(161 days)
Not Found
No
The description focuses on standard oscillometric blood pressure measurement and does not mention any AI or ML techniques.
No
The device is a monitor that measures noninvasive blood pressure and pulse rate, which are diagnostic functions, not therapeutic ones.
No
The device measures noninvasive blood pressure and pulse rate, which are physiological parameters, but it is not explicitly stated that it provides a diagnosis or aids in making a medical diagnosis. Instead, it seems to monitor vital signs.
No
The device description explicitly states it is a "compact, battery-powered monitor" and measures blood pressure using a cuff and oscillometric method, indicating it is a hardware device with integrated software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The BioCare vTrust 701DH NIBP monitor measures noninvasive blood pressure and pulse rate. This is a direct measurement of physiological parameters from the body, not an analysis of a sample taken from the body.
- Intended Use: The intended use describes measuring blood pressure and pulse rate, which are vital signs, not analyzing a biological sample for diagnostic purposes.
Therefore, the function and intended use of this device clearly fall outside the scope of an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
The BioCare vTrust 701DH NIBP monitor is a compact, battery-powered monitor which measures noninvasive blood pressure (systolic and mean arterial pressure) and pulse rate based on the principle of plethysmography (oscillometric on inflation).
This equipment is intended for use only by qualified medical providers in conjunction with established medical protocols. It is not intended for use in an MRU environment, or in an air transport environment.
Product codes
DXN
Device Description
The BioCare vTrust 701DH NIBP monitor is a compact, battery-powered monitor which measures noninvasive blood pressure (systolic, diastolic and mean arterial pressure) and pulse rate based on the principle of plethysmography (oscillometric on inflation).
The oscillometric method detects volume displacements within the artery and senses pressure variations within the blood pressure cuff during inflation. The measurement is more quickly and with less patient discomfort than with monitors which measure NIBP during cuff deflation. Cycle time function enables you to pre-set measurement intervals. User can set alarm limits by selectable for high or low range.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Upper arm or limb
Indicated Patient Age Range
Not Found
Intended User / Care Setting
qualified medical providers
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The performance test protocols and data analysis are following "ANSVAAMI SP10: 2002" standard. In this part, the substantial equivalence is demonstrated by showing conformance to performance requirements in the SP-10 standard. The pressure transducer accuracy and overall system effectiveness were determined and compared to the preset criteria.
Performance Characteristics and Results:
- Environmental performance and stability: Passed.
- Storage conditions: ≤± 3 mmHg (±0.4 kPa) for storage, ≤± 4 mmHg for in use. Max deviation -3.4 mmHg.
- Operating conditions: ≤± 3 mmHg (±0.4 kPa) for storage, ≤± 4 mmHg for in use. Max deviation ± 3 mmHg.
- Vibration and shock: ISTA-1A (14,200 vibratory impacts, 10 drop test) - no visible damage, SD ≤8 mmHg.
- Stability - Voltage range: Red light for low battery (25 mmHg/sec.
- Electrical safety: Complied with IEC 60601-1, 1988; Amendment 1, 1991; & Amendment 2, 1995.
- Performance: Passed.
- Electronic manometers - Range: 0 mmHg to at least 260 mmHg by visual inspection. Cleared for 0~300 mmHg.
- Resolution: 1 mmHg by visual inspection. Cleared for 1 mmHg.
- Accuracy of pressure measurement: Ambient temp 10°C~40°C; RH: 15%~90%. Pressure stages: 50, 100, 150, 200, 250, 300 mmHg. Cleared for ≤±3 mmHg (±0.4 kPa) or 2% of the reading; ≤±4 mmHg for in use. Max deviation ±3 mmHg.
- Repeatability: Cuff-Link simulator test. Adult: SYS120/DIA80/MAP90 mmHg, 80 BPM. Neo: SYS80/DIA50/MAP62 mmHg, 120 BPM. Average difference ≤± 5 mmHg. Max deviation 5 mmHg.
