K Number
K082055
Device Name
MODEL PC-860 FETAL DOPPLER
Date Cleared
2009-02-06

(200 days)

Product Code
Regulation Number
884.2660
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.
Device Description
This Doppler tests the fetal heart rate through non-invasive ultrasonic Doppler Effect. As is known, ultrasonic wave propagating at a given frequency will be reflected when encountering an obstacle. If it is still an obstacle, the back wave will share the same frequency with the transmitted wave. Once the obstacle moves, the frequency of the back wave will be changed. The higher rate the object moves at, the bigger frequency change will take place. This is the so-called Doppler Effect. With the apparatus, the ultrasonic probe is placed on the abdomen of the pregnant woman. The ultrasonic probe can perceive the fetal heartbeat. When the transmitted wave encounters the fetal heart, the back wave will develop offset frequency. With the offset frequency, the fetal heart rate and frequency can be worked out.
More Information

Not Found

No
The description details the use of the Doppler effect to detect fetal heart rate and does not mention any AI or ML components or algorithms.

No.
The device is used to detect and listen to fetal heartbeats, providing diagnostic information rather than actively treating a condition.

No
The device detects fetal heart rate using the Doppler Effect but the description does not indicate that it is used to diagnose any disease or condition. It is for monitoring and listening to the fetal heartbeat.

No

The device description explicitly details the use of an ultrasonic probe and a built-in speaker, which are hardware components necessary for its function. The description of the Doppler effect and how the device works further confirms its reliance on physical hardware to transmit and receive ultrasonic waves.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
  • Device Function: This device uses ultrasonic technology applied externally to the abdomen of a pregnant woman to detect and measure the fetal heart rate. It does not analyze samples taken from the body.
  • Mechanism of Action: The description clearly explains the use of the Doppler effect with external ultrasound waves, not the analysis of biological samples.

Therefore, this device falls under the category of a medical device, specifically an ultrasonic fetal heart beat detector, rather than an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.

Product codes (comma separated list FDA assigned to the subject device)

KNG

Device Description

This Doppler tests the fetal heart rate through non-invasive ultrasonic Doppler Effect. As is known, ultrasonic wave propagating at a given frequency will be reflected when encountering an obstacle. If it is still an obstacle, the back wave will share the same frequency with the transmitted wave. Once the obstacle moves, the frequency of the back wave will be changed. The higher rate the object moves at, the bigger frequency change will take place. This is the so-called Doppler Effect. With the apparatus, the ultrasonic probe is placed on the abdomen of the pregnant woman. The ultrasonic probe can perceive the fetal heartbeat. When the transmitted wave encounters the fetal heart, the back wave will develop offset frequency. With the offset frequency, the fetal heart rate and frequency can be worked out.

As discussed in the General description, the Shenzhen Creative Industry Co., Ltd Fetal Doppler, model PC-860 is relatively simple to use. The Doppler is placed on the abdomen of the patient by first feeling out the fetal contour with hand to find the approximate position of fetal heart. Generally, fetal heart is at a location 113 of the lower abdomen (below the navel) during short pregnant weeks, and along with the pregnant weeks increasing, it moves upwards and lean to right or lean to left. To scan the abdomen, the user must first daub the acoustics surface of Doppler probe uniformly with the appropriate ultrasonic gel, and then put the probe on pregnant woman's abdomen (a location near fetal heart). It is important to ensure that probe contacts surface completely.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

ultrasonic

Anatomical Site

abdomen (pregnant woman)

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Clinical testing was used to validate the effectiveness and accuracy of the device. Specific testing relating to Ultrasound equipment was use to verify performance to recognized standards. All test results were satisfactory.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K040480

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 884.2660 Fetal ultrasonic monitor and accessories.

(a)
Identification. A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.(b)
Classification. Class II (performance standards).

0

K082055 page 1/3

FEB - 6 2009

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: May 14, 2008

  1. Company and Correspondent making the submission:

Name - Shenzhen Creative Industry Co., Ltd.

