K Number
K083921
Manufacturer
Date Cleared
2009-05-20

(140 days)

Product Code
Regulation Number
886.5916
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SynergEyes® SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses for daily wear are indicated for use in the correction of hyperopic, myopic and astigmatic refractive error including presbyopia, in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

Device Description

The SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Daily Wear Contact Lens is a combination of high Dk rigid gas permeable contact lens corneal option surrounded by a soft hydrophilic silicone-hydrogel skirt (periphery) that straddles the limbus of the eye :

  • in the power range of -20.00 to +20.00 diopters for sphere,
  • . cylinder powers up to 6.00 diopters for astigmatism
  • with center thickness from 0.12mm to 0.30mm .
  • with base curves of 7.10mm to 9.00mm
  • with diameter of 14.50mm
  • Add Powers 1.00 to 4.00 (multifocal)
    This lens material for the rigid portion is petrafocon A, an upgrade high Dk silicone base RGP material, lathe cut, surrounded by soft hydrophilis silicone-hydrogel copolymer (hem-larafilizon A), sterilized by means of e-beam sterilization. When placed on the human cornea, the SynergEyes™ SiH Hybrid Contact Lens acts as a refracting medium to focus light rays onto the The device is available in the following indications: spherical, toric, and multifocal, with retina. spherical or aspheric surfaces in violet visibility tim the RGP center, and multing, will may device is substantially equivalent to the SyneryEyes™ A and M (paflufocon D hem-iberfilion A) Hybrid Daily Wear Contact Lens predicate devices manufactured by the sponsor, Synerglyes, Inc.
    The SynergEyes™ Sill Hybrid Daily Wear Contact Lens center is a highly oxygen permeable (Dk 130) rigid gas material of (petrafocon A). The soft skirt is comprised of silicone-hydrogel copolymer (hem-larafilcon A) of 33% water and 73% polymer.
    The junction between the rigid material is bound by a proprietary chemical bonding method.
AI/ML Overview

Acceptance Criteria and Study for SynergEyes™ SiH Hybrid Contact Lens

This document describes the acceptance criteria and the study conducted to demonstrate the substantial equivalence of the SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lens to its predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by demonstrating "no worse than" performance compared to the predicate device across various safety and effectiveness measures. The study aimed to show substantial equivalence, meaning the new device performs at least as well as the existing one.

Acceptance Criteria (Implicit: No Worse Than Predicate)Reported Device Performance (vs. Control Lens)
Primary Safety Variables:
Adverse Event RateTest Cohort: 4 AEs (2 SAEs, 1 abrasion, 1 microbial keratitis)
Control Cohort: 1 AE (1 SAE, liver failure - unrelated to lens)
Slit Lamp Findings (e.g., staining, other signs)Test cohort eyes presented with more staining overall.
All other slit lamp findings were similar rates and severities.
Effectiveness Variables:
Discontinuation RateTest Cohort: 15 (73.2% completed)
Control Cohort: 4 (87.1% completed)
Visual Acuity Stability (Snellen)Stable for both cohorts throughout the study.
Only 2 Test cohort eyes and 2 Control cohort eyes reported 2 or more line drops at final visit.
Average Daily Wearing TimeSimilar (Test: 11.6-13.0 hrs/day; Control: 12.1-12.8 hrs/day)
Lens Deposit EvaluationsSimilar results between cohorts.
Fitting EvaluationsSimilar results between cohorts.
Lens ReplacementsTest lenses replaced more for parameter change.
Control lenses replaced more for discomfort.
Symptoms/Complaints (e.g., discomfort, dryness, halos)Test cohort reported proportionately greater symptoms (e.g., 8.6% for halos, 16.1% for dryness).
Differences were generally small (3.7% or less), except for halos and dryness.

Note: The submission frames the "acceptance criteria" as demonstrating that "the test lens performed no worse than the control lens on a daily basis," thereby establishing substantial equivalence. Specific quantitative thresholds for "no worse than" are not explicitly stated in the provided text for each outcome.

