(51 days)
Not Found
Not Found
No
The document describes a calibration verification/linearity test set for in vitro diagnostic use and does not mention any AI or ML components.
No
This device is for in vitro diagnostic use, specifically for linearity, calibration, and reportable range verification in chemistry systems. It is used to test the performance of other devices, not to directly treat or diagnose patients.
No
The device is a quality control material intended for in vitro diagnostic use to verify linearity, calibration, and reportable range in chemistry systems. It is not used to diagnose a patient's condition.
No
The device description clearly states it is "Quality control material (assayed and unassayed)" and "human based calibration verification materials containing multiple levels". This indicates a physical product (solutions in bottles) and not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The "Intended Use / Indications for Use" section explicitly states that the solutions are "intended for in vitro diagnostic use".
- Device Description: The device is described as "Quality control material (assayed and unassayed)" and "human based calibration verification materials". These are typical descriptions for materials used in laboratory testing to ensure the accuracy of diagnostic assays.
- Performance Studies: The document mentions performance studies comparing the device to a predicate device, which is a common requirement for IVD submissions to regulatory bodies.
- Predicate Device: A predicate device is listed, which is another strong indicator that this device is being considered within the regulatory framework for IVDs.
- Intended User: The intended user is "trained laboratory professionals", which aligns with the use of IVDs in a laboratory setting.
All of these points collectively indicate that the VALIDATE® CM2 Calibration Verification / Linearity Test Set is an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The VALIDATE® CM2 Calibration Verification / Linearity Test Sets are used by trained laboratory professionals for quantitatively verifying calibrating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for the analytes listed on the package insert. These materials are not intended for use as routine quality control materials or as calibration materials.
VALIDATE® CM2 Calibration Verification / Linearity Test Set solutions are intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi automated, and manual chemistry systems. Each VALIDATE® CM2 Calibration Verification / Linearity Test Set consists of two sets of bottles. Set 1 contains BNP, hs-CRP, Troponin-1, and MPO. Set 2 contains NT-proBNP, hs-CRP, Troponin-T and MPO.
Product codes (comma separated list FDA assigned to the subject device)
JJY
Device Description
VALIDATE® CM2 Calibration / Linearity Test Sets are human based calibration verification materials containing multiple levels used to establish the relationship between theoretical operation and actual performance of the included analytes. There exists a linear relationship among each set of solutions.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
trained laboratory professionals
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
The performance of VALIDATE® CM2 Calibration Verification / Linearity Test Set solutions on the Beckman Access II (Tnl, BNP), the Beckman Immage 800 (hsCRP), the Beckman Synchron DxC with Diazyme reagent (MPO), the Roche e411 (TNT, NTproBNP), and the Roche Integra 400+ (hsCRP) has been shown to be substantially equivalent using pre-production lots of VALIDATE® CM2 Calibration Verification / Linearity Test Sets to the VALIDATE® CM1 Calibration Test Set.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Not Found
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
Image /page/0/Picture/0 description: The image shows the logo for Maine Standards Company. The logo consists of the word "maine" above a horizontal line that is intersected by a diagonal line. Below the horizontal line is the word "standards" and below that is the word "COMPANY" in smaller letters.
510(k) SUMMARY
APR 1 6 2009
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: k 090475."
Submitter: Maine Standards Company Address: 765 Roosevelt Trail Windham, ME 04062 Telephone: 207-892-1300 Fax: 207-892-2266 Contact:
Summary prepared on: February 23, 2009
Device classification name: | Multi-Analyte Controls, All Kinds (Assayed and Unassayed) |
---|---|
Device description: | Quality control material (assayed and unassayed)* |
Proprietary Name: | VALIDATE® CM2 Calibration Verification / Linearity Test |
Set | |
Regulation Number: | 21 CFR 862.1660 |
Product Code: | JJY* |
*Note: There is no FDA product code for calibration / linearity materials. Therefore, as with previous submissions by Maine Standards and other calibration verification / linearity manufacturers, JJY has been selected as the "best fit" FDA code for this product.
Regulatory Class: Class I
Predicate Device:
VALIDATE® CM1 Calibration Verification / Linearity Test Set, Maine Standards Company, Windham, ME.
Device description: VALIDATE® CM2 Calibration / Linearity Test Sets are human based calibration verification materials containing multiple levels used to establish the relationship between theoretical operation and actual performance of the included analytes. There exists a linear relationship among each set of solutions.
1
Intended use: The VALIDATE® CM2 Calibration Verification / Linearity Test Sets are used by trained laboratory professionals for quantitatively verifying calibrating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for the analytes listed on the package insert. These materials are not intended for use as routine quality control materials or as calibration materials.
Comparison of VALIDATE® CM2 Calibration Verification / Linearity Test Set to the predicate device:
Table 1 compares characteristics of the VALIDATE® CM2 Calibration Verification / Linearity Test Set with those of the VALIDATE® CM1 Calibration Verification / Linearity Test Set.
| | VALIDATE®
CM1 | VALIDATE®
CM2 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Catalog # | 401 | 402 |
| Intended
Use | For in vitro diagnostic use
in quantitatively verifying
calibration, validating
reportable ranges, and
determining linearity in
automated, semi-
automated and manual
chemistry analyzers. | For in vitro diagnostic use
in quantitatively verifying
calibration, validating
reportable ranges, and
determining linearity in
automated, semi-
automated and manual
chemistry analyzers. |
| Analytes | CK-MB, MYO | Tnl, TnT, BNP, NT-
proBNP,hsCRP,MPO |
| Matrix | Human Serum | Human Serum |
| Number of
Levels | 5 | 2 sets of 6 including base
matrix |
| Preparation | Liquid, ready to use | Liquid, ready to use |
| Packaging | 5 x 3.0mL | 12 x 2.0 mL |
| Stability | Until expiration | Until expiration |
| Storage | -10 to -20°C | -10 to -20°C |
TABLE 1 - Comparison of Devices
The performance of VALIDATE® CM2 Calibration Verification / Linearity Test Set solutions on the Beckman Access II (Tnl, BNP), the Beckman Immage 800 (hsCRP), the Beckman Synchron DxC with Diazyme reagent (MPO), the Roche e411 (TNT, NTproBNP), and the Roche Integra 400+ (hsCRP) has been shown to be substantially equivalent using pre-production lots of VALIDATE® CM2 Calibration Verification / Linearity Test Sets to the VALIDATE® CM1 Calibration Test Set.
2
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The seal is presented in black and white.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Maine Standards Company c/o Ms. Holly A. Cressman Manager QA/RA 765 Roosevelt Trail Windham, ME 04062
APR ] 6 2009
K090475 Re:
Trade/Device Name: VALIDATE CM2 Calibration Verification/Linearity Test Set Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (assayed and unassayed) Regulatory Class: Class I, reserved Product Code: JJY Dated: February 23, 2009 Received: February 24, 2009
Dear Ms. Cressman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
3
Page - 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Corg C. He
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
4
Indications for Use
K090475 510(k) Number:
VALIDATE® CM2 Calibration Verification / Linearity Test Set Device Name:
Indications For Use:
VALIDATE® CM2 Calibration Verification / Linearity Test Set solutions are intended for in vitro diagnostic use in the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi automated, and manual chemistry systems. Each VALIDATE® CM2 Calibration Verification / Linearity Test Set consists of two sets of bottles. Set 1 contains BNP, hs-CRP, Troponin-1, and MPO. Set 2 contains NT-proBNP, hs-CRP, Troponin-T and MPO.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Lucy Phillips
Division City Off
Division Sian-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 199 0475
Page 1 of