(99 days)
To provide CPAP to spontaneously breathing adult patients in the hospital and pre-hospital (EMS) environment.
The Pulmodyne CHF Flow Generator is a venturi type oxygen / air mixture delivery device which provides CPAP pressure at high flows to a spontaneously breathing patient. It can deliver up to 150 Lpm flow at a FiO2 between 28- 100%. It utilizes standard in-line PEEP valves to set the prescribed pressure and interfaces with the patient via a face mask or ET tube. The Pulmodyne CHF Flow Generator incorporates several components:
Flow generator (two styles - fixed flow and variable flow) .
Patient interface mask .
PEEP valve .
Circuit / tubing and connectors .
Air entrainment filter .
The Pulmodyne CHF Flow Generator is multi-patient, reusable and can be cleaned while the other components: circuit, mask, entrainment filter, and PEEP valve are disposable, single patient use.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Pulmodyne CHF Flow Generator:
This document is a 510(k) Premarket Notification Summary for a medical device, specifically the Pulmodyne CHF Flow Generator. A 510(k) submission generally aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness through extensive clinical trials. Therefore, the "study" referenced here is primarily a comparison to a predicate device to establish substantial equivalence.
Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly defined by the characteristics of the predicate device, the Caradyne - Whisperflow - K982283. The Pulmodyne CHF Flow Generator demonstrates substantial equivalence by matching or exceeding these characteristics.
| Attribute | Acceptance Criteria (Predicate: Whisperflow WF 8500 K982283) | Reported Device Performance (Proposed: Pulmodyne CHF Flow Generator) |
|---|---|---|
| Use | ||
| Intended for delivery of CPAP | Yes | Yes |
| Used with PEEP valves, CPAP mask and circuit tubing | Yes | Yes |
| Environment - Hospital, sub-acute and pre-hospital (EMS) | Yes | Yes |
| Design | ||
| Works by a venturi method to create a vacuum to provide high flows | Yes | Yes |
| Has oxygen inlet fitting which attach to wall oxygen source standard CGA or DISS fitting | Yes | Yes |
| Has an On / Off valve | Yes | Yes |
| Can adjust oxygen flow through the venturi port | Yes | Yes |
| Has an air entrainment port with 22 mm ID inlet | Yes | Yes |
| Air entrainment port with particulate filter | Yes | Yes |
| Has a valve which adjusts the flow to the venturi nozzle | Yes | Yes |
| Outlet port (22 mm) | Yes | Yes |
| Option for an in-line oxygen analyzer | Yes | Yes |
| Circuit is standard 22 mm tubing | Yes | Yes |
| Can have a humidifier placed in-line | Yes | Yes |
| Connects to patient interface - mask or ET tube | Yes | Yes |
| Utilizes a standard PEEP valve to establish the circuit pressure | Yes | Yes |
| Patient can entrain room air should oxygen flow fail | Yes | Yes |
| One-way valve to prevent rebreathing, if no gas flow | Yes (in mask) | Yes (in elbow) |
| Flow generator can be cleaned and is reusable | Yes | Yes |
| Other components -- circuit, mask, PEEP valve, entrainment filter are Disposable, single patient use | Yes | Yes |
| Accessories required | ||
| CPAP mask | Yes | Yes |
| Particulate filter at air entrainment port | Yes | Yes |
| 22 mm tubing | Yes | Yes |
| Head strap for mask | Yes | Yes |
| Various connectors | Yes | Yes |
| PEEP valves | Yes | Yes |
| Performance Standards / Specifications | ||
| Oxygen input pressure 60 psi | Yes | Yes |
| Output flow range up to 140 Lpm | Yes | Yes |
| Fittings - 22 and 30 mm | Yes | Yes |
| Range of Oxygen (FiO₂) | 28-100% | 28-100% |
| Materials | ||
| Generator | Stainless steel and PVC | Stainless steel and PVC |
| CPAP mask | PVC and silicone | PVC and silicone |
| Circuit tubing and connectors | PE | PE |
Study Details
Given this is a 510(k) submission based on substantial equivalence, the "study" is primarily a comparison of the new device's specifications and features to those of the predicate device. There is no mention of a traditional efficacy or performance study involving patients or a test set in the provided summary.
