(83 days)
The OrthoPediatrics Plating System is intended for use in the treatment of pelvic, small and long bone fractures.
The OrthoPediatrics Pediatric Plating system is a system of bone plates and screws to provide the pediatric and general orthopedic surgeon with necessary devices to achieve optimal bone fixation when treating fractures, osteotomies, mal-unions, and mal-alignments.
- Materials: The devices are manufactured from 316L stainless steel, which . meet ASTM F138 and F139, and ISO-5832 standards.
- Eunction: The system functions to provide immediate stability and . temporary fixation during the natural healing process.
The system includes cortical and cannulated screws, and specialty plates. The specialty plates will include three distinct shapes and each shape will be offered in multiple sizes to accommodate the individual requirements of the patient anatomy to aid the surgeon in achieving optimal fixation. The three specialty plates' shapes resemble the letters H and I, and the infinity symbol (similar to the number 8, lying on its side).
The provided text is related to a 510(k) premarket notification for a medical device (OrthoPediatrics Plating System). It primarily focuses on demonstrating substantial equivalence to predicate devices based on design, materials, and indications for use.
Crucially, it does not contain any information about acceptance criteria, device performance studies, sample sizes, ground truth establishment, expert adjudication, or comparative effectiveness studies (MRMC or standalone AI performance tests). These types of studies and their results are typically required for AI/ML-driven devices to demonstrate their clinical validity and utility.
Therefore, I cannot populate the requested table and answer the study-related questions based on the provided text.
The document is a standard 510(k) clearance letter for a traditional medical device (bone plates and screws), where substantial equivalence is typically established through comparisons of material properties, design specifications, and intended use with already marketed devices, rather than through performance metrics of an AI algorithm.
{0}------------------------------------------------
K073344 (pg. 1 of 2)
SUMMARY OF SAFETY AND EFFECTIVENESS
| NAME OF FIRM: | OrthoPediatrics, Corp.210 N. Buffalo StreetWarsaw, Indiana 46580Establishment Registration No.: 9102640 |
|---|---|
| FEB 19 2003 | |
| 510(K) CONTACT: | Gary BarnettVP-Regulatory & QualityTel: (574) 268-6379Fax: (574) 269-3692 |
| TRADE NAME: | OrthoPediatrics Plating System |
| COMMON NAME: | Plate, Fixation, Bone |
| CLASSIFICATION: | Single/multiple component metallic bone fixationappliances and accessories: Class II per 21 CFR§888.3030 |
| DEVICE PRODUCT CODE(S): | HRS and HWC |
| SUBSTANTIALLYEQUIVALENT DEVICES: | Precimed Trauma System (K002486)Precimed Cannulated Screw System (K050754)Smith & Nephew Bone Plate System (TC-100Plating and Screw System) (K993106) |
DEVICE DESCRIPTION:
The OrthoPediatrics Pediatric Plating system is a system of bone plates and screws to provide the pediatric and general orthopedic surgeon with necessary devices to achieve optimal bone fixation when treating fractures, osteotomies, mal-unions, and mal-alignments.
- Materials: The devices are manufactured from 316L stainless steel, which . meet ASTM F138 and F139, and ISO-5832 standards.
- Eunction: The system functions to provide immediate stability and . temporary fixation during the natural healing process.
{1}------------------------------------------------
K073344 (pg. 2 of 2)
The OrthoPediatrics Plating System is intended for use in the treatment of pelvic, small and long bone fractures. The system includes cortical and cannulated screws, and specialty plates. The specialty plates will include three distinct shapes and each shape will be offered in multiple sizes to accommodate the individual requirements of the patient anatomy to aid the surgeon in achieving optimal fixation. The three specialty plates' shapes resemble the letters H and I, and the infinity symbol (similar to the number 8, lying on its side).
INDICATIONS FOR USE:
The OrthoPediatrics Plating System is intended for use in the treatment of pelvic, small and long bone fractures.
BASIS FOR SUBSTANTIAL EQUIVALENCE:
OrthoPediatrics believes that this system is substantially equivalent to the legally marketed predicate devices based on similarities in design, materials, and indications. Indications include fixation for pelvic, small and long bone fractures.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. The eagle is depicted in black and white.
Public Health Service
FEB 19 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OrthoPediatrics, Corporation % Mr. Gary Barnett Vice President, Regulatory & Quality 210 N. Buffalo Street Warsaw, Indiana 46580
Re: K073344 Trade/Device Name: OrthoPediatrics Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories. Regulatory Class: Class II Product Code: HRS, HWC Dated: November 19, 2007 Received: November 28, 2007
Dear Mr. Barnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Mr. Gary Barnett
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K073344
Device Name: OrthoPediatrics Plating System
The OrthoPediatrics Plating System is intended for use in the treatment of pelvic, small and long bone fractures.
Prescription Use × Over-The-Counter Use Preschption Ose
(Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Mark A. Mulhern
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
510(k) Number K07 33 44
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.