K Number
K082073
Manufacturer
Date Cleared
2008-09-11

(50 days)

Product Code
Regulation Number
888.3045
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Actifuse™ Flow Bone Graft Substitute is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse™ Flow Bone Graft Substitute can be injected into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral spinal fusion procedures with appropriate stabilizing hardware. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.

Device Description

Actifuse™ Flow is phase-pure silicon-substituted calcium phosphate osteoconductive bone void filler, comprising a single-phase calcium phosphate scaffold delivered in a matrix of resorbable polymer. The interconnected and open porous structure inherent of the source material for the mineral phase of Actifuse is similar to human cancellous bone. Actifuse Flow is available as a hydrated putty.

AI/ML Overview

This is a submission for a medical device (Actifuse™ Flow Bone Graft Substitute) and not an AI/ML device, so many of the requested fields (e.g., sample sizes for test/training sets, expert qualifications, MRMC studies) are not applicable. The provided document is a 510(k) Premarket Notification summary and an FDA clearance letter, which focuses on demonstrating substantial equivalence to a predicate device rather than performance metrics of an AI.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Bench Testing)Reported Device Performance (Bench Testing)
Meet requirements of all relevant standards for Calcium Salt Bone Void FillersConfirmed to meet requirements of all relevant standards for Calcium Salt Bone Void Fillers
Safe and effective in providing a scaffold for rapid bone repair via bony infiltration of the porous scaffoldConfirmed to be safe and effective in providing a scaffold for rapid bone repair via bony infiltration of the porous scaffold

2. Sample size used for the test set and the data provenance

Not applicable. This is a medical device (bone graft substitute), not an AI/ML device. The testing referenced is bench testing on the physical properties of the material, not a clinical study on human data in the context of AI.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable (see point 2).

4. Adjudication method for the test set

Not applicable (see point 2).

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is not an AI.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is not an AI.

7. The type of ground truth used

For bench testing, the "ground truth" would be established by the physical and chemical properties of the material itself and its performance against established engineering or biological standards for bone graft substitutes. The document states "Bench testing has shown Actifuse™ Flow to meet the requirements of all relevant standards for Calcium Salt Bone Void Fillers" and that it was "confirmed...safe and effective in providing a scaffold for rapid bone repair via bony infiltration of the porous scaffold." This suggests compliance with pre-defined criteria rather than an expert consensus or pathology review of a specific clinical outcome dataset.

8. The sample size for the training set

Not applicable. This device is not an AI.

9. How the ground truth for the training set was established

Not applicable. This device is not an AI.

{0}------------------------------------------------

K082073

Premarket Notification [510(K)] Summary (per 21 CFR 807.92)

SEP 1 1 2008

Submitter:ApaTech Limited370 Centennial AvenueElstree, Hertfordshire WD6 3TJUnited Kingdom
Contact Person:James Cassidy, PhD508 543 0700 (phone)508 543 5551 (fax)
Date Prepared:July 14, 2008
Classification:Resorbable calcium salt bone void filler devices have beenclassified by the Orthopedics Device Panel as Class II SpecialControls per 21 CFR 888.3045, Product Code MQV.
Trade Name:Actifuse™ Flow Bone Graft Substitute
Common Name:Bone Void Filler
Predicate Devices:Actifuse™ ABX E-Z-fil Putty Bone Graft Substitute, K071206

Intended use:

Actifuse™ Flow Bone Graft Substitute is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse™ Flow Bone Graft Substitute can be injected into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral spinal fusion procedures with appropriate stabilizing hardware. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.

Device Description

Actifuse™ Flow is phase-pure silicon-substituted calcium phosphate osteoconductive bone void filler, comprising a single-phase calcium phosphate scaffold delivered in a matrix of resorbable polymer. The interconnected and open porous structure inherent of the source material for the mineral phase of Actifuse is similar to human cancellous bone. Actifuse Flow is available as a hydrated putty.

Technological Characteristics and Substantial Equivalence

Actifuse Flow is composed of a porous calcium phosphate, equivalent to that contained in both predicate devices and to that in routine clinical use. The technologies employed in Actifuse Flow and its predicate devices are therefore substantially equivalent. Actifuse Flow has the same indications, contraindications, risks and potential adverse events as the predicate devices, and thus substantial equivalence is claimed for the device.

Testing

Bench testing has shown Actifuse™ Flow to meet the requirements of all relevant standards for Calcium Salt Bone Void Fillers. Testing has confirmed Actifuse Flow to be safe and effective in providing a scaffold for rapid bone repair via bony infiltration of the porous scaffold.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, with a staff and intertwined snakes.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 1 1 2008

ApaTech Limited % Ms. Candace F. Cederman Consultant 15058 Armel Drive Oregon City, Oregon 97045

Re: K082073

Trade/Device Name: Actifuse™ Flow Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV Dated: July 22, 2008 Received: July 23, 2008

Dear Ms. Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Candace F. Cederman

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR-Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Mullen

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Statement of Indications for Use

510(k) Number (if known): KOS 207 3

Device Name: Actifuse™ Flow

Indications for Use:

Actifuse™ Flow Bone Graft Substitute is a bone void filler intended only for orthopaedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. Actifuse™ Flow Bone Graft Substitute can be injected into bony voids or gaps of the skeletal system, i.e., extremities, pelvis, and spine, including use in posterolateral spinal fusion procedures with appropriate stabilizing hardware. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced by bone during the healing process.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of General, Restorative. and Neurological Devices

510(k) Numb

Page 1 of 1

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.