K Number
K090989

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-09-23

(169 days)

Product Code
Regulation Number
870.5310
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ZOLL R Series® External Defibrillator is intended for the defibrillation, Noninvasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including ECG Monitoring, Pulse Oximetry, Non-Invasive Blood Pressure measurement (proposed), End Tidal CO2 (proposed), CPR performance and data printing and recording for resting patients in critical care and in-hospital transport. The ZOLL R Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment.

Defibrillator Function:
Manual Operation: Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: Unconsciousness, Absence of breathing, Absence of pulse. This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate. The advisory function should be used to confirm ventricular fibrillation or wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (listed above).
Semiautomatic Operation (AED): The R Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient. They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically approved patient care protocol. Use of the R Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: Unconsciousness, Absence of breathing, Absence of pulse.

ECG Monitoring: The unit is intended for use when ECG monitoring is indicated to evaluate the patient's heart rate or ECG morphology.

CPR Monitoring: The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.

External Pacemaker (Pacer Version Only):
Pacemaker: This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. The purposes of pacing include: Resuscitation from standstill or bradycardia of any etiology; As a standby when standstill or bradycardia might be expected; Suppression of tachycardia.

SpO2 Monitoring: The R Series pulse oximeter, with the Masimo® SET® technology and the LNCS® series of oximeter sensors, is indicated for the continuous, monitoring of arterial oxygen saturation (SpO2) and pulse rate during both no motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital or prehospital environment.

NIBP Monitoring: The ZOLL R Series NIBP option is indicated for the non-invasive measurement of arterial blood pressure for resting patients in critical care and in-hospital transport.

EtCO2 Monitoring: The ZOLL R Series EtCO2 option with Respironics Novametrix technology is indicated for the continuous noninvasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in patients requiring ventilator support, in-hospital transport, or anesthesia.

Device Description

The ZOLL R Series® External Defibrillator is intended for the defibrillation, Noninvasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including ECG Monitoring, Pulse Oximetry, Non-Invasive Blood Pressure measurement (proposed), End Tidal CO2 (proposed), CPR performance and data printing and recording for resting patients in critical care and in-hospital transport. The ZOLL R Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The design facilitates table top use while still providing a light weight and easy to carry device for transport.

AI/ML Overview

The provided text describes ZOLL R Series® NIBP and EtCO2 Options, a multi-parameter monitor, but does NOT contain specific acceptance criteria, detailed performance study results, or information about AI/ML algorithms.

The document states:

"Extensive performance testing ensures that the ZOLL R Series with NIBP and EtCO2 Options performs as well as the indicated predicate device and meets all of its functional requirements and performance specifications. Safety testing assures that the device complies with applicable sections of recognized industry and safety standards."

This is a general statement about testing without providing quantifiable acceptance criteria or detailed results. The submission focuses on demonstrating substantial equivalence to a predicate device (ZOLL E Series NIBP and EtCO2 Options).

Therefore, I cannot fulfill the detailed request for acceptance criteria, device performance, sample sizes, expert qualifications, or details specific to AI/ML comparative effectiveness or standalone studies.

Based on the provided text, I can only provide the following limited information:


1. A table of acceptance criteria and the reported device performance

The document does not provide a table of acceptance criteria or specific reported device performance metrics. It generally states that "Extensive performance testing ensures that the ZOLL R Series with NIBP and EtCO2 Options performs as well as the indicated predicate device and meets all of its functional requirements and performance specifications."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the document.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the document, as the device is a medical monitor, not an AI/ML diagnostic tool requiring expert ground truth for its performance validation in the typical sense.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention an AI/ML component or an MRMC study. This device is a monitor, not a diagnostic AI/ML tool intended to assist human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This document does not describe an AI/ML algorithm or a standalone performance study for such an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The document focuses on the measurement capabilities (NIBP, EtCO2) and functionality of a monitor, not on an algorithm requiring ground truth in the context of diagnostic interpretation. Device performance would typically be validated against reference standards for physiological measurements. No details on this are provided beyond the general statement of "extensive performance testing."

