K Number
K091588
Date Cleared
2009-09-04

(94 days)

Product Code
Regulation Number
862.3100
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The UCP Home Drug Screening Test Cards and UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the following drugs and their metabolites in human urine: Amphetamine, Barbiturates, Benzodiazepines, Cocaine, Marijuana, Methadone, Methamphetamine, MDMA, Morphine, Opiate 2000, Oxycodone, Phencyclidine, Tricyclic Antidepressant. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users. The tests only provide a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method for most drugs (HPLC is the preferred confirmatory method for Tri-cyclic Antidepressants). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

Device Description

UCP Home Drug Screening Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates. Benzodiapines. Cocaines. Marijuana. Methamphetamine. MDMA. Opiates, Morphine, Oxycodone, Phencyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument. The tests contain two formats: 1) Test Card; 2) Test Cup. The test configuration comes with single drug screening test or any combinations of multiple drug screening tests.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text, structured as requested:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are implicitly tied to the agreement rate with confirmatory laboratory methods, as the device provides preliminary results. The reported performance focuses on the ability of lay users to obtain results that agree with these reference methods.

Drug TypeCut-off Level (ng/mL)Acceptance Criteria (Implicit)Reported Device Performance (Agreement Rate with GC/MS or HPLC)
Marijuana50High agreement with reference96.7% or above
Cocaine300High agreement with reference96.7% or above
Amphetamine1000High agreement with reference96.7% or above
Methamphetamine1000High agreement with reference96.7% or above
Opiates2000High agreement with reference96.7% or above
Morphine300High agreement with reference96.7% or above
Phencyclidine25High agreement with reference96.7% or above
Barbiturates300High agreement with reference96.7% or above
Benzodiazepines300High agreement with reference96.7% or above
Methadone300High agreement with reference96.7% or above
Oxycodone100High agreement with reference96.7% or above
MDMA500High agreement with reference96.7% or above
Tricyclic Antidepressants1000 pg/mLHigh agreement with reference96.7% or above

Note: The text states "96.7% or above agreement rate" for all tests, implying this is the acceptance level for lay user performance against the confirmed drug concentrations.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set (Consumer Study):
    • Test Cards: 115 lay persons (58 females, 57 males)
    • Test Cups: 110 lay persons (55 females, 55 males)
    • The total number of urine samples tested is not explicitly stated in terms of unique samples per drug, but rather concentrations were prepared in samples. For each participant, various prepared urine samples were likely used.
  • Data Provenance: The studies were conducted in three geographic regions: Texas, Pennsylvania, and California. The study appears to be prospective as lay persons were recruited to perform the tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for the prepared urine samples was established by laboratory methods, specifically:

  • GC/MS (Gas Chromatography/Mass Spectrometry): Used for confirming final drug concentrations in most urine samples.
  • HPLC (High-Performance Liquid Chromatography): Used for confirming TCA (Tricyclic Antidepressant) concentrations in urine samples.

The text does not mention the use of human experts (e.g., toxicologists, lab technicians) in establishing the ground truth beyond the "confirmation by GC/MS" statement. It implies the methods themselves are the gold standard for confirmation. Therefore, the number of experts is not specified as being directly involved in the establishment of ground truth, but rather the accredited laboratory procedures.

4. Adjudication Method for the Test Set

The text does not describe an adjudication method involving multiple human readers comparing their interpretations. Instead, the "test results performed by the lay users" were compared to the "GC/MS results" (or HPLC for TCA). This implies a direct comparison rather than an adjudication process among different readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study was not done. This device is a rapid, qualitative immunochromatographic assay intended for direct use by consumers (lay persons) and healthcare professionals, not an AI-assisted diagnostic tool.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

No, a standalone algorithm performance study was not done. This device is a physical test kit that requires human interaction (collecting urine, performing the test, and visually interpreting the results). It does not involve an algorithm separate from human-in-the-loop performance. The "standalone" aspect in this context would be the test kit's inherent chemical reaction, which is then interpreted by a human. The consumer studies directly evaluate human users' ability to correctly interpret these reactions.

7. The Type of Ground Truth Used

The type of ground truth used for the test set was laboratory confirmation using established analytical methods:

  • GC/MS (Gas Chromatography/Mass Spectrometry) for most drugs.
  • HPLC (High-Performance Liquid Chromatography) for Tricyclic Antidepressants.

This is considered a highly objective and quantitative gold standard for drug concentration.

8. The Sample Size for the Training Set

The document does not specify a sample size for a training set. This is not an AI/machine learning device that typically involves distinct training and testing sets. The studies described are performance evaluation studies for a diagnostic device.

9. How the Ground Truth for the Training Set Was Established

Since there is no mention of a "training set" in the context of an algorithm, this question is not applicable to the information provided. The performance studies used urine samples with known concentrations established by GC/MS or HPLC, effectively serving as the reference for evaluating user accuracy.

