K Number
K083247

Validate with FDA (Live)

Date Cleared
2009-01-29

(87 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KD-7906 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 6.1023 inches to 9.8425 inches.

Device Description

KD-7906 Fully Automatic Electronic Blood Pressure Monitor is Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, the device is used to monitor systolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Buckling a cuff around the wrist, which cuff circumference is limited to 6.1023 inches to 9.8425 inches. the device can analyze the signals promptly and display the test results. Besides, It can store and show 2*60 times measuring result with the day and time. And it has the function of blood pressure level classification and calculating the average of the last 3 readings.

AI/ML Overview

Here's an analysis of the KD-7906 Fully Automatic Electronic Blood Pressure Monitor based on the provided 510(k) submission, specifically addressing acceptance criteria and supporting studies where information is available:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document refers to meeting specific standards as the primary acceptance criteria for the KD-7906. The performance itself is largely claimed to be "Identical" to a predicate device, K070826.

Acceptance Criteria (Standard Met)Reported Device Performance (Claimed based on the provided document)
ANSI/AAMI SP-10 standardMet standard; Performance "Identical" to predicate (K070826)
IEC 60601-1 (Safety)Met standard
IEC 60601-1-2 (EMC)Met standard

Note: The document does not provide specific numerical performance metrics (e.g., mean difference, standard deviation between device readings and a reference method) from a dedicated clinical study for the KD-7906 itself. Instead, it relies on the claim of "Identical" performance to its predicate device, K070826, which would have had to meet these standards in its own submission.

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not specify a sample size for a test set used to evaluate the KD-7906's performance directly. The claim is that its performance is "Identical" to the predicate device (K070826). Therefore, any test set data would refer to the studies conducted for the predicate.

The document does not mention the country of origin of any data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. As the submission relies on demonstrating substantial equivalence to a predicate device and meeting general standards, specific details about expert-established ground truth for a new clinical study are absent.

4. Adjudication Method for the Test Set

This information is not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or described in this 510(k) submission for the KD-7906. This type of study is more common for diagnostic imaging devices where human interpretation is involved. For an automatic blood pressure monitor, the performance is typically evaluated against direct physiological measurements.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

The device itself is a "Fully Automatic Electronic Blood Pressure Monitor," implying it operates in a standalone manner without human intervention for the measurement process itself. While a "standalone" study in the AI sense might not be explicitly stated, the nature of the device means its performance evaluation inherently reflects its standalone capability in generating blood pressure readings. The document does not describe specific "algorithm only" performance separate from the device.

7. The Type of Ground Truth Used

The document does not explicitly state the type of ground truth used for the KD-7906's verification. However, for a device claiming to meet the ANSI/AAMI SP-10 standard, the ground truth for blood pressure measurement studies typically involves:

  • Auscultatory reference method: Measurements taken by trained observers using a mercury sphygmomanometer or highly accurate aneroid sphygmomanometer, often with dual observers for consensus.

8. The Sample Size for the Training Set

The provided document does not specify a sample size for a training set. This 510(k) submission describes an electronic blood pressure monitor, not an AI/ML algorithm that typically undergoes a distinct training phase with a dedicated dataset. The device operates based on oscillometric principles, which are well-established.

9. How the Ground Truth for the Training Set Was Established

Since this is not an AI/ML device with a distinct "training set" in the modern sense, the concept of establishing ground truth for a training set is not applicable here. The oscillometric algorithm's development and calibration would have been based on extensive physiological data and established principles, likely incorporating a form of "ground truth" through direct arterial pressure measurements or expert auscultation during its initial design and validation. However, this is not detailed in the 510(k) for the KD-7906.

{0}------------------------------------------------

510K submission of KD-7906 Fully Automatic Electronic Blood Pressure Monitor

510(k) Summary

ldentification of the submitter:

JAN 2 9 2009

Submitter: Andon Health Co., Ltd. No 31, Changjiang Road, Nankai District, Tianjin, P.R. China, 300193 86-22-6052 6161 Telephone number: Fax number: 86-22-6052 6162 Liu Yi Contact: Date of Application: 10/29/08

Identification of the product:

Device proprietary Name: Fully Automatic Electronic Blood Pressure Monitor Common name: Noninvasive blood pressure measurement systems Classification name: Noninvasive blood pressure measurement system Class II per 21 CFR 870.1130

Marketed Devices to which equivalence is claimed:

Device KD-795

manufacture Andon Health Co., Ltd 510(k) number K070826

Device description:

KD-7906 Fully Automatic Electronic Blood Pressure Monitor is Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, the device is used to monitor systolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Buckling a cuff around the wrist, which cuff circumference is limited to 6.1023 inches to 9.8425 inches. the device can analyze the signals promptly and display the test results. Besides, It can store and show 2*60 times measuring result with the day and time. And it has the function of blood pressure level classification and calculating the average of the last 3 readings.

Intended use:

KD-7906 Fully Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with the cuff around the wrist

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the text "K083247 p2/2" in a handwritten style. The text is black and the background is white. The text is slightly slanted to the right.

according to the instruction in the user's guide manual, which is same as predicated device.

Summary comparing technological characteristics with predicate device:

KD-7906 Fully Automatic Electronic Blood Pressure Monitor has the same principle with predicated device, which utilizes Oscilliometric measurement method to monitor the blood pressure and the result can be shown on the LCD.

The modifications that were made are:

    1. No voice
    1. Appearance
    1. Change 120 memory recall / 2 users (2x60) from 1x60
    1. Memory average function: calculating the average of last 3 readings

Please find the following tabulated comparison supporting that the proposed device is substantially equivalent to the predicated device.

FDA file reference number510K# K070826
Technological CharacteristicsComparison result
Indications for useIdentical
Target populationIdentical
DesignSimilar
MaterialsIdentical
PerformanceIdentical
SterilityNot Applicable
BiocompatibilityIdentical
Mechanical safetyIdentical
Chemical safetyNot Applicable
Energy used and/or deliveredIdentical
Where usedIdentical
Standards metIdentical
Electrical safetyldentical

Device testing:

KD-7906 Fully Automatic Electronic Blood Pressure Monitor meet the following standards:

  • · ANSI/AAMI SP-10 standard
  • IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety
  • IEC 60601-1-2 Electromagnetic Compatibility

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized wing-like shapes.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 9 2009

Andon Health Co., Ltd. c/o Ms. Liu Yi No 31, Changjiang Road, Nankai District, Tianjin, P.R. China, 300193

Re: K083247

Trade/Device Name: KD-7906 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 19, 2008 Received: December 30, 2008

Dear Ms. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

Page 2 - Ms. Liu Yi

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

$

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510K submission of KD-7906 Fully Automatic Electronic Blood Pressure Monitor

(08324)

Statement of Indications for Use

510(k) Number :

Applicant: Andon Health Co., Ltd

Device name: KD-7906 Fully Automatic Electronic Blood Pressure Monitor

Indications for use:

KD-7906 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 6.1023 inches to 9.8425 inches.

Prescription use Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

YES

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

hf

(Division Sigm-Off) Division of Cardiovascular Devices 510(k) Number Koszá 47

Page 1 of

. 9

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).