8. The sample size for the training set

This information is not provided as the document does not describe an AI/ML algorithm that would undergo a training phase.

9. How the ground truth for the training set was established

This information is not applicable as the document does not describe an AI/ML algorithm that would require a training set with established ground truth.


Summary of Device and Testing Approach from the Document:

The ZOLL R Series® NIBP and EtCO2 Options is a multi-parameter monitor. The regulatory submission emphasizes substantial equivalence to a predicate device (ZOLL E Series NIBP and EtCO2 Options) rather than comprehensive de novo performance studies with detailed statistical results. The testing described is "extensive performance testing" and "safety testing" to ensure it meets functional requirements, performance specifications, and complies with industry and safety standards, similar to the predicate device.

{0}------------------------------------------------

510(k) Summary:

SEP 2 3 2009

Submitter's Name and Address:

ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655

Contact Person:

Charles W. Kolifrath

(978) 421-9786

Date Summary Prepared:

March 24, 2009

Device:

ZOLL R Series® NIBP and EtCO2 Options

Classification:

Defibrillator, Low-energy - DC : Class II (21 CFR 870.5300 - LDD)

Automatic External Defibrillators: Class III (21 CFR 870.5310 - MKJ)

Cardiopulmonary Resuscitation Aid: Class III (21 CFR 870.5200 - DRM)

Cardiac Monitors (including Cardiotachometers and Rate Alarms): Class II (21CFR 870.2300 - DRT)

External Transcutaneous Cardiac Pacemakers (Non-invasive): Class II (21 CFR 870.5550 - DRO)

Oximeters: Class II (21 CFR 870.2700 - DQA)

System, Measurement, Blood-Pressure, Non-Invasive: Class II (21 CFR 870.1130 - DXN)

Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase: Class II (21 CFR 868.1400 -CCK)

Description:

The ZOLL R Series® External Defibrillator is intended for the defibrillation, Noninvasive Transcutaneous Pacing, multi-parameter monitoring of patient vital signs, including ECG Monitoring, Pulse Oximetry, Non-Invasive Blood Pressure measurement (proposed), End Tidal CO2 (proposed), CPR performance and data printing and recording for resting patients in critical care and in-hospital transport. The ZOLL R

{1}------------------------------------------------

Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The design facilitates table top use while still providing a light weight and easy to carry device for transport.

Intended Use:

Defibrillator Function

Intended Use - Manual Operation

Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • . Unconsciousness
  • Absence of breathing

  • . Absence of pulse.

This product should be used only by qualified medical personnel for converting ventrioular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.

In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate.

The advisory function should be used to confirm ventricular fibrillation or wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (listed above).

Intended Use - Semiautomatic Operation (AED)

The R Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.

They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically approved patient care protocol.

Use of the R Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • . Unconsciousness
  • . Absence of breathing
  • . Absence of pulse

Specifications for the ECG rhythm analysis function are provided in the section, "ECG Rhythm Analysis Algorithm Accuracy" on page A-29.

When the patient is less than 8 years of age or weighs less than 55lbs (25Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.

{2}------------------------------------------------

Intended Use - ECG Monitoring

The unit is intended for use when ECG monitoring is indicated to evaluate the patient's heart rate or ECG morphology. In ECG monitoring mode, the unit is intended to be used by personnel who are qualified by training in the R Series defibrillator, basic life and/or advanced life support, or other physician authorized emergency medical training.

Intended Use - CPR Monitoring

The CPR monitoring function provides visual and audio feedback designed to encourage resouers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.

The CPR monitoring function is not intended for use on patients under 8 years of age.

External Pacemaker (Pacer Version Only)

Intended Use - Pacemaker

This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.

The purposes of pacing include:

Resuscitation from standstill or bradycardia of any etiology:

Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.

As a standby when standstill or bradycardia might be expected:

Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.

Suppression of tachycardia:

Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.

Pediatric Pacing

{3}------------------------------------------------

Pacing can be performed on pediatric patients weighing 33lb. (15kg) or less using ZOLL pediatric hands-free therapy electrodes. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.