- Leakage: Ambient temp 15°C
25°C; RH: 2085%. Pressure stages: 50, 100, 150, 200, 250 mmHg, wait at least 60 s before reading the values. Cleared for ≤6 mmHg/min. Max deviation 1.2 mmHg/min. - Pressure transducer accuracy: Ambient temp 10℃~40℃; RH: 15%~90%. Pressure stages: 50, 100, 150, 200, 250, 300 mmHg. Cleared for ≤±3 mmHg (±0.4 kPa) or 2% of the reading; ≤±4 mmHg for in use. Max deviation ±3 mmHg.
- Overall system efficacy-clinical testing (auscultatory method as the reference Std.): No fewer than 85 subjects with a minimum of 255(=3*85) observations. For systolic and diastolic: mean≤±5 mmHg; SD ≤8 mmHg. Mean: Sys/Dia 3.4/-4.5. SD: Sys/Dia 5.23/4.64 mmHg.
- Requirements for inflation source and pressure control valves: Passed.
- Inflation source: pressure~300mmHg≤10s. Cleared for inflate at 5 mmHg/sec, Speed test: Inflation > 5 mmHg/sec.
- Automated valves-Release rate: rapid exhaust with fully-opened valve. 260mmHg → 15mmHg ≤10 sec or 150mmHg → 5mmHg ≤5 sec. Cleared for deflate at 25 mmHg/sec.
- IEC 60601-1-2-30 Automatic cycling function. 1. In long term automatic mode (normal condition): Adult 120/80(90)[SYS/DIA(MAP)]--long term; Neonate 80/50(62)[SYS/DIA(MAP)]--long term. Deflated time ≥30 sec; mean≤± 5 mmHg. Results: average for adults 34 sec / for neonates 42 sec; - 4mmHg. 2. In long term automatic mode (single fault condition): Deflated time ≤30 sec; release cuff pressure to below 15mmHg(adult)/ 5mmHg(neo.). Results:
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
01/6 Koq 1414
510(k) Summary
MAR 11 2009
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 8807.92.
1. Submitter's Name: BioCare Corporation
Address: 4F, No.12, Lane 5, Sec. 2, Nanshan Rd., Luihu Township, Taoyuan County 33852, Taiwan.
OCT 2 1 2009
Phone: +886-3-2616678
Fax: +886-3-2616679
Contact person: Judy Wei / Administrator
2. Device Name and Classification
Trade name: vTrust " Noninvasive Blood Pressure Monitor, Model 701DH series Common/Usual name: Blood Pressure Monitor
Classification name/Code: Noninvasive Blood Pressure Measurement System-DXN
Classification Panel: Circulatory System Devices Panel (74)
Device Class: II
3. Predicate Device
The predicate device is CSI 507 non-invasive vital signs monitor(K910852) marketed by Criticare Systems, Inc.
4. Device Description
The BioCare vTrust 701DH NIBP monitor is a compact, battery-powered monitor which measures noninvasive blood pressure (systolic, diastolic and mean arterial pressure) and pulse rate based on the principle of plethysmography (oscillometric on inflation).
The oscillometric method detects volume displacements within the artery and senses pressure variations within the blood pressure cuff during inflation. The measurement is more quickly and with less patient discomfort than with monitors which measure NIBP during cuff deflation. Cycle time function enables you to pre-set measurement intervals. User can set alarm limits by selectable for high or low range.
ર-।
1
5. Intended Use
This equipment is intended to be use only by qualified medical providers in conjunction with established medical protocols. It is not intended for use in an MRI environment, or in an air transport environment.
6. Summary of Technological Characteristics of the Device Compared to the Predicate Device
Both vTrust 701DH blood pressure monitor and 507 non-invasive vital signs monitor is intended to be used in measuring human's Systolic, Diastolic and Pulse rate by oscillometric method. They share similar features and specifications. Performance characteristics of both devices are in accordance with ANSI/AAMI SP-10: 2002.
The substantial equivalence between vTrust 701DH and 507 non-invasive vital signs monitor can be evaluated from several aspects as listed in below Table.
