K Number
K071880
Manufacturer
Date Cleared
2007-10-05

(88 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OPTIMA™ ZS Spinal System is a posterior pedicle screw fixation system indicated for the treatment of severe spondylolisthesis, (Grade 3 and 4), of the L5-S1 vertebra in skeletally-mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of implants after the attainment of a solid fusion.

In addition, The OPTIMA™ ZS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

  • Degenerative spondylolisthesis with objective evidence of neurological impairment
  • Fracture of the vertebral body
  • Dislocation
  • Scoliosis
  • Kyphosis
  • Spinal tumor
  • Failed previous fusion (pseudoarthrosis).

When the Zimmer GmbH Dynesys® Spinal System and the OPTIMA™ ZS Spinal System are used on contiguous levels, they must be used with the Zimmer GmbH Zimmer DTO™ are acod on combination implant, and the OPTIMA™ ZS Transition Screw. The indications for use for each level is as specified for each system.

Device Description

The OPTIMA™ ZS Transition Screw is a posterior pedicle screw that is used as a transition device when the Zimmer GmbH Dynesys® Spinal System and the OPTIMA ZS Spinal System are used on contiquous levels. The OPTIMA ZS Transition Screw interfaces with the rod section of the Zimmer DTO™ Implant, cord-rod combination implant. The cord portion of the Zimmer DTO™ Implant interfaces with the Dynesys Spinal System.

AI/ML Overview

I am sorry, but the provided text does not contain information about acceptance criteria or a study proving the device meets acceptance criteria. The document is a 510(k) summary for a medical device (OPTIMA™ ZS Transition Screw and OPTIMA™ ZS Spinal System), which focuses on establishing substantial equivalence to predicate devices and outlining the intended use and regulatory classifications.

Therefore, I cannot generate the requested table and study details based on this input.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the number 6671880 in large, bold font. Below the number is the text "U&I CORPORATION" in a smaller, bold font. The text is slightly blurred, but still legible. The image appears to be a scan of a document.

510(k) Summary

OCT 5 " 2007

Manufacturer:

U & I Corporation 529-1, Yonghyun-dong, Uijungbu Kvonagi-Do, Korea 480-050 Dae-Kyoon, Rah, Requlatory Affairs Manager

Sponsor:

Zimmer Spine, Inc. 7375 Bush Lake Road Minneapolis, MN 55439

Sponsor Contact: Nicole Bowden Requlatory Affairs Associate

July 2, 2007 Date Prepared:

Trade Name: OPTIMA™ ZS Transition Screw Device Name:

Common Name: Spinal Fixation System

  • Classification Name: Orthosis, Spondylolisthesis Spinal Fixation (MNH), per 21 CFR 888.3070
    Orthosis, Spinal Pedicle Fixation (MNI), per 21 CFR 888.3070

Spinal Interlaminal Fixation Orthosis (KWP), per 21 CFR 888.3050

Product Code: MNH, MNI and KWP

OPTIMA™ ZS Spinal System Predicate Devices:

Description of Device:

The OPTIMA™ ZS Transition Screw is a posterior pedicle screw that is used as a transition device when the Zimmer GmbH Dynesys® Spinal System and the OPTIMA ZS Spinal System are used on contiquous levels. The OPTIMA ZS Transition Screw interfaces with the rod section of the Zimmer DTO™ Implant, cord-rod combination implant. The cord portion of the Zimmer DTO™ Implant interfaces with the Dynesys Spinal System.

{1}------------------------------------------------

Intended Use:

The OPTIMA™ ZS Spinal System is a posterior pedicle screw fixation system indicated for the treatment of severe spondylolisthesis, (Grade 3 and 4), of the L5-S1 vertebra in skeletally-mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of implants after the attainment of a solid fusion.

In addition, The OPTIMA™ ZS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

  • Degenerative spondylolisthesis with objective evidence of neurological impairment
  • · Fracture of the vertebral body
  • · Dislocation
  • · Scoliosis
  • · Kyphosis
  • · Spinal tumor
  • · Failed previous fusion (pseudoarthrosis).

When the Zimmer GmbH Dynesys® Spinal System and the OPTIMA™ ZS Spinal System are used on contiguous levels, they must be used with the Zimmer GmbH Zimmer DTO™ are acod on combination implant, and the OPTIMA™ ZS Transition Screw. The indications for use for each level is as specified for each system.

Substantial Equivalence:

The OPTIMA ZS Transition Screw is substantially equivalent to the pedicle screws of the OPTIMA ZS Spinal System in design, materials, function and intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, and three horizontal lines above it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

U & I Corporation % Zimmer Spine, Inc. Ms. Nicole Bowden Regulatory Affairs Associate 7375 Bush Lake Road Minneapolis, MN 55441

OCT 5 2007

Re: K071880

Trade/Device Name: OPTIMA™ ZS Transition Screw Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: II Product Code: MNI, MNH, KWP Dated: July 2, 2007 Received: July 9, 2007

Dear Ms. Bowden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the includions for use stated in the enclosure) to legally marketed predicate devices marketed in intestate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendmounts, or to devices that have been reclassified in accordance with the provisions of the Federal Foend, Or to and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls grovisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such uns could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Ant than solor user user limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for the indication of spinal stabilization without fusion have not been established.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

{3}------------------------------------------------

Page 2 - Nicole Bowden

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Donna Rae Tillman

Donna-Bea Tillman Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known):

Device Name: OPTIMA™ ZS Transition Screw

Indications for Use:

The OPTIMA™ ZS Spinal System is a posterior pedicle screw fixation system indicated for the treatment of severe spondylolisthesis, (Grade 3 and 4), of the L5-S1 vertebra in skeletally-mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine with removal of implants after the attainment of a solid fusion.

In addition, The OPTIMA™ ZS Spinal System is intended to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine:

  • · Degenerative spondylolisthesis with objective evidence of neurological impairment
  • · Fracture of the vertebral body
  • · Dislocation
  • · Scoliosis
  • · Kyphosis
  • · Spinal tumor
  • · Failed previous fusion (Pseudoarthrosis).

When the Zimmer GmbH Dynesys® Spinal System and the OPTIMA™ ZS Spinal System are used on contiguous levels, they must be used with the Zimmer GmbH Zimmer DTO™ Implant, rod-cord combination implant, and the OPTIMA™ ZS Transition Screw. The indications for use for each level is as specified for each system.

Prescription Use Over-The-Counter Use - AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hubare buchy Norman
(Division Sign-Off)

Division of General, Restorative, and Neurological Devices

510(k) Number K071880

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.