K Number
K091452
Date Cleared
2009-08-21

(95 days)

Product Code
Regulation Number
870.2700
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FORE-SIGHT® Cerebral Oximeter, Model MC-2000 Series is indicated for the continuous noninvasive monitoring of regional hemoglobin oxygen saturation of blood in the brain (SctO2). It is intended for use in any individual at risk for reducedflow or no-flow ischemic states.

When used with FORE-SIGHT large sensors, the FORE-SIGHT MC-2000 Cerebral Oximeter Monitor is indicated for use with adults and children over 40Kg. When used with the FORE-SIGHT medium sensors, the FORE-SIGHT MC-2000 Cerebral Oximeter is indicated for use with small adults and children between 4 kg and 80 kg. When used with FORE-SIGHT small sensors the FORE-SIGHT MC-2000 Series Cerebral Oximeter Monitor is indicated for infants and neonates ≤ 8Kg.

Device Description

The Cerebral Oximeter Monitor measures cerebral tissue oxygen saturation allowing the clinician to accurately determine absolute levels of brain tissue blood oxygen saturation and brain venous oxygen saturation in the brain. This measurement can be of significant value in numerous acute care (OR ICU, ER) situations, providing health care professionals with information to guard against neurological injuries due to compromised brain oxygenation, which can occur during many surgical and clinical procedures.

The Cerebral Oximeter Monitor consists of an optical transducer containing a laser light source and photodiode detectors, and a graphic display monitor with user interface. The non-invasive, reflection mode, optical transducer is placed on the forehead of the subject via a disposable sensor attachment to determine cerebral oxygenation. The Cerebral Oximeter Monitor is safe to use, because it is designed to operate as a Class I laser product, the safest FDA laser classification. Additional safety features include a laser interlock system designed to prevent laser operation in case the optical transducer is not securely attached to the subject. A patent-protected algorithm optimizes accuracy of the device for measurements of absolute cerebral tissue oxygen saturation.

AI/ML Overview

Here's a summary of the acceptance criteria and the study details for the CASMED FORE-SIGHT® Cerebral Oximeter Monitor, Model MC-2000, based on the provided 510(k) summary:

Acceptance Criteria and Reported Device Performance

The acceptance criterion for the device's performance is not explicitly stated as a target value in the document. Instead, the "Precision (1 Standard deviation)" is reported for different patient populations. This precision value reflects the agreement between the device's SctO2 measurement and the reference SctO2 derived from co-oximetry of blood samples.

Acceptance Criteria (Implicit)Reported Device Performance (Precision/1 Std Dev)
Precision of SctO2 for Adult subjects± 3.7 %
Precision of SctO2 for Pediatric subjects± 4.86 %
Precision of SctO2 for Infant & Neonate subjects± 5.0 %

Study Details

  1. Sample sizes used for the test set and the data provenance:

    • Adult Subject Validation: Healthy adult volunteers. The exact number of subjects is not provided. Data collected at Duke University Medical Center in Durham, North Carolina. Prospective study design.
    • Infant & Neonate Subject Validation: 2044 hours of clinical data collected from subjects undergoing venous Extracorporeal Membrane Oxygenation (VV-ECMO) with cephalad catheterization. The exact number of subjects is not provided, but the duration of data collection suggests a significant number of data points. Data collected at Children's National Medical Center in Washington, DC, and the Children's Hospital of Atlanta (CHOA), Emory University, Atlanta, GA. Prospective study design.
    • Pediatric Subject Validation: Subjects undergoing cardiac catheterization. The exact number of subjects is not provided. Data collected at Boston Children's Hospital in Boston, MA. Prospective study design.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not mention the use of experts to establish ground truth. The ground truth was established through direct physiological measurements from blood samples analyzed by co-oximetry.
  3. Adjudication method for the test set:

