K Number
K083383
Manufacturer
Date Cleared
2009-02-18

(93 days)

Product Code
Regulation Number
866.1645
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae:

VITEK® 2 Gram Positive Erythromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Gram positive Streptococcus pneumoniae and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae is a qualitative test.

This submission is for the addition of an intermediate interpretive criteria.

Ervthromcvin has been shown to be active against the microorganisms listed below according to the FDA label for the antimicrobial.

Active in vitro and in clinical infections: Streptococcus pneumoniae

Device Description

VITEK® 2 Gram Positive Erythromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococus pneumoniae. It is intended for use with the VITEK 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mog/ml. The VITEK 2 AST Cards are essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance

MetricAcceptance Criteria (Not explicitly stated, but implied by regulatory guidance)Reported Device Performance (Category Agreement)
Overall Category AgreementSubstantially equivalent to CLSI broth microdilution reference method (implied by "acceptable performance" of 98.4%)98.4%
ReproducibilityAcceptable (implied)Acceptable
Quality ControlAcceptable (implied)Acceptable

Note: The document states that the performance must be "substantially equivalent" to the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003. While specific numerical acceptance criteria (e.g., minimum percentage for category agreement, essential agreement) are often detailed within such guidance documents, they are not explicitly listed in this 510(k) summary. The summary highlights a "98.4% overall Category Agreement" as demonstrating "acceptable performance," implying this met their internal and regulatory expectations.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as an exact number in the provided text. The text mentions "fresh and stock clinical isolates and stock challenge strains" were used for an "external evaluation." The original 510(k) (K063492) that this submission refers to would contain the detailed sample size.
  • Data Provenance: The data was from a retrospective analysis of clinical trial data that was part of the previous 510(k) (K063492). "There was no new clinical data gathered for this 510(k) notification." The country of origin is not specified but implicitly within the US given the submission to the FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not specified. The ground truth was established by the "CLSI broth microdilution reference method," which is a standard laboratory procedure, not typically involving human adjudication in the same way as, for example, image interpretation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Done: No. This device is an automated antimicrobial susceptibility testing system, not an imaging device requiring human reader interpretation or assistance. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.

6. Standalone Performance Study

  • Standalone Study Done: Yes. The study focused on the "algorithm only" performance (i.e., the VITEK 2 system) compared to the CLSI reference method. The VITEK 2 System is an automated device designed to provide a result without human intervention for interpretation beyond loading samples and reviewing the final report. The "overall Category Agreement" of 98.4% represents this standalone performance.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth used was the CLSI broth microdilution reference method. This is a recognized standard laboratory method for determining minimum inhibitory concentrations (MICs) of antimicrobial agents.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable or specified. This submission is for a software algorithm change to utilize existing FDA breakpoints, re-analyzing previously collected clinical trial data. While the original device (K063492) would have had a development dataset, this specific submission does not involve new training of a machine learning model but rather a modification to how results are interpreted based on established breakpoints. The term "training set" is usually reserved for machine learning models, which is not what is being described here in the conventional sense.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable or specified for this submission. As mentioned above, this involves a re-analysis of existing data with a modified algorithm to apply FDA breakpoints, not the training of a new model. The ground truth for the original data was established using the CLSI broth microdilution reference method.

{0}------------------------------------------------

1083383

BIOMÉRIEUX

FEB 1 8 2009

510(k) SUMMARY

VITEK® 2 Gram Positive Erythromycin for Streptococcus pneumoniae

510(k) Submission Information:

Submitter's Name:bioMérieux, Inc.
Address:595 Anglum RoadHazelwood, MO 63042
Contact Person:Jolyn TenlladoAssociate Staff Regulatory Affairs Specialist
Phone Number:314 -731-8386
Fax Number:314-731-8689
Date of Preparation:November 14, 2008
Device Name:
Formal/Trade Name:VITEK® 2 Gram Positive Erythromycin for Streptococcuspneumoniae
Classification Name:Fully Automated Short-Term Incubation CycleAntimicrobial Susceptibility Device, 21 CFR 866.1645
Common Name:VITEK 2 AST-GP Erythromycin for Streptococcuspneumoniae
Predicate Device:VITEK 2 Gram Positive Telithromycin for Streptococcus

D, 510(k) Summary:

в.

C.

