(44 days)
No
The summary describes a standard vital signs monitor with basic data display, alarms, and trending capabilities, with no mention of AI or ML.
No.
This device is described as a monitor that measures and displays physiological parameters, providing data and alarms to the healthcare provider, which classifies it as a diagnostic or monitoring device, not a therapeutic one.
No
The device monitors physiological parameters and displays real-time data to help professional healthcare providers interpret the monitored data. It does not perform a diagnosis itself, but rather provides information for a human to make a diagnosis.
No
The device description explicitly states it "measures and displays real time physiological data" and monitors parameters like NIBP and SpO2, which requires hardware sensors and components to acquire the data. It is a vital signs monitor, not a software-only device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "monitor physiological parameters of patients within any healthcare environment." This involves measuring vital signs directly from the patient's body.
- Device Description: The device measures and displays real-time physiological data like NIBP and SpO2. These are measurements taken from the patient, not from a sample of bodily fluid or tissue.
- Lack of IVD Characteristics: There is no mention of analyzing samples (blood, urine, tissue, etc.) or providing information about a disease or condition based on the analysis of such samples.
IVD devices are specifically designed to examine specimens derived from the human body to provide information for the diagnosis, monitoring, or treatment of a disease or condition. This device directly monitors the patient's physiological state.
N/A
Intended Use / Indications for Use
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
Product codes
DRT, DQA
Device Description
The 506CN monitor measures and displays real time physiological data of the patient, including a graphical plethysmogram and numerical data. The 506CN can be used to monitor one or more of the following parameters: Noninvasive BP (NIBP) and SpO2. For all these vital parameters, the 506CN will be capable of limit alarms and alerts, printing of strip chart recordings and storing trends for retrospective review.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
professional health care provider.
any healthcare environment.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The 506CN monitor performance for each monitoring modality has been confirmed to be equivalent to the predicate device. Additionally, the 506CN complies with applicable safety and performance standards (detailed below) for each monitoring modality and verification of compliance has been completed. The patient monitoring technologies present in the 506CN monitor have been in clinical use for at least six years in the 506 monitor and it's predicates. CSI's field experience with these modalities in the predicate devices has been satisfactory.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.2300 Cardiac monitor (including cardiotachometer and rate alarm).
(a)
Identification. A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.(b)
Classification. Class II (performance standards).
0
10 April 2009
510K Summary
MAY 2 7 2009
Model 506CN Patient Monitor
Contact:
Alex Kaplan Director of QA & RA Criticare Systems, Inc. 20925 Crossroads Circle Waukesha, WI 53186 USA 262-798-8282 Voice 262-798-8290 FAX
Trade Name: 506CN Patient Monitor
Vital Signs Monitor Common Name:
Classification Name: Monitor, Physiological, Patient (74D-RT)
Substantial Equivalence is claimed to : 506 Vital Signs Monitoring System (K051038).
Device Description:
The 506CN monitor measures and displays real time physiological data of the patient, including a graphical plethysmogram and numerical data. The 506CN can be used to monitor one or more of the following parameters: Noninvasive BP (NIBP) and SpO2. For all these vital parameters, the 506CN will be capable of limit alarms and alerts, printing of strip chart recordings and storing trends for retrospective review.
Intended Use:
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
Comparison with predicate device:
Criticare Systems Inc. has developed and distributed physiological monitoring devices worldwide since its inception in 1984. The 506CN monitor utilizes existing core technologies from the predicate 506 monitor for patient monitoring of NIBP and SpO2. The patient data collected by the 506CN monitor is displayed for the user on a graphic
@2009 Criticare Systems, Inc All Rights Reserved
page 1 of 2
1
LCD equivalent to the predicate device. Membrane key panels provide a user interface equivalent to the predicate device. The packaging design of the 506CN monitor is molded plastic and allows for it to be either a stationary monitor or to be used during patient translocation within the healthcare facility, as did the predicate 506.
Determination of Substantial Equivalence:
The 506CN monitor performance for each monitoring modality has been confirmed to be equivalent to the predicate device. Additionally, the 506CN complies with applicable safety and performance standards (detailed below) for each monitoring modality and verification of compliance has been completed. The patient monitoring technologies present in the 506CN monitor have been in clinical use for at least six years in the 506 monitor and it's predicates. CSI's field experience with these modalities in the predicate devices has been satisfactory. This combination of equivalence testing, applicable objective standards compliance and field experience substantiates a high level of confidence in the safety and efficacy of the 506CN monitor.
Therefore, the 506CN monitor is substantially equivalent to the predicate devices.
Compliance to standards and regulations:
The 506CN Vital Signs Monitor complies with the following national and international standards:
Safety
IEC 60601-1 Medical Electrical Safety IEC 60601-1-2 EMC Compliance ISO 10993-5.10-11 Biocompatibility IEC 60601-1-8 Alarms
Performance
ISO 9919 Oximetry Performance IEC 60601-2-30 NIBP Safety EN 1060-1 NIBP Performance EN 1060-3 NIBP Performance AAMI SP-10 NIBP Performance
@2009 Criticare Systems, Inc All Rights Reserved
Page 2 of 2
2
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.
Public Health Service
MAY 2 7 2009
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Criticare Systems, Inc. c/o Mr. Alex Kaplan Director of QA & QC 20925 Crossroads Circle Waukesha, WI 53186
Re: K091050
506CN Vital Signs Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II (two) Product Code: DRT, DQA Dated: May 12, 2009 Received: May 12, 2009
Dear Mr. Kaplan:
We have reviewed your Section 510(k) premarket notification of intent to market the device w o nave roviewed your co determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the energy of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 - Mr. Alex Kaplan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Indications for Use
510(k) Number (if known):
Device Name: 506CN Vital Signs Monitor
Indications For Use:
This system is intended to monitor physiological parameters of patients within any healthcare environment. The user, responsible to interpret the monitored data made available, will be a professional health care provider. Physiological data, system alarms and patient data analysis will be available to the care provider from the monitor.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division/Sign-Off)
Division of Cardiovascular Devices
510(k) Number K091050