K Number
K083318

Validate with FDA (Live)

Date Cleared
2009-01-29

(80 days)

Product Code
Regulation Number
870.1130
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KD-5904 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Description

KD-5904 Fully Automatic Electronic Blood Pressure Monitor is Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, the device is used to monitor systolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Buckling a cuff around the left upper arm, which cuff circumference is limited to 22cm-48cm, the device can analyze the signals promptly and display the test results. It has some other functions, such as 6 buttons for operation.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information for the KD-5904 Fully Automatic Electronic Blood Pressure Monitor, based on the provided text:

Acceptance Criteria and Device Performance

The provided text only explicitly mentions that the device "meets the following standards: ANSI/AAMI SP-10 standard, EN60601-1 Medical electrical equipment Part 1: General requirements for safety, EN60601-1-2 Electromagnetic Compatibility."

The ANSI/AAMI SP-10 standard is the primary standard for non-invasive sphygmomanometers and outlines the accuracy requirements for blood pressure devices. While the exact numerical criteria from ANSI/AAMI SP-10 are not detailed in the provided documents, adherence to this standard implies meeting its specified accuracy requirements. Without the full standard document, the specific numerical acceptance criteria (e.g., mean difference and standard deviation between device and reference measurements) cannot be extracted.

Based on the provided text, the table below reflects what is stated:

Acceptance Criteria (Standard)Reported Device Performance
ANSI/AAMI SP-10 standardMet
EN60601-1Met
EN60601-1-2Met

Study Information

The document states that the device "meets the following standards: ANSI/AAMI SP-10 standard, EN60601-1 Medical electrical equipment Part 1: General requirements for safety, EN60601-1-2 Electromagnetic Compatibility." This implies that a study (or studies) was conducted to demonstrate compliance with these standards. However, the provided text does not offer specific details about the design, methodology, or results of such a study beyond the blanket statement of compliance.

Therefore, many of the requested details about the study are not available in the given document.

  1. Sample size used for the test set and the data provenance: Not specified in the provided text.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not specified in the provided text. Blood pressure device validation typically involves comparing the device's readings to those of trained observers using a reference method (e.g., auscultation with a mercury sphygmomanometer).
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not specified in the provided text.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a standalone blood pressure monitor, not an AI-assisted diagnostic tool for human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Yes, the device itself is a standalone system, and its performance would have been evaluated directly. The statement "KD-5904 Fully Automatic Electronic Blood Pressure Monitor meets the following standards: ANSI/AAMI SP-10 standard" implies a standalone performance evaluation against clinical reference measurements.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For blood pressure monitors, the "ground truth" (or reference standard) in clinical validation studies is typically simultaneous measurements taken by trained observers using a calibrated Riva-Rocci/Korotkoff auscultatory method (often with a mercury sphygmomanometer) as defined by standards like ANSI/AAMI SP-10. However, this specific detail is not explicitly stated in the provided text.
  7. The sample size for the training set: Not applicable/Not specified. For a device like this, which likely uses established oscillometric algorithms, there isn't typically a "training set" in the sense of machine learning models. The algorithm is developed and then validated against a test set.
  8. How the ground truth for the training set was established: Not applicable/Not specified. See point 7.

{0}------------------------------------------------

510(k) Summary

JAN 2 9 2009

ldentification of the submitter:

Submitter:

Fax number:

Contact:

Telephone number:

Date of Application:

Andon Health Co., Ltd. No 31, Changjiang Road, Nankai District, Tianjin, P.R. China, 300193 86-22-6052 6161 86-22-6052 6162 Liu Yi 10/29/08

Identification of the product:

Device proprietary Name: Fully Automatic Electronic Blood Pressure Monitor Common name: Noninvasive blood pressure measurement systems Classification name: Noninvasive blood pressure measurement system Class II per 21 CFR 870.1130

Marketed Devices to which equivalence is claimed:

Devicemanufacture510(k) number
KD-595Andon Health Co., LtdK070828

Device description:

KD-5904 Fully Automatic Electronic Blood Pressure Monitor is Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, the device is used to monitor systolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Buckling a cuff around the left upper arm, which cuff circumference is limited to 22cm-48cm, the device can analyze the signals promptly and display the test results. It has some other functions, such as 6 buttons for operation.

Intended use:

KD-5904 Fully Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate of adult each time, with the cuff around the left upper arm according to the instruction in the user's guide manual, which is same as predicated device.

{1}------------------------------------------------

K083318
P2/2

Summary comparing technological characteristics with predicate device:

KD-5904 Fully Automatic Electronic Blood Pressure Monitor has the same principle with predicated device, which utilizes Oscilliometric measurement method to monitor the blood pressure and the result can be shown on the LCD.

The modifications that were made are:

    1. Appearance
    1. Change into 6 buttons for operation from original two buttons. 6 buttons include,,,,<M+>,.

Pressing : turn on the device and measure/turn off the device: Pressing: begin memory and browse the item:

Pressing: begin to adjust the time, press <M+> to increase, press to decrease.

Pressing : turn on /turn off the voice.

    1. Larger cuff (cuff circumference is limited to 22cm-48cm).
  1. Change microprocessor from MB89485 to H8/38124.

Please find the following tabulated comparison supporting that the proposed device is substantially equivalent to the predicated device.

FDA file reference number510K# K070828
Technological CharacteristicsComparison result
Indications for useIdentical
Target populationIdentical
DesignSimilar
MaterialsIdentical
PerformanceIdentical
SterilityNot Applicable
BiocompatibilityIdentical
Mechanical safetyIdentical
Chemical safetyNot Applicable
Energy used and/or deliveredIdentical
Where usedIdentical
Standards metIdentical
Electrical safetyIdentical

Device testing:

KD-5904 Fully Automatic Electronic Blood Pressure Monitor meets the following standards:

  • . ANSI/AAMI SP-10 standard
  • EN60601-1 Medical electrical equipment Part 1: General requirements for safetv
  • EN60601-1-2 Electromagnetic Compatibility

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 9 2009

Andon Health Co., Ltd. c/o Ms. Liu Yi No 31, Changjiang Road, Nankai District, Tianjin, P.R. China, 300193

Re: K083318

Trade/Device Name: KD-5904 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 19, 2008 Received: December 29, 2008

Dear Ms. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Liu Yi

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

h

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510K submission for KD-5904 Fully Automatic Electronic Blood Pressure Monitor

(083319

Statement of Indications for Use

510(k) Number: K083318

Andon Health Co., Ltd Applicant:

Device name:

KD-5904 Fully Automatic Electronic Blood Pressure Monitor

Indications for use:

KD-5904 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Prescription use Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use YES (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sidh-Off)

Division of Cardiovascular Devices 4083316 510(k) Number

Page 1 of

  • 9 -

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).