{0}------------------------------------------------

UCP 510K Submission (K091588) UCP Home Drug Screening Test

10. 510(K) SUMMARY

SEP 0 4 2009

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92.

The Assigned 510(k) number is K091588

Submitter:

UCP Biosciences, Inc 1445 Koll Circle, Ste 111 San Jose, CA 95014 Tel: 408-392-0064 Fax: 408-392-0163

Date:

May 28, 2009

Contact Person:

Dr. Nancy Chen

Trade Name:

UCP Home"M Drug Screening Test Card UCP Home™ Drug Screening Test Cup

Common Name:

Amphetamine Test System Methamphetamine Test System Cocaine Test System Barbiturate Test System Benzodiazepine Test System Methamphetamine Test System (MDMA) Opiates Test System Methadone Test System Opiates Test System (Oxycodone) Phencyclidine Test (unclassified) Cannabinoid Test System Tricyclic Antidepressant Test System

Product Code:

DKZ, LAF, DIO, DIS, JXM, DJC, DMB, DJG, LCM, LDJ, LFG

Regulation Section:

CFR 21 § 862.3650 CFR 21 § 862.3100

  • 1 -

{1}------------------------------------------------

UCP 510K Submission (K091588) UCP Home Drug Screening Test CFR 21 § 862.3100 CFR 21 § 862.3250 CFR 21 § 862.3150 CFR 21 § 862.3610 CFR 21 § 862.3620 CFR 21 § 862.3250

CFR 21 § 862.3250 CFR 21 § 862.3100 CFR 21 § 862.3870 CFR 21 § 862.3910

Panel: Toxicology (91)

Device Classification: II

Substantially Equivalent Devices:

UCP Rapid Drug Screening Tests UCP Rapid Drug Screening Tricyclic Antidepressants Tests UCP Multiple Drug Screen Test Cups

Product Description:

UCP Home Drug Screening Tests are competitive binding, lateral flow immunochromatographic assays for qualitatively the detection of Amphetamine, Barbiturates. Benzodiapines. Cocaines. Marijuana. Methamphetamine. MDMA. Opiates, Morphine, Oxycodone, Phencyclic Antidepressants and their metabolites at the cut-off levels as indicated. The tests can be performed without the use of an instrument.

Intended Use:

2. UCP Home™ Drug Screening Tests:

UCP Home "10 Drug Screening Tests are rapid, qualitative, competitive binding immunoassays for qualitatively the detection of the following drugs and their metabolites in human urine at the following cut-off concentrations:

Marjuana:50 ng/mL
Cocaine:300 ng/mL
Amphetamine:1000 ng/mL
Methamphetamine:1000 ng/mL
Opiates:2000 ng/mL

{2}------------------------------------------------

UCP 510K Submission (K091588) UCP Home Drug Screening Test

Morphine:300 ng/mL
Phencyclidine:25 ng/mL
Barbiturares:300 ng/mL
Benzodiazepines:300 ng/mL
Methadone:300 ng/mL
Oxycodone:100 ng/mL
MDMA:500 ng/mL
Tricyclic Antidepressants:1000 ng/mL

The tests are intended for over-the-counter (OTC) consumer use as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory test in the second step of the two-step process, is provided in the package labeling.

The tests will yield preliminary positive results when the prescription drugs Barbiturates, Benzodiazepines, Oxycodone, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Oxycodone, Tricyclic Antidepressant in urine. The tests provide only preliminary test results, which should be confirmed by other methods such as gas chromatography/mass spectrometry (GC/MS). Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring the drug levels.

For Over-The-Counter (OTC) use For In Vitro Diagnostics only

Comparison to Predicate Devices:

When compared to the predicates, UCP Home Drug Screening Tests provide the qualitative determination of the same drugs in the same matrix, and utilizes the same cutoff concentrations. Both tests are immunochromatographic, lateral flow assays for the qualitative detection of drugs with visual, qualitative end results. Both tests are intended to provide preliminary analytical test results. The predicates are intended for professional use only, whereas UCP Home Drug Screening Tests are intended for OTC consumers use.

Safety and Effectiveness Data:

Accuracy

  • 3 -

{3}------------------------------------------------

UCP 510K Submission (K091588) UCP Home Drug Screening Test

A clinical comparison study for UCP Home Drug Screening Tests can be found in 510k submissions (K050540, K061457, K072062).