Address - 2/F, Block 3, Nanyou Tian'an Industry Town, Shenzhen, China

Telephone - +86-755-2643 4955

Fax - +86-755-2643 5433

Contact - Jianbin Liu

Email - liujianbin@creative-sz.com

  1. Device :

Trade/proprietary name: Fetal Doppler, Model PC-860 Common Name : Fetal Doppler Classification Name : Monitor, Ultrasonic, Fetal

Predicate Device: Edan Instruments, Inc., Ultrasonic Pocket Doppler Model: Sonotrax 510(K) number: K040480

  1. Classifications Names & Citations : 21CFR 884.2660, KNG, Fetal Doppler, Class II

1

4. Description

4.1 General

This Doppler tests the fetal heart rate through non-invasive ultrasonic Doppler Effect. As is known, ultrasonic wave propagating at a given frequency will be reflected when encountering an obstacle. If it is still an obstacle, the back wave will share the same frequency with the transmitted wave. Once the obstacle moves, the frequency of the back wave will be changed. The higher rate the object moves at, the bigger frequency change will take place. This is the so-called Doppler Effect. With the apparatus, the ultrasonic probe is placed on the abdomen of the pregnant woman. The ultrasonic probe can perceive the fetal heartbeat. When the transmitted wave encounters the fetal heart, the back wave will develop offset frequency. With the offset frequency, the fetal heart rate and frequency can be worked out.

4.2 Directions

As discussed in the General description, the Shenzhen Creative Industry Co., Ltd Fetal Doppler, model PC-860 is relatively simple to use. The Doppler is placed on the abdomen of the patient by first feeling out the fetal contour with hand to find the approximate position of fetal heart. Generally, fetal heart is at a location 113 of the lower abdomen (below the navel) during short pregnant weeks, and along with the pregnant weeks increasing, it moves upwards and lean to right or lean to left. To scan the abdomen, the user must first daub the acoustics surface of Doppler probe uniformly with the appropriate ultrasonic gel, and then put the probe on pregnant woman's abdomen (a location near fetal heart). It is important to ensure that probe contacts surface completely.

5. Indication for use :

The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.

2

6. Comparison with predicate device :

Shenzhen Creative Industry Co., Ltd., believes that the Fetal Doppler, Model PC-860 is substantially equivalent to the Edan Instruments, Inc., Ultrasonic Pocket Doppler, Model: Sonotrax, 510(K) number: K040480

    1. Safety and Performance Data :
      Electrical, mechanical, environmental safety and performance testing according to standard EN/IEC 60601-1. Clinical testing was used to validate the effectiveness and accuracy of the device. Specific testing relating to Ultrasound equipment was use to verify performance to recognized standards. All test results were satisfactory.
  1. Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Shenzhen Creative Industry, Inc., Ltd concludes that Fetal Doppler, Model PC-860 is safe and effective and substantially equivalent to predicate devices as described herein. ·

    1. Shenzhen Creative Industry Co., Ltd. will update and include in a summary any other information deemed seasonably necessary by the FDA.
      END

3

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH &" is vertically oriented on the left side of the logo, and the text "HUMAN SERVICES" is vertically oriented on the right side of the logo.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB - 6 2009

Shenzhen Creative Industry Co., Ltd. c/o Mr. Mark Job Reviewer Regulatory Technology Services LLC 1394 25th Street, N.W. BUFFALO MN 55313

Re: K082055

Trade/Device Name: PC-860 Fetal Doppler Regulation Number: 21 CFR §884.2660 Regulation Name: Fetal ultrasonic monitor and accessories Regulatory Class: II Product Code: KNG Dated: January 22, 2009 Received: January 23, 2009

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

4

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter,

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

Indications for Use

Indications for Use

510(k) Number (if known):

K082055

Device Name: PC-860 Fetal Doppler

Indications For Use:

The device is an ultrasonic fetal heart beat detector, which can detect the Fetal Heart Rate. The built in speaker of the device allows for listening of the fetal heartbeat.

Prescription Use ... . AND/OR S (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Aogn M. Whing


(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Device 510(k) Number