Intended Use - SpO2 Monitoring

The R Series pulse oximeter, with the Masimo® SET® technology and the LNCS® series of oximeter sensors, is indicated for the continuous, nonitoring of arterial oxygen saturation (SpO2) and pulse rate during both no motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital or prehospital environment.

Intended Use - NIBP Monitoring

The ZOLL R Series NIBP option is indicated for the non-invasive measurement of arterial blood pressure for resting patients in critical care and in-hospital transport.

The R Series NIBP option is designed to measure blood pressure for adult patients (21 years of age and older) and for pediatric patients, as described in the following table:

Pediatric SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age

Intended Use - EtCO2 Monitoring

The ZOLL R Series EtCO2 option with Respironics Novametrix technology is indicated for the continuous noninvasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in patients requiring ventilator support, in-hospital transport, or anesthesia.

This option uses the CAPNOSTAT 5 Mainstream CO2 sensor attached to an airway adapter that connects to an endotracheal tube. mask or disposable mouthpiece.

The R Series EtCO2 option is designed to monitor adult, pediatric, and neonatal patients.

The following substances can influence CO2 measurements made with the CAPNOSTAT 5 CO2 sensor:

  • · elevated oxygen levels
  • · nitrous oxide
  • · halogenated agents

The R Series EtCO2 option provides settings for high oxygen and/or nitrous oxide compensation. Halogenated anesthetic agents alter CO2 readings, but the R Series unit will monitor CO2 within specifications when these agents are present at normal clinical levels. The presence of Desflurane in the exhaled breath beyond normal values (5%) may positively bias measured carbon dioxide values.by up to an additional 3 mmHg.

The R Series EtCO2 option is intended for use only with the ZOLL/Respironics Novametrix CAPNOSTAT 5 Mainstream CO2 Sensor and mainstream airway adapters.

The R Series EtCO2 option can be used on adult patients (21 years of age and older) and on pediatric patients, as described in the following table:

{4}------------------------------------------------

Pediatric SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age

Substantial Equivalence:

The features and functions of the proposed ZOLL R Series NIBP and EtCO2 Options are substantially equivalent to the ZOLL E Series NIBP and EtCO2 Options (K042007, cleared for use on 4/7/2005).

Comparison of Technological Characteristics

The ZOLL R Series with NIBP and EtCO2 Options features design characteristics that are very similar to those of the indicated predicate device, and utilizes the same interpretive features and functions as those of the currently marketed ZOLL R Series (K081574). The ZOLL R Series NIBP and EtCO2 Options utilize technologies very similar to those used by the ZOLL E Series.

Performance Testing:

Extensive performance testing ensures that the ZOLL R Series with NIBP and EtCO2 Options performs as well as the indicated predicate device and meets all of its functional requirements and performance specifications. Safety testing assures that the device complies with applicable sections of recognized industry and safety standards.

Conclusion

Performance and safety testing of the ZOLL R Series with NIBP and EtCO2 Options demonstrates that its features and functions are substantially equivalent to those of the indicated commercially distributed predicate device with regard to performance, safety and effectiveness.

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight. The symbol is composed of curved lines and shapes, giving it a dynamic and modern appearance.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

SEP 2 8 2009

Zoll Medical Corporation Worldwide Headquarters c/o Mr. Charles W. Kolifrath Regulatory Affairs Manager 269 Mill Road Chelmsford, MA 01824-4105

Re: K090989

Trade/Device Name: Zoll R Series with NIBP & EtCO2 Options Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ, LDD, DRM, DRT, DRO, DQA, DXN, CCK Dated: September 4, 2009 Received: September 8, 2009

Dear Mr. Kolifrath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Charles W. Kolifrath

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,
W. Mote

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

SECTION 4 - Indications for Use

Interfire 510(k) Number (if known): ZOLL R Series Device Name:

Defibrillator Function

Intended Use - Manual Operation

Use of the R Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • ◆ Unconsciousness
  • Absence of breathing .
  • Absence of pulse.

This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac thythms capable of producing hemodynamically significant heart beats.

In manual mode, the unit can also be used for synchronized cardioversion of certain atrial or ventricular arrhythmias. A qualified physician must decide when synchronized cardioversion is appropriate.