Items | Proposed device | Predicate device | Similarities | Differences |
---|---|---|---|---|
Device name | vTrust 701DH non-invasive | |||
blood pressure monitor | 507 non-invasive vital signs | |||
monitor | ||||
Device | ||||
Classification | Class II | Class II | Equivalent | Negligible |
Classification | ||||
Panel | Cardiovascular | Cardiovascular | Equivalent | Negligible |
Intended Use | ||||
(Indications for use) | Monitor non-invasive blood | |||
pressure and pulse rates | Monitor non-invasive blood | |||
pressure and pulse rates | Equivalent | Negligible | ||
Anatomical sites | Upper arm or limb | Upper arm or limb | Equivalent | Negligible |
Device Description | ||||
Technology - | ||||
measurement | ||||
methodology | Oscillometric measure upon | |||
inflation | Oscillometric measure upon | |||
inflation | Equivalent | Negligible | ||
Materials | Cuff: neoprene (non-latex) | Cuff: neoprene (non-latex) | Equivalent | Negligible |
Energy used | Recharge NiMH battery or | |||
AC adapter | Recharge NiMH battery or | |||
AC adapter | Equivalent | Negligible | ||
Weight | 650 g | 630 g | ||
Dimensions | ||||
(HWD) mm | 190(H) × 90(W) × 50(D) | 190(H) × 95(W) × 45(D) | ||
Memory | 100 measurements | 100 measurements | Equivalent | Negligible |
Items | Proposed device | Predicate device | Similarities | Differences |
Device name | vTrust 701DH non-invasive | |||
blood pressure monitor | ||||
Image: proposed device | 507 non-invasive vital signs | |||
monitor | ||||
Image: predicate device | ||||
Descriptive characteristics | ||||
Inflation & deflation | Automatic control | Automatic control | Equivalent | Negligible |
Display | LCD / LED | LED | ||
Parameters | SYS DIA MAP HR | SYS DIA HR | ||
STAT mode | 5 min of continuous readings | 5 min of continuous readings | Equivalent | Negligible |
Automatic | ||||
Measurement Cycles | 1, 2, 3, 5, 10, 15, 30, 60min | 5, 10, 15min | ||
Auto shutoff | 20 min after last key operation | 20 min after last key operation | Equivalent | Negligible |
System output | RS 232-compatible; digital; | |||
mini-din connector | RS 232-compatible; digital; | |||
mini-din connector | Equivalent | Negligible | ||
Performance specifications | ||||
Standards met | ANSI/AAMI SP-10: 2002, | |||
IEC-60601-1, IEC-60601-1-2, | ||||
ISO 10993 | ANSI/AAMI SP-10: 1992, | |||
IEC-60601-1, IEC-60601-1-2, | ||||
ISO 10993 | Equivalent | Negligible | ||
Operating | ||||
environment | Temp: 0° | |||
RH:15% to 90%, non-condensing | Temp: 0° | |||
RH:15% to 90%, non-condensing | Equivalent | Negligible | ||
Storage | ||||
environment | Temp: -10° | |||
RH:15% to 90%, non-condensing | Temp: -5° | |||
RH: 15% to 90%, non-condensing | ||||
Measurement range | Pressure: 30-300 mmHg; | |||
Pulse: 30-300 beats/min | Pressure: 30-300 mmHg; | |||
Pulse: 30-300 beats/min | Equivalent | Negligible | ||
Resolution | 1 mmHg | 1 mmHg | Equivalent | Negligible |
Accuracy | Pressure: ±2mmHg or 2% | |||
of reading; Pulse: ±2 bpm | ||||
or ±2% of reading | Pressure: ±2mmHg or 2% | |||
of reading; Pulse: ±2 bpm | ||||
or ±2% of reading | Equivalent | Negligible |
2
The differences between the two devices list as: weight, dimensions, display, . parameters, cycle time. There are no significant differences that affect the safety and effectiveness. Therefore, BioCare vTrust 701DH non-invasive blood pressure monitor is substantially equivalent to the legally marketed device CSI 507 non-invasive vital signs monitor, K910852.
