(318 days)
The Bionexia OptiCup is a rapid chromatographic immunoassay for the qualitative and simultaneous detection of elevated levels of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline in human urine at or above the following concentrations:
| Marijuana | 50 ng/mL |
|---|---|
| Cocaine | 300 ng/mL |
| Phencyclidine | 25 ng/mL |
| Morphine | 2,000 ng/mL |
| Methamphetamine | 1,000 ng/mL |
| Methadone | 300 ng/mL |
| Amphetamine | 1,000 ng/mL |
| Barbiturates | 300 ng/mL |
| Benzodiazepines | 300 ng/mL |
| Nortriptyline | 1,000 ng/mL |
The Bionexia OptiCup is used to obtain visual qualitative test results and is intended for healthcare professionals only. It is not intended for over the counter sale to non-professionals.
Configurations of the Bionexia OptiCup can consist of any combination of the above listed drug analytes.
This device provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
Bionexia OptiCup is derived from a previously FDA-cleared device, ADT's ACCUSTEP Single and Multi-Strip Cassette/Dipstick DOA Screen Panels (K061005).
The unmodified test is made of various number of single drug-of-abuse (DOA) test strips (from 1 to 10 strips) placed in an open faced housing. These test strips are fixed in position with an adhesive label covering over the absorbent end of the test card with dipping end protruding. When running the test, the dipping end of the test panel is dipped into urine sample and test results read at 5 minutes.
The test strips used in Bionexia OptiCup are made in the exact same manner as the strips used in the cleared ACCUSTEP Single and Multi-Strip Dipstick DOA Screen Panels from one to ten single test strips placed in an open-faced housing inside a plastic cup. In Bionexia OptiCup, these test strips are assembled and fixed in position with an adhesive label covering over the absorbent end of the test card with dipping end protruding. The dipstick assembly is then sealed, (integrated) into a plastic urine cup, which also serves as a urine collection and specimen delivering device to run the test in one integrated device.
Bionexia OptiCup is a user-friendly modification of the unmodified device. To use an unmodified device, user has to collect urine sample in a container, tear open the dipstick device pouch, take off the device cover, dip the device into urine sample and hold it for 5-10 seconds, cover the dipstick device cover, lay device flat and wait for 5 minutes to read the preliminary qualitative test results. To use a modified device, user can open the cup pouch and take out the integrated OptiCup, simultaneously collect, dispense urine for dipping, run the test and obtain preliminary qualitative test results in 5 minutes. The leftover urine specimen in the cup, which is not adulterated or tampered with in any way, can be sent in to a toxicology lab for confirmation especially when a presumptive positive test result is obtained. This modified device allows user to administrate multiple drug test from urine specimen collection, testing, and result confirmation with one single device.
Bionexia OptiCup presents an ideal and convenient for various testing environments such as hospitals, clinics, workplace pre-employment screening, justice system or court-ordered compliance monitoring and random drug testing. This integrated Cup is designed to accompany available line of drugs of abuse tests from ADT. Customers can choose any combination of the cleared drug of abuse tests and integrated into this cup.
The Bionexia OptiCup is a rapid chromatographic immunoassay for the qualitative and simultaneous detection of elevated levels of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine. Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriotyline in human urine at or above the following concentrations:
| Marijuana | 50 ng/mL |
|---|---|
| Cocaine | 300 ng/mL |
| Phencyclidine | 25 ng/mL |
| Morphine | 2,000 ng/mL |
| Methamphetamine | 1,000 ng/mL |
| Methadone | 300 ng/mL |
| Amphetamine | 1,000 ng/mL |
| Barbiturates | 300 ng/mL |
| Benzodiazepines | 300 ng/mL |
| Nortriptyline | 1,000 ng/mL |
These tests can be performed without the use of an instrument.
A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing Marijuana, Cocaine, Phencyclidine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline below the designated cutoff levels will generate a colored line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
Here's an analysis of the acceptance criteria and study information for the Bionexia OptiCup, based on the provided text:
Based on the provided 510(k) summary, the device is a qualitative drug screening immunoassay, and as such, the "performance" is typically measured by its ability to correctly identify the presence or absence of the target analytes (drugs) at or above specified cutoff concentrations. The document does not explicitly state "acceptance criteria" in a numerical, performance-based format (e.g., sensitivity > X%, specificity > Y%), which is common for descriptive performance summaries of qualitative tests. Instead, it defines the intended performance by listing the cutoff concentrations for each drug and implies that the device should accurately detect drugs at or above these levels and not detect them below these levels.
