K Number
K073046

Validate with FDA (Live)

Date Cleared
2008-02-25

(119 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Articulating Needle Knife is a monopolar electrosurgical instrument intended for cutting, dissecting and cauterizing soft tissue during endoscopic electrosurgical procedures. The Articulating Needle Knife is not intended for use in the central nervous system or in the central circulatory system.

Device Description

The Ethicon Endo Surgery® (EES) Articulating Needle Knife is an endoscopic instrument intended for cutting, dissecting and cauterizing tissue via highfrequency electrical current waveform. The device consists of a flexible wire cable and needle knife electrode, which can be extended, rotated, articulated, and retracted from the flexible outer shaft using a three-finger actuator. When connected to an electrosurgical generator and activated, the Needle Knife delivers a monopolar electrical current to the surgical site. This device passes through endoscopes having a 3.2 mm or larger working channels. This device is supplied sterile for single-patient use.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "Ethicon Endo Surgery Articulating Needle Knife". This device is an electrosurgical instrument. The information given focuses on bench testing and compliance with electrical safety standards, rather than a clinical study involving human readers or sophisticated AI algorithms.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Materials BiocompatibilityComplies with ISO 10993-1
Electrical Safety (Electrosurgical Devices)Complies with AAMI / ANSI HF 18:2001
Electrical Safety (High Frequency Surgical Equipment)Complies with IEC/EN 60601-2-2:2000
Electrical Safety (Specific Safety Requirements)Complies with IEC/EN 60601-2-18:1996
Functionality (cutting, dissecting, cauterizing via RF current)Bench testing performed to demonstrate device performs as intended

2. Sample size used for the test set and the data provenance

  • Test Set Sample Size: Not applicable. The "test set" here refers to the device itself being tested in a benchtop environment, not a dataset of patient images or information.
  • Data Provenance: Not applicable. The performance data is derived from direct electrical and material testing of the manufactured device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. "Ground truth" in this context would typically refer to expert annotations on medical images or clinical outcomes. For this device, the "ground truth" for its performance is established by objective engineering standards and direct physical/electrical measurements, not human expert interpretation of data.

4. Adjudication method for the test set

Not applicable. Adjudication methods like "2+1" or "3+1" are used in studies involving human interpretation or uncertain clinical outcomes. The performance of this device is assessed against objective engineering standards.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No. This document describes a medical device clearance for an electrosurgical instrument, not an AI or imaging diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant and was not performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

Not applicable. This device is an electrosurgical instrument, not a standalone algorithm. Its performance is always predicated on human use connected to an electrosurgical generator.

7. The type of ground truth used

The "ground truth" for the device's performance is compliance with established electrical safety standards (AAMI / ANSI HF 18:2001, IEC/EN 60601-2-2:2000, IEC/EN 60601-2-18:1996) and biocompatibility standard (ISO 10993-1). Additionally, bench testing demonstrated the device performs its intended functions (cutting, dissecting, cauterizing).

8. The sample size for the training set

Not applicable. This device does not involve a "training set" as it is not an AI or machine learning model.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

510(k) Summary

CompanyEthicon Endo-Surgery, Inc.
4545 Creek Road
Cincinnati, OH 45242

Contact Glenda C Marsh QS/RA Project Manager Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-7659 Fax: (513) 337-2860 Email: gmarsh@eesus.ini.com

FEB 25 2003

Date Prepared October 26, 2007

  • New Device Name Trade Namc: Ethicon Endo Surgery Articulating Needle Knife Common or Usual Name: Electrosurgical Needle Electrodes Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400, Product Code GEI)
  • Predicate Devices KSEA Monopolar Coagulating Needle Electrode (K972497) Colorado MicroDissection Needle (K000348) Boston Scientific RX Needle Knife (K970053)

Device Description The Ethicon Endo Surgery® (EES) Articulating Needle Knife is an endoscopic instrument intended for cutting, dissecting and cauterizing tissue via highfrequency electrical current waveform. The device consists of a flexible wire cable and needle knife electrode, which can be extended, rotated, articulated, and retracted from the flexible outer shaft using a three-finger actuator. When connected to an electrosurgical generator and activated, the Needle Knife delivers a monopolar electrical current to the surgical site. This device passes through endoscopes having a 3.2 mm or larger working channels. This device is supplied sterile for single-patient use.

Indications for Use The Articulating Needle Knife is a monopolar electrosurgical instrument intended for cutting, dissecting and cauterizing soft tissue during endoscopic electrosurgical procedures. The Articulating Needle Knife is not intended for use in the central nervous system or in the central circulatory system.

Technological Characteristics The EES device has similar technologic characteristics to the predicate devices in that it contains a metal electrode tip that is used to deliver monopolar energy to the surgical site. In all devices the electrode tip is in the shape of a needle. All devices are designed to be connected to electrosurgical generators, and utilize RF monopolar energy for operation.

The EES device is similar to the Boston Scientific RX Needle Knife in that it consists of an clongated flexible wire shaft and a handle. These devices allow for the manipulation

{1}------------------------------------------------

of the needle knife via the handle of the device. In addition, the EES device features rotation and articulation of the end-effector to provide the clinician with improved tissue targeting capability.

Performance Data. Bench testing was performed to demonstrate that the EES device performs as intended. The device materials have been evaluated for biocompatibility and comply with the requirements of ISO 10993-1. The device was tested to demonstrate compliance with the following standards:

  • AAMI / ANSI HF 18:2001, Electrosurgical Devices - IEC/EN 60601-2-2:2000, Medical electrical Equipment --- Part 2-2: Particular requirements for the safety of high frequency surgical equipment - IEC/EN 60601-2-18:1996, Medical electrical equipment - Part 2: Particular requirements for safety ------------------------------------------------------------------------------------------------------------------------------------------------------

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread.

Public Health Service

FEB 25 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ethicon End-Surgery, Inc. % Ms. Glenda C. Marsh Project Manager, QS/RA 4545 Creek Road Cincinnati, Ohio 45242

Re: K073046

Trade/Device Name: Ethicon Endo Surgery Articulating Needle Knife Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories. Regulatory Class: II Product Code: GEI Dated: February 11, 2008 Received: February 12, 2008

Dear Ms. Marsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 – Ms. Glenda C. Marsh

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Mulhern

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K073046

Device Name: Ethicon Endo Surgery® Articulating Needle Knife

Indications for Use:

The Articulating Needle Knife is a monopolar electrosurgical instrument intended for cutting, dissecting and cauterizing soft tissue during endoscopic electrosurgical procedures. The Articulating Needle Knife is not intended for use in the central nervous system or in the central circulatory system.

Prescription Use(Part 21 CFR 801 Subpart D)X
AND/OROver-The-Counter Use(21 CFR 801 Subpart C)

(Division Sign-Off)

(PLEASE DO NOT WRITE BELOW THIS LINE DIVISION OF CONTINUE ON ANOTHER PAGE IF NEEDED)

Division of General, Restorative, and Neurological Devices (ODE)

Page 1 of 1Concurrence of CDRH,
510(k) NumberIL0B046

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.