3
7. Determination of Substantial Equivalence is based on an assessment of Performance Data
The performance test protocols and data analysis are following "ANSVAAMI SP10: 2002" standard. In this part, the substantial equivalence is demonstrated by showing conformance to performance requirements in the SP-10 standard. The pressure transducer accuracy and overall system effectiveness were determined and compared to the preset criteria.
Clause | Performance Characteristics | Max. permissible error or acceptance criteria | Max. deviation | Passed/ Failed | Reference to documentation |
---|---|---|---|---|---|
4.2 | Environmental performance and stability | ||||
4.2.1 | Storage conditions: | ||||
[23°F(-5°C)] for 24 hrs; | |||||
[122°F(50°C)] for 24 hrs; | |||||
RH: 90% (noncondensing) | ≤± 3 mmHg(±0.4 kPa) | ||||
≤± 4 mmHg for in use | -3.4 mmHg | Passed | T-0406-004 | ||
4.2.2 | Operating conditions: | ||||
Temp: 50°F (10°℃)~104°F(40℃); | |||||
RH: 15~90% (noncondensing); | |||||
Bar. 105kPa | ≤± 3 mmHg(±0.4 kPa) | ||||
≤±4 mmHg for in use | ± 3 mmHg | Passed | T-0406-004 | ||
4.2.3 | Vibration and shock: | ||||
ISTA-standard drop and vibration test | |||||
IEC 60601-1, sections 21.5 & 21.6 | ISTA-1A: 14,200 vibratory impacts; drop test-10 drops; before and after transportation test within ±5 mmHg of the average reading. | no visible damage; | |||
SD ≤8 mmHg | Passed | T-0406-012 | |||
4.2.4 | Stability | ||||
4.2.4.1 | Voltage range: | ||||
Battery-powered devices-indication of battery condition | Red light: low battery; | ||||
Blue light: batteries are charging; | |||||
Green light: fully charged | 25 mmHg/sec | mmHg/sec | Passed | PS-1040245-T | |
4.3.2 | Electrical safety: | IEC 60601-1, 1988; Amendment | |||
1, 1991; & Amendment 2, 1995. | Passed | Section 17 | |||
4.4 | Performance | ||||
4.4.3 | Electronic manometers | ||||
4.4.3. | |||||
1.A | Range: 0 mmHg to at least 260 mmHg by | ||||
visual inspection | NIBP pressure calibration | 0~300 mmHg | Passed | PS-1040245-T | |
4.4.3. | |||||
2.A | Resolution: 1 mmHg by visual inspection | NIBP pressure calibration | 1 mmHg | Passed | PS-1040245-T |
4.4.3. | |||||
3.A | Accuracy of pressure measurement: | ||||
Ambient temp 10°C~40°C; RH: 15%~90% | |||||
Pressure stages: 50, 100, 150, 200, 250, 300 mmHg | ≤±3 mmHg(±0.4 kPa) | ||||
or 2% of the reading; | |||||
≤±4 mmHg for in use | ±3 mmHg | Passed | T-0406-003 | ||
4.4.3. | |||||
4.A | Repeatability | ||||
Cuff-Link simulator test for stability and | |||||
reproduction of performance | |||||
Adult: SYS120/DIA80/MAP90 mmHg, 80 BPM | |||||
Neo: SYS80/DIA50/MAP62 mmHg, 120 BPM | Average difference≤± | ||||
5 mmHg | 5 mmHg | Passed | T-0406-003 | ||
4.4.3. | |||||
6.A | Leakage: | ||||
Ambient temp 15°C | |||||
Pressure stages: 50, 100, 150, 200, 250 mmHg, | |||||
wait at least 60 s before reading the values. | ≤6 mmHg/min | 1.2 mmHg/min | Passed | T-0406-003 | |
4.4.4. | |||||
B | Pressure transducer accuracy: | ||||
Ambient temp 10℃~40℃; RH: 15%~90% | |||||
Pressure stages: 50, 100, 150, 200, 250, 300 mmHg | ≤±3 mmHg(±0.4 kPa) | ||||
or 2% of the reading | |||||
≤±4 mmHg for in use | ±3 mmHg | Passed | T-0406-003 | ||
4.4.5. | |||||
B | Overall system efficacy-clinical testing | ||||
(auscultatory method as the reference Std.) | |||||
no fewer than 85 subjects with a minimum | For systolic and diastolic: | ||||
mean≤±5 mmHg; | |||||
SD ≤8 mmHg | Mean: | ||||
Sys/Dia | |||||