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: Not applicable in the context of a clinical test set. The "test" is a comparison of product specifications and design to a predicate device.
- Data Provenance: The data comes from the design specifications and materials of the proposed Pulmodyne CHF Flow Generator and is compared directly against the documented characteristics of the predicate device (Caradyne - Whisperflow - K982283), as presented by the manufacturer. This is likely an internal engineering and design comparison.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable. Ground truth, in this context, would be the established specifications and performance of the predicate device, which are typically determined by the predicate device manufacturer and accepted by the FDA. The submission itself is prepared by engineering/regulatory personnel from Pulmodyne.
-
Adjudication method for the test set:
- Not applicable. There's no clinical data or expert review of cases described. The comparison is based on objective device characteristics and features.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done. This device is a mechanical CPAP flow generator, not an AI-powered diagnostic or assistive technology. Therefore, this type of study is entirely irrelevant to this submission.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No standalone algorithm performance study was done. This device is not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The "ground truth" for this substantial equivalence submission is the established design, performance specifications, and intended use of the legally marketed predicate device (Caradyne - Whisperflow - K982283). The new device's characteristics are compared directly to these established features.
-
The sample size for the training set:
- Not applicable. This is not an AI/machine learning device requiring a training set.
-
How the ground truth for the training set was established:
- Not applicable. No training set is involved.
In summary: The provided document is a 510(k) summary, which demonstrates substantial equivalence of a new medical device to a previously cleared predicate device. It primarily relies on a direct comparison of design, materials, and performance specifications rather than a clinical trial with patient data or extensive "studies" in the traditional sense of diagnostic or therapeutic efficacy. The "acceptance criteria" are the features and performance of the predicate device, and the "study" is the detailed comparison showing that the new device meets or is equivalent to these criteria.
{0}------------------------------------------------
MAY - 9 2008
Non-Confidential Summary of Safety and Effectiveness
Page 1 of 4 6-May-08
| Pulmodyne2055 Executive Dr.Indianapolis, IN 46241 | Tel - (317) 246-5505 |
|---|---|
| Official Contact: | Tami Lefevers, Quality Manager |
| Proprietary or Trade Name: | Pulmodyne CHF Flow Generator |
| Common/Usual Name: | CPAP flow generator |
| Classification Name: | Breathing Attachment Positive End Expiratory PressureBYE - 868.5965 |
| Predicate Devices: | Caradyne - Whisperflow - K982283 |
Device Description:
The Pulmodyne CHF Flow Generator is a venturi type oxygen / air mixture delivery device which provides CPAP pressure at high flows to a spontaneously breathing patient. It can deliver up to 150 Lpm flow at a FiO2 between 28- 100%. It utilizes standard in-line PEEP valves to set the prescribed pressure and interfaces with the patient via a face mask or ET tube. The Pulmodyne CHF Flow Generator incorporates several components:
Flow generator (two styles - fixed flow and variable flow) .
- Patient interface mask .
- PEEP valve .
- Circuit / tubing and connectors .
- Air entrainment filter .
The Pulmodyne CHF Flow Generator is multi-patient, reusable and can be cleaned while the other components: circuit, mask, entrainment filter, and PEEP valve are disposable, single patient use.
| Indications for Use: | To provide CPAP to spontaneously breathing adult patients in thehospital and pre-hospital (EMS) environment. |
|---|---|
| Patient Population: | Adults |
| Environment of Use: | Hospital, sub-acute institution, or pre-hospital (EMS) |
{1}------------------------------------------------
Non-Confidential Summary of Safety and Effectiveness
Page 2 of 4 6-May-08 ·
There are several conditions were therapeutic CPAP is Contraindications: contraindicated:
Patients who have:
.
- Facial lacerations .
- Laryngeal trauma .
- Recent tracheal or esophageal anastomosis .
- Gastrointestinal bleeding or ileus .
-
- Recent gastric surgery
- Basilar skull fracture .
