(60 days)
Not Found
No
The summary describes standard signal processing and calculation of indices, with no mention of AI or ML terms, training data, or complex algorithms typically associated with AI/ML.
No
The device is used to measure and present data for interpretation, and it explicitly states it should be used "only as an adjunct to clinical history and the results of other noninvasive and /or invasive tests," indicating it is for diagnostic or informational purposes, not therapeutic intervention.
Yes
The device is used to measure and present data for interpretation by a trained physician, and its use is described as an "adjunct to clinical history and the results of other noninvasive and /or invasive tests," which aligns with the purpose of a diagnostic device.
No
The device description explicitly states that the device consists of hardware components including an IBM compatible PC, electrodes, an electronics board, and a patient interface module, in addition to software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The MI-1000 system measures electrical activity of the heart using skin surface electrodes. It also applies low-level electrical stimulation externally.
- No Sample Analysis: The device does not analyze any biological samples taken from the patient. It interacts with the patient's body externally.
Therefore, the MI-1000 system falls under the category of a medical device that interacts with the patient's body directly, rather than an IVD that analyzes samples taken from the body.
N/A
Intended Use / Indications for Use
The MI-1000 system is used to measure Wavelet Surface Residuum indices at rest and during low level electrical stimulation during the cardiac refractory period. This data is presented in a graphical format for interpretation by a trained physician. The MI-1000 system is also used for the measurement of SAECG indices. The MI-1000 system should be used only as an adjunct to clinical history and the results of other noninvasive and /or invasive tests.
Product codes (comma separated list FDA assigned to the subject device)
DPS, DPX
Device Description
The device consists of an IBM compatible PC, nine skin surface electrodes, one electronics board, a patient interface module and software. Seven of the electrodes are used for sensing cardiac electrical activity. The remaining two electrodes are used to apply low level current pulses (maximum 40 milliamps) to the patient. This technology uses noninvasive, externally applied, subthreshold (low amplitude) far-field stimulus while acquiring electrocardiogram data.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
trained physician
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
A complete software validation and verification was completed to verify that the acquired data was able to be correctly analyzed and the Wedensky indices were correctly calculated and displayed. All testing successfully met the required criteria.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
No clinical testing was required.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.2340 Electrocardiograph.
(a)
Identification. An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.(b)
Classification. Class II (performance standards).
0
Pg 1 of 2
APR - 3 2009
510(k) Summary
バー シ
| Submitter: | Harbinger Medical, Inc.
10125 Crosstown Circle, Suite 105
Eden Prairie, Minnesota 55344 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Harold Hoium
Telephone: (952) 943-1686
Fax: (952) 943-1686 |
| Date Prepared: | January 30, 2009 |
| Trade Name: | Micro-Induction 1000 system (MI-1000 System) |
| Classification Name | 21 CFR 870.2340 (Electrocardiograph) and |
| And Regulation Number | 21CFR 870.2360 (Electrode Electrocardiography) |
| Class: | Class II |
| Product Code: | DPS and DPX |
| Predicate Devices: | Micro-Induction 1000 System (MI-1000) (K013615) |
| Device Description: | The device consists of an IBM compatible PC, nine skin surface electrodes, one electronics board, a patient interface module and software. Seven of the electrodes are used for sensing cardiac electrical activity. The remaining two electrodes are used to apply low level current pulses (maximum 40 milliamps) to the patient. This technology uses noninvasive, externally applied, subthreshold (low amplitude) far-field stimulus while acquiring electrocardiogram data. |
| Intended Use: | The MI-1000 system is used to measure Wavelet Surface
Residuum indices at rest and during low level electrical
stimulation during the cardiac refractory period. This data
is presented in a graphical format for interpretation by a
trained physician. The MI-1000 system is also used for the
measurement of SAECG indices. The MI-1000 system
should be used only as an adjunct to clinical history and
the results of other noninvasive and /or invasive tests. |
The MI-1000 system is a tool used to identify heart cells that are significantly less stable than normal or conduct
1
K090249
pg 1 of 2
slower than normal and, thus, are an indication of cardiac electrical problems.
The modified MI-1000 system is substantially equivalent to Technological Characteristics: the original MI-1000 system and utilizes all of the same fundamental technology to capture the ECG signal from the patient. The only change for the current model is in the signal post-processing and display. All of the displayed information is subject to physician interpretation.
A complete software validation and verification was Performance Data-non-clinical: completed to verify that the acquired data was able to be correctly analyzed and the Wedensky indices were correctly calculated and displayed. All testing successfully met the required criteria. No clinical testing was required. Performance Data - clinical:
The modified Micro-Induction 1000 system is substantially Conclusion: equivalent to the previously cleared version of the Micro-Induction 1000 system.
2
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 3 2009
Harbinger Medical, Inc. C/O Harold Hoium 10125 Crosstown Circle, Suite 105 Eden Prairie, Minnesota 55344
Re: K090249
Trade/Device Name: Micro-Induction 1000 System Regulation Number: 21 CFR 870.2340 (Electrocardiograph) Regulatory Class: Class II Product Code: DPS and DPX Dated: March 18, 2009 Received: March 20, 2009
Dear Mr. Hoium:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
3
Page 2 - Mr. Harold Hoium
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
[signature]
0 / Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
. Enclosure
4
Indications for Use
510(k) Number (if known): K090249
Device Name: _ Micro-Induction (MI) 1000 System
Indications For Use:
The MI-1000 is used to measure Wavelet Surface Residuum indices at rest and during low level electrical stimulation during the cardiac refractory period. This data is presented in a graphical format for interpretation by a trained physician. The MI-1000 is also used for the measurement of SAECG indices. The MI-1000 should be used only as an adjunct to clinical history and the results of other noninvasive and/or invasive tests.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for B Zuckerman
(Division Sign-Off) | 4/3/09 |
---|---|
Division of Cardiovascular Devices | |
510(k) Number | K090249 |
Page 1 of 1