K Number
K081757
Date Cleared
2008-08-15

(56 days)

Product Code
Regulation Number
892.1100
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Verista Imaging ISOCAM II Gamma Camera is intended to image the distribution of radionuclides in the body by means of photon radiation detector(s).

Device Description

The ISOCAM II Gamma Camera is a medical imaging device used in nuclear medicine. It produces gamma ray distribution emitting radionuclides. It consists of one or more detectors mounted on a gantry that is connected to an acquisition system for operating the camera and storing the images. The base unit is a nuclear imaging device which utilizes gamma ray scintillation technologies. The modification to the device is limited to an upgrade to the mechanical operation of the device. The modifications improve patient safety through changes to the arm drive mechanical components and the addition of a weighted counter balance system.

AI/ML Overview

The provided text describes a 510(k) summary for the Verista Imaging ISOCAM II Gamma Camera. It focuses on demonstrating substantial equivalence to predicate devices, primarily due to a mechanical upgrade, rather than a clinical performance study involving acceptance criteria for diagnostic accuracy. Therefore, several requested tables and sections are "Not Applicable" or cannot be filled due to the nature of the submission.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Not specified. This submission focuses on mechanical upgrades and substantial equivalence, not clinical performance metrics with defined acceptance criteria.Not specified. The submission states that modifications improve patient safety through changes to mechanical components.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample size for test set: Not applicable. No clinical test set involving patient data was used for this submission. The testing conducted was non-clinical, focusing on mechanical changes.
  • Data provenance: Not applicable.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No expert review of a test set for ground truth establishment was conducted or mentioned, as the submission is for mechanical modifications, not clinical performance.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. No test set requiring adjudication was used.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a gamma camera, not an AI-powered diagnostic tool, and the submission is for mechanical upgrades. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a medical imaging device (gamma camera), not an algorithm or AI. The submission focuses on device mechanics.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable. No ground truth for diagnostic accuracy was established or used, as the submission is for mechanical modifications to an existing device. The software "remains identical" to the predicate device.

8. The sample size for the training set

Not applicable. No training set was used. This submission is for mechanical modifications to an existing device, not for training a new algorithm or AI model.

9. How the ground truth for the training set was established

Not applicable. No training set was used.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: ¥08 /.7 57

1. Submitter's Identification:

Verista Imaging, Inc. 22111 230th Avenue Centerville, lowa 52544

Telephone: (641) 856-5779

Contact: Mr. Kevin Wiskus

Date Summary Prepared: June 6, 2008

    1. Name of the Device: Verista Imaging ISOCAM II Gamma Camera
    1. Common or Usual Name: Scintillation (Gamma Camera)
    1. Predicate Device Information:

K940487, ISOCAM II (Dual Head Gamma Camera), Independent Scintillation Imaging Systems, Inc. K964834, ISOCAM II (Dual Head Gamma Camera), Park Medical Systems, Inc.

Lachine, Quebec, Canada K970128, ISOCAM I (Single Head Gamma Camera)/ISOCAM II (Dual Head Gamma Camera), Park Medical Systems, Inc., Lachine, Quebec, Canada

5. Device Description:

The modified Verista Imaging ISOCAM II Gamma Camera and the predicate devices are intended to image the distribution of radionuclides in the body by means of photon radiation detector(s).

AUG 1 5 2008

{1}------------------------------------------------

The ISOCAM II Gamma Camera is a medical imaging device used in nuclear medicine. It produces gamma ray distribution emitting radionuclides. It consists of one or more detectors mounted on a gantry that is connected to an acquisition system for operating the camera and storing the images. The base unit is a nuclear imaging device which utilizes gamma ray scintillation technologies. The modification to the device is limited to an upgrade to the mechanical operation of the device. The modifications improve patient safety through changes to the arm drive mechanical components and the addition of a weighted counter balance system.

The base unit is a nuclear imaging device which utilizes gamma ray scintillation technologies. The modification to the device is limited to an upgrade to the mechanical operation of the device. The modifications improve patient safety through changes to the arm drive mechanical components and the addition of a weighted counter balance svstem.

6. Intended Use:

Verista Imaging ISOCAM II Gamma Camera is intended to image the distribution of radionuclides in the body by means of photon radiation detector(s).

7. Comparison to Predicate Devices:

The base device pf the modified Verista Imaging ISOCAM II Gamma Camera has the same technological characteristics as the predicate gamma cameras. The software contained in the subject device has not been modified as a result of the modifications made to the device and remains identical to the software contained in the predicate device(s). The modifications made to the subject device do not raise any new issues of safety or effectiveness.

Discussion of Non-Clinical Tests Performed for Determination of 8. Substantial Equivalence are as follows:

Verista Imaging, Inc. has conducted testing to verify and validate the changes made to the previously cleared Park Medical Systems ISOCAM Gamma Camera.

9. Discussion of Clinical Tests Performed: Not Applicable

10. Conclusions:

Verista Imaging's ISOCAM II Gamma Camera that is the subject of this submission is a modification to the previously marketed Park Medical System's ISOCAM Gamma Cameras, to re-design an arm drive mechanism. The modified subject device has the

{2}------------------------------------------------

same intended use and indications for use, similar principles of operation, and similar technological characteristics as the previously cleared predicate device(s). Thus, the modified subject device is substantially equivalent to the predicate device(s).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 5 2008

Verista Imaging, Inc % Ms. Susan D. Goldstein-Falk Official Correspondent mdi Consultants, Inc. 55 Northern Blvd., Suite 200 GREAT NECK NY 11021

Rc: K081757

Trade/Device Name: Verista Imaging ISOCAM II Gamma Camera Regulation Number: 21 CFR 892.1100 Regulation Name: Scintillation (gamma) camera Regulatory Class: II Product Code: IYX Dated: July 31, 2008 Received: August 1, 2008

Dear Ms. Glodstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Exhibit D

Indications for Use

Page _ 1

510(k) Number (if known): _

Verista Imaging ISOCAM II Gamma Camera Device Name

Indications For Use:

Verista Imaging ISOCAM II Gamma Camera is intended to image the distribution of radionuclides in the body by means of photon radiation detector(s).

Prescription Use X (Per 21 CFR 801 Subpart D) OR

Over-The Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Healey Revers

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number __

N/A