2. Sample Size and Data Provenance for Test Set

  • Sample Size:
    • Total Subjects Randomized and Dispensed: 87
    • Test Cohort (dispensed): 56 subjects (eyes implicitly, given contact lens context)
    • Control Cohort (dispensed): 31 subjects (eyes implicitly)
    • Completed 90-day study: 41 Test cohort subjects (73.2%), 27 Control cohort subjects (87.1%)
  • Data Provenance: The study was conducted at 8 investigational sites. The country of origin is not explicitly stated, but the submission is to the FDA (United States), suggesting the study likely took place in the USA. The study was prospective and unmasked, open-label.

3. Number of Experts and Qualifications for Ground Truth

This type of study (clinical trial for contact lenses) does not typically involve external experts establishing a "ground truth" for the test set in the way an imaging AI study would. Instead, the ground truth is clinical data collected by investigators/clinicians at the 8 investigational sites.

Their qualifications are not explicitly stated but would be assumed to be licensed ophthalmologists or optometrists experienced in fitting and evaluating contact lenses and ocular health, as is standard for clinical trials. They performed slit-lamp examinations, visual acuity assessments, and recorded adverse events and patient complaints.

4. Adjudication Method for Test Set

There is no mention of an independent adjudication method (like 2+1 or 3+1 consensus) for the collected clinical data. The data interpretation and reporting appear to be based on the findings of the individual investigators at the 8 sites, compiled and analyzed statistically.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This is a clinical trial comparing a new device (contact lens) to a predicate device in human subjects, not an assessment of AI's assistance to human readers. Therefore, an effect size of human readers improving with/without AI assistance is not applicable.

6. Standalone (Algorithm Only) Performance

No, a standalone (algorithm only) performance assessment was not done. This device is a physical contact lens, not a software algorithm.

7. Type of Ground Truth Used

The "ground truth" in this context refers to the clinical observations and outcomes data collected directly from the study subjects. This includes:

  • Adverse events (e.g., abrasions, microbial keratitis)
  • Slit lamp findings (e.g., staining, other ocular signs)
  • Visual acuity measurements
  • Patient-reported symptoms and complaints (e.g., discomfort, dryness, itching/burning, halos)
  • Wearing times
  • Lens deposit and fitting evaluations
  • Reasons for discontinuation

These are directly observed or reported clinical realities in the study population.

8. Sample Size for the Training Set

This submission describes a clinical trial for a physical medical device (contact lens). There is no "training set" or "algorithm" involved in the development or evaluation of this contact lens, as it is not an AI/ML device. Therefore, this question is not applicable.

9. How Ground Truth for Training Set Was Established

As there is no training set for an algorithm, this question is not applicable.

{0}------------------------------------------------

K083921

510 (K) SUMMARY

MAY 20 2009

Applicant's Name and Address

SynergEyes, Inc. 2232 Rutherford Road Carlsbad, CA 92008 (760) 476 9410 (760) 476-9340 FAX

Richard Lippman, OD FAAO Vice President Ophthalmic Regulatory Products R.P. Chiacchierini & Associates, LLC 15825 Shady Grove Road Suite 30 Rockville, Maryland 20850 (240) 683-3738 rel@rpcaconsulting.com

1. Identification of device

Common Name: Trade Name:

Contact Person

Classification: Device classification:

Contact Lens SynergEyes™ SiH(petrafocon A hem-larafilcon A) Hybrid Daily Wear Lens Daily Wear Rigid Gas Permeable Contact Lens Class II (21 CFR 886.5916)

2. Description of device

The SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Daily Wear Contact Lens is a combination of high Dk rigid gas permeable contact lens corneal option surrounded by a soft hydrophilic silicone-hydrogel skirt (periphery) that straddles the limbus of the eye :

  • in the power range of -20.00 to +20.00 diopters for sphere,
  • . cylinder powers up to 6.00 diopters for astigmatism
  • with center thickness from 0.12mm to 0.30mm .
  • with base curves of 7.10mm to 9.00mm
  • � with diameter of 14.50mm
  • Add Powers 1.00 to 4.00 (multifocal)

