K Number
K090389

Validate with FDA (Live)

Date Cleared
2009-07-14

(147 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AP-1000 Blood Glucose Monitoring System (BGMS) is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Viro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

The AP-1000 Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The AP-1000 Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

MAJOR Level I/Level II Control Solutions are for use with the AP-1000 meter and AP-1000 Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Device Description

Based on an electrochemical biosensor technology and the principle of capillary action, AP-1000 Blood Glucose Monitoring System only needs a small amount of blood. Capillary action at the end of the test strip draws the blood into the action chamber and your blood glucose result is precisely and displayed in 6 seconds.

AI/ML Overview

The provided text describes the AP-1000 Blood Glucose Monitoring System and mentions pre-clinical and clinical data without detailing the specifics of the study, acceptance criteria, or performance against specific metrics. Therefore, a comprehensive answer to all points of your request cannot be fully extracted from the provided document.

However, based on the information available, here's what can be gathered regarding the study and acceptance criteria:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state the acceptance criteria for specific performance metrics (e.g., accuracy percentages within certain glucose ranges) or fully report the device's performance against such criteria. It only indicates that "Pre-clinical and clinical data are employed upon submission of this 510(K) premarket notification according to the Guidance Document for In Vitro Diagnostic Test System; Guidance for Industry and FDA document provided by CDRH/ FDA." This implies that the study was conducted to meet the requirements outlined in the FDA guidance for in vitro diagnostic test systems, but the specific numerical acceptance criteria and a table of reported performance against these criteria are not included in this document.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin of the data, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide information on the number of experts used to establish ground truth or their qualifications.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

The document does not mention an MRMC comparative effectiveness study or any effect size of AI assistance on human readers, as the device is a standalone blood glucose monitoring system and not an AI-assisted diagnostic tool for image interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The AP-1000 Blood Glucose Monitoring System is a standalone device intended for quantitative measurement of glucose. Its performance would inherently be evaluated in a standalone manner (algorithm/device only) without a "human-in-the-loop" as it's a direct measurement device rather than an interpretive aid. The document implies a standalone evaluation was performed by stating "Pre-clinical and clinical data are employed," but details are missing.

7. The Type of Ground Truth Used

For a blood glucose monitoring system, the "ground truth" for accuracy studies would typically be established by comparing the device's readings to laboratory reference methods (e.g., YSI analyzer) known for their high accuracy. The document does not explicitly state the specific type of ground truth used but it would be expected to be a laboratory reference method.

8. The Sample Size for the Training Set

The document does not specify any training set sample size. For a blood glucose monitoring system based on electrochemical biosensor technology, the development would involve calibration and optimization using blood samples, but these are generally not referred to as "training sets" in the same way as machine learning models.

9. How the Ground Truth for the Training Set Was Established

Not applicable as a distinct "training set" in the context of machine learning is not mentioned or relevant for this type of device. The calibration and development would have been based on comparison to laboratory reference methods.

In summary, while the document indicates that pre-clinical and clinical data were used to demonstrate safety and effectiveness in accordance with FDA guidance, it lacks the detailed information requested regarding specific acceptance criteria, study sizes, ground truth establishment, and expert involvement.

{0}------------------------------------------------

K09038-9

Image /page/0/Picture/1 description: The image shows the logo for BESTGEN Biotech Corp. The word BESTGEN is in large, bold, black letters. Below that, in smaller letters, is the text Biotech Corp.

技股份有限公司
技股份有限公司
技股份有限公司

235 6 北棒中彩市道 --Jhonghe Gity, Taipoi Coun 086 2 8227 2188

JUL 1 4 2009

510(K) Summary of Safety and Effectiveness II.

(Per 21 CFR 807.92)

2.1. General Information Establishment

  • BESTGEN Biotech Corp. Manufacturer: 를
  • 7F, No.186, Jian 1st Rd. Jhonghe City, Taipei County, 23511, Taiwan, ROC Address: 통
  • 9102780 Owner Number: 월
  • E-mail: ceirs.jen@msa.hinet.net Dr. Jen, Ke-Min ■ Contact Person: 886-3-5208829 (Tel); 886-3-5209783 (Fax)

. No.58, Fu Chiun Street, Hsin Chu City, 30067, Taiwan, ROC Address:

  • May 10, 2008 Date Prepared: ●
    Device

.