The advisory function should be used to confirm ventricular fibrillation or wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (listed above).

Intended Use - Semiautomatic Operation (AED)

The R Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.

They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically approved patient care protocol.

Use of the R Series in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:

  • Unconsciousness
  • Absence of breathing .
  • Absence of pulse .

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

and the country of the country of the county of the county of the county of the county of the county of the county of the country of the country of the county of the countryConcurrence of CDRH, Office of Device Evaluation (ODE)
(

W.M.st.

Number K6909

{8}------------------------------------------------

Specifications for the ECG rhythm analysis function are provided in the section, "ECG Rhythm Analysis Algorithm Accuracy" on page A-29.

When the patient is less than 8 years of age or weighs less than 55lbs (25Kg), you must use ZOLL pediatric defibrillation electrodes. Do not delay therapy to determine the patient's exact age or weight.

Intended Use - ECG Monitoring

The unit is intended for use when ECG monitoring is indicated to evaluate the patient's heart rate or ECG morphology. In ECG monitoring mode, the unit is intended to be used by personnel who are qualified by training in the R Series defibrillator, basic life and/or advanced life support, or other physician authorized emergency medical training.

Intended Use - CPR Monitoring

The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.

The CPR monitoring function is not intended for use on patients under 8 years of age.

External Pacemaker (Pacer Version Only)

Intended Use - Pacemaker

This product can be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.

The purposes of pacing include:

Resuscitation from standstill or bradycardia of any etiology:

Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.

As a standby when standstill or bradycardia might be expected:

Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative

{9}------------------------------------------------

to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.

Suppression of tachycardia:

Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.

Pediatric Pacing

Pacing can be performed on pediatric patients weighing 33lb. (15kg) or less using ZOLL pediatric hands-free therapy electrodes. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.

Intended Use - SpO2 Monitoring

The R Series pulse oximeter, with the Masimo® SET® technology and the LNCS® series of oximeter sensors, is indicated for the continuous, nonitoring of arterial oxygen saturation (SpO2) and pulse rate during both no motion and patient motion conditions for adult patients, and no motion conditions for pediatric and neonatal patients in a hospital or prehospital environment.

Intended Use - NIBP Monitoring

The ZOLL R Series NIBP option is indicated for the non-invasive measurement of arterial blood pressure for resting patients in critical care and in-hospital transport.

The R Series NIBP option is designed to measure blood pressure for adult patients (21 years of age and older) and for pediatric patients, as described in the following table:

Pediatric SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age

Intended Use - EtCO2 Monitoring

The ZOLL R Series EtCO2 option with Respironics Novametrix technology is indicated for the continuous noninvasive monitoring of end tidal carbon dioxide (EtCO2) and respiration rate in patients requiring ventilator support, in-hospital transport, or anesthesia.

This option uses the CAPNOSTAT 5 Mainstream CO2 sensor attached to an airway adapter that connects to an endotracheal tube, mask or disposable mouthpiece.

The R Series EtCO2 option is designed to monitor adult, pediatric, and neonatal patients.

The following substances can influence CO2 measurements made with the CAPNOSTAT 5 CO2 sensor:

  • · elevated oxygen levels
    · nitrous oxide

  • · halogenated agents

{10}------------------------------------------------

The R Series EtCO2 option provides settings for high oxygen and/or nitrous oxide compensation. Halogenated anesthetic agents alter CO2 readings, but the R Series unit will monitor CO2 within specifications when these agents are present at normal clinical levels. The presence of Desflurane in the exhaled breath beyond normal values (5%) may positively bias measured carbon dioxide values by up to an additional 3 mmHg.

The R Series EtCO2 option is intended for use only with the ZOLL/Respironics Novametrix CAPNOSTAT 5 Mainstream CO2 Sensor and mainstream airway adapters.

The R Series EtCO2 option can be used on adult patients (21 years of age and older) and on pediatric patients, as described in the following table:

Pediatric SubpopulationApproximate Age Range
Newborn (neonate)Birth to 1 month of age
Infant1 month to 2 years of age
Child2 to 12 years of age
Adolescent12-21 years of age

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.