This lens material for the rigid portion is petrafocon A, an upgrade high Dk silicone base RGP material, lathe cut, surrounded by soft hydrophilis silicone-hydrogel copolymer (hem-larafilizon A), sterilized by means of e-beam sterilization. When placed on the human cornea, the SynergEyes™ SiH Hybrid Contact Lens acts as a refracting medium to focus light rays onto the The device is available in the following indications: spherical, toric, and multifocal, with retina. spherical or aspheric surfaces in violet visibility tim the RGP center, and multing, will may device is substantially equivalent to the SyneryEyes™ A and M (paflufocon D hem-iberfilion A) Hybrid Daily Wear Contact Lens predicate devices manufactured by the sponsor, Synerglyes, Inc.

The SynergEyes™ Sill Hybrid Daily Wear Contact Lens center is a highly oxygen permeable (Dk 130) rigid gas material of (petrafocon A). The soft skirt is comprised of silicone-hydrogel copolymer (hem-larafilcon A) of 33% water and 73% polymer.

{1}------------------------------------------------

SynergEyes™, Inc. Carlsbad, CA 92008

The junction between the rigid material is bound by a proprietary chemical bonding method.

3. Intended use

SynergEyes® SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses for daily wear are indicated for use in the correction of hyperopic, myopic and astigmatic refractive error including presbyopia, in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

4. Predicate device

For Design: Predicate- K051035 SynergEyes™ A & M (paflufocon D hem-iberfilcon A) Hybrid Contact Lens

The SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lens design is based on the SynergEyes™ (paflufocon D hem-iberfilcon A) Hybrid Contact Lens. The designs of both lenses are based on a rigid gas permeable center optic and a soft hydrogel skirt with a special proprietary bonding process to join the center and the peripheral portions at the junction of the optic edge where it joins the peripheral ring.

For Material Similarities: Predicate- K051035, K052560, K033919, K970746

The SynergEyes™ SiH (petrafocon A hem-larafiicon A) Hybrid Contact Lens has a rigid gas permeable center optic portion which is a high Dk (Dk 130) silicone based polymer (petrafocon A) surrounded by a silicone-hydrogel high Dk (Dk 84.2) soft skirt. The predicate lens, the SynergEyes™ A (paflufocon D hem-iberfilcon A) Hybrid Contact Lens has the center materialfluorosilicone acrylate, Paragon HDS material (Dk 100) permeability, surrounded by a hemabased polymer soft ring with Dk 9.1. The materials in both the subject device and predicate device are characterized as Group I low water, non-ionic in nature for their soft portions as the water content for the subject device is 33% and the predicate device is 25.0%. The water content for the rigid portion of both lenses is < 1%.

For Physical and Optical Properties: Predicate- K050135

The lens being cited as the predicate lens for the subject SynergEves™ SiH (petrafocon A. hemlarafilcon A) Hybrid Contact Lens is the SynergEyes™ A and M (paflufocon D hem iberfilcon A) Hybrid Contact Lens (K051035). The physical and optical properties that characterize both the subject and predicate device are shown in a side-by-side companson chart. Properties such as oxygen permeability for both the center portion and peripheral skirt demonstrate the subject lens material to have a substantially higher permeability in both the RGP center and the soft skirt. Other properties that are compared for similarities and differences include refractive index, wetting angle, specific gravity, Shore Hardness, luminance transmittance, surface charge, surface treatment, water content, base curve, diameter, power range, chord diameter, UV additive,

5. Characteristics

The physical and dimensional characteristics of the SynergEyes™ SiH Hybrid Contact Lens are compared to the characteristics of the predicate device SynergEyes™ A and M Hybrid Contact Lens in the following table.

