K Number
K081381

Validate with FDA (Live)

Manufacturer
Date Cleared
2008-06-13

(28 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Non-cervical spinal fixation devices intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis; and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion.

Non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (1,3 to sacrum ) with removal of the implants after the attainment of a solid fusion.

Device Description

The Range Spinal System is a top-loading, multiple component, posterior system which consists of pedicle screws, rods, hooks, and transverse connectors. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

Materials: The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO standards,

Function: The system functions as an adjunct to fusion to provide immobilization of spinal segments of the thoracic, lumbar and / or sacral spine.

AI/ML Overview

The medical device being described is the Range Spinal System.

Here's an analysis of the provided text regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implicit)Reported Device Performance
Mechanical/Biomechanical Performance: The device (Range Spinal System) must perform at least equally to, or better than, predicate devices and other currently marketed systems in ASTM F1717 testing, which likely encompasses various biomechanical tests related to spinal fixation systems (e.g., strength, stiffness, fatigue). Design Features and Sizing: The design features and sizing of components must be substantially the same as predicate systems, implying that it fits within established anatomical and surgical norms for spinal fixation. Material Composition: Must be manufactured from FDA-recognized materials for spinal implants (specifically Ti6Al-4V ELI alloy per ASTM and ISO standards). Intended Use: The device must be indicated for the same intended uses as legally marketed predicate devices, encompassing various spinal conditions requiring fixation (e.g., degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis, failed previous fusion, and severe spondylolisthesis at L5-S1 for pedicle screw fixation).Mechanical/Biomechanical Performance: "The Range Spinal System was biomechanically tested and compared to the predicate systems and other currently marketed systems and performed equal to or better than these systems in ASTM testing to F1717." Design Features and Sizing: "The design features and sizing of the components were also compared and the Range Spinal System found to be substantially the same as these systems." Material Composition: "It is manufactured from the same FDA recognized materials [Ti6Al-4V ELI alloy per ASTM and ISO standards]." Intended Use: "It is indicated for the same intended uses as these systems." (The specific indications are listed in section 5 of the 510(k) summary and reiterated in the Indications for Use document.)

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not specify a sample size for a "test set" in the context of clinical data or patient outcomes. The evaluation is primarily based on biomechanical testing against established ASTM standards and comparison to predicate devices, and not on a clinical trial with a defined patient test set.

The data provenance is related to the biomechanical testing results of the Range Spinal System itself, compared against predicate devices. It is therefore prospective in the sense that the new device was tested, but the comparison is against established, already-marketed devices. The country of origin for the data is not specified, but the testing was done according to ASTM standards, which are international.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

This information is not applicable to this document. The "ground truth" for this medical device's acceptance is based on adherence to biomechanical performance standards (ASTM F1717) and substantial equivalence to legally marketed predicate devices in terms of design, materials, and intended use. There is no mention of expert consensus on patient data or clinical outcomes being used to establish ground truth in this submission.

4. Adjudication Method for the Test Set

This information is not applicable as there is no human-adjudicated "test set" described in the context of clinical performance or diagnostic accuracy. The evaluation revolves around engineering and material properties, and comparison to predicate devices.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data, and their performance is compared with and without AI assistance. The Range Spinal System is an implantable spinal fixation device, not a diagnostic tool requiring interpretation by human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

N/A. This question is primarily relevant to AI/diagnostic algorithms. The Range Spinal System is a physical medical implant. The determination of its performance is through physical testing (biomechanical) and comparison to existing physical devices, not through an algorithm operating in isolation.

7. The Type of Ground Truth Used

The "ground truth" for the Range Spinal System's substantial equivalence determination is based on:

  • Biomechanical performance standards: Compliance with ASTM F1717.
  • Material standards: Use of Ti6Al-4V ELI alloy per ASTM and ISO standards.
  • Comparison to predicate devices: Demonstration of "substantially the same" design, function, material, and intended use as other legally marketed spinal systems.

