K Number
K081086

Validate with FDA (Live)

Date Cleared
2008-08-13

(119 days)

Product Code
Regulation Number
868.5730
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Portex® Blue Line® SACETT™ is intended for oral intubation only for airway management of patients anticipated to require prolonged mechanical ventilation in critical care units. The product may be inserted in the emergency room or pre-hospital to avoid the need for re-intubation, SACETT™'s dorsal suction lumen allows drainage by continuous or intermittent suctioning of contaminated mucous and subglottis secretion that accumulate above the cuff.

Device Description

Single use Tracheal tube with integral suction line for lavage and evacuation above the cuff. The product design is based upon a tracheal tube with the addition of a third (integral) lumen within the tube. The lumen terminates above the cuff via a 'notch' which enables the entrance (via suction) of secretions which have pooled above the cuff into the third (suction) lumen. Approximately half way along the tube length the suction lumen is joined to a suction line which is external to the main tube. The suction line is joined to the suction lumen in a similar manner to that of the joint between the inflation line and the inflation lumen. The distal (clinician) end of the suction line terminates in a connector which can be connected to either suction tubing or a syringe.

AI/ML Overview

This is a 510(k) summary for a medical device called the Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube. It is a premarket notification to the FDA to demonstrate substantial equivalence to legally marketed predicate devices.

Here's an analysis of the provided information concerning acceptance criteria and study data:

1. Table of Acceptance Criteria and Reported Device Performance:

Based on the provided text, there are no explicit acceptance criteria or reported device performance metrics that would typically be found in a study proving the device meets specific performance thresholds (e.g., sensitivity, specificity, accuracy).

The document is a 510(k) premarket notification, which focuses on demonstrating "substantial equivalence" to existing, legally marketed predicate devices, rather than proving a new device's performance against predefined acceptance criteria from a clinical trial.

Instead, the submission states:

  • "The Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube has the same intended use as the Mallinckrodt Medical Hi-Lo™ Evac Endotracheal Tube."
  • "The Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube is composed essentially of the same materials as the Portex Tracheal Tube and the Portex® Blue Line® Ultra Suctionaid (BLUS) Tracheostomy Tube."
  • "The Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube incorporates a number of design and performance characteristics from the Mallinckrodt Hi-Lo™ Evac Endotracheal Tube, the Portex Tracheal Tube and the Portex® Blue Line® Ultra Suctionaid Tracheostomy Tube, including similar sterilisation and packaging processes."
  • "The determination of substantial equivalence of the Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube to the predicate devices was based on a comparison of device technological characteristics and materials of composition."

Therefore, the "acceptance criteria" are implied to be that the device's technological characteristics, materials, intended use, and performance characteristics are substantially equivalent to the cited predicate devices. There are no quantitative performance metrics for the SACETT™ device itself presented in this summary to fill a table.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size for Test Set: Not applicable. The document does not describe a study with a test set in the conventional sense of a clinical trial or performance evaluation with patient data. The basis for equivalence is a comparison of design, materials, and intended use, not observed performance on a sample of patients or specimens.
  • Data Provenance: Not applicable, as there's no study explicitly described providing new data. The information is based on the design and characteristics of the new device in comparison to existing predicate devices.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

  • Not applicable. There is no mention of experts establishing a ground truth for a test set. The assessment is a regulatory review by the FDA based on the provided technical documentation and comparison to predicate devices, not an expert panel reviewing performance data.

4. Adjudication Method for the Test Set:

  • Not applicable. There is no test set or adjudication method described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

  • No. The document presents a 510(k) summary for a physical medical device (tracheal tube), not an AI-powered diagnostic tool. Therefore, an MRMC study and effects of AI assistance are not relevant or described.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • No. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used:

  • Not applicable in the context of device performance data. The "ground truth" for the 510(k) submission is the established safety and effectiveness of the legally marketed predicate devices, against which the new device is compared for substantial equivalence in terms of design, materials, and intended use.

8. The Sample Size for the Training Set:

  • Not applicable. As a physical medical device, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. There is no training set for a physical medical device.

