K Number
K083623
Device Name
CARPENTIER-MCCARTHY-ADAMS IMR ETLOGIX ANNULOPLASTY RING, MODEL 4100
Date Cleared
2009-04-15

(128 days)

Product Code
Regulation Number
870.3800
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Carpentier-McCarthy-Adams IMR ETlogix Mitral Annuloplasty Ring is indicated for the correction of mitral valvular insufficiency where the lesions are not so severe as to require total valve replacement. The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendineae. The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus. For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures of chordae tendineae, or chordal hypertrophy.
Device Description
The Carpentier-McCarthy-Adams IMR ETlogix annuloplasty ring, model 4100 (IMR ETlogix) is constructed of titanium alloy and has a sewing ring margin that consists of a layer of silicone rubber covered with a woven polyester cloth.
More Information

No
The document describes a physical annuloplasty ring and its intended use in surgical procedures. There is no mention of software, algorithms, or data processing that would indicate the use of AI/ML.

Yes
The device is described as an annuloplasty ring for the correction of mitral valvular insufficiency, which is a medical condition, making it a therapeutic device.

No
The device is an annuloplasty ring used for the surgical correction of mitral valvular insufficiency, not for diagnosing medical conditions.

No

The device description explicitly states it is constructed of titanium alloy and silicone rubber, indicating it is a physical implantable device, not software.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In Vitro Diagnostics are medical devices used to perform tests on samples taken from the human body (like blood, urine, or tissue) to provide information about a person's health.
  • Device Description and Intended Use: The Carpentier-McCarthy-Adams IMR ETlogix Mitral Annuloplasty Ring is a physical implant designed to correct mitral valvular insufficiency by remodeling the mitral annulus. It is surgically implanted directly into the patient's body.
  • Lack of IVD Characteristics: The description does not mention any testing of samples, analysis of biological markers, or any function related to diagnosing a condition based on in vitro analysis.

Therefore, this device falls under the category of a surgically implanted medical device, not an In Vitro Diagnostic.

N/A

Intended Use / Indications for Use

The IMR ETlogix Annuloplasty Ring is indicated for the correction of mitral valvular insufficiency where the lesions are not so severe as to require total valve replacement.

The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendineae.

The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus.

For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures of chordae tendineae, or chordal hypertrophy.

Product codes

KRH

Device Description

The Carpentier-McCarthy-Adams IMR ETlogix annuloplasty ring, model 4100 (IMR ETlogix) is constructed of titanium alloy and has a sewing ring margin that consists of a layer of silicone rubber covered with a woven polyester cloth.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

mitral valvular

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Functional/Safety Testing: The Carpentier-McCarthy-Adams IMR ETlogix mitral annuloplasty ring has successfully completed design verification testing.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

Carpentier-Edwards Classic® ring (K912554), Carpentier-Edwards Physio® annuloplasty ring (K926138), GeoForm® annuloplasty ring (K032250)

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.3800 Annuloplasty ring.

(a)
Identification. An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

0

K083623

Image /page/0/Picture/1 description: The image shows a logo with the word "Edwards" written below it. The logo is a square grid with a stylized letter "E" in the center. The letter "E" is surrounded by circular shapes and lines, creating a complex and abstract design.

510(k) Summary

APR ] 5 2009

Submitter:Edwards Lifesciences LLC
Contact Person:Martin A. Kaufman
Director - Regulatory Affairs
Edwards Lifesciences
One Edwards Way
Irvine, CA 92614
martin_kaufman@edwards.com
(ph) 949.250.2708
(fax) 949.250.3630
Date Prepared:December 3, 2008
Trade Name:Carpentier-McCarthy-Adams IMR ETlogix™ Mitral Annuloplasty Ring, or IMR ETlogix™
Mitral Annuloplasty Ring
Classification Name:Class II, CFR 870.3800 Annuloplasty Ring, KRH
Predicate Device(s):Carpentier-Edwards Classic® ring (K912554)
Carpentier-Edwards Physio® annuloplasty ring (K926138)
GeoForm® annuloplasty ring (K032250)
Device Description:The Carpentier-McCarthy-Adams IMR ETlogix annuloplasty ring, model 4100 (IMR
ETlogix) is constructed of titanium alloy and has a sewing ring margin that consists of a
layer of silicone rubber covered with a woven polyester cloth.
Indications for Use:The IMR ETlogix Annuloplasty Ring is indicated for the correction of mitral valvular
insufficiency where the lesions are not so severe as to require total valve replacement.
The decision to undertake valvuloplasty can be made only after visual analysis of the
lesion present. The most favorable conditions for valvuloplasty using an annuloplasty
ring are a combination of a distended natural valve ring associated with supple valve
cusps and normal chordae tendineae.
The remodeling valvuloplasty technique with a prosthetic ring may be used in all
acquired or congenital mitral insufficiencies with dilatation and deformation of the
fibrous mitral annulus.
For Type I mitral insufficiencies with no subvalvular lesions and normal valvular
movements, this ring technique used alone is sufficient. However, the ring technique
must be associated with mitral valvuloplasty repair in Type II insufficiencies with
prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III
insufficiencies with limitation of valvular movements due to fusion of the commissures
of chordae tendineae, or chordal hypertrophy.
Comparative Analysis:It has been demonstrated that the Carpentier-McCarthy-Adams IMR ETlogix mitral
annuloplasty ring is comparable to the predicate devices in design, intended use,
materials, and principal of operation.
Functional/
Safety Testing:The Carpentier-McCarthy-Adams IMR ETlogix mitral annuloplasty ring has successfully
completed design verification testing.
Conclusion:The Carpentier-McCarthy-Adams IMR ETlogix mitral annuloplasty ring is substantially
equivalent to the predicate devices.

1

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract caduceus or a bird-like figure with three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 5 2009

Edwards Lifesciences, LLC. Martin A. Kaufman Director of Regulatory Affairs One Edwards Way Irvine CA 92614

Re: K083623

Trade/Device Name: IMR ETlogix annuloplasty ring, model 4100 Regulation Number: 21 CFR 870.3800 Regulatory Class: Class II Product Code: KRH Dated: March 13, 2009 Received: March 16, 2009

Dear Mr. Kaufman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (DMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

2

Mr. Kaufman - Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

3

Indications for Use

510(k) Number (if known): K083623

Device Name: Carpentier-McCarthy-Adams IMR ETlogix™ annuloplasty ring, model 4100, or IMR ETlogix™ annuloplasty ring, model 4100

Indications for Use:

The Carpentier-McCarthy-Adams IMR ETlogix Mitral Annuloplasty Ring is indicated for the correction of mitral valve insufficiency where the lesions are not so severe as to require total valve replacement.

The decision to undertake valvuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for valvuloplasty using an annuloplasty ring are a combination of a distended natural valve ring associated with supple valve cusps and normal chordae tendinae.

The remodeling valvuloplasty technique with a prosthetic ring may be used in all acquired or congenital mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus.

For Type I mitral insufficiencies with no subvalvular lesions and normal valvular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repair in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae and in Type III insufficiencies with limitation of valvular movements due to fusion of the commissures or chordae tendineae, or chordal hypertrophy.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office Of Device Evaluation (ODE)

(Division Sign-Off)
Vision of Cardiovascular Devices
10(k) Number ko 83623

Page of