard - Quarters M2 4601 South Broad Street Philadelphia, PA 19112

Contact Person: Edward A. Kroll Representative Consultant for Freedom Sciences, LLC

Date Prepared: June 8, 2009

Name of Device and Name/Address of Sponsor

Freedom Sciences Version Motion 1.5 Control Module (MCM)

Freedom Sciences, LLC The Navy Yard - Quarters M2 4601 South Broad Street Philadelphia, PA 19112

Common or Usual Name

Power Wheelchair

Classification Name

Wheelchair, Powered

Predicate Devices

Freedom Sciences LLC Motion Control Module for Powered Wheelchairs (K073330)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black and white drawing of a cluster of grapes. The grapes are tightly packed together and appear to be ripe. The stem of the grapes is visible at the top of the cluster. The drawing is simple and realistic.

FreedomScience

Special 510(k) Premarket Notification

Version 1.5 Motion Control Module

Intended Use

The intended use of the Freedom Sciences Version 1.5 Motion Control Module is to allow for remote motion control of a powered wheelchair that uses the Invacare MK6 Electronics while the wheelchair is unoccupied. It is not intended for use when a person is seated in the wheelchair.

Technological Characteristics and Substantial Equivalence

A. Device Description

The Freedom Sciences Version 1.5 Motion Control Module for powered wheelchairs is a wireless, remote control product designed for use with powered wheelchairs which use the Invacare MK6 Electronics . Its intended function and use is to allow for remote motion control of the powered wheelchair only while the wheelchair is unoccupied. It is not intended for use when a person is seated in the wheelchair.

The MCM v1.5 allows for remote motion control using high level motion control commands. It interfaces to the power wheelchair using the host wheelchair attendant control interface on the motor controller. This method of interfacing with the powered wheelchair retains all configured safety interlocks inherent to the host motor controller itself and is the standard means for integrating auxiliary input devices.

B. Substantial Equivalence

The Freedom Sciences MCM v1.5 Motion Control Module is substantially equivalent to the Freedom Sciences Motion Control Module (MCM) for powered wheelchairs (K073330).

Performance Data

The MCM v1.5 was tested as required by FDA's July 26, 1995, draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three- Wheeled Vehicles". The Freedom Sciences MCM v1.5 met the applicable required performance criteria and functioned as intended.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 6 2009

Freedom Sciences, LLC % Spectre Solutions, Incorporation Mr. Edward A. Kroll President 5905 Fawn Lane Cleveland, Ohio 44141

Re: K091723

Trade/Device Name: Freedom Sciences, LLC Version 1,5. Motion Control

Module (MCM)

Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: II Product Code: ITI Dated: June 8, 2009 Received: June 11, 2009

Dear Mr. Kroll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration

{3}------------------------------------------------

Page 2-Mr. Edward A. Kroll

and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

for Dr. Doodr x/clor

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): TBD

Device Name: Freedom Sciences, LLC Version 1.5 Motion Control Module (MCM)

Indications for Use:

The intended use of the Freedom Sciences version 1.5 Motion Control Module is to allow for remote motion control of a powered wheelchairs while the wheelchair is unoccupied. It is not intended for use when a person is seated in the wheelchair.

The MCM Version 1.5 is compatible with powered wheelchairs that use Invacare MK6 Electronics.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091723

Page 1 of 1

§ 890.3860 Powered wheelchair.

(a)
Identification. A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class II (performance standards).