    • Not applicable as the ground truth was based on objective physiological measurements (co-oximetry of blood samples) rather than expert interpretation.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not performed. This device is a diagnostic/monitoring device, not an AI-assisted interpretation tool for human readers.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, the performance reported (Precision of SctO2) is the standalone performance of the algorithm/device. The device directly measures and calculates SctO2, and its output is compared against a reference standard.
  6. The type of ground truth used:

    • Physiological Measurement/Co-Oximetry of Blood Samples:
      • For adults and pediatric subjects: Reference SctO2 derived from co-oximetry of arterial (SaO2) and jugular bulb (SjvO2) blood samples.
      • For infant & neonate subjects: Reference SctO2 derived from pulse oximetry measured arterial oxygen saturation (SaO2) and co-oximetry measured internal jugular vein venous oxygen saturation (SjvO2) from blood samples.
  7. The sample size for the training set:

    • The document does not explicitly mention a separate "training set" or its size. In the context of medical devices like oximeters, the algorithm is often developed and refined using a dataset that precedes the formal validation studies presented here. The provided clinical studies are validation studies demonstrating the device's performance against a reference standard. If algorithm development involved clinical data, it's not detailed as a distinct training set.
  8. How the ground truth for the training set was established:

    • As no separate "training set" is explicitly described, the method for establishing ground truth for any potential internal development/training data is not detailed in this document. However, it can be inferred that any data used for algorithm development would likely rely on similar physiological measurement techniques as those used for validation.

{0}------------------------------------------------

K091452

:

AUG 21 2009

.

Image /page/0/Picture/2 description: The image shows the logo for CASMED. The logo is in black and white, with the word "CASMED" in large, bold letters. Below the word "CASMED" is the phrase "FOR WHAT'S VITAL" in smaller letters. There is a registered trademark symbol to the right of the word CASMED.

1 /

!

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Submitter:CAS Medical Systems, Inc.
Address:44 East Industrial Rd. Branford CT. 06405 USA
Contact:Ron Jeffrey - Director, Regulatory Affairs Phone - (203) 488-6056Fax - (203) 488-9438Email - rjeffrey@casmed.com
Prepared:May 15, 2009
Trade Name:FORE-SIGHT® Cerebral Oximeter Monitor
Common Name:Model MC-2000 Series
Classification Name:Cerebral Oximeter (870.2700)

{1}------------------------------------------------

EQUIVALENCE (Predicate Device)

The FORE-SIGHT® Cerebral Oximeter Monitor, Model MC-2000 is equivalent to the following devices:

DESCRIPTION

The Cerebral Oximeter Monitor measures cerebral tissue oxygen saturation allowing the clinician to accurately determine absolute levels of brain tissue blood oxygen saturation and brain venous oxygen saturation in the brain. This measurement can be of significant value in numerous acute care (OR ICU, ER) situations, providing health care professionals with information to guard against neurological injuries due to compromised brain oxygenation, which can occur during many surgical and clinical procedures.

The Cerebral Oximeter Monitor consists of an optical transducer containing a laser light source and photodiode detectors, and a graphic display monitor with user interface. The non-invasive, reflection mode, optical transducer is placed on the forehead of the subject via a disposable sensor attachment to determine cerebral oxygenation. The Cerebral Oximeter Monitor is safe to use, because it is designed to operate as a Class I laser product, the safest FDA laser classification. Additional safety features include a laser interlock system designed to prevent laser operation in case the optical transducer is not securely attached to the subject. A patent-protected algorithm optimizes accuracy of the device for measurements of absolute cerebral tissue oxygen saturation ..

Cerebral Oximeter Monitor Intended Use

The FORE-SIGHT® Cerebral Oximeter, Model MC-2000 Series is indicated for the continuous noninvasive monitoring of regional hemoglobin oxygen saturation of blood in the brain (SctO2). It is intended for use in any individual at risk for reduced-flow or no-flow ischemic states.

