(202 days)
No
The description focuses on standard oscillometric blood pressure measurement and data transmission, with no mention of AI or ML algorithms for analysis or interpretation.
No.
The device is intended for measuring blood pressure and pulse rate, not to treat or cure a condition.
Yes
The device is intended for measuring blood pressure and pulse rate, which are diagnostic indicators of health.
No
The device description clearly states it is a "wrist non-invasive blood pressure device" and describes a physical blood pressure meter with a cuff and a radio module for data transmission. This indicates a hardware component is central to the device's function.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections.
- Device Function: The tenso-comfort BPM 105 / BPM 205 is a non-invasive blood pressure monitor. It measures blood pressure and pulse rate directly from the body (at the wrist) without requiring any samples to be taken and analyzed outside the body.
Therefore, based on the provided information, this device falls under the category of a non-invasive medical device, not an In Vitro Diagnostic.
N/A
Intended Use / Indications for Use
The tenso-comfort BPM 105 /BPM 205 is a wrist non-invasive blood pressure device which is intended for use in measuring blood pressure and pulse rate in adult patient population. The measuring method is an oscillometric blood pressure measurement with automatic sequence and refers to the auscultatory method as the reference standard. The model version BPM 105 is equipped with a radio module to transmit the measurement data to a PC.
The device is not intended for neonatal use.
Product codes
DXN
Device Description
The non-invasive wrist blood pressure meter BPM105 and BPM205 determine the arterial blood pressure by means of the oscillometric blood pressure measuring method. With this method the pressure fluctuations are measured, which develop when depressing pulse-cyclic blood pulses in the compressed arteria under the blood pressure cuff put on. The blood pressure apparatus in the model variant BPM105 is equipped with a radio module, with which the transmission of the stored measured values on a PC is optionally possible (radio interface for PC and software is offered as accessory).
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
wrist
Indicated Patient Age Range
adult patient population.
The device is not intended for neonatal use.
Intended User / Care Setting
Lay users
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
510(k) Summary
The assigned 510(k) number is:
Date Prepared:
Submitter Information:
Official Correspondent:
Mrs. Marion Otto Quality Manager Biocomfort GmbH & Co. KG Phone: +49 7158 98016-48 +49 7158 98016-40 Fax: E-mail: otto@biocomfort.de
Burlington, MA 01803 USA Phone: +1 866 294 8267 E-mail: schill@biocomfort.com
Mr. Dieter Schill
President Biocomfort Inc. 23 Third Avenue
Biocomfort Diagnostics GmbH & Co. KG
Registration Number: 3006493236 Owner Operator Number: 10023755
Ko 81498
Bernhaeuser Strasse 17 73765 Neuhausen a.d.F.
May 19, 2008
Germany
US Agent (Contact):
Device Trade Name: Common Name: Device Classification Name: Product Code: Device Classification No.: Regulatory Status:
Predicate Devices: Device Trade Name: 510(k) Number: Device Classification Name: Product Code: Device Classification No .: Regulatory Status:
Device Description:
tenso-comfort BPM 105 / 205 Blood pressure meter System, Test, Non- invasive Blood Pressure meter, Over The Counter DXN Part 870.1130 Class II
Clever TD-3018A Clever TD-3018A K051703 Blood pressure meter, Over The Counter DXN Part 870.1130 Class II
The non-invasive wrist blood pressure meter BPM105 and BPM205 determine the arterial blood pressure by means of the oscillometric blood pressure measuring method. With this method the pressure fluctuations are measured, which develop when depressing pulse-cyclic blood pulses in the compressed arteria under the blood pressure cuff put on. The blood pressure apparatus in the model variant BPM105 is equipped with a radio module, with which the transmission of the stored
06 - 510(k) Summary - Page 1 of 3
DEC 1 7 2008
1
measured values on a PC is optionally possible (radio interface for PC and software is offered as accessory).
Intended Use:
The tenso-comfort BPM 105 /BPM 205 is a wrist non-invasive blood pressure device which is intended for use in measuring blood pressure and pulse rate in adult patient population. The measuring method is an oscillometric blood pressure measurement with automatic sequence and refers to the auscultatory method as the reference standard. The model version BPM 105 is equipped with a radio module to transmit the measurement data to a PC.
The device is not intended for neonatal use..
