K Number
K082666
Manufacturer
Date Cleared
2008-10-15

(33 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ArthroCare System 12000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

Arthroscopic and Orthopedic ProceduresJoint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist)
Ablation and Debridement
• ACL/PCLKnee
• AcromioplastyShoulder
• Articular CartilageAll Joints
• BursectomyAll Joints
• ChondroplastyAll Joints
• FasciaAll Joints
• LigamentAll Joints
• NotchplastyKnee
• Scar TissueAll Joints
• Soft TissueAll Joints
• Subacromial DecompressionShoulder
• SynovectomyAll Joints
• TendonAll Joints
Excision and Resection
• Acetabular LabrumHip
• Articular LabrumAll Joints
• CapsuleAll Joints
• Capsular ReleaseKnee
• Cartilage FlapsKnee
• CystsAll Joints
• Discoid MeniscusKnee
• Frozen Shoulder ReleaseShoulder
• Glenoidale LabrumShoulder
• Lateral ReleaseKnee
• LigamentAll Joints
• Loose BodiesAll Joints
• Meniscal CystectomyKnee
• MeniscectomyKnee
• Plica RemovalAll Joints
• Scar TissueAll Joints
• Soft TissueAll Joints
• Synovial MembraneAll Joints
• TendonAll Joints
• Triangular Fibrocartilage (TFCC)Wrist
• VillusectomyKnee
Coagulation
• ACL/PCLKnee
• Articular CartilageAll Joints
• Carpal LigamentsWrist
• Glenohumeral CapsuleShoulder
• LigamentAll Joints
• Medial RetinaculumKnee
• Rotator CuffShoulder
• TendonAll Joints
• Wrist TendonsWrist
Device Description

The ArthroCare® System 12000 is a bipolar, high frequency, electrosurgical generator called the Controller that is intended to be used with a family of disposable, binolar, single use Wands.

AI/ML Overview

This 510(k) summary for the ArthroCare System 12000 does not describe any acceptance criteria or a study that uses a quantitative test set to prove the device meets said criteria.

The document states:
"This Special 510(k) proposes a modification in the performance specifications and labeling for the ArthroCare System 12000, which was previously cleared in K071709 on August 7, 2007. The indications for use, technology, principle of operation, materials, packaging, and sterilization parameters of the ArthroCare System 12000 remain the same as in the predicate cleared 510(k). The modified ArthroCare System 12000, as described in this Special 510(k), is substantially equivalent to the predicate device. The proposed modifications in performance specifications and labeling are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the System."

This indicates that the submission is for a modification to an already cleared device, and the basis for clearance is substantial equivalence to a predicate device (K071709), rather than a new study demonstrating performance against specific acceptance criteria using a test set. Therefore, the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies is not present in the provided text.

{0}------------------------------------------------

OCT 1 5 2008

82666

Page 173

510(k) Summary

ArthroCare Corporation ArthroCare® System 12000

General Information

Submitter Name/Address:ArthroCare Corporation680 Vaqueros AvenueSunnyvale, CA 94085-3523
Establishment Registration Number:2951580
Contact Person:Valerie Defiesta-NgDirector, Regulatory Affairs
Date Prepared:September 11, 2008

Device Description

Trade Name:ArthroCare® System 12000
Generic/Common Name:Electrosurgical Device and Accessories
Classification Name:Electrosurgical Cutting and CoagulationDevice and Accessories (Class II, 21 CFR878.4400, Product Code GEI)

Predicate Devices

ArthroCare® System 12000K071709
-----------------------------------

Product Description

The ArthroCare® System 12000 is a bipolar, high frequency, electrosurgical generator called the Controller that is intended to be used with a family of disposable, binolar, single use Wands.

{1}------------------------------------------------

Ko82666

Page 2 of 3

Intended Uses

The ArthroCare System 12000 is indicated for resection, ablation, and coagulation of soft
tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

Arthroscopic and Orthopedic ProceduresJoint Specific or AllJoints (ankle, elbow,hip, knee, shoulder, andwrist)
Ablation and Debridement
ACL/PCLKnee
AcromioplastyShoulder
Articular CartilageAll Joints
BursectomyAll Joints
ChondroplastyAll Joints
FasciaAll Joints
LigamentAll Joints
NotchplastyKnee
Scar TissueAll Joints
Soft TissueAll Joints
Subacromial DecompressionShoulder
SynovectomyAll Joints
TendonAll Joints
Excision and Resection
Acetabular LabrumHip
Articular LabrumAll Joints
CapsuleAll Joints
Capsular ReleaseKnee
Cartilage FlapsKnee
CystsAll Joints
Discoid MeniscusKnee
Frozen Shoulder ReleaseShoulder
Glenoidale LabrumShoulder
Lateral ReleaseKnee
LigamentAll Joints
Loose BodiesAll Joints
Meniscal CystectomyKnee
MeniscectomyKnee