The study that proves the device meets the acceptance criteria is described as a modification of a previously FDA-cleared device. The 510(k) process for a modified device often relies heavily on demonstrating substantial equivalence to the predicate device, especially if the modification is primarily in form factor (as is the case here, from a dipstick to an integrated cup). The provided text describes the changes and claims equivalence in the test strips themselves.
Here's the breakdown of the information you requested:
Acceptance Criteria and Reported Device Performance
The document defines the performance by the cutoff concentrations for each drug. The device's "performance" is its ability to qualitatively detect these drugs at or above these specific concentrations. The provided text does not include quantitative performance metrics like sensitivity, specificity, or accuracy percentages, which are often reported for diagnostic devices. Instead, it implicitly states that the device is intended to detect at these cutoffs.
| Drug Analyte | Cutoff Concentration (ng/mL) | Reported Device Performance |
|---|---|---|
| Marijuana | 50 | Detects at/above 50 ng/mL |
| Cocaine | 300 | Detects at/above 300 ng/mL |
| Phencyclidine | 25 | Detects at/above 25 ng/mL |
| Morphine | 2,000 | Detects at/above 2,000 ng/mL |
| Methamphetamine | 1,000 | Detects at/above 1,000 ng/mL |
| Methadone | 300 | Detects at/above 300 ng/mL |
| Amphetamine | 1,000 | Detects at/above 1,000 ng/mL |
| Barbiturates | 300 | Detects at/above 300 ng/mL |
| Benzodiazepines | 300 | Detects at/above 300 ng/mL |
| Nortriptyline | 1,000 | Detects at/above 1,000 ng/mL |
(Note: The "Reported Device Performance" column reflects the intended qualitative claim based on the cutoff concentrations provided. The document does not provide a separate table of actual measured performance results (e.g., sensitivity, specificity tables) against these cutoffs for the Bionexia OptiCup specifically, but rather relies on the equivalence to the predicate device.)
Study Details
-
Sample size used for the test set and the data provenance:
The provided text does not specify the sample size for a test set or the data provenance (e.g., country of origin, retrospective/prospective) for this specific device's validation. The submission focuses on the Bionexia OptiCup being a "user-friendly modification" of a previously cleared device (K061005), implying that much of the core performance data comes from the predicate device. For a modified device with largely unchanged core technology (the test strips themselves), detailed new performance studies on a large test set might not be explicitly detailed in the 510(k) summary if the substantial equivalence argument is strong. -
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the given text. For drug screening tests, ground truth is typically established by more specific confirmatory laboratory methods rather than expert consensus. -
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
This information is not provided in the given text. Adjudication methods are typically used in imaging or subjective interpretation tasks, not usually for objective chemical assays like this. -
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable to this device. The Bionexia OptiCup is a rapid chromatographic immunoassay, not an AI-powered diagnostic device, and does not involve human "readers" in the context of MRMC studies typical for AI in radiology. The results are visual and qualitative, read directly from the test strip. -
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
This is not applicable as it is not an algorithm or AI device. The device is a standalone chemical assay. The performance is the "standalone" performance of the assay itself. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
For drug screening tests, the established ground truth typically involves Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography Mass Spectrometry (LC/MS). The text explicitly states: "A more specific alternative chemical method must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography Mass Spectrometry (LC/MS) is the preferred confirmatory method." This indicates that these laboratory methods are considered the gold standard for confirmation. -
The sample size for the training set:
The document does not specify a "training set" sample size. This term is primarily relevant for machine learning algorithms. For an immunoassay, the development typically involves assay optimization and validation, not "training" in the ML sense. -
How the ground truth for the training set was established:
As there is no "training set" in the context of machine learning for this device, this question is not applicable. Assay development and validation would use samples spiked with known concentrations of analytes or clinically collected samples confirmed by a reference method (like GC/MS or LC/MS), but these aren't referred to as a "training set."
{0}------------------------------------------------
510(k) SUMMARY
AUG 2-0 2009
510(k) SUMMARY
This summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92.
