(78 days)
Not Found
No
The summary describes a software that generates optimized IMRT plans based on a reference dose distribution, but it does not mention or imply the use of AI or ML techniques for this optimization process. The description focuses on the input and output data and the software's function without referencing learning or adaptive algorithms.
No
RayAutoplan is a software for generating treatment plans, not a device that directly administers therapy or has a therapeutic effect. It aids in the planning phase of treatment.
No
Explanation: The device is a software for generating optimized IMRT treatment plans, not for diagnosing a medical condition. It takes an existing dose distribution and patient information to produce treatment plans.
Yes
The device description explicitly states "RayAutoplan is a stand-alone software program" and details its function as generating treatment plans based on input data and exporting them electronically. There is no mention of accompanying hardware components being part of the device itself.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are devices intended for use in the collection, preparation, and examination of specimens taken from the human body (such as blood, urine, or tissue) to provide information for the diagnosis, treatment, or prevention of disease.
- RayAutoplan's Function: RayAutoplan is a software program used in radiation therapy planning. It takes existing patient data (CT images, regions of interest) and a reference dose distribution to generate optimized treatment plans. It does not analyze biological specimens from the patient's body.
- Intended Use: The intended use is to assist clinically qualified staff in creating and selecting radiation therapy treatment plans for subsequent treatment of the patient. This is a therapeutic planning function, not a diagnostic one based on in vitro analysis.
Therefore, RayAutoplan falls under the category of medical devices used in treatment planning, not In Vitro Diagnostics.
N/A
Intended Use / Indications for Use
RayAutoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI Among the generated plans, the user selects and climically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the pattent
The intended users of RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
Product codes
MUJ
Device Description
RayAutoplan is a stand-alone software program that generates a number of treatment plans, based on the dose distribution exported from a TomoTherapy treatment planning system
To execute, RayAutoplan needs the following unput
- . A patient description, normally CT images and regions of interests
- A dose distribution serving as a reference dose, which should be benchmarked ●
As output, RayAutoplan produces a inac IMRT treatment plan and the corresponding dose distribution from an accurate calculation of the dose engine
plan to DICOM archive or R&V system
The software runs on a Windows XP or Windows Vista platform
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
CT images
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.5050 Medical charged-particle radiation therapy system.
(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
0
Title | Document ID | Version |
---|---|---|
510(k) Application - RayAutoplan | RSL-D-58-10 | 1.0 |
5. 510(k) Summary RayAutoplan
- 5 1 510(k) owner
RaySearch Laboratories AB Sveavägen 25, plan 9 111 34 Stockholm Sweden
+46 (8) 54506130 Tel Fax +46 (8) 54506139
-
5 2 Contact person
Anders Murman -
5 3 Preparation date
09/01/2008
5 4 Trade name
RayAutoplan, alias t-RayAutoplan, alias SharePlan
RayAutoPlan is the product name RaySearch will use moving this device into the world market
t-RayAutoplan was the name of the product during the development project, and when seen in the documentation related to this application, it is the same as RayAutoplan, by definition
SharePlan is the trade name TomoTherapy Inc will use on the world market, as being the exclusive distributor of this device This name is not further used in the documentation of this application
5 5 Common name
Radiation treatment planning system
5 6 Classification name
Medical charged-particle radiation therapy system (21 CFR 892 5050, Product Code MUJ)
5 7 Predicate devices
Oncentra MasterPlan v3 1 510(k) number K081281
5 8 Device description
RayAutoplan is a stand-alone software program that generates a number of treatment plans, based on the dose distribution exported from a TomoTherapy treatment planning system
Detailed information of RayAutoplan can be found in reference 1, Customer requrement specifications
However, the most important workflow is described below
Flow of Events | ||
---|---|---|
User | System | |
1 | The user launches RayAutoplan | 3 The system imports the data and checks |
consistency of in-data | ||
2 | The user selects the transferred | |
TomoTherapy case from a list of | ||
TomoTherapy cases | ||
4 | The user selects machine and treatment | |
energy | ||
5 | The user selects MLC delivery mode | |
(SMLC/DMLC) | ||
6 | The user specifies beam configuration | |
7 | The user specifies which plan family or | |
families that should be generated |
KOB3uey
JAN 2 2 2009
1
Title | Document ID | |
---|---|---|
510(k) Application - RayAutoplan | RSL-D-58-10 | 10 |
8 The user starts the automatic plan generation | 9 The system generates plans | |
10 The system displays plans as |
- 2D and 3D dose and patient displays
- Dose difference maps
- DVHs (several overlaid in same graph)
- Dose statistics
- Plan data (machine, beams, segments and so on) | |
| 11 The user reviews the Linac plans
12 The user selects a plan for approval and generates plan documentation | 13 The system creates a report containing - Patient info
- Plan info
- Dose info | |
| 14 The user approves the plan for delivery | 15 Prints plan documentation and exports | |
Alternative workflows exist, including automatic plan generation and beam commissioning
To execute, RayAutoplan needs the following unput
- . A patient description, normally CT images and regions of interests
- A dose distribution serving as a reference dose, which should be benchmarked ●
As output, RayAutoplan produces a inac IMRT treatment plan and the corresponding dose distribution from an accurate calculation of the dose engine
plan to DICOM archive or R&V system
The software runs on a Windows XP or Windows Vista platform
5 9 Intended use
Ray Autoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI Among the generated plans, the user selects and climically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the pattent
The intended users of RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
5 10 Technological characteristics summary
The technological characteristics are the same for Ray Autoplan as for Oncentra MasterPlan v3 1 Both devices produce IMRT treatment plans with corresponding dose distributions computed using a firee dimensional collapsed cone dose engine Both devices have a function of electronic clinical approval of treatment plans by trained and authorized staff, and finally export in DICOM format for commencing treatment or archiving
2
Image /page/2/Picture/1 description: The image is a black and white logo. The logo features a stylized bird-like symbol with three curved lines representing its wings. The bird is oriented towards the right. Encircling the bird is text that reads "DEPARTMENT OF HUMAN SERVICES (USA)" in a circular arrangement. The text is in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
RaySearch Laboratories AB % Mr Jay Y Kogoma Senior Staff Engineer Intertek Testing Services 2307 E Aurora Rd . Unit B7 TWINSBURG OH 44087
JAN 2 2 2009
Re K083264
Trade/Device Name RayAutoplan 1 0 Regulation Number 21 CFR 892 5050 Regulation Name Medical charged-particle radiation therapy system Regulatory Class II Product Code: MUJ Dated January 5, 2008 Received January 6, 2008
Dear Mr Kogoma
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA)_ You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteratıon
If your device is classified (see above) into etther class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register
3
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801, good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter
21 CFR 876 xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
---|---|---|
21 CFR 884 xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
21 CFR 892 xxx | (Radiology) | (240) 276-0120 |
Other | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry suppor/index.html
Sincerely yours,
Lamme M. Morre
anıne M N Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
Title | Document ID | Version |
---|---|---|
510(k) Application - RayAutoplan | RSL-D-58-10 | 1.0 |
4. Indications for Use Statement
510(k) Number (if known) - Device Name RayAutoplan 1 0
2083264
Indications For Use
RayAutoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI Among the generated plans, the user selects and clinically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the pattent
The intended users of RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hales Reem
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number