  • AP-1000 Blood Glucose Monitoring System Proprietary Name: ●
    • Blood Glucose Monitoring System
    • SYSTEM, TEST, BLOOD GLUCOSE, OVER Classification Name: THE COUNTER, Class II

Product Code:

NBW

2.2. Safety and Effectiveness Information

  • Predicate Device: ●
    Common Name:

Claim of Substantial Equivalence (SE) is made to EASY CHECK Blood Glucose Monitoring System (K062538)

Device Description: Based on an electrochemical biosensor technology and . the principle of capillary action, AP-1000 Blood Glucose Monitoring System only needs a small amount of blood. Capillary action at the end of the test strip draws the blood into the action chamber and your blood glucose result is precisely and displayed in 6 seconds.

{1}------------------------------------------------

..... 235 台北縣中和市建一路166號七樓(社)

No 186, Jian 1st Rd., Jhongne City, Taipei County 235, Taiwan R.O.C.

TE:+886 2 8227 2188 Fax +006262272178

Intended Use: .

The AP-1000 Blood Glucose Monitoring System (BGMS) is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Viro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

The AP-1000 Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The AP-1000 Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

MAJOR Level I/Level II Control Solutions are for use with the AP-1000 meter and AP-1000 Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Synopsis of Test Methods and Results .

Pre-clinical and clinical data are employed upon submission of this 510(K) premarket notification according to the Guidance Document for In Vitro Diagnostic Test System; Guidance for Industry and FDA document provided by CDRH/ FDA.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "BESTGEN" in large, bold, black letters. Below the word "BESTGEN" is the text "Bestgen Biotech Corp." in a smaller font size. The text is also in black. The background of the image is white.

235 0 北 科 中 中 文 第 1665 元 元 166 元 元 ( ) ( ) ( ) ( ) ( ) ( ) ) ( ) ) ( ) ( ) ) ( ) ) ( ) ) ( ) ) ( ) ( ) ) ( ) ( ) ) ( ) ( ) ) ( ) ( ) ) ( ) ( ) ( ) ( ) ) ( ) ( ) ( ) ( ) ( ) ( F. No. 188. Jian 1st Rd. Jhonano City. Taipei County 235. Taiwan R.O.G. : +886 2 8227 2183 +806 2 8227 2171

Substantial Equivalence (SE) .

A claim of substantial equivalence is made to EASY CHECK Blood Glucose Monitoring System (K062538). Both of them have the same working principle and technologies. The differences are test time, coding method, test range, sample volume, meter dimension, weight, HCT range, and memory data Besides, the subject device is intended for the quantitative number. measurement of glucose in fresh capillary whole blood from the finger but the predicate device can test whole blood from the finger, forearm and palm. Thus the differences are due to the feature design aspects, not relating to the safety or effectiveness aspects. They are substantially equivalent.

Keaton

5

Dr. Jen, Ke-Min official correspondent for BESTGEN BIOTECH CORP.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

Bestgen Biotech Corp. c/o Dr. Ke-Min Jen 58 Fu Chuin Street Hsin Chu City, China (Taiwan) 30067

JUL 1 4 2009

Re: K090389

Trade/Device Name: AP-1000 Blood Glucose Monitoring System Regulation Number: 21 CFR § 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: NBW, CGA, JJX Dated: June 14, 2009 Received: June 18, 2009

Dear Ke-Min Jen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical devicerelated adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'' (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

C. H.

Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): K090389

Device Name: AP-1000 Blood Glucose Monitoring System

Indication For Use:

The AP-1000 Blood Glucose Monitoring System (BGMS) is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home {over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates.

The AP-1000 Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Meter. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The AP-1000 Blood Glucose Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. AP-1000 Blood Glucose Test Strips must be used with the AP-1000 Blood Glucose Meter. Testing is done outside the body (In Vitro diagnostic use). They are indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

MAJOR Level I/Level II Control Solution are for use with the AP-1000 meter and AP-1000 Blood Glucose Test Strips as a quality control check to verify the accuracy of blood glucose test results.

Prescription Use And/Or (21 CFR Part 801 Subpart D)

/Or

Over the Counter Use V (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Carol C. Benson

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K090389

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.