(432 days)
The Portable Oxygen Unit SCA900 is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults; a pressure regulator, flow meter and oxygen cylinder fully integrated into single unit. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
The Portable Oxygen Unit SCA900 is a portable oxygen delivery system, consisting of a fully integrated cylinder, valve, regulator, nasal cannula and mask. The oxygen is delivered through the mask or nasal cannula with a range of user-selectable flow settings. This unit is suitable for use in all healthcare settings. including hospital and home healthcare.
This submission is for a Class I medical device, a Portable Oxygen Unit, which typically does not require extensive clinical studies to demonstrate effectiveness or safety. The FDA filing indicates a "substantial equivalence" determination to existing predicate devices based on design, materials, indications, intended use, packaging, labeling, and performance. Therefore, a formal study with defined acceptance criteria and detailed performance reporting as one might expect for a higher-risk device with novel technology is not present in this document.
The documentation provided does not contain specific acceptance criteria or a study designed to prove the device meets those criteria in the way a clinical trial would. Instead, the submission relies on demonstrating substantial equivalence to already legally marketed predicate devices.
Here's an analysis based on the provided text, addressing the requested points where possible, and noting where information is not available:
1. Table of Acceptance Criteria and Reported Device Performance
As this is a 510(k) submission for a substantially equivalent device of Class I, specific, quantifiable acceptance criteria and corresponding performance metrics from a formal study are not explicitly stated or provided in the document. The acceptance is based on demonstrating equivalence in general aspects.
| Acceptance Criteria (Implied) | Reported Device Performance (Implied from "Substantially Equivalent") |
|---|---|
| Equivalent Design | Design is comparable to predicate devices. |
| Equivalent Materials | Materials are comparable to predicate devices. |
| Equivalent Indications | Indications for Use are comparable to predicate devices. |
| Equivalent Intended Use | Intended Use is comparable to predicate devices. |
| Equivalent Packaging | Packaging is comparable to predicate devices. |
| Equivalent Labeling | Labeling is comparable to predicate devices. |
| Equivalent Performance | Performance is comparable to predicate devices (e.g., oxygen delivery, safety features). |
The phrase "Testing and other comparisons have established that the subject of Portable Oxygen Unit SCA900 is substantially equivalent in design, materials, indications and intended use, packaging, labeling, and performance to other predicate devices of the type currently marketed in the U.S." (page 2) is the primary "proof" of meeting the (implied) acceptance criteria.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not applicable/not provided. This type of submission does not detail a "test set" in the context of clinical or diagnostic performance assessment.
- Data Provenance: Not applicable/not provided. The comparison is against predicate devices and likely relies on design specifications and engineering bench testing, rather than patient data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Number of Experts: Not applicable/not provided. Ground truth in the context of expert review of images or clinical outcomes is not relevant for this type of device submission.
- Qualifications of Experts: Not applicable/not provided.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable/not provided. There is no mention of a "test set" requiring adjudication in the provided documentation.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size
- MRMC Study: No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices to assess human reader performance with and without AI assistance. It is not relevant for a portable oxygen unit.
- Effect Size: Not applicable.
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done
- Standalone Performance: No, a standalone performance assessment (in the context of an algorithm) was not done. This device is a mechanical system, not an AI or algorithmic device.
7. The Type of Ground Truth Used
- Type of Ground Truth: Not applicable in the traditional sense of clinical or diagnostic studies. The "ground truth" for this submission is based on the established safety and effectiveness of the predicate devices and the demonstration that the SCA900 is functionally equivalent to them. This would primarily involve engineering specifications, performance testing (e.g., flow rate accuracy, pressure regulation, durability), and compliance with relevant standards, rather than expert consensus on medical images or patient outcomes.
8. The Sample Size for the Training Set
- Sample Size: Not applicable/not provided. This device relies on established engineering principles and materials, not a "training set" for a machine learning algorithm.
