K Number
K081451
Device Name
RIGHTEST BLOOD GLUCOSE MONITORING SYSTEM, MODEL GM100
Date Cleared
2008-09-18

(118 days)

Product Code
Regulation Number
862.1345
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or people with diabetes at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from fingertip, palm and forearm by using Rightest Blood Glucose Monitoring System. This test device is not intended for testing neonate blood samples.
Device Description
Our Blood Glucose Monitoring System consists of a Meter, Blood Glucose Test Strips, Two Control Solutions, Lancing Device and lancets. The Rightest Meter, Blood Glucose Test Strips are manufactured by BIONIME Corporation. The Rightest Meter, when used with the Rightest Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System is verified by the Control Solution.
More Information

Not Found

No
The summary describes a standard blood glucose monitoring system using test strips and a meter, with no mention of AI/ML capabilities or related concepts like image processing, training/test sets, or complex algorithms beyond basic glucose measurement.

No
The device is a monitoring system intended for diagnostic use to measure glucose concentration, not to provide therapy.

Yes
The "Intended Use / Indications for Use" section states: "The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or people with diabetes at home to measure the glucose concentration for aiding diabetes management." This clearly identifies it as a diagnostic device.

No

The device description explicitly states that the system consists of a Meter, Blood Glucose Test Strips, Two Control Solutions, Lancing Device, and lancets, which are all hardware components.

Yes, this device is an IVD (In Vitro Diagnostic).

The "Intended Use / Indications for Use" section explicitly states: "The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body)."

N/A

Intended Use / Indications for Use

The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or people with diabetes at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from fingertip, palm and forearm by using Rightest Blood Glucose Monitoring System. This test device is not intended for testing neonate blood samples.

Product codes

NBW, CGA

Device Description

Our Blood Glucose Monitoring System consists of a Meter, Blood Glucose Test Strips, Two Control Solutions, Lancing Device and lancets. The Rightest Meter, Blood Glucose Test Strips are manufactured by BIONIME Corporation. The Rightest Meter, when used with the Rightest Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System is verified by the Control Solution.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Fingertip, palm and forearm

Indicated Patient Age Range

Not intended for testing neonate blood samples.

Intended User / Care Setting

professional healthcare personnel or people with diabetes at home

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The clinical test was designed in Alternative site testing study as below
Test capillary blood Study:
It shows similarly slope and intercept for difference position of capillary blood test.

TechnicianRightest fingerstick vs Olympus-PlasmaRightest palmstick vs Olympus-PlasmaRightest armstick vs Olympus-Plasma
Test range (mg/dL)35~56834~54239~572
Test number164164164
Slope0.990.960.96
Intercept0.746.350.46
r0.9930.9900.987

The "Alternative Site Test" clinical evaluation shows substantial equivalence to Rightest used in finger, palm and arm position. They all have similar slope and intercept of Rightest value versus Olympus AU2700. So the result tells us Rightest blood glucose monitoring system, model GM100 is suitable to be used in fingertip, palm and forearm.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K042678, K053635, K062567

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

0

Exhibit 1

510(K) SUMMARY

This summary of 5l0(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is:_KQ8 145

  • Submitter's Identification: 1.
    BIONIME CORPORRATION BIONME CONT ORFORD, DALI CITY, TAICHUNG COUNTY, TAIWAN 412 Contact Person: Mr. George Chi Phone Number: 886-4-24951268 FAX Number: 886-4-24952568

Date Summary Prepared: July 13, 2007

  • Name of the Device: Rightest Blood Glucose Monitoring System, Model GM100 2.
  • Common or Usual Name: Glucose test system 3. Panel: Clinical Chemistry 75 Product Code: NBW, System, Test, Blood Glucose, Over-the-Counter. Classification: Class II
    1. Device Description:

Our Blood Glucose Monitoring System consists of a Meter, Blood Glucose Test Strips, Two Control Solutions, Lancing Device and lancets. The Rightest Meter, Blood Glucose Test Strips are manufactured by BIONIME Corporation. The Rightest Meter, when used with the Rightest Test Strips, quantitatively measures glucose in fresh capillary whole blood. The performance of the Rightest Blood Glucose Monitoring System is verified by the Control Solution.

  • Intended Use: 5.
    The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or people with diabetes at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from fingertip, palm and forearm by using Rightest Blood Glucose Monitoring System. This test device is not intended for testing neonate blood samples.

1

Special condition for use statement(s): Rightest system provides plasma equivalent results.

Predicate Device Information: ે.

The Rightest Blood Glucose Monitoring System, model GM100 is substantially equivalent to the brand of Rightest Blood Glucose Monitoring System (Alternative Site Testing) noted below.

