(120 days)
The AMS Pelvic Floor Repair System is intended for use where the connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, gynecological, and gastroenterological anatomy. This includes but is not limited to the following procedures: pubourethral support, including urethral slings, vaginal wall prolapse repairs including anterior and posterior wall repairs, vaginal suspension, reconstruction of the pelvic floor and tissue repair.
The Apogee and Perigee prolapse repair systems are part of the AMS Pelvic Floor Repair System family of devices. They consist of a permanently-implanted mesh assembly and non-implanted surgical instruments (needle passers) that are used as aids to place the mesh assembly in the pelvic floor.
This K081710 510(k) submission describes a modification to an existing surgical mesh device. The core of the submission revolves around demonstrating that the modified device (with a PC coating) is "substantially equivalent" to its predicate device (without the coating). This type of submission rarely includes detailed studies with acceptance criteria in the way a new device would. Instead, the "acceptance criteria" are implicitly met if the new device performs similarly enough to the predicate device.
Here's the breakdown of the information requested, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Given that this is a 510(k) for a modified device, the "acceptance criteria" are not explicitly stated for clinical performance as they might be for a novel device. Instead, the study aimed to show "substantial equivalence" to a predicate device.
| Acceptance Criteria (Implied for Substantial Equivalence) | Reported Device Performance |
|---|---|
| Biocompatibility: Demonstrated no adverse biological reactions due to the new PC coating. | Components of Apogee and Perigee systems were tested for biocompatibility and found to be substantially equivalent to the predicate device. |
| Mechanical Performance (Tensile Strength): Maintained similar tensile strength in both weft and warp directions compared to the predicate. | Tensile testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Mechanical Performance (Cycle): Maintained similar cyclic loading performance in both weft and warp directions compared to the predicate. | Cycle testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Mechanical Performance (Elongation): Maintained similar elongation characteristics in both weft and warp directions compared to the predicate. | Elongation testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Mechanical Performance (Stiffness): Maintained similar stiffness characteristics in both weft and warp directions compared to the predicate. | Stiffness testing was performed in both the weft and warp directions of the mesh, comparing the mesh with and without the PC coating. Results indicated substantial equivalence. |
| Dimensional Characteristics (Fiber Diameter): Maintained similar fiber diameter compared to the predicate. | A comparison of the dimensional characteristics was completed and included fiber diameter. Results indicated substantial equivalence. |
| Dimensional Characteristics (Pore Size): Maintained similar pore size compared to the predicate. | A comparison of the dimensional characteristics was completed and included pore size. Results indicated substantial equivalence. |
| Dimensional Characteristics (Mesh Density): Maintained similar mesh density compared to the predicate. | A comparison of the dimensional characteristics was completed and included mesh density. Results indicated substantial equivalence. |
| Dimensional Characteristics (Mesh Thickness): Maintained similar mesh thickness compared to the predicate. | A comparison of the dimensional characteristics was completed and included mesh thickness. Results indicated substantial equivalence. |
| Indications for Use: The modified device has the same intended use as the predicate device. | The Indications for Use for the Apogee® and Perigee® Systems with PC Coated IntePro® Lite™ are identical to the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
The provided text does not specify sample sizes for mechanical testing or biocompatibility studies. The studies were likely bench-top (mechanical) and potentially in-vitro/in-vivo (biocompatibility) tests.
- Sample Size for Test Set: Not specified.
- Data Provenance: Not specified, but generally, such tests are conducted in a laboratory setting, typically in the country of the manufacturer or a contracted testing facility. It's retrospective in the sense that the data is generated for the submission, not from a clinical trial.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
This type of 510(k) submission, focused on material and mechanical equivalence for a modified surgical mesh, does not involve expert-established ground truth in the way a diagnostic AI device would. The "ground truth" for the mechanical tests would be the objective measurements themselves against established engineering standards or direct comparison to the predicate device's measured properties. For biocompatibility, it refers to standardized testing results.
4. Adjudication Method for the Test Set
Not applicable. This is not a study involving human interpretation or clinical outcomes requiring adjudication. The tests described are objective, physical, and biological assessments.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images with and without AI assistance. This submission is for a surgical mesh, not a diagnostic device.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was done
Not applicable. This submission is for a surgical mesh, which is a physical implant, not an algorithm or AI system.
7. The Type of Ground Truth Used
The "ground truth" for the tests mentioned (tensile, cycle, elongation, stiffness, dimensional characteristics, and biocompatibility) would be based on:
- Objective physical measurements: Using standardized testing methodologies and equipment to quantify mechanical properties and dimensions.