. A 510(k) for VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae (K063492) was cleared by FDA on 21Dec2006. This premarket notification is for the same AST with the only modification being a software algorithm change to utilize the existing FDA intermediate breakpoint for Erythromycin and S. pneumoniae, and thereby lift the VITEK 2 Systems limitation regarding the inability of the Systems to report intermediate results for Erythromycin and S. pneumoniae strains. There has been no change to the VITEK 2 AST-GP Erythromcyin for S. pneumoniae formulation since K063492 was cleared. The analysis was modified to conform to standard practice of reporting results for all available FDA breakpoints for an antibiotic/organism combination. The clinical trial data submitted as part of K063492 was re-analyzed using the modified algorithm analysis. There was no new clinical data gathered for this 510(k) notification. The modified algorithm for this AST will be included in a future software release (VITEK 2 Systems software version PC 4.01).

pneumoniae (K053186).

VITEK® 2 Gram Positive Erythromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Streptococus pneumoniae. It is intended for use with the VITEK 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. The antimicrobial presented in VITEK 2 AST Cards is in concentrations equivalent by efficacy to standard method concentrations in mog/ml. The VITEK 2 AST Cards are

bioMérieux, Inc.

Fax: 314/731-800/638-4835 595 Anglum Road, Hazelwood, Missouri 63042-2320, USA Phone: 314/731-8500 8700

{1}------------------------------------------------

essentially miniaturized versions of the doubling dilution technique for determining the minimum inhibitory concentration (MIC) microdilution methodology.

The bacterial isolate to be tested is diluted to a standardized concentration in 0.45% saline before being used to rehydrate the antimicrobial medium within the card. The VITEK 2 System automatically fills, seals and places the card into the incubator/reader. The VITEK 2 Compact has a manual filling and sealing operation. The VITEK 2 monitors the growth of each well in the card over a defined period of time (up to 18 hours). At the completion of the incubation cycle, a report is generated that contains the MIC value along with the interpretive category result for each antibiotic contained on the card.

VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae demonstrated substantially equivalent performance when compared with the CLSI broth microdilution reference method, as defined in the FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA, Issued Feb. 5, 2003.

The Premarket Notification (510[k]) presents data in support of VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae. An external evaluation was conducted with fresh and stock clinical isolates and stock challenge strains. The external evaluations were designed to confirm the acceptability of VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae by comparing its performance with the CLSI broth microdilution reference method. The data is representative of performance on both the VITEK 2 and VITEK 2 Compact instrument platforms, as evidenced in the AST equivalency study presented in the VITEK 2 Compact 510(k), K050002. VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae demonstrated acceptable performance of 98.4% overall Category Agreement. Reproducibility and Quality Control demonstrated acceptable results.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with outstretched wings, symbolizing protection and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

FFB 1 8 2009

Ms. Jolyn Tenllado Associate Staff Regulatory Affairs Specialist bioMérieux 595 Anglum Road Hazelwood. MO 63042

Re: K083383

Trade/Device Name: VITEK® 2 Gram Positive Erythromycin (≤ 0.25 .- ≥ 1. jig/til) Regulation Number: 21 CFR § 866.1645 Regulation Name: Antimicrobial Susceptibility Test System - Short Incubation Regulatory Class: II Product Code: LON Dated: November 14, 2008 Received: November 17, 2008

Dear Ms. Tenllado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at 240-276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041, or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sally attaymto

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Fealth

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K083383

Device Name: VITEK® 2 Gram Positive Erythromycin (< 0.25 - > 1 ug/ml)

Indications For Use:

:

The VITEK® 2 Antimicrobial Susceptibility Test (AST) is intended to be used with the VITEK® 2 and VITEK® 2 Compact Systems for the automated quantitative or qualitative susceptibility testing of isolated colonies for the most clinically significant aerobic gramnegative bacilli, Staphylococcus spp., Enterococcus spp., Streptococcus agalactiae, and S. pneumoniae:

VITEK® 2 Gram Positive Erythromycin for Streptococcus pneumoniae is designed for antimicrobial susceptibility testing of Gram positive Streptococcus pneumoniae and is intended for use with the VITEK® 2 and VITEK® 2 Compact Systems as a laboratory aid in the determination of in vitro susceptibility to antimicrobial agents. VITEK 2 Gram Positive Erythromycin for Streptococcus pneumoniae is a qualitative test.

This submission is for the addition of an intermediate interpretive criteria.

Ervthromcvin has been shown to be active against the microorganisms listed below according to the FDA label for the antimicrobial.

Active in vitro and in clinical infections: Streptococcus pneumoniae

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart.C)

(PEEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Freddie In Cooke
Division Sign Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K08 3383

Page 1 of 1

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”