Rightest Blood Glucose Monitoring System, model GM100 Name: Device Company: Bionime Corporation K042678, K053635 and K062567 510(K) Number:

Similarities
ItemSubject DevicePredicate Device(s)
Rightest BGMS GM100
(Alternative Site Testing)Rightest BGMS
(Alternative Site Testing)
Detection methodAmperometryAmperometry
EnzymeGlucose Oxidase
(Aspergillus niger)Glucose Oxidase
(Aspergillus niger)
MediatorPotassium ferricyanidePotassium ferricyanide
Hematocrit Range30 - 55%30 - 55%
Temperature range50 - 104° F
10 - 40° C50 - 104° F
10 - 40° C
Humidity range10 - 90%10 - 90%
Warranty(meter)3 years3 years
Open use time
(strip)3 months3 months
ElectrodeNoble metal electrodeNoble metal electrode
Test range20 - 600 mg/dL20 - 600 mg/dL
Test Time8 seconds8 seconds
Sample Volume1.4 uL1.4 uL
Sample SourceThe glucose
concentration is
measured with
quantitative capillary
whole blood from the
fingertip, palm and
forearm by using Rightest
Blood Glucose
Monitoring System.The glucose
concentration is
measured with
quantitative capillary
whole blood from the
fingertip, palm and
forearm by using Rightest
Blood Glucose
Monitoring System.
The unit of
measurement dataFix on mg/dLFix on mg/dL

Comparison to Predicate Devices: 7.

2

| Annual Property of Children Commenssion in Children of Children to Children

Battery lifeRunning 1,000 testRunning 1,000 test
Acres of the research division for the
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Differences
ItemSubject Device(s)Predicate Device(s)
Rightest BGMS GM100
(AST)Rightest BGMS(AST)
Description and
LabellingWe mention the
information about
modification in user's
Instruction. We also show
a diagrammatic
explanation about
alternative test sites in
user's Instruction.We mention the
information about
modification in user's
manual. We also show a
diagrammatic explanation
about alternative test sites
in user's manual.
Coding
InterferenceNo codingCode key
Uric acid > 9.0 mg/dL
L-Doga > 1.5 ml/dL
Methyldopa > 1.5 mg/dL
Cholesterol > 250 mg/dL
Memory Capacity10 blood glucose test
results with date and time300 blood glucose test
results with date and time
Power SavingTurn off automatically after
2 minutesTurn off automatically after
3 minutes
Power SupplyOne CR2032 batteryTwo 1.5V(AAA) batteries

Discussion of Non-Clinical Tests Performed for Determination of Substantial 8. Equivalence are as follows:

Verification and validation of test results were evaluated to establish the performance, functionality and reliability of The Rightest Blood Glucose Monitoring System.

The evaluation included precision, linearity, interference, hematocrit and control solution.

  • Discussion of Clinical Tests Performed: ்.
    The clinical test was designed in Alternative site testing study as below

Test capillary blood Study:

It shows similarly slope and intercept for difference position of capillary blood test.

3

| Technician | Rightest fingerstick vs
Olympus-Plasma | Rightest palmstick vs
Olympus-Plasma | Rightest armstick vs
Olympus-Plasma |
|--------------------|-------------------------------------------|-----------------------------------------|----------------------------------------|
| Test range (mg/dL) | 35568 | 34542 | 39~572 |
| Test number | 164 | 164 | 164 |
| Slope | 0.99 | 0.96 | 0.96 |
| Intercept | 0.74 | 6.35 | 0.46 |
| r | 0.993 | 0.990 | 0.987 |

Fig 1 J inear regression from Rightest versus Olympus AU2700
------------------------------------------------------------------

The "Alternative Site Test" clinical evaluation shows substantial equivalence to Rightest used in finger, palm and arm position. They all have similar slope and intercept of Rightest value versus Olympus AU2700. So the result tells us Rightest blood glucose monitoring system, model GM100 is suitable to be used in fingertip, palm and forearm.

10. Conclusions:

Results of clinical testing demonstrate that the performance of the Rightest Blood Glucose Monitoring System, model GM100 testing capillary whole blood is substantial equivalence of Rightest Blood Glucose Monitoring System (AST). The precision and accuracy of Rightest is suitable for its in monitoring the effectiveness of diabetes management at home and in clinical settings.

4

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle faces right. Encircling the eagle is text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 1 8 2008

Bionime Corporation, MDI Consultant, Inc. c/o Ms. Susan D. Goldstein-Falk Official Correspondent 55 Northern Blvd., Suite 200 Great Neck, NY 11021

Re: K081451

Trade Name: Rightest Blood Glucose Monitoring System, Model GM100 Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW, CGA Dated: August 15, 2008 Received: August 18, 2008

Dear Ms. Goldstein-Falk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

5

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Jean M. Cooper, M.S., D.V.M.

Yéan M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

6

Indications for Use

510(k) Number (if known): K081451

Device Name: Rightest Blood Glucose Monitoring System, Model GM100

Indications For Use:

The Rightest Blood Glucose Monitoring System is intended for in vitro diagnostic use (outside of body). It is indicated to be used by professional healthcare personnel or diabetics at home to measure the glucose concentration for aiding diabetes management. The glucose concentration is measured with quantitative capillary whole blood from the fingertip, palm, and forearm by using Rightest Blood Glucose Monitoring System. This device is not intended for testing neonate blood samples.

Special conditions for use statement(s): Rightest System provides plasma equivalent results.

Prescription Use × AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Ann Chappie
Division Sign-Off

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k)________________________________________________________________________________________________________________________________________________________________________

Page 1 of