{0}------------------------------------------------

Koriore

AUG 1 3 2008

SECTION 5: 510(k) SUMMARY

Trade Name:Portex® Blue Line® SACETT™ Suction Above the Cuff TrachealTube
Common Name:™Endotracheal Tube
Classification Name:Tube, Tracheal (21 CFR 888.5730, Product Code BTR)
Contact Person:John TulletRegulatory Affairs Manager (International)Smiths Medical International Ltd.Military Road,Hythe,Kent,England.CT21 6JLPhone: 00 44 (0)1303 260551Fax: 00 44 (0)303 262798
Equivalent to:Hi-Lo™ Evac Endotracheal Tube - K965132Portex® Tracheal Tube - already marketed in the USA from 1976under pre-amendment arrangement.Portex® Blue Line® Ultra Suctionaid Tracheostomy Tube -
K030570
Device Description:Single use Tracheal tube with integral suction line for lavage andevacuation above the cuff.The product design is based upon a tracheal tube with the additionof a third (integral) lumen within the tube. The lumen terminatesabove the cuff via a 'notch' which enables the entrance (viasuction) of secretions which have pooled above the cuff into thethird (suction) lumen. Approximately half way along the tubelength the suction lumen is joined to a suction line which isexternal to the main tube. The suction line is joined to the suctionlumen in a similar manner to that of the joint between the inflationline and the inflation lumen. The distal (clinician) end of thesuction line terminates in a connector which can be connected toeither suction tubing or a syringe.

{1}------------------------------------------------

Intended Use: Portex® Blue Line® SACETT™ is intended for oral intubation only for airway management of patients anticipated to require prolonged mechanical ventilation in critical care units. The product may be inserted in the emergency room or pre-hospital to avoid the need for re-intubation, SACETT™'s dorsal suction lumen allows drainage by continuous or intermittent suctioning of contaminated mucous and subglottis secretion that accumulate above the cuff.

Substantial Equivalence:

The Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube has the same intended use as the Mallinckrodt Medical Hi-Lo™ Evac Endotracheal Tube.

The Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube is composed essentially of the same materials as the Portex Tracheal Tube and the Portex® Blue Line® Ultra Suctionaid (BLUS) Tracheostomy Tube. The Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube incorporates a number of design and performance characteristics from the Mallinckrodt Hi-Lo™ Evac Endotracheal Tube, the Portex Tracheal Tube and the Portex® Blue Line® Ultra Suctionaid Tracheostomy Tube, including similar sterilisation and packaging processes.

The determination of substantial equivalence of the Portex® Blue Line® SACETT™ Suction Above the Cuff Tracheal Tube to the predicate devices was based on a comparison of device technological characteristics and materials of composition.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 3 2008

Smiths Medical ASD, Incorporated C/O Mr. John Tullett Regulatory Affairs Manager Smiths Medical International Military Road Hythe, Kent, United Kingdom CT216JL

Re: K081086

Trade/Device Name: Portex® Blue Line® SACETT™ Suction above the Cuff Tracheal Tube Regulation Number: 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: August 1, 2008 Received: August 1, 2008

Dear Mr, Tullett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Tullett

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

JAhrmuels Lend. mis for 11

Chiu S. Lin, Ph. D Division Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

SECTION 4: INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

Portex® Blue Line® SACETT™ Suction Above the Cuff Device Name: Tracheal Tube

Portex® Blue Line® SACETT™ is intended for oral Indications for Use: intubation only for airway management of patients anticipated to require prolonged mechanical ventilation in critical care units. The product may be inserted in the emergency room or pre-hospital to avoid the need for reintubation, SACETT™'s dorsal suction lumen allows drainage by continuous or intermittent suctioning of contaminated mucous and subglottis secretion that accumulate above the cuff.

Prescription Use______________________________________________________________________________________________________________________________________________________________

AND/ OR

Over-The-Counter Use

(Per 21 CFR 801 Subpart D)

(Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K091086

510(k) Number: _______________________________________________________________________________________________________________________________________________________________

§ 868.5730 Tracheal tube.

(a)
Identification. A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.(b)
Classification. Class II (performance standards).