{2}------------------------------------------------

Kof 2666

Continued
Arthroscopic and Orthopedic ProceduresJoint Specific or AllJoints (ankle, elbow,hip, knee, shoulder, andwrist)
• Plica RemovalAll Joints
• Scar TissueAll Joints
• Soft TissueAll Joints
• Synovial MembraneAll Joints
• TendonAll Joints
• Triangular Fibrocartilage (TFCC)Wrist
• VillusectomyKnee
Coagulation
• ACL/PCLKnee
• Articular CartilageAll Joints
• Carpal LigamentsWrist
• Glenohumeral CapsuleShoulder
• LigamentAll Joints
• Medial RetinaculumKnee
• Rotator CuffShoulder
• TendonAll Joints
• Wrist TendonsWrist

Substantial Equivalence

This Special 510(k) proposes a modification in the performance specifications and labeling for the ArthroCare System 12000, which was previously cleared in K071709 on August 7, 2007. The indications for use, technology, principle of operation, materials, packaging, and sterilization parameters of the ArthroCare System 12000 remain the same as in the predicate cleared 510(k).

Summary of Safety and Effectiveness

The modified ArthroCare System 12000, as described in this Special 510(k), is substantially equivalent to the predicate device. The proposed modifications in performance specifications and labeling are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the System.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the caduceus symbol.

Public Health Service

Image /page/3/Picture/3 description: The image shows a date, "OCT 1 5 2008". The date is written in a simple, sans-serif font. The letters and numbers are all capitalized and evenly spaced. The text is clear and easy to read.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ArthroCare Corporation % Ms. Valerie Defiesta-Ng Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-3523

Re: K082666

Trade/Device Name: ArthroCare System 12000 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 25, 2008 Received: September 29, 2008

Dear Ms. Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Valerie Defiesta-Ng

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Prosmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark N. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Kw82666

Page 1 of 2

Indications for Use Statement

ArthroCare System 12000 Device Name

K_____________________________________________________________________________________________________________________________________________________________________________ 510(k) Number:

Indications for Use:

The ArthroCare System 12000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

Arthroscopic and Orthopedic ProceduresJoint Specific or AllJoints (ankle, elbow,hip, knee, shoulder, andwrist)
Ablation and Debridement
• ACL/PCLKnee
• AcromioplastyShoulder
• Articular CartilageAll Joints
• BursectomyAll Joints
• ChondroplastyAll Joints
• FasciaAll Joints
• LigamentAll Joints
• NotchplastyKnee
• Scar TissueAll Joints
• Soft TissueAll Joints
• Subacromial DecompressionShoulder
• SynovectomyAll Joints
• TendonAll Joints
Excision and Resection
• Acetabular LabrumHip
• Articular LabrumAll Joints
• CapsuleAll Joints
• Capsular ReleaseKnee
• Cartilage FlapsKnee
• CystsAll Joints
• Discoid MeniscusKnee
• Frozen Shoulder ReleaseShoulder
• Glenoidale LabrumShoulder
• Lateral ReleaseKnee
• LigamentAll Joints
• Loose BodiesAll Joints
• Meniscal CystectomyKnee
• MeniscectomyKnee

əqmini (1)01 gift uoi

{6}------------------------------------------------

s.666

Continued

Arthroscopic and Orthopedic ProceduresJoint Specific or AllJoints (ankle, elbow,hip, knee, shoulder, andwrist)
• Plica RemovalAll Joints
• Scar TissueAll Joints
• Soft TissueAll Joints
• Synovial MembraneAll Joints
• TendonAll Joints
• Triangular Fibrocartilage (TFCC)Wrist
• VillusectomyKnee
Coagulation
• ACL/PCLKnee
• Articular CartilageAll Joints
• Carpal LigamentsWrist
• Glenohumeral CapsuleShoulder
• LigamentAll Joints
• Medial RetinaculumKnee
• Rotator CuffShoulder
• TendonAll Joints
• Wrist TendonsWrist

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-the-Counter Use

D.A.

X

(Division Sign-Off) (Division of General, Restorative, and Neurological Devices

510(k) Number K082666

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.