9. How the Ground Truth for the Training Set Was Established
- Ground Truth Establishment: Not applicable/not provided. As there is no training set, there is no ground truth establishment for it.
{0}------------------------------------------------
K080243
page 1 of 2
Portable Oxygen Unit SCA900 510(K) Submission
Sancheong Co., Ltd.
APR - 7 2009
510(K) Summary
Submitter Sancheong Co., Ltd. Chris Kim 185 Songmoon-ri, Yangji-Myunyongin-City Gyoung gi-do, South Korea Tel: 82-031-321-4077 Fax: 82-031-321-4088
Official Correspondent Kodent Inc. Jung Bae Bang 13340 E. Firestone Blvd. Suite J Santa Fe Springs, CA 90670 Email: kodentinc@kodent.co.kr Phone: 562-404-8466 Fax: 562-404-2757
Device Information
Product Name: Portable Oxygen Unit SCA900
Common Name: Portable Oxygen Delivery System
Classification Name: Cylinder, Compressed Gas, and Valve
Product Code: CAN, KGA, ECX
Regulation Number: 868.2700
Device Class: Class I
General Description
The Portable Oxygen Unit SCA900 is a portable oxygen delivery system, consisting of a fully integrated cylinder, valve, regulator, nasal cannula and mask. The oxygen is delivered through the mask or nasal cannula with a range of user-selectable flow settings. This unit is suitable for use in all healthcare settings. including hospital and home healthcare.
Indication for Use
The Portable Oxygen Unit SCA900 is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults; a pressure regulator, flow meter and oxygen cylinder fully integrated into single unit. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
{1}------------------------------------------------
RU80243
page 2 of 2
Sancheong Co., Ltd.
Predicate Devices
The subject device is substantially equivalent to the following predicate devices:
- . Cyl-Fil Oxygen System (K061785) manufactured by Responsive Respiratory Inc.
- EZ-OX Plus (K053117) manufactured by Air Liquide Healthcare America. .
- LIV (Linde Integrated Valve) (K063354) manufactured by Linde Gas Therapeutics. .
Comparison to Predicate Devices
Testing and other comparisons have established that the subject of Portable Oxygen Unit SCA900 is substantially equivalent in design, materials, indications and intended use, packaging, labeling, and performance to other predicate devices of the type currently marketed in the U.S.
2
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, facing to the right. The eagle is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the border of the circle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
San Cheong Company, Limited C/o Mr. Jung Bae Bang Kodent, Incorporated 13340 East Firestone Boulevard, Suite J Santa Fe Springs, California 90670
APR - 7 2009
K080243 Re:
Trade/Device Name: Portable Oxygen Unit SCA900 Regulation Number: 21 CFR 868.2700 Regulation Name: Pressure Regulator Regulatory Class: 1 Product Code: CAN, ECX, KGA Dated: April 3. 2009 Received: April 3, 2009
Dear Mr. Bang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Bang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Shula A. Murphy, MD, Susan Reuma DDS, MD
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indication for Use
510(K) Number (if known): K080243
Device Name: Portable Oxygen Unit SCA900
Indication for Use:
The Portable Oxygen Unit SCA900 is an integrated portable oxygen delivery system intended to provide supplemental oxygen to adults; a pressure regulator, flow meter and oxygen cylinder fully integrated into single unit. For emergency use only when administered by properly trained personnel for oxygen deficiency and resuscitation. For all other medical applications, Rx only. Compressed gas cylinders in service or in storage shall be stabilized or otherwise secured to prevent falling and rolling.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter_ X
(Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Aria Cles
(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
510(k) Number: K080243
§ 868.2700 Pressure regulator.
(a)
Identification. A pressure regulator is a device, often called a pressure-reducing valve, that is intended for medical purposes and that is used to convert a medical gas pressure from a high variable pressure to a lower, more constant working pressure. This device includes mechanical oxygen regulators.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 868.9.