(260 days)
EZvue(Ocufilcon D)1-day Soft (hydrophilic) Contact Lens for daily wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
Eyecare practitioners may prescribe the contact lens for single-use disposable daily wear. The EZvue(Ocufilcon D)1-day Soft (hydrophilic) Contact Lenses are not intended to be disinfected and should be discarded after a single use.
EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear is available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 55% water. The hydrogel lens' material is a copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photopolymerization. The EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline.
Here's a breakdown of the acceptance criteria and study information for the EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The submission primarily focuses on demonstrating substantial equivalence to a predicate device, rather than establishing specific quantitative acceptance criteria for de novo performance. The "acceptance criteria" are implied by the performance of the predicate device and the non-inferiority demonstrated through testing.
| Acceptance Criteria (Implied by Predicate/Testing) | Reported Device Performance (EZvue Ocufilcon D) |
|---|---|
| Technological Characteristics: | |
| Diameter range: Equivalent to predicate | 13.90 ~ 14.30mm |
| Power range: Equivalent to predicate | +8.00D ~ -12.00D |
| Center thickness: Equivalent to predicate | Varies with power (0.080 ~ 0.150mm) |
| Refractive index (hydrated): Equivalent to predicate | 1.410 |
| Light transmittance: >93% | >93% |
| Water content: 53-57% (Ocufilcon D) | 53 to 57 % |
| Oxygen permeability (Dk/t): Equivalent to predicate | 16 x 10^-14 [(cm²/sec)(mlO₂/ml-mmHg)]@ 35℃ |
| Biocompatibility: | |
| Non-toxic and non-irritating | Irritation test (rabbit), Systemic toxicity (non-toxic, non-irritating) |
| Non-cytotoxic | Cytotoxicity testing (not cytotoxic) |
| Microbiology: | |
| Sterility Assurance Level (SAL): 10^-6 | Validated SAL of 10^-6 |
| Shelf-life stability for sterility | Shelf-life stability data supports sterility through expiration date |
| Leachability: | |
| No detectable leachable monomers or additive residues | No leachable monomers and addictive residues at detection levels |
| Clinical Performance: | |
| Vision correction and patient satisfaction comparable to standard/predicate | Nearly 100% participants' vision corrected; nearly all satisfied with lens wearing |
| Safety comparable to standard/predicate | Products are "good and safe for customers" |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: "over 60 human eyes"
- Data Provenance: The study was conducted on human eyes. The submitter company is I-SEE VISION TECHNOLOGY INC. based in Hisnchu, Taiwan, so it is highly probable the data originated from Taiwan. The study is prospective as it involved testing the device on human eyes.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not explicitly state the number of experts or their qualifications used to establish ground truth for the clinical study. It mentions that "Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses." This suggests assessment by eye care practitioners, but specific expert count or qualifications are not provided.
4. Adjudication Method for the Test Set
The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The statement implies direct observation and assessment by eye care professionals, rather than a separate adjudication process for ground truth.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The study described is a clinical study on the device's performance, not a comparison of human reader performance with and without AI assistance.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
This device is not an AI algorithm. It is a soft contact lens. Therefore, the concept of a "standalone" algorithmic performance study is not applicable.
7. The Type of Ground Truth Used
For the clinical study, the "ground truth" was based on:
- Observational Clinical Outcomes: Corrected vision (presumably measured by visual acuity tests)
- Patient Reported Outcomes: Patient satisfaction with lens wearing and care.
For the nonclinical studies, ground truth was established by direct physical and chemical measurements (e.g., diameter, power, refractive index, water content, oxygen permeability) and standardized biological testing (irritation tests, cytotoxicity assays, sterility validation, leachability analysis).
8. The Sample Size for the Training Set
This device is not an AI algorithm, so there is no "training set." The materials used for manufacturing the lens were characterized.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as this is not an AI device.
{0}------------------------------------------------
Nu'"
K082880
510 (K) Summary
SUMMARY OF SAFETY AND EFFECTIVENESS FOR EZvue (Ocufilcon D) Soft (hydrophilic) Contact Lens for Daily Wear JUN 1 6 2009
Submitter Information:
| Company: | I-SEE VISION TECHNOLOGY INC. |
|---|---|
| 1F, No. 45, Lane 2, Section 2, Kuang-Fu Rd., Hisnchu, Taiwan, | |
| 30071 | |
| Contact Person: | Mr. Tom Lin, |
| Special Assistant of President Office | |
| Phone: | 886-3-5752822 |
| FAX: | 886-3-5752823 |
| E-Mail: | tomlin@iseevision.com |
Identification of Device:
| Classification Name: | Lens, Contact, (Disposable) |
|---|---|
| Trade Name: | EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens( or EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) ContactLens for daily wear) |
| Common or usual Name: | Soft (hydrophilic) Contact lens (daily wear) |
| FDA Classification: | Class II |
Predicate Device:
Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens cleared via K972303 Ocular Sciences / American Hydron Inc.
