K Number
K073062
Device Name
OMNI-VUE SYSTEM
Date Cleared
2008-03-10

(132 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
Omni-Vue™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc.; and teleconferencing. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images must only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. The Omni-Vue device allows limited image processing capability: "sharpening preset" and window leveling function. For image editing; the only an "L" or "R" marker can be added to the image but it must be already defined in the original image DICOM header as received by the system. "L" or "R" can not be added to the image if left & right is not defined in the original image. The device provides scout line which is a reference line and is drawn on the AP image to display the location of the slice being viewed. In multi series images, whenever one image is clicked, scout line will be drawn on other series which is crossed direction.
Device Description
Omni-Vue™ System makes possible the capturing, storage, distribution, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available, the System can acquire medical images using a DICOM gateway, which generates DICOM-type files. For example, film digitizers obtain images from old film and convert them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from an imaging workstation.
More Information

No
The document describes a standard PACS system with basic image processing and viewing capabilities. There is no mention of AI, ML, or related concepts.

No
The device is described as a software device for receiving, storing, communicating, processing, and displaying digital images and data, primarily for viewing and manipulation of medical images, rather than for direct treatment or diagnosis.

No
The document describes a software device for managing and displaying medical images, not for diagnosing medical conditions. It explicitly states that mammographic images must be interpreted using an FDA approved monitor and that lossy compressed mammographic images should not be reviewed for primary image interpretation, implying that the device itself is not for primary diagnosis.

Yes

The device is explicitly described as a "software device" in the Intended Use section and its function is centered around the handling, processing, and display of digital medical images and data. While it interacts with hardware (imaging sources, monitors, workstations), the device itself is presented as the software system managing these interactions and data.

Based on the provided information, the Omni-Vue™ System is not an IVD (In Vitro Diagnostic).

Here's why:

  • IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to provide information for diagnosis, monitoring, or screening.
  • Omni-Vue System Function: The Omni-Vue System is described as a software device that receives, stores, communicates, processes, and displays digital images and data from various imaging modalities (CT, MR, ultrasound, etc.). It is used for viewing and manipulating these images for diagnostic purposes.
  • Lack of Biological Sample Analysis: The system does not perform any analysis on biological samples from the human body. Its function is centered around the handling and display of medical images.

Therefore, the Omni-Vue System falls under the category of medical imaging software or a Picture Archiving and Communication System (PACS), not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

Omni-Vue™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc.; and teleconferencing. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images must only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

The Omni-Vue device allows limited image processing capability: "sharpening preset" and window leveling function.

For image editing; the only an "L" or "R" marker can be added to the image but it must be already defined in the original image DICOM header as received by the system. "L" or "R" can not be added to the image if left & right is not defined in the original image.

The device provides scout line which is a reference line and is drawn on the AP image to display the location of the slice being viewed. In multi series images, whenever one image is clicked, scout line will be drawn on other series which is crossed direction.

Product codes

LLZ

Device Description

Omni-Vue™ System makes possible the capturing, storage, distribution, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available, the System can acquire medical images using a DICOM gateway, which generates DICOM-type files. For example, film digitizers obtain images from old film and convert them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from an imaging workstation.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K042311

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

K073062

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared: October 6, 2007

Submitter's Information: 21 CFR 807.92(a)(1) Mr. Alain Tabares, Chief Technical Officer Genesis Digital Imaging, Inc. 12921 W. Washington Blvd. Los Angeles, CA 90066

MAR 1 0 2008

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) Omni-Vue™ System Trade Name: Common Name: Picture Archiving Communications System Device Classification: 892 2050 System, Image Processing Name:

Predicate Device: 21 CFR 807. 92(a)(3)

Device Classification Namesystem, image processing, radiological
510(k) NumberK042311
Regulation Number892.2050 Class II
Device NamePACSPartner™
ApplicantMedical Standard Co. Ltd.,
Product CodeLLZ
Decision Date09/09/2004
DecisionSubstantially equivalent (SE)
Classification Advisory CommitteeRadiology
Review Advisory CommitteeRadiology
TypeTraditional

Device Description: 21 CFR 807 92(a)(4)

Omni-Vue™ System makes possible the capturing, storage, distribution, manipulation, and networking of medical images at distributed locations. In cases where DICOM images are not directly available, the System can acquire medical images using a DICOM gateway, which generates DICOM-type files. For example, film digitizers obtain images from old film and convert them to meet DICOM standards and stored. Stored files are transmitted using a network and can be viewed or manipulated from an imaging workstation.

Indications for Use: 21 CFR 807 92(a)(5)

Omni-Vue™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer

1

510(k) Summary of Safety and Effectiveness

networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc.; and teleconferencing. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images must only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

The Omni-Vue device allows limited image processing capability: "sharpening preset" and window leveling function.

For image editing; the only an "L" or "R" marker can be added to the image but it must be already defined in the original image DICOM header as received by the system. "L" or "R" can not be added to the image if left & right is not defined in the original image.

The device provides scout line which is a reference line and is drawn on the AP image to display the location of the slice being viewed. In multi series images, whenever one image is clicked, scout line will be drawn on other series which is crossed direction.

Technological Characteristics: 21 CFR 807 92(a)(6)

Omni-Vue™ System is a software product that handles and manipulates digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed.

Conclusion: 21 CFR 807 92(b)(1)

The 510(k) Pre-Market Notification for Omni-Vue™ System contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device.

Omni-Vue™ System has been and will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The submission contains the results of a hazard analysis and the "Level of Concern for potential hazards has been classified as "Minor".

2

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a person with outstretched arms, composed of three curved lines. Below the symbol are two wavy lines.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

MAR 1 0 2008

Genesis Digital Imaging, Incorporated % Mr. Carl Alletto Consultant OTech, Incorporated 1600 Manchester Way CORINTH TX 76210

Re: K073062

Trade/Device Name: Omni-Vue™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 4, 2008 Received: February 6, 2008

Dear Mr. Alletto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular seal with the text "ISOE-2005 FDA Centennial". The seal also contains three stars at the bottom. The text is arranged in a circular fashion around the central FDA logo and the word "Centennial".

Protecting and Promoting Public Health

3

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxxx(Gastroenterology/Renal/Urology)240-276-0115
21 CFR 884.xxxx(Obstetrics/Gynecology)240-276-0115
21 CFR 892.xxxx(Radiology)240-276-0120
Other240-276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You max obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

inclosure

4

510(k) Number:

K073062

Device Name:

Omni-Vue™

Indications for Use:

Omni-Vue™ System is a software device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Device options make possible reading (including mammography), telecommunications; fast demonstration; etc.; and teleconferencing. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images must only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA.

The Omni-Vue device allows limited image processing capability: "sharpening preset" and window leveling function.

For image editing; the only an "L" or "R" marker can be added to the image but it must be already defined in the original image DICOM header as received by the system. "L" or "R" can not be added to the image if left & right is not defined in the original image.

The device provides scout line which is a reference line and is drawn on the AP image to display the location of the slice being viewed. In multi series images, whenever one image is clicked, scout line will be drawn on other series which is crossed direction.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Arnitk. War
(Division Sign Off)

Division of Reproductive, Abde 510(k) Number