K Number
K083214
Manufacturer
Date Cleared
2009-05-12

(193 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AMPLATZER TorqVue 45°x45° Delivery Sheath is intended to provide a pathway through which devices are introduced within the chambers of the heart.

Device Description

The AMPLATZER® TorqVue® 45°x45° Delivery Sheath is a sterile, singleuse, general-purpose sheath that is an extension of the TorqVue Delivery System product line. It is designed to provide a pathway through which a device may be delivered to the chambers of the heart. The AMPLATZER® TorqVue® 45°x45° Delivery Sheath has a usable length of 100 cm and is available in 9 Fr, 10 Fr, 12 Fr, and 13 Fr sizes (inner diameter). A dilator, which eases penetration of tissue, is packaged with each sheath. The distal end of both the sheath and dilator are curved approximately 45° in two dimensions, resulting in a three-dimensional geometry. The 10 Fr, 12 Fr, and 13 Fr sizes also include a flush adapter that facilitates connection of accessories to the proximal end of the sheath. The body of each sheath is radiopaque for visibility under fluoroscopy. The sheath is constructed of two segments of different Pebax materials to provide support for device advancement and tip flexibility for vessel engagement. The distal end of the sheath has a soft tip composed of 55-durometer Pebax® to minimize vessel trauma.

AI/ML Overview

The provided text describes a 510(k) submission for the AMPLATZER® TorqVue® 45°x45° Delivery Sheath. My analysis will focus on extracting information related to acceptance criteria and the study that proves the device meets them, based on the provided document.

Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Functional Characteristics (Substantial equivalence to predicate)The functional characteristics were found to be "substantially equivalent to the predicate device cited."
Distal Curve Dimensions after SterilizationVerified (implied to meet criteria, as the device was deemed substantially equivalent).
Distribution SimulationVerified (implied to meet criteria, as the device was deemed substantially equivalent).
Six-Month Accelerated AgingVerified (implied to meet criteria, as the device was deemed substantially equivalent).

Note: The document explicitly states that "Bench testing of the AMPLATZER TorqVue® 45°x45° Delivery Sheath included verification of the distal curve dimensions after sterilization, distribution simulation, and six-month accelerated aging." While specific quantitative acceptance criteria values are not provided, the conclusion is that these tests demonstrated substantial equivalence, meaning the performance met the specified (though unquantified in this summary) criteria.

Study Information

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not explicitly stated. The document refers to "Bench testing" without specifying the number of units tested for each criterion.
    • Data Provenance: Not specified. The testing described is "Bench testing," which typically occurs in a controlled laboratory environment. There is no information on human or animal data, country of origin, or whether it was retrospective or prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable. The study involved "Bench testing" of physical device characteristics, not clinical assessment requiring expert interpretation for ground truth.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. This type of adjudication is typically used for clinical interpretation or subjective assessments, not for objective bench testing of device characteristics.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The study performed was "Bench testing" of a medical device (delivery sheath), not an AI-assisted diagnostic or interpretive system. Therefore, MRMC studies and effect sizes for human reader improvement are not relevant.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithmic device; it is a physical medical device.
  6. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • The ground truth for the "Bench testing" was derived from predefined engineering specifications and performance standards to establish substantial equivalence with the predicate device. These standards would dictate acceptable ranges for dimensions, stability after sterilization, and aging performance.
  7. The sample size for the training set

    • Not applicable. This device does not involve machine learning or AI, and therefore does not have a "training set."
  8. How the ground truth for the training set was established

    • Not applicable, as there is no training set for this device.

{0}------------------------------------------------

MAY 12 2009

510(K) Summary (per 21 CFR 807.92)

    1. Applicant Information
      Date Prepared:

October 30, 2008

Submitter:

AGA Medical Corporation 5050 Nathan Lane North Plymouth, MN 55442 USA

Regulatory Affairs Associate

Establishment Registration Number:

Contact person:

Telephone Number: Fax Number:

763.531.3161 763.647.5932

Diane Carroll

2135147

2. General Device Information

Trade Name:AMPLATZER® TorqVue® 45°x45° DeliverySheath
Common Name:Catheter, Delivery Sheath
Classification Name:Catheter, Percutaneous
Classification:Class II, 21 CFR 870.1250
Product Code:DQY
Predicate Device:AMPLATZER® TorqVue® Delivery System

AMPLATZER® TorqVue® Delivery System, K072313, cleared November 2, 2007 Reg. No. 870.1250

3. Device Description

The AMPLATZER® TorqVue® 45°x45° Delivery Sheath is a sterile, singleuse, general-purpose sheath that is an extension of the TorqVue Delivery System product line. It is designed to provide a pathway through which a device may be delivered to the chambers of the heart.

{1}------------------------------------------------

The AMPLATZER® TorqVue® 45°x45° Delivery Sheath has a usable length of 100 cm and is available in 9 Fr, 10 Fr, 12 Fr, and 13 Fr sizes (inner diameter). A dilator, which eases penetration of tissue, is packaged with each sheath. The distal end of both the sheath and dilator are curved approximately 45° in two dimensions, resulting in a three-dimensional geometry. The 10 Fr, 12 Fr, and 13 Fr sizes also include a flush adapter that facilitates connection of accessories to the proximal end of the sheath.

The body of each sheath is radiopaque for visibility under fluoroscopy. The sheath is constructed of two segments of different Pebax materials to provide support for device advancement and tip flexibility for vessel engagement. The distal end of the sheath has a soft tip composed of 55-durometer Pebax® to minimize vessel trauma.

4. Intended Use

The AMPLATZER® TorqVue® 45°x45° Delivery Sheath is intended to provide a pathway through which devices are introduced within the chambers of the heart.

5. Comparison to Predicate

The AMPLATZER® TorqVue® 45°x45° Delivery Sheath has the following similarities when compared to the predicate device:

  • . Indications for Use are similar; the AMPLATZER® TorqVue® 45°x45° Delivery Sheath indications are a subset of the predicate's indications
  • . Operating principal is the same
  • Manufacturing materials and processes are the same .
  • ·Sterilization method is the same .
  • Sterile package materials are the same .

{2}------------------------------------------------

6. Test Data

The functional characteristics of the AMPLATZER TorqVue® 45°x45° Delivery Sheath are substantially equivalent to the predicate device cited. Bench testing of the AMPLATZER TorqVue® 45°x45° Delivery Sheath included verification of the distal curve dimensions after sterilization, distribution simulation, and six-month accelerated aging.

7. Summary

Based on the technical information, intended use, and laboratory verification test information provided, the AMPLATZER TorqVue® 45°x45° Delivery Sheath is substantially equivalent to the currently marketed predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Public Health Service

MAY 12 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AGA Medical Corporation c/o Ms. Diane Carroll Regulatory Affairs Associates 5050 Nathan Lane North Plymouth, MN 55442-3209

Re: K083214

Trade/Device Name: AMPLATZER® TorqVue® 45° x 45°Delivery Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DOY Dated: May 4, 2009 Received: May 5, 2009

Dear Ms. Carroll:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

Page 2 - Ms. Diane Carroll

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Diredtor Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number:K083214
Device Name:AMPLATZER TorqVue 45°x45° Delivery Sheath
Indications for Use:The AMPLATZER TorqVue 45°x45° Delivery Sheath is intended to provide a pathway through which devices are introduced within the chambers of the heart.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/ OR

Over-The-Counter-Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices
$10(k) Number K083214

Page 1 of 1

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).