The Assigned 510(k) number is K082977
Submitter:
MK Bio Inc 10239 Flanders Court San Diego, California 92121 Tel: 858-875-1900 Fax: 858-875 1999
FDA Establishment Registration Number: 3002752660
Date:
October 2, 2008
Contact Person:
Jane Zhang
Product Name:
Bionexia OptiCup
Common Name:
Single-drug lateral flow immunochromatographic test for the simultaneous and qualitative detection of Marijuana, Cocaine, Phencyclidine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline in urine
Regulation Name:
Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline test systems
l
{1}------------------------------------------------
Product Code:
| Marijuana | LDJ |
|---|---|
| Cocaine | DIO |
| Phencyclidine Hydrochloride | LCM |
| Morphine | DNK |
| Methamphetamine | DKZ |
| Methadone | DJR |
| Amphetamine | DJC |
| Barbiturates | DIS |
| Benzodiazepines | JXM |
| Nortriptyline | LFG |
Classification Number:
| Marijuana | 21 CFR § 862.3870 |
|---|---|
| Cocaine | 21 CFR § 862.3250 |
| Phencyclidine Hydrochloride | 21 CFR § 862.3100 |
| Morphine | 21 CFR § 862.3640 |
| Methamphetamine | 21 CFR § 862.3610 |
| Methadone | 21 CFR § 862.3620 |
| Amphetamine | 21 CFR § 862.3100 |
| Barbiturates | 21 CFR § 862.3150 |
| Benzodiazepines | 21 CFR § 862.3170 |
| Nortriptyline | 21 CFR § 862.3910 |
Device Classification:
Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, The Marijuana, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline test systems have been classified as Class II devices with moderate complexity.
The Bionexia OptiCup is similar to other FDA-cleared devices for the qualitative and simultaneous detection of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline in human urine.
{2}------------------------------------------------
Intended Use:
The Bionexia OptiCup is a rapid chromatographic immunoassay for the qualitative and simultaneous detection of elevated levels of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline in human urine at or above the following concentrations:
| Marijuana | 50 ng/mL |
|---|---|
| Cocaine | 300 ng/mL |
| Phencyclidine | 25 ng/mL |
| Morphine | 2,000 ng/mL |
| Methamphetamine | 1,000 ng/mL |
| Methadone | 300 ng/mL |
| Amphetamine | 1,000 ng/mL |
| Barbiturates | 300 ng/mL |
| Benzodiazepines | 300 ng/mL |
| Nortriptyline | 1,000 ng/mL |
The Bionexia OptiCup is used to obtain visual qualitative test results and is intended for healthcare professionals only. It is not intended for over the counter sale to non-professionals.
Configurations of the Bionexia OptiCup can consist of any combination of the above listed drug analytes.
This device provides only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical test result. Gas Chromatography/Mass Spectrometry (GC/MS) or Liquid Chromatography Mass Spectrometry (LC/MS) is the preferred confirmatory method. Clinical considerations and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.
Description:
Bionexia OptiCup is derived from a previously FDA-cleared device, ADT's ACCUSTEP Single and Multi-Strip Cassette/Dipstick DOA Screen Panels (K061005).
The unmodified test is made of various number of single drug-of-abuse (DOA) test strips (from 1 to 10 strips) placed in an open faced housing. These test strips are fixed in position with an adhesive label covering over the absorbent end of the test card with dipping end protruding. When running the test, the dipping end of the test panel is dipped into urine sample and test results read at 5 minutes.
{3}------------------------------------------------
The test strips used in Bionexia OptiCup are made in the exact same manner as the strips used in the cleared ACCUSTEP Single and Multi-Strip Dipstick DOA Screen Panels from one to ten single test strips placed in an open-faced housing inside a plastic cup. In Bionexia OptiCup, these test strips are assembled and fixed in position with an adhesive label covering over the absorbent end of the test card with dipping end protruding. The dipstick assembly is then sealed, (integrated) into a plastic urine cup, which also serves as a urine collection and specimen delivering device to run the test in one integrated device.
Bionexia OptiCup is a user-friendly modification of the unmodified device. To use an unmodified device, user has to collect urine sample in a container, tear open the dipstick device pouch, take off the device cover, dip the device into urine sample and hold it for 5-10 seconds, cover the dipstick device cover, lay device flat and wait for 5 minutes to read the preliminary qualitative test results. To use a modified device, user can open the cup pouch and take out the integrated OptiCup, simultaneously collect, dispense urine for dipping, run the test and obtain preliminary qualitative test results in 5 minutes. The leftover urine specimen in the cup, which is not adulterated or tampered with in any way, can be sent in to a toxicology lab for confirmation especially when a presumptive positive test result is obtained. This modified device allows user to administrate multiple drug test from urine specimen collection, testing, and result confirmation with one single device.
Bionexia OptiCup presents an ideal and convenient for various testing environments such as hospitals, clinics, workplace pre-employment screening, justice system or court-ordered compliance monitoring and random drug testing. This integrated Cup is designed to accompany available line of drugs of abuse tests from ADT. Customers can choose any combination of the cleared drug of abuse tests and integrated into this cup.