When used with FORE-SIGHT large sensors, the FORE-SIGHT MC-2000 Cerebral Oximeter Monitor is indicated for use with adults and children over 40Kg. When used with the FORE-SIGHT medium sensors, the FORE-SIGHT MC-2000 Cerebral Oximeter is indicated for use with small adults and children between 4 kg and 80 kg. When used with FORE-SIGHT small sensors the FORE-SIGHT MC-2000 Series Cerebral Oximeter Monitor is indicated for infants and neonates ≤ 8Kg.

{2}------------------------------------------------

Cerebral Oximeter Monitor Technology Compared to Predicate Devices

The FORE-SIGHT Cerebral Oximeter Monitor compares substantially to one or more of the cited predicate devices in that they use fundamentally the same optical operating principle, called diffuse reflectance spectroscopy. All cited monitors use light to probe a cross-section tissue microvasculature. (mixed bed of arterioles, capillaries and venules). The Cerebral Oximeter Monitor and predicate devices analyze light returning from tissue, after having passed through tissues, for hemoglobin in its oxygenated and deoxygenated forms in the optically sampled region. All cited monitors calculate oxygen saturation. This value reflects the percentage of oxygenated hemoglobin in the sampled tissue.

Non-Clinical Performance Testing to Demonstrate Substantial Equivalence

The Cerebral Oximeter Monitor has been tested to the following standards in accordance with CAS Medical Systems Product Performance Specifications. The following non-clinical tests have been performed:

  • UL 60601-1 Safety testing for use of the UL Classified mark; .
  • CAN/CSA C22.2 No. 601.1-M90
  • IEC 60601-1 Safety of Medical Electrical Equipment; ●
  • EN 60601-1 Safety of Medical Electrical Equipment; ◆
  • ◆ IEC 60601-1-1 Safety of Medical Electrical Systems:
  • IEC 60601-1-2: 2001 Safety of Medical Electrical Equipment with regard to EMC Emissions � and EMC Immunity;
  • IEC 60601-1-4 Safety of Programmable Electrical Medical Systems; ●
  • IEC 60601-1-8 Safety of Alarm Systems for Medical Equipment/Systems; .
  • IEC 60825-1: Safety of Laser Products (with amendments A1 and A2); .

In addition to the above laboratory tests, CAS has conducted a full compliment of individual hardware, software and systems monitor and sensor verification and validation studies.

Clinical Testing to Show Substantial Equivalence

Adult Subject Validation: Clinical data on adult subjects was collected at the Duke University Medical Center in Durham, North Carolina. In this study, healthy adult volunteers were subjects for comparison using an internal jugular bulb catheter on the subject's right side and a radial arterial line on the left. Two sensors from the FORE-SIGHT Cerebral Oximeter Monitor were placed bilaterally on the patient's forehead. Hypoxic mixtures of gas were delivered and data was collected in 5 minute intervals during periods of ascending and descending concentrations. At each data collection point, blood samples were drawn simultaneously from the jugular bulb and the radial arterial catheters and analyzed for hemoglobin oxygen saturation using a co-oximeter. The patient was monitored and the protocol stopped if SpO2 values from a pulse oximeter reached 70%.

Infant & Neonate Subject Validation: 2044 hours of clinical data was collected at the Children's National Medical Center in Washington, DC, and the Children's Hospital of Atlanta (CHOA), Emory University, Atlanta, GA, from subjects undergoing venous Extracorporeal Membrane Oxygenation (VV-ECMO) with cephalad catheterization. In this study, cerebral venous oxygen saturation (SjvO2) measured from blood samples obtained from the internal jugular vein via the cephalad catheter, along with pulse oximetry arterial oxygen saturation (SaO2) data, were recorded from VV-ECMO neonates without alteration to patient care or blood oxygenation levels while being monitored by the FORE-SIGHT Cerebral Oximeter Monitor over a period of several days for each subject.