SE Discussion: | ||||
---|---|---|---|---|
Substantial Equivalent | ||||
Device | Predicate Devices | |||
Biocomfort tenso- | ||||
comfort BPM 105/ BPM | ||||
205 | TaiDoc Clever TD-3018A | Discussion of differences | ||
[01] | Indication for | |||
use | The tenso-comfort BPM | |||
105 /BPM 205 is a wrist | ||||
non-invasive blood | ||||
pressure device which is | ||||
intended for use in | ||||
measuring blood pressure | ||||
and pulse rate in adult | ||||
patient population. The | ||||
measuring method is an | ||||
oscillometric blood | ||||
pressure measurement | ||||
with automatic sequence | ||||
and refers to the | ||||
auscultatory method as | ||||
the reference standard. | ||||
The model version BPM | ||||
105 is equipped with a | ||||
radio module to transmit | ||||
the measurement data to | ||||
a PC. | The Clever TD-3018A | |||
Blood Pressure Monitor | ||||
provide intended to use | ||||
non-invasive measure the | ||||
systolic and diastolic | ||||
blood pressure and pulse | ||||
rate of an adult | ||||
individual, over age 18, | ||||
at home by using a non- | ||||
invasive technique in | ||||
which an inflatable cuff is | ||||
wrapped around the | ||||
wrist. The cuff | ||||
circumference is limited | ||||
to 5.25"~ 7.75". | The BPM 105 is equipped | |||
with a transmitter to send | ||||
the measuring results to a | ||||
PC. | ||||
[02] | Target | |||
population | Adults, | |||
Lay users | Adults | |||
Lay users | Equivalent | |||
[03] | Measuring | |||
principle | Oscillometric method | Oscillometric method | Equivalent | |
[04] | Type of results | Pressure: mmHg | ||
Puls: beats/minute | Pressure: mmHg | |||
Puls: beats/minute | Equivalent | |||
[05] | Presentation of | |||
results | LCD Digital Display | LCD Digital Display | Equivalent | |
[06] | Measurement | |||
range | Pressure: 0 - 300 mmHg | |||
Pulse: 40 - 199 | ||||
beats/minute | Pressure: 0 - 300 mmHg | |||
Pulse: 40 - 199 | ||||
beats/minute | Equivalent | |||
[07] | Measuring | |||
accuracy | Pressure: ±3 mmHg | |||
Pulse: ±5% of the value | Pressure: ±3 mmHg or | |||
2% of reading | ||||
Pulse: ±4% of the reading | In the same area and | |||
considered equivalent | ||||
Substantial Equivalent | ||||
Device | Predicate Devices | |||
Biocomfort tenso- | ||||
comfort BPM 105/ BPM | ||||
205 | TaiDoc Clever TD-3018A | Discussion of differences | ||
[08] | Inflation | Automatic inflation | Automatic inflation | Equivalent |
[09] | Deflation | Electric Valve | Electric Valve | Equivalent |
[10] | Pressure | |||
release | Automatic exhaust valve | Automatic exhaust valve | Equivalent | |
[11] | Pressure | |||
detection | Piezo-resistive silicon | |||
pressure transducer | Piezo-resistive silicon | |||
pressure transducer | Equivalent | |||
[12] | Measuring | |||
period | App. 30 seconds | App. 20 seconds | Equivalent | |
[13] | Operation | |||
environment | 10°C - 40°C | |||
50°F - 104°F | 10°C - 40°C | |||
50°F - 104°F | Equivalent | |||
[14] | Storage | |||
environment | -20°C - 60°C | |||
-4°F - 140°F | ||||
10% - 95% relative | ||||
humidity | -20°C - 60°C | |||
-4°F - 140°F | ||||
10% - 95% relative | ||||
humidity | Equivalent | |||
[15] | Battery life | App. 300 measurements | App. 200 uses | Improved energy |
management and | ||||
considered equivalent | ||||
[16] | Cuff size | 135mm - 220mm | ||
5 ½ in - 8 ⅜ in | 135mm - 195mm | |||
5 ¼ in - 7 ¾ in | Slightly larger cuff and | |||
considered equivalent | ||||
[17] | Dimensions | 70 x 90 x 26 mm | 76 x 64 x 29 mm | Slightly different |
dimension due to | ||||
corporate design. The | ||||
devices are considered | ||||
equivalent. | ||||
[18] | Weight | Ca. 140 g (without | ||
batteries) | 132 g (with batteries) | Equivalent | ||
[19] | Mobility | Hand-held | Hand-held | Equivalent |
[20] | Memory | |||
Capability | 110 measurements per | |||
user (up to 8 users) with | ||||
date and time | 352 sets of reading with | |||
date and time | Different number of | |||
measurement due to the | ||||
user management. | ||||
It is considered | ||||
equivalent. | ||||
[21] | Power supply | Two 1.5 alkaline batteries | ||
type AAA/LR03 | Two 1.5 alkaline batteries | |||
type AAA/LR03 | Equivalent | |||
[22] | Connectivity | Wireless interface with | ||
2.4 GHz according to | ||||
IEEE802.15.4, with 10 m | ||||
indoor range | None | The model BPM 105 | ||
provides the option to | ||||
communicate via wireless | ||||
technology to a PC which | ||||
is set up for the interface | ||||
protocol |
06 – 510(k) Summary – Page 2 of 3
2
Discussion of the Substantial Equivalence Decision:
The only difference between the BPM 105/205 and the predicate device is the ability to communicate rit only unceless connection. This feature does neither affect nor even represent the measurement of with a r e vireless connection is for the upload of measurement data and the set up of basic device functions.
3
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's emblem in the center. The emblem is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 7 2008
Mr. Dieter Schill President Biocomfort Inc. 23 Third Avenue Burlington, MA 01803
Re: K081498
Trade/Device Name: tenso-comfort BPM 105/205 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: December 12, 2008 Received: December 17, 2008
Dear Mr. Schill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part
4
Page 2 - Mr. Dieter Schill
807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
07 Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Indications for Use
510(k) Number (if known): K081498
Device Name: tenso-comfort BPM 105 / 205
The tenso-comfort BPM 105 /BPM 205 is a wrist non-invasive blood pressure device which is intended for use in measuring blood pressure and pulse rate in adult patient population. The measuring method is an oscillometric blood pressure measurement with automatic sequence and refers to the auscultatory method as the reference standard. The model version BPM 105 is equipped with a radio module to transmit the measurement data to a PC.
The device is not intended for neonatal use.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(Please Do not Write Below This Line-Continue on another Page OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) | |
---|---|
(Division Sign-Off) | |
12/17/08 | |
Division of Cardiovascular Devices | |
4081498 |
510(k) Number