{2}------------------------------------------------

SynergEyes™, Inc. Carlsbad, CA 92008

SynergEyes™Hybrid ContactLensK051035SynergEyes™HybridSilicone/HydrogelLens(Subject Device)
Lens Characteristics
ManufacturerSynergEyes™ Inc.SynergEyes, Inc.
Base Curves7.10-9.00mm7.10-9.00mm
RGP Center8.40mm8.40mm
Optic Zone Diameter7.80mm7.80mm
Lens DesignsSphere, Aspheric,Front Surface Toric,MultifocalSphere, Aspheric,Front Surface Toric,Multifocal
Diameters:14.5mm14.5mm
Power Range-20.00 to + 20.00D-20.00 to +20.00
Cylinder Range0.50 to 6.00D0.50 to 6.00D
Refractive Index (RGP)RGP Center:1.442RGP Center:1.442S/H Skirt:1.435
Wetting (Contact) AngleRGP Center:42°RGP Center:43.19°Soft Skirt:29.40°
Specific Gravity (RGP)1.101.15±0.025
HardnessRGP:74.8Soft Skirt:RGP:76.0Soft Skirt:53.4Dry-Hydrated- 8.9(Shore A)
Indications for UseDaily WearDaily Wear
UV BlockingNoYes
MaterialRGP Center:Paflufocon DSoft Skirt: iberfilconA (HEMA, MEMA)RGP Center:petrafocon ASoft Skirt:hem-larafilcon A
TintVisibility BlueVisibility violet
Soft Skirt Water ContentHEMA Skirt:25%S/H Skirt (Finished):27.99%
Core (RGP) Water Content< 1%<1%

6. Non clinical studies

A battery of pre-clinical studies were performed on the subject device in accordance with the Guidance for Daily Wear Contact Lenses, May 1994.

Toxicology Studies were conducted under GLP and determined that the SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses were biocompatible for the intended use in accordance with ISO 10993 Standard for the following tests:

  • . Cytotoxicity
  • . Acute Ocular Irritation
  • . Systemic Injection
  • . Guinea Pig Maximization
  • . 22 Day Ocular Irritation Study in Rabbits

3

{3}------------------------------------------------

Physico-chemical Studies including physical, optical and chemical properties as identified in Daily Wear Guidance for Contact Lenses demonstrated equivalent performance to the predicate device as witnessed in the chart listed above.

All pre-clinical and non-clinical studies demonstrated that the final lens subject to extractions and simulated in the above tests performed as expected with no findings related to safety that would be considered as unsafe for human use in the indications set forth for daily wear.

8. Clinical data

A three month daily wear clinical study was performed on the subject device. The study was a prospective, unmasked, open label study of a total of 98 subjects, in an approximate 2:1 ratio of test to control.

The test lens was the SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lens, and the control lens was the SynergEyes™ A (paflufocon D hem-iberficon A) Hybrid Contact Lens. The study was performed at 8 investigational sites, and was designed to demonstrate that the test lens performed no worse than the control lens on a daily basis.

Eighty seven (87) subjects were randomized and dispensed lenses. The ratio of dispensed Test cohort subjects to dispensed Control cohort subjects was 1.81 to 1. The 90 day study period was completed by 73.2% (41 out of 56) dispensed Test cohort subjects and 87.1% (27 out of 31) dispensed Control cohort subjects.

Nineteen (19) subjects discontinued from the study (15 Test and 4 Control) with the most common reason for discontinuation reported as Subject Decision for the Test cohort (40% or 6/15) and Discomfort for the Control cohort (75% or 3/4). One Test cohort subject was discontinued for , Adverse Event and 1 Test cohort subject was discontinued for Positive Slit Lamp.

Reason for DiscontinuationTest CohortControl CohortOverall
Adverse Event16.7%00.0%5.3%
Positive Slit Lamp16.7%00.0%15.3%
Comfort Related213.3%375.0%526.3%
Subject Decision640.0%125.0%736.8%
Protocol Violation213.3%00.0%210.5%
Unacceptable Visual Acuity213.3%00.0%210.5%
Lost to Follow-up16.7%00.0%15.3%
Total Discontinued1519

Reasons for Subiect Discontinuation

The demographics of the study subjects enrolled showed a slightly older population (average age: Control 40.6 / Test 37.3) with a greater proportion of female to male ratio: Control 3.9:1. Test 2.9:1) in the Control cohort as compared to the Test cohort. The majority (greater than 87%) of the subjects in the study were Caucasian for both cohorts.