- ◆ Patients at high risk of vomiting
- Emphysematous Bulla when an area of the lung may be brittle and present a risk . of bursting
- Hypovolaemia low blood volume .
| Attribute | Variable / Adjustable | |
|---|---|---|
| Whisper flowWF 8500K982283 | ProposedPulmodyne CHFFlow Generator | |
| Use | ||
| Intended for delivery of CPAP | Yes | Yes |
| Used with PEEP valves, CPAP mask andcircuit tubing | Yes | Yes |
| Environment - Hospital, sub-acute andpre-hospital (EMS) | Yes | Yes |
| Design | ||
| Works by a venturi method to create avacuum to provide high flows | Yes | Yes |
| Has oxygen inlet fitting which attach to walloxygen source standard CGA or DISS fitting | Yes | Yes |
| Has an On / Off valve | Yes | Yes |
| Can adjust oxygen flow through theventuri port | Yes | Yes |
| Has an air entrainment port with22 mm ID inlet | Yes | Yes |
| Attribute | Variable / Adjustable | |
| Whisper flowWF 8500K982283 | ProposedPulmodyne CHFFlow Generator | |
| Air entrainment port with particulate filter | Yes | Yes |
| Has a valve which adjusts the flow to theventuri nozzle | Yes | Yes |
| Outlet port (22 mm) | Yes | Yes |
| Option for an in-line oxygen analyzer | Yes | Yes |
| Circuit is standard 22 mm tubing | Yes | Yes |
| Can have a humidifier placed in-line | Yes | Yes |
| Connects to patient interface - maskor ET tube | Yes | Yes |
| Utilizes a standard PEEP valve toestablish the circuit pressure | Yes | Yes |
| Patient can entrain room air shouldoxygen flow fail | Yes | Yes |
| One-way valve to preventrebreathing, if no gas flow | Yes in mask | Yes in elbow |
| Flow generator can be cleaned and is reusable | Yes | Yes |
| Other components -- circuit, maskPEEP valve, entrainment filter areDisposable, single patient use | Yes | Yes |
| Accessories required -CPAP mask | Yes | Yes |
| Particulate filter at air entrainment port | Yes | Yes |
| 22 mm tubing | Yes | Yes |
| Head strap for mask | Yes | Yes |
| Various connectors | Yes | Yes |
| PEEP valves | Yes | Yes |
| Attribute | Variable / Adjustable | |
| Whisper flowWF 8500K982283 | ProposedPulmodyne CHFFlow Generator | |
| Performance Standards / Specifications | ||
| Oxygen input pressure 60 psi | Yes | Yes |
| Output flow range up to 140 Lpm | Yes | Yes |
| Fittings - 22 and 30 mm | Yes | Yes |
| Range of Oxygen (FiO₂) | 28-100% | 28-100% |
| Materials | ||
| Generator -Stainless steel and PVC | Yes | Yes |
| CPAP mask - PVC and silicone | Yes | Yes |
| Circuit tubing and connectors - PE | Yes | Yes |
{2}------------------------------------------------
Non-Confidential Summary of Safety and Effectiveness
Page 3 of 4
6-May-08
:
·
{3}------------------------------------------------
Non-Confidential Summary of Safety and Effectiveness Page 4 of 4 6-May-08
Conclusion:
There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tails, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 9 2008
Pulmodyne C/O Mr. Paul E. Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134-2958
Re: K080256
Trade/Device Name: Pulmodyne CHF Flow Generator Regulation Number: 21 CFR 868.5965 Regulation Name: Positive End Expiratory Pressure Breathing Atachment Regulatory Class: II Product Code: BYE Dated: May 6, 2008 Received: May 7, 2008
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sutte Y. Michael Omd.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use Statement
Page 1 of 1
510(k) Number: K080256 (To be assigned)
Pulmodyne CHF Flow Generator Device Name:
Indications for Use:
To provide CPAP to spontaneously breathing adult patients in the hospital and prehospital (EMS) environment.
Prescription Use XX (Part 21 CFR 801 Subpart D) or
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off
(Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number. K080256
§ 868.5965 Positive end expiratory pressure breathing attachment.
(a)
Identification. A positive end expiratory pressure (PEEP) breathing attachment is a device attached to a ventilator that is used to elevate pressure in a patient's lungs above atmospheric pressure at the end of exhalation.(b)
Classification. Class II (performance standards).