The Bionexia OptiCup is a rapid chromatographic immunoassay for the qualitative and simultaneous detection of elevated levels of Marijuana, Cocaine, Phencyclidine, Morphine, Methamphetamine. Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriotyline in human urine at or above the following concentrations:
| Marijuana | 50 ng/mL |
|---|---|
| Cocaine | 300 ng/mL |
| Phencyclidine | 25 ng/mL |
| Morphine | 2,000 ng/mL |
| Methamphetamine | 1,000 ng/mL |
| Methadone | 300 ng/mL |
| Amphetamine | 1,000 ng/mL |
| Barbiturates | 300 ng/mL |
| Benzodiazepines | 300 ng/mL |
| Nortriptyline | 1,000 ng/mL |
These tests can be performed without the use of an instrument.
A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing Marijuana,
4
{4}------------------------------------------------
Cocaine, Phencyclidine, Methamphetamine, Methadone, Amphetamine, Barbiturates, Benzodiazepines and Nortriptyline below the designated cutoff levels will generate a colored line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
Unmodified Applied DNA Technologies (ADT) Test Device:
The Bionexia OptiCup is derived from a previously FDA cleared device: K061005. This legally marketed but unmodified device and 510(k) number under which it was previously cleared is listed below in Table 1.
| Table 1: Unmodified ADT device with K number and Product Codes | ||||
|---|---|---|---|---|
| -- | ---------------------------------------------------------------- | -- | -- | -- |
| Previously Cleared ADT Drug of Abuse Test | 510(k) Number | Product Code |
|---|---|---|
| ACCUSTEP Single and Multi-Strip DOA Cassetteand Dipstick Screen Panels | K061005 | LDI, DIO, LCM, LFG, DKZ,DJR, DKZ, DIS, JXM, DJG |
Modified ADT Test Device:
The Bionexia OptiCup is a modified form of a previously FDA-cleared device, K061005. The difference between the modified devices and the unmodified device is that the former has the test strips assembled in a cup format, while the unmodified device has test strips assembled in a dipstick format.
With a cup format, this device can provide many "user friendly" features for the urine-based drug tests. User will be able to collect urine sample, administrate drug test, obtain preliminary test result and send split unadulterated urine sample for confirmation, all with one single drug testing device, the Bionexia OptiCup.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with three lines representing its wings.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
MK Bio. Inc. c/o Ms. Jane Zhang Director of QA/RA 10239 Flanders Court San Diego, CA 92121
AUG 20 2009
Re: K082977
Trade Name: Bionexia OptiCup Regulation Number: 21 CFR §862.3100 Regulation Name: Amphetamine Test System Regulatory Class: Class II Product Codes: DIO, DIS, DJC, DJR, DNK, DKZ, JXM, LCM, LDJ, LFG Dated: July 16, 2009 Received: July 21, 2009
Dear Ms. Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{6}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at . (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or ( 301 ) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
G.C.H
Courtney C. Harper, Ph.D. Acting Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
INDICATIONS FOR USE:
510(K) Number: K082977
Device Name: Bionexia OptiCup
Indications for Use: The Bionexia OptiCup is a rapid chromatographic immunoassay for the qualitative and simulaneous detection of elevated levels of Marijuana, Cocaine, Pheney, for the
qualitative and simultaneous detection of elevated levels of Marjiuna, Cocaine Methamphetamine, Methadone, Amphetamine, Barbiturates, Carbiturates, Nortriptyline in human urine at or above the following concentrations: Benzodiazepines and
| Marijuana | 50 ng/mL |
|---|---|
| Cocaine | 300 ng/mL |
| Phencyclidine | 25 ng/mL |
| Morphine | 2,000 ng/mL |
| Methamphetamine | 1,000 ng/mL |
| Methadone | 300 ng/mL |
| Amphetamine | 1,000 ng/mL |
| Barbiturates | 300 ng/mL |
| Benzodiazepines | 300 ng/mL |
| Nortriptyline | 1,000 ng/mL |
The test strips with an integrated cup is intended for healthcare professionals only. It is not intended for
over the counter sale to non-professionals. over the counter sale to non-professionals.
(Please do not write below this point)
Concurrence of CDRH, Office of In With Delow this point)
Concurrence of CDRH, Office of In Vitro Device Evaluation and Safety
Prescription Use
X
Division Sign-Off
Over-the-counter Use Prescription Use
(Per 21 CFR 801.109)
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K082977
§ 862.3870 Cannabinoid test system.
(a)
Identification. A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds includedelta -9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.(b)
Classification. Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).