3.4/-4.5 | Passed | Section 20 | |||
Clause | Performance Characteristics | Max. permissible error | |||
or acceptance criteria | Max. | ||||
deviation | Passed/Failed | Reference to | |||
documentation | |||||
of 255(=3*85) observations | SD: | ||||
Sys/Dia | |||||
5.23/4.64 | |||||
mmHg | |||||
4.5 | Requirements for inflation source and pressure control valves | ||||
4.5.1 | Inflation source: pressure~300mmHg≤10s | NIBP pressure calibration: inflate at 5 | |||
mmHg/sec | |||||
Speed test: | |||||
Inflation > 5 mmHg/sec | Passed | PS-1040245-T | |||
4.5.4. | |||||
3.B | Automated valves-Release rate: rapid | ||||
exhaust with fully-opened valve | 260mmHg $\rightarrow$ 15mmHg | ||||
≤10 sec or | |||||
150mmHg $\rightarrow$ 5mmHg | |||||
≤5 sec | deflate at 25 | ||||
mmHg/sec | Passed | PS-1040245-T | |||
IEC | |||||
6060 | |||||
1-2- | |||||
30 | Automatic cycling function |
- In long term automatic mode (normal condition)
Adult 120/80(90)[SYS/DIA(MAP)]--long term
Neonate 80/50(62)[SYS/DIA(MAP)]--long term - In long term automatic mode (single fault condition)
- In short term automatic mode (stat mode) | 1. deflated time ≥30 sec; mean≤± 5 mmHg;
- deflated time ≤30 sec; release cuff pressure to below 15mmHg(adult)/
5mmHg(neo.); - deflated time≥2sec.
after each
determination | 1. average for adults 34 sec / for
neonates 42 sec; -
4mmHg; - Summary of performance testing-Bench & Clinical results
4
:
5
8. Conclusions
The BioCare vTrust 701DH non-invasive blood pressure monitor has the same intended use and similar technological characteristics as the CSI 507 non-invasive vital signs monitor (K910852) marketed by Criticare Systems, Inc. Moreover, performance testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the BioCare vTrust 701DH non-invasive blood pressure monitor is substantially equivalent to the predicate devices.
6
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight. The logo is rendered in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
BioCare Corporation c/o Ms. Judy Wei Administrator 4F, No.12, Lane 5, Sec. 2, Nanshan Rd. Lujhu Township Taoyuan County 33852, Taiwan
OCT 2 1 2009
Re: K091414
Trade/Device Name: vTrust™ Noninvasive Blood Pressure Monitor Model 701DH series Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: August 24, 2009 Received: August 27, 2009
Dear Ms. Wei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
7
Page 2 - Ms. Judy Wei
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
W.M.T.
erBram D. Zuckerman, M Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
8
Indications for Use
510(k) Number (if known): KO91414
Device Name: vTrust Blood Pressure Monitor, Model: 701DH Series
Indications For Use:
The BioCare vTrust 701DH NIBP monitor is a compact, battery-powered monitor which measures noninvasive blood pressure (systolic and mean arterial pressure) and pulse rate based on the principle of plethysmography (oscillometric on inflation).
This equipment is intended for use only by qualified medical providers in conjunction with established medical protocols. It is not intended for use in an MRU environment, or in an air transport environment.
Prescription Use (Part 21 CFR 801 Subpart D)
AND / OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
--- (PLEASE NO NOT WRITE BELOW THIS LINE-CONTINUE ON-AMOTHER ------PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
W.M.R.
Page 1 of
ion Slan-Off ivision of Cardiovascular Devices
510(k) Number K0914