Consumer Studies

The consumer study for UCP Home Drug Screening Test Cards was conducted among 115 lay persons in three geographic regions: Texas, Pennsylvania and California. Fifty eight females and fifty seven males from ages between 18 and 75 years have participated the studies. Fifty eight participants had high school education or less, fifty seven participants had finished college courses. None of the participants had experiences using drug testing products before. A separately consumer study for UCP Home Drug Screening Test Cups was conducted among 110 lay persons in three geographic regions: Texas, Pennsylvania and California. Fifty five females and fifty five males from ages between 18 and 77 years participated the study. Fifty five participants had high school education or less, fifty five participants had finished college courses. None of the participants had experiences using drug testing products before. The urine samples were prepared to contain strong negative (0% of cutoff), a very weak negative (50% of cutoff), a weak negative (75% of cutoff), a very weak positive (125% of cutoff), a weak positive (150% of cutoff) and high positive (300% of Cutoff), The urine samples with various drug concentrations were prepared by spiking pure drugs or drug metabolites into drug free human urine, the final drug concentrations in each urine sample were confirmed by GC/MS but TCA, TCA concentrations in the urine samples was confirmed by HPLC. The test results performed by the lay users showed 96.7% or above agreement rate with GC/MS results and indicate the lay users can perform UCP Home Drug Screening Test Card or Cup satisfactorily by following the test instruction. The post-study survey was conducted to determine if the lay users can understand the test instruction, the meaning of the test results and how to interpret the test results. Consumers were asked 9 questions including whether the test was easy to run, the results was easy to read, how to interpret the test results, importance of confirmatory test and some medicines and foods may affect the test results. Participant responses support that the lay users can understand how to run the test, interpret the test results, the importance of confirmatory test, and some issues concerning certain prescription medicines and foods may affect the test results.

Other Information about Performance Characteristics:

The performance characteristics of UCP Home Drug Screening Tests including the precision study, sensitivity study, specificity and cross reactivity study, interference study and stability study can be found in 510k submissions (K050540, K061457, K072062)

Conclusion:

{4}------------------------------------------------

UCP Home Drug Screening Test

UCP Home Drug Screening Test Cards and UCP Home Drug Screening Test Cups are substantially equivalent to UCP Rapid Drug Screening Tests, UCP Rapid Drug Screening Tricyclic Antidepressants, UCP Multiple Drug Screen Test Cups. .

  • 5 -

{5}------------------------------------------------

Image /page/5/Picture/12 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is partially visible along the left side of the logo.


SEP 0 4 2009

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

UCP Biosciences Inc. c/o Nancy Chen 1445 Koll Circle Suite 111 San Jose, CA 95112

Re: K091588

Trade Name: UCP Home Drug Screening Test Cards and UCP Home Drug Screening Test Cups Regulation Number: 21 CFR 862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Codes: DKZ, DIS, JXM, DIO, DJC, DMB, DJG, LCM, LDJ, LFG Dated: August 13, 2009 Received: August 13, 2009

Dear Ms. Chen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{6}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

G.C.H

Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known): K091588

Device Name: UCP Home Drug Screening Test Cards and UCP Home Drug Screening Test Cups

Indications For Use:

The UCP Home Drug Screening Test Cards and UCP Home Drug Screening Test Cups are rapid, qualitative, competitive binding immunoassays for the following drugs and their metabolites in human urine:

TestCalibratorCut-off
AmphetamineD-Amphetamine1000 ng/mL
BarbituratesSecobarbital300 ng/mL
BenzodiazepinesOxazepam300 ng/mL
CocaineBenzoylecgonine300 ng/mL
MarijuanaDelta-9-THC-COOH50 ng/mL
MethadoneMethadone300 ng/mL
MethamphetamineD-Methamphetamine1000 ng/mL
MDMAMDMA500 ng/mL
MorphineMorphine300 ng/mL
Opiate 2000Morphine2000 ng/mL
OxycodoneOxycodone100 ng/mL
PhencyclidinePhencyclidine25 ng/mL
Tricyclic AntidepressantNortriptyline1000 pg/mL

The tests contain two formats: 1) Test Card; 2) Test Cup. · The test configuration comes with single drug screening test or any combinations of multiple drug screening tests. The test is intended for over-the-counter (OTC) users as the first step in a two step process to provide consumers, including but not limited to concerned parents, with information concerning the presence or absence of the above stated drugs or their metabolites in a urine sample. Information regarding confirmatory testing - the second step in the process, along with the materials for shipping the urine specimen to the laboratory, is provided. The test is also intended for health care professional users.

The tests will yield preliminary positive results when the prescription drugs Barbiturates. Benzodiazepines, Oxycodone, Tricyclic Antidepressants are ingested, even at or above therapeutic doses. There are no uniformly recognized drug levels for Barbiturate, Benzodiazepines, Oxycodone, Tricyclic Antidepressant in urine. The tests only provide a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method for most drugs (HPLC is the preferred confirmatory method for Tri-cyclic Antidepressants). Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated. The tests are not intended to be used in monitoring drug levels.

Prescription Ues × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use × (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Division Sign-Off
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Office of In Vitro Diagnostic
Device Evaluation and Safety
510(k)K091588

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).