Indications for Use
EZvue(Ocufilcon D)1-day Soft (hydrophilic) Contact Lens for daily wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
Eyecare practitioners may prescribe the contact lens for single-use disposable daily wear. The EZvue(Ocufilcon D)1-day Soft (hydrophilic) Contact Lenses are not intended to be disinfected and should be discarded after a single use.
Description of Device
EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear is available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 55% water. The hydrogel lens' material is a copolymer
{1}------------------------------------------------
composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) via UV photopolymerization. The EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens with visible tint is tinted blue using Reactive Blue Dye #19 to make the lens more visible for handling. Lenses are supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline.
Summary of Clinical Study:
The EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lenses for Daily Wear were tested over 60 human eyes separately within 6 months. Nearly a hundred percent of the participants' vision was corrected and nearly all were satisfied with the lens wearing and care of lenses. In general, these products are good and safe for customers.
Nonclinical Studies:
A series of nonclinical performance tests were performed to demonstrate the safety and effectiveness of the EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear, and establish substantial equivalence to predicate lenses - Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens (K972303). All testing was conducted in accordance with the May 1994 FDA guideline titled Premarket Notification 510(K) Guidance Document for Class IV Contact lenses, and in conformance to applicable device regulations. The evidence of substantial equivalent to the predicate lens described as follow:
- a) Technological characteristics studies
EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear designs in the following parameter ranges:
Diameter range; 13.90~14.30mm
Power range: +8.00D~-12.00D
. Center thickness: varies with power (0.080~0.150mm)
Lenses have the following properties:
Refractive index: 1.410 (hydrated)
Light transmittance: >93%
Water content: 53 to 57 %
Oxygen permeability (edged corrected): 16×10-14(cm²/sec)(mlQz/ml-mmHg)]@ 35℃
There characterizations of EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lenses are equivalent and comparable to those of predicate lenses.
b) Biocompatibility
In accordance with the May 1994 Guidance Document for daily wear contact lenses,
{2}------------------------------------------------
toxicity studies have been conducted on the model: EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear. The Irritation test in the rabbit eye and Systemic toxicity studies indicate the extracts would be considered as non-toxic and nor irritated. The Cytotoxicity testing demonstrates the lens is not cytotoxic under the conditions of the study.
c) Microbiology
Steam sterilization process has been validated to deliver a minimum SAL of 10°, thereby complying with the requirement of FDA Group IV. There is shelf-life stability data supporting that the lens remains sterile through the expiration date claimed for the product.
d) Leachability
Studies were conducted to determine the leachable materials from the finished lens. The results show that, at the levels of the detection reported, there are no leachable monomers and addictive residues.
Substantial equivalence Statement:
Testing performed on the EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear indicated that it can support the efficiency and security use as well as the predicate devices- Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens (K972303), when used in accordance with the instructions for use. It is due to the facts that The risks and benefits of the subject device are the same as soft contact lenses for to the daily wear.
In conclusion, it is I-SEE's conviction that data submitted in this 510(k) to validate the claim of substantial equivalency, substantiates our ability to manufacture a soft contact lens, the EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear, with the same established safety profile and effectiveness as the predicate device-- Hydron Biomedics 55 (Ocufilcon D) Visibility Tint Molded daily wear soft contact lens (K972303).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol.
2009 SEP 9
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
I-See Vision Technology Inc. c/o Ms. Jennifer Reich Senior Consultant Harvest Consulting Corp. 2904 N. Boldt Drive Flagstaff, AZ 86001
Re: K082880
Trade/Device Name: EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lenses Regulatory Class: Class II Product Code: LPL, MVN Dated: June 9, 2009 Received: June 11, 2009
Dear Ms. Reich:
This letter corrects our substantially equivalent letter of June 16, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other
{4}------------------------------------------------
Page 2 - Ms. Jennifer Reich
requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Kesia Alexander for
Malvina B. Eydelman. N Director Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K082880
Device Name: EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear I-SEE Vision Technology, Inc.
Indications For Use:
EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lens for Daily Wear is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic and/or not-aphakic persons with non-diseased eyes who may exhibited refractive astigmatism up to 2.00 diopters that does not interfere with visual acuity.
Eyecare practitioners may prescribe the contact lens for single-use disposable daily wear. The EZvue (Ocufilcon D) 1-Day Soft (hydrophilic) Contact Lenses are not intended to be disinfected and should be discarded after a single use.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Afhoang
(Division Sign-Off)
Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K082880
Page 1 of 1
N/A