There is no mention of pathology, expert consensus on clinical cases, or outcomes data being used as ground truth for this 510(k) submission.

8. The Sample Size for the Training Set

This information is not applicable. The Range Spinal System is a physical medical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a sequence of alphanumeric characters, specifically 'K081381'. The characters are rendered in a bold, sans-serif font, giving them a clear and distinct appearance. The black color of the characters contrasts sharply against the white background, enhancing their legibility.

510(k) Summary for the Range Spinal System,

This safety and effectiveness summary for the Mesa Spinal per Section 513(i)(3) of the Food, Drug and Cosmetic Act.

  1. Submitter : K2M, Inc. 751 Miller Drive SE, Suite F1

Leesburg, VA 20175

Date Prepared: May 07, 2008 Contact Person : Richard W. Woods K2M, LLC 751 Miller Drive SE, Suite F1 Leesburg, VA 20175 Telephone: 703-777-3155

2. Tradename:Range Spinal System
Common Name:Spine Fixation System
Classification Name:Pedicle Screw Spinal System (21 CFR 888.3070(b)(1))Spinal Interlaminal fixation orthosis (21 CFR 888.3050)

3. Predicate or legally marketed devices which are substantially equivalent :

  • . Range Spinal System ( K2M, Inc.)
  • . Expedium (DcPuy)

4. Description of the device :

The Range Spinal System is a top-loading, multiple component, posterior system which consists of pedicle screws, rods, hooks, and transverse connectors. All of the components are available in a variety of sizes to match more closely the patient's anatomy.

Materials: The devices are manufactured from Ti6Al-4V ELI alloy per ASTM and ISO standards,

Function: The system functions as an adjunct to fusion to provide immobilization of spinal segments of the thoracic, lumbar and / or sacral spine.

5. Intended Use:

The Range Spinal System is a:

Non-cervical spinal fixation device intended for posterior, non-pedicle fixation for the following indications; degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis, kyphosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The Mesa Spinal System is also intended for:

Non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum ) with removal of the implants after the attainment of a solid fusion.

6. Comparison of the technological characteristics of the device to predicate and legally marketed

devices :

The Range Spinal System was biomechanically tested and compared to the predicate systems and other currently marketed systems and performed equal to or better than these systems in ASTM testing to F1717. The design features and sizing of the components were also compared and the Range Spinal System found to be substantially the same as these systems. It is manufactured from the same FDA recognized materials and is indicated for the same intended uses as these systems. There are no significant differences between the Range Spinal Systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.

page 1 of )

{1}------------------------------------------------

Public Health Service

Image /page/1/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling a bird or a human figure with outstretched arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 19 2009

K2M, Inc. % Ms. Nancy Giezen 751 Miller Drive, Southeast, Suite F1 Leesburg, Virginia 20175

Re: K081381

Trade/Device Name: Range Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNH, MNI, KWP Dated: May 16, 2008 Received: May 16, 2008

Dear Ms. Giezen:

This letter corrects our substantially equivalent letter of June 13, 2008.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2- Ms. Nancy Giezen

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Sincerely yours,

Daniel Krane

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

KOQ138 |

Indications for Use

510(k) Number: 300000000

Device Name: Range Spinal System

Indications for Use:

Non-cervical spinal fixation devices intended for posterior, non-pedicle fixation for the following indications: degenerative disc disease ( DDD ) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies ); spondylolisthesis; trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures ( i.e. scoliosis; and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion.

Non-cervical pedicle screw fixation for the following indications: trauma ( i.e. fracture or dislocation ); spinal stenosis; curvatures (i.e. scoliosis; and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. It is also indicated for the treatment of severe spondylolisthesis ( grades 3 and 4 ) of the LS-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (1,3 to sacrum ) with removal of the implants after the attainment of a solid fusion.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-the-counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS-LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R. Odlum

Division of General, Restorat and Neurological Devices

510(k) Number K081391

1 of 1

N/A