- Biocompatibility standards and assays: Results from validated in-vitro and potentially in-vivo tests comparing the new material's biological response to known standards or the predicate material.
8. The Sample Size for the Training Set
Not applicable. This submission is for a physical medical device (surgical mesh), not an AI algorithm, so there is no concept of a "training set."
9. How the Ground Truth for the Training Set Was Established
Not applicable. No training set was used.
{0}------------------------------------------------
PCT 1 5 2008
న. 510(k) Summary
| Submitter: | American Medical Systems (AMS)10070 Bren Road WestMinnetonka, MN 55343Phone: 952.933.6489FAX: 952.930.5785 |
|---|---|
| Contact Person: | Denise Thompson |
| Device Common Name: | Surgical Mesh |
| Device Trade Name: | Apogee® and Perigee® Systems withPC Coated IntePro® Lite |
| Device Classification Name: | Surgical Mesh, polymeric (OTP) |
| Predicate Devices: | AMS Pelvic Floor Repair System (K051485) |
Device Description
The Apogee and Perigee prolapse repair systems are part of the AMS Pelvic Floor Repair System family of devices. They consist of a permanently-implanted mesh assembly and non-implanted surgical instruments (needle passers) that are used as aids to place the mesh assembly in the pelvic floor.
Indications for Use
The AMS Pelvic Floor Repair System is intended for use where the connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, gynecological, and gastroenterological anatomy. This includes but is not limited to the following procedures: pubourethral support, including urethral slings. vaginal wall prolapse repairs including anterior and posterior wall repairs, vaginal suspension, reconstruction of the pelvic floor and tissue repair.
Comparison to Predicate Devices
The Apogee and Perigee System are identical to the previous with the exception of a synthetic phosphorylcholine (PC) polymer coating being added to the center section of the mesh assembly.
Supporting Information
The components of the Apogee and Perigee systems have been tested for biocompatibility and performance requirements and found to be substantially equivalent to the predicate device. A comparison of the mesh with and without the PC coating was made. Tensile, cycle, elongation and stiffness testing were performed in both the weft and warp directions of the mesh. In addition, a comparison of the dimensional characteristics was completed and included fiber diameter, pore size, mesh density and mesh thickness.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three abstract human profiles facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Denise Thompson Senior Regulatory Specialist American Medical Systems 10700 Bren Road West MINNETONKA MN 55343
SEP 2 8 2012
K081710 Re: Trade/Device Name: Apogee® System with PC Coated IntelPro® Lite™ Perigree® System with PC Coated IntelPro® Lite™ Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP Dated: October 3, 2008 Received: October 6, 2008
Dear Ms. Thompson:
This letter corrects our substantially equivalent letter of October 15, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{2}------------------------------------------------
Page 2 -
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Sincerely yours,
Benjamin R. Weeks
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known):
Device Name: Apogee® System with PC Coated IntePro® Lite™ Perigee® System with PC Coated IntePro® Lite™
Indications For Use:
The AMS Pelvic Floor Repair System is intended for use where the connective tissue has ruptured or for implantation to reinforce soft tissucs where weakness exists in the urological, gynecological, and gastroenterological anatomy. This includes but is not limited to the following procedures: pubourethral support, including urethral slings, vaginal wall prolapse repairs including anterior and posterior wall repairs, vaginal suspension, reconstruction of the pelvic floor and tissue repair.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED
Neil RPO sple for mxn
(Division Sion. General. Restorative. Division and Neurologica! Beri
510(k) Number
i.
§ 884.5980 Surgical mesh for transvaginal pelvic organ prolapse repair.
(a)
Identification. Surgical mesh for transvaginal pelvic organ prolapse repair is a prescription device intended to reinforce soft tissue in the pelvic floor. This device is a porous implant that is made of synthetic material, non-synthetic material, or a combination of synthetic and non-synthetic materials. This device does not include surgical mesh for other intended uses (§ 878.3300 of this chapter).(b)
Classification. Class III (premarket approval).(c)
Date premarket application approval or notice of completion of a product development protocol is required. A premarket application approval or notice of completion of a product development protocol for a device is required to be filed with the Food and Drug Administration on or before July 5, 2018, for any surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976, or that has, on or before July 5, 2018, been found substantially equivalent to a surgical mesh described in paragraph (a) of this section that was in commercial distribution before May 28, 1976. Any other surgical mesh for transvaginal pelvic organ prolapse repair shall have an approved premarket application or declared completed product development protocol in effect before being placed in commercial distribution.