A total of 5 adverse events were reported for 5 eyes during the study with 4 adverse events reported for the Test cohort and 1 adverse event reported for the Control cohort. Three (3) of the 5 adverse events (2 Test/ 1 Control) were reported as serious adverse events. Of the 2 Test cohort serious adverse events, 1 was a minor abrasion where the investigator noted a grade 1 (trace) infiltrate and 1+ cells in the anterior chamber, all of which had cleared in 4 days. The second serious adverse event was a microbial keratitis which was treated and cleared. The

4

{4}------------------------------------------------

SynergEyes™, Inc. Carlsbad, CA 92008

single serious adverse event reported for the Control cohort was liver failure related to the use of homeopathic products.

Slit lamp findings reported during the study were compared between the Test and the Control cohorts and related to the baseline rates as well. Results of the slit lamp examinations showed the Test cohort eyes presenting with more staining overall when compared to the Control cohort examinations. All other slit lamp findings were reported a similar rates and severities when looking at the overall visit combined findings for both the Test and the Control cohorts.

Symptoms problems and complaints were compared between the Test and the Control cohorts and reviewed against the baseline proportions as well. The Test cohort eyes reported proportionately greater symptoms (1.1% for itching/burning to 16.1% for dryness) when compared to the Control cohort eyes except for excessive tearing (essentially equal) and variable vision (3.6% Control). Most of the differences in symptoms rates were small (3.7% or less) except for halos (8.6% Test) and dryness (16.1% Test).

Snellen visual acuity with contact lenses remained stable throughout the study for both the test and the control cohorts with only 2 Test cohort eyes and 2 Control cohort eyes reporting 2 or line drops of Snellen lens visual acuity with the contact lenses at the final visit.

Average daily wearing times were similar between the two cohorts and remained stable during the study and averaged between 11.6 to 13.0 hours per day for the Test cohort and 12.1 to 12.8 hours per day for the Control cohort.

Lens deposit evaluations and fitting evaluations provided similar results between the two cohorts. Lens replacements indicated that the Test lenses were replaced more frequently for parameter change and the Control lenses were replaced more frequently for discomfort.

In conclusion, the evaluation of primary safety variables and effectiveness of the test product SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lens, support a conclusion of substantial equivalence between the test and control lens evaluated in the study.

9. Risk and Benefits:

The risks of the subject device are the same as those normally attributed to the wearing of RGP and soft (hydrophilic) contact lenses on a daily wear base. The benefits to the patient are the same as those for other RGP and soft (hydrophilic) contact lenses. The subject device has characteristics associated with it that should favor the risk benefit ratio in favor of the new device. Overall, the risks and benefits associated with daily wear contact lenses are the same as for other daily wear contact lenses and raise no additional concerns for safety or effectiveness.

5

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's logo, which is a stylized caduceus, a symbol of medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.

Public Health Service

MAY 20 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SynergEyes, Inc. c/o Richard E. Lippman, O.D., F.A.A.O. Vice President for Ophthalmic Product Regulatory Affairs R.P. Chiacchierini & Associates, LLC 15825 Shady Grove Road, Suite 30 Rockville, Maryland 20850

Re: K083921

Trade/Device Name: SynergEyes™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HQD Dated: May 8, 2009 Received: May 11, 2009

Dear Dr. Lippman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Malvina B. Eglehmis, m.s.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known) K083921

Device Name: SynergEyes® SiH (petrafocon A hem-larafilcon A)) Hybrid Contact Lens

Indication for Use

SynergEyes® SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses for daily wear are indicated for use in the correction of hyperopic, myopic and astigmatic refractive error including presbyopia, in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

Over-the -counter-use OR

Thyphith

(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices

510(k) Number K083921

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.