{3}------------------------------------------------

Pediatric Subject Validation: Clinical data on pediatric subjects were collected at Boston Children's Hospital in Boston, MA from subjects undergoing cardiac catheterization. In this study, cereball vein along with at with a to the cared from blood samples obtained from the internal jugular vein, along with arterial oxygen saturation (SaO)) data, were recorded from cardiac catherization subjects without alteration to patient care while being monitored by the FORE-SIGHT Cerebral Oximeter Monitor

Conclusions Drawn from Clinical and Non-Clinical Testing

The data is presented as Precision measured as one standard deviation for the cerebral tissue oxygen saturation (SctO2) parameter to determine the accuracy of the monitor.

Adult SctO: Using the FORE-SIGHT Large sensor, the Cerebral Oximeter SctO₂ showed a strong correlation with the reference SctO2 over the spectrum of values between 45 to 95%. The Precision (1 Standard deviation) for the Cerebral Oximeter Monitor SctO2 compared to reference SctO2 derived from co-oximetry of arterial (SaQ) and jugular bulb (SjvO2) blood samples was ± 3.7 %, based on Equation 1 below.

Pediatric SctO₂: Using the FORE-SIGHT Medium sensor, the Cérebral Oximeter SctO₂ showed a strong correlation with the reference SctO2 over the spectrum of values between 50 to 99%. The Precision (1 Standard deviation) for the Cerebral Oximeter Monitor SctO2 compared to reference SctO-derived from co-oximeral (SaO-) and jugular bulb (SjvQ) blood samples was ± 4.86 %, based on Equation 1 below.

Infant & Neonate SctO2: Using the FORE-SIGHT Small sensor, the Cerebral Oximeter SctO2 showed strong agreement with the reference SctO2 over the spectrum of values between 50to, The Precision (1 Standard deviation) for the Cerebral Oximeter Monitor SctO2 compared to the reference SctO2 derived from pulse oximetry measured arterial oxygen saturation SaC2 and cooximetry measured internal jugular vein venous oxygen saturation (SivO) from blood samples was ± 5.0 %, based on Equation 1.

$$\text{References: Set}\O_2% = \text{Sa}\O_2 \ge 0.3 + \text{Sj}\\nu O_2 \ge 0.7 \tag{Eequation:} 1.6$$

The Cerebral Oximeter Monitor SctO2 value represents oxygen saturation in the brain tissue microvasculature containing venous and arterial blood volume at a ratio of 70:30,

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

..

AUG 2 1 2009

CAS Medical Systems, Inc. c/o Ron Jeffrey Director of Regulatory Affairs 44 East Industrial Rd. Branford, Connecticut 06405

Re: K091452

Trade/Device Name: Fore-Sight, Model MC-2000 SERIES Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: MUD Dated: July 29, 2009 Received: July 30. 2009

Dear Mr. Jeffrey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

Page 2 - Mr. Ron Jeffrey

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdaggv/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,
Kesia Alexander for

Malvina B. Eydelman, M.D Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number:

K091452

Device Name:

FORE-SIGHT™ Cerebral Oximeter Monitor, Model 2000 Series.

Indications for Use:

The FORE-SIGHT® Cerebral Oximeter, Model MC-2000 Series is indicated for the continuous noninvasive monitoring of regional hemoglobin oxygen saturation of blood in the brain (SctO2). It is intended for use in any individual at risk for reducedflow or no-flow ischemic states.

When used with FORE-SIGHT large sensors, the FORE-SIGHT MC-2000 Cerebral Oximeter Monitor is indicated for use with adults and children over 40Kg. When used with the FORE-SIGHT medium sensors, the FORE-SIGHT MC-2000 Cerebral Oximeter is indicated for use with small adults and children between 4 kg and 80 kg. When used with FORE-SIGHT small sensors the FORE-SIGHT MC-2000 Series Cerebral Oximeter Monitor is indicated for infants and neonates ≤ 8Kg